ACL TOP Family 50 Series Models as follows: In vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis (1) ACL TOP 550 CTS Model Number: 00000280045; (2) ACL TOP 750 CTS Model Number: 00000280055; (3) ACL TOP 350 CTS Model Number: 00000280065; (4) ACL TOP 750 Model Number: 00000280015; (5) ACL TOP 750 LAS Model Number: 00000280035 (6) ACL TOP 970 CL(only available ROW)Model Number: 00000280097

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Names UDI (1)ACL TOP 550 CTS 08426950729242; (2)ACL TOP 750 CTS 08426950784067; (3)ACL TOP 350 CTS 08426950784081; (4) ACL TOP 750 08426950784074; (5) ACL TOP 750 LAS 08426950784050 (6) ACL TOP 970 CL 08430793045476 Software: 1)ACL TOP 550 CTS v6.3.0 or later; (2)ACL TOP 750 CTS v6.3.0 or later; (3)ACL TOP 350 CTS v6.3.0 or later; (4) ACL TOP 750 v6.3.0 or later; (5) ACL TOP 750 LAS v6.3.0 or later; (6) ACL TOP 970 CL Version 1.1.0 (OUS). UPDATE: 10/16/23 Affected Serial Number Range Start SN End SN ACL TOP 350 CTS 21113058 23074369; ACL TOP 550 CTS 21112046 23072623; ACL TOP 750 22010516 23060671; ACL TOP 750 CTS 21110860 23071015; ACL TOP 750 LAS 21120879 23071077
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Instrumentation Laboratory
Reason for Recall:
Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

ACL TOP Family 50 Series Models as follows: In vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis (1) ACL TOP 550 CTS Model Number: 00000280045; (2) ACL TOP 750 CTS Model Number: 00000280055; (3) ACL TOP 350 CTS Model Number: 00000280065; (4) ACL TOP 750 Model Number: 00000280015; (5) ACL TOP 750 LAS Model Number: 00000280035 (6) ACL TOP 970 CL(only available ROW)Model Number: 00000280097

Product Codes/Lot Numbers:

Model Names UDI (1)ACL TOP 550 CTS 08426950729242; (2)ACL TOP 750 CTS 08426950784067; (3)ACL TOP 350 CTS 08426950784081; (4) ACL TOP 750 08426950784074; (5) ACL TOP 750 LAS 08426950784050 (6) ACL TOP 970 CL 08430793045476 Software: 1)ACL TOP 550 CTS v6.3.0 or later; (2)ACL TOP 750 CTS v6.3.0 or later; (3)ACL TOP 350 CTS v6.3.0 or later; (4) ACL TOP 750 v6.3.0 or later; (5) ACL TOP 750 LAS v6.3.0 or later; (6) ACL TOP 970 CL Version 1.1.0 (OUS). UPDATE: 10/16/23 Affected Serial Number Range Start SN End SN ACL TOP 350 CTS 21113058 23074369; ACL TOP 550 CTS 21112046 23072623; ACL TOP 750 22010516 23060671; ACL TOP 750 CTS 21110860 23071015; ACL TOP 750 LAS 21120879 23071077

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2578-2023

Related Recalls

Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details →