HemosIL PT-Fibrinogen HS PLUS, IVD --- Instrumentation Laboratory IL Company. A high sensitivity thromboplastin reagent based on recombinant human tissue factor (RTF) for the quantitative determination in human citrated plasma of Prothrombin Time (PT) and Fibrinogen on IL Coagulation Systems. The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy (OAT).

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lots: N1122175, exp. date 11/30/2014; N0132936, exp. date 1/31/2015; N0233302, exp. date 2/28/2015; N0333382, exp. date 3/31/2015; E0333382, exp. date 3/31/2015; N0333626, exp. date 3/31/2015; N0433932, exp. date 4/30/2015; N0634953, exp. date 6/30/2015; N0735317, exp. date 5/31/2015; N0735599, exp. date 7/31/ 2015; N0836106, exp. date 8/31/2015; N0936197, exp. date 9/30/2015; N0349325, exp. date 12/31/2015; N1222324, exp. date 12/31/2015; N0148094, exp. date 12/31/2015; N0248652, exp. date 2/29/2016; N0249017, exp. date 2/29/2016; N0449790, exp. date 4/30/2016; N0440009, exp. date 4/30/2016.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Instrumentation Laboratory Co.
Reason for Recall:
Some vials of HemosIL PT-Fibrinogen HS PLUS, Part No. 0008469810 (various lots manufactured prior to June 2014) exhibit a yellow/brownish color after reconstitution (yellow color or darker) instead of the typical white to off-white color. Vials exhibiting this yellow/brownish color were observed to cause prolonged prothrombin (PT) clotting times.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

HemosIL PT-Fibrinogen HS PLUS, IVD --- Instrumentation Laboratory IL Company. A high sensitivity thromboplastin reagent based on recombinant human tissue factor (RTF) for the quantitative determination in human citrated plasma of Prothrombin Time (PT) and Fibrinogen on IL Coagulation Systems. The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy (OAT).

Product Codes/Lot Numbers:

Lots: N1122175, exp. date 11/30/2014; N0132936, exp. date 1/31/2015; N0233302, exp. date 2/28/2015; N0333382, exp. date 3/31/2015; E0333382, exp. date 3/31/2015; N0333626, exp. date 3/31/2015; N0433932, exp. date 4/30/2015; N0634953, exp. date 6/30/2015; N0735317, exp. date 5/31/2015; N0735599, exp. date 7/31/ 2015; N0836106, exp. date 8/31/2015; N0936197, exp. date 9/30/2015; N0349325, exp. date 12/31/2015; N1222324, exp. date 12/31/2015; N0148094, exp. date 12/31/2015; N0248652, exp. date 2/29/2016; N0249017, exp. date 2/29/2016; N0449790, exp. date 4/30/2016; N0440009, exp. date 4/30/2016.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0832-2015

Related Recalls

Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details →