Instrumentation Laboratory

47 recalls β€’ β€’ Categories: Other Medical Devices, Diagnostic Equipment, Infusion Pumps, Surgical Instruments

Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.

Feb 3, 2025 Other Medical Devices Nationwide View Details β†’

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Sep 29, 2023 Other Medical Devices View Details β†’

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Sep 29, 2023 Other Medical Devices View Details β†’

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Sep 29, 2023 Other Medical Devices View Details β†’

Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result

Jul 13, 2023 Diagnostic Equipment Nationwide View Details β†’

Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.

Apr 8, 2022 Other Medical Devices Nationwide View Details β†’

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

Feb 22, 2022 Other Medical Devices Nationwide View Details β†’

The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.

Oct 13, 2021 Diagnostic Equipment Nationwide View Details β†’

Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days

Aug 25, 2021 Infusion Pumps Nationwide View Details β†’

Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days

Aug 25, 2021 Infusion Pumps Nationwide View Details β†’

There is a potential for increased imprecision, out of range quality controls and prolonged sample results.

Aug 1, 2018 Diagnostic Equipment Nationwide View Details β†’

Some vials of HemosIL PT-Fibrinogen HS PLUS, Part No. 0008469810 (various lots manufactured prior to June 2014) exhibit a yellow/brownish color after reconstitution (yellow color or darker) instead of the typical white to off-white color. Vials exhibiting this yellow/brownish color were observed to cause prolonged prothrombin (PT) clotting times.

Nov 21, 2014 Diagnostic Equipment Nationwide View Details β†’

Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

Jul 19, 2013 Diagnostic Equipment Nationwide View Details β†’

Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

Jul 19, 2013 Diagnostic Equipment Nationwide View Details β†’

Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

Jul 19, 2013 Diagnostic Equipment Nationwide View Details β†’

Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

Jul 19, 2013 Diagnostic Equipment Nationwide View Details β†’

Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

Jul 19, 2013 Diagnostic Equipment Nationwide View Details β†’

Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

Jul 19, 2013 Diagnostic Equipment Nationwide View Details β†’

Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

Jul 19, 2013 Diagnostic Equipment Nationwide View Details β†’

Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

Aug 3, 2012 Diagnostic Equipment Nationwide View Details β†’

Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

Aug 3, 2012 Diagnostic Equipment Nationwide View Details β†’

Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

Aug 3, 2012 Diagnostic Equipment Nationwide View Details β†’

Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

Aug 3, 2012 Diagnostic Equipment Nationwide View Details β†’

Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

Aug 3, 2012 Diagnostic Equipment Nationwide View Details β†’

Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

Aug 3, 2012 Diagnostic Equipment Nationwide View Details β†’