HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Part Number: 0020300600; UDI-DI: 08426950472483; Lot Numbers: N0128933, N0220450, N0522479, N0724070, N1026509, N0239584, N0531891, N0733658, N0249207, N0442161, N0744673;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Instrumentation Laboratory
Reason for Recall:
Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic

Product Codes/Lot Numbers:

Part Number: 0020300600; UDI-DI: 08426950472483; Lot Numbers: N0128933, N0220450, N0522479, N0724070, N1026509, N0239584, N0531891, N0733658, N0249207, N0442161, N0744673;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1020-2025

Related Recalls

Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Jan 20, 2026 Other Medical Devices Nationwide View Details β†’