Exactech

210 recalls • • Categories: Other Medical Devices, Implants & Prosthetics, Surgical Instruments, Infusion Pumps

AcuMatch XLE Item Number / Product Description: 140-28-27 AM EHXL 0¿ LNR G 28MM 140-32-27 AM EHXL 0¿ LNR G 32MM 140-32-28 AM EHXL 0¿ LNR H 32MM 140-28-29 AM EHXL 0¿ LNR J 28MM 140-32-29 AM EHXL 0¿ LNR J 32MM 140-36-29 AM EHXL 0¿ LNR J 36MM 140-32-30 AM EHXL 0¿ LNR K 32MM 140-36-30 AM EHXL 0¿ LNR K 36MM 140-28-25 AM EHXL 0¿LNR E 28MM 140-28-26 AM EHXL 0¿LNR F 28MM 140-28-28 AM EHXL 0¿LNR H 28MM 140-36-28 AM EHXL 0¿LNR H 36MM 140-28-30 AM EHXL 0¿LNR K 28MM 142-28-24 AM EHXL 15¿ LNR D 28MM 142-28-26 AM EHXL 15¿ LNR F 28MM 142-32-27 AM EHXL 15¿ LNR G 32MM 142-32-28 AM EHXL 15¿ LNR H 32MM 142-36-28 AM EHXL 15¿ LNR H 36MM 142-36-29 AM EHXL 15¿ LNR J 26MM 142-28-29 AM EHXL 15¿ LNR J 28MM 142-32-29 AM EHXL 15¿ LNR J 32MM 142-28-30 AM EHXL 15¿ LNR K 28MM 142-32-30 AM EHXL 15¿ LNR K 32MM 142-36-30 AM EHXL 15¿ LNR K 36MM 142-28-25 AM EHXL 15¿LNR E 28MM 142-28-27 AM EHXL 15¿LNR G 28MM 142-28-28 AM EHXL 15¿LNR H 28MM 144-28-28 AM EHXL EXT LNR H 28MM 144-28-29 AM EHXL EXT LNR J 28MM 144-28-25 AM EHXL EXTLNR E 28MM 144-28-26 AM EHXL EXTLNR F 28MM 144-28-27 AM EHXL EXTLNR G 28MM 144-28-30 AM EHXL EXTLNR K 28MM 148-28-26 AM EHXL LAT LNR F 28MM 148-28-27 AM EHXL LAT LNR G 28MM 148-36-28 AM EHXL LAT LNR H 36MM 148-36-29 AM EHXL LAT LNR J 36MM 148-28-30 AM EHXL LAT LNR K 28MM 148-36-30 AM EHXL LAT LNR K 36MM 148-28-25 AM EHXL LATLNR E 28MM 148-28-28 AM EHXL LATLNR H 28MM 148-28-29 AM EHXL LATLNR J 28MM All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment.

Class I - Dangerous

Due to the lack of the Ethylene Vinyl Alcohol (EVOH) barrier layer.

Dec 31, 2024 Implants & Prosthetics Nationwide View Details →

Novation XLE Item Number / Product Description: 140-22-70 NV EHXL NTL LNR G00 22MM 140-28-50 NV EHXLNTRL LNR G0 28MM 140-28-51 NV EHXLNTRL LNR G1 28MM 140-28-52 NV EHXL NTRL LNR G228MM 140-32-51 NV EHXL NTRL LNR G132MM 140-32-52 NV EHXL NTRL LNR G232MM 140-32-53 NV EHXL NTRL LNR G332MM 140-32-54 NV EHXLNTRL LNR G4 32MM 140-32-55 NV EHXLNTRL LNR G5 32MM 140-36-52 NV EHXLNTRL LNR G2 36MM 140-36-53 NV EHXL NTRL LNR G336MM 140-36-54 NV EHXLNTRL LNR G4 36MM 140-36-55 NV EHXLNTRL LNR G5 36MM 140-40-53 NV EHXLNTRL LNR G3 40MM 140-40-54 NV EHXLNTRL LNR G4 40MM 140-40-55 NV EHXLNTRL LNR G5 40MM 142-22-68 NV EHXL ALPLNR G00 22MM 142-22-70 NV EHXL LIPLNR G00 22MM 142-28-50 NV EHXL LIP LNR G0 28MM 142-28-51 NV EHXL LIP LNR G1 28MM 142-28-52 NV EHXL LIP LNR G2 28MM 142-28-60 NV EHXL ALIPLNR G0 28MM 142-28-61 NV EHXL ALIP LNR G1 28MM 142-28-62 NV EHXL ALIP LNR G2 28MM 142-32-51 NV EHXL LIP LNR G1 32MM 142-32-52 NV EHXL LIP LNR G2 32MM 142-32-53 NV EHXL LIP LNR G3 32MM 142-32-54 NV EHXL LIP LNR G4 32MM 142-32-55 NV EHXL LIP LNR G5 32MM 142-32-61 NV EHXL ALIP LNR G1 32MM 142-32-62 NV EHXL ALIP LNR G2 32MM 142-32-63 NV EHXL ALIP LNR G3 32MM 142-32-64 NV EHXL ALIP LNR G4 32MM 142-32-65 NV EHXL ALIP LNR G5 32MM 142-36-52 NV EHXL LIP LNR G2 36MM 142-36-53 NV EHXL LIP LNR G3 36MM 142-36-54 NV EHXL LIP LNR G4 36MM 142-36-55 NV EHXL LIP LNR G5 36MM 142-36-62 NV EHXL ALIP LNR G2 36MM 142-36-63 NV EHXL ALIP LNR G3 36MM 142-36-64 NV EHXL ALIP LNR G4 36MM 142-36-65 NV EHXL ALIP LNR G5 36MM 142-40-53 NV EHXL LIP LNR G3 40MM 142-40-54 NV EHXL LIP LNR G4 40MM 142-40-55 NV EHXL LIP LNR G5 40MM 142-40-63 NV EHXL ALIP LNR G3 40MM 142-40-64 NV EHXL ALIP LNR G4 40MM 142-40-65 NV EHXL ALIP LNR G5 40MM 146-22-70 NV EHXL LATLNR G00 22MM 146-28-51 NV EHXL LAT LNR G1 28MM 146-32-51 NV EHXL LAT LNR G1 32MM 146-32-52 NV EHXL LAT LNR G2 32MM 146-32-53 NV EHXL LAT LNR G3 32MM 146-32-54 NV EHXL LAT LNR G4 32MM 146-32-55 NV EHXL LAT LNR G5 32MM 146-36-52 NV EHXL LAT LNR G2 36MM 146-36-53 NV EHXL LAT LNR G3 36MM 146-36-54 NV EHXL LAT LNR G4 36MM 146-36-55 NV EHXL LAT LNR G5 36MM 146-40-53 NV EHXL LAT LNR G3 40MM 146-40-54 NV EHXL LAT LNR G4 40MM 146-40-55 NV EHXL LAT LNR G5 40MM 148-32-51 NV EHXL 10¿ LNR G1 32MM 148-36-52 NV EHXL 10¿ LNR G2 36MM 148-36-53 NV EHXL 10¿ LNR G3 36MM 148-36-54 NV EHXL 10¿ LNR G4 36MM 148-36-55 NV EHXL 10¿ LNR G5 36MM 148-40-53 NV EHXL 10¿ LNR G3 40MM 148-40-54 NV EHXL 10¿ LNR G4 40MM 148-40-55 NV EHXL 10¿ LNR G5 40MM All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment.

Class I - Dangerous

Due to the lack of the Ethylene Vinyl Alcohol (EVOH) barrier layer.

Dec 31, 2024 Implants & Prosthetics Nationwide View Details →

Alteon XLE Item Number / Product Description: 140-22-70 NV EHXL NTL LNR G00 22MM 140-28-50 NV EHXLNTRL LNR G0 28MM 140-28-51 NV EHXLNTRL LNR G1 28MM 140-28-52 NV EHXL NTRL LNR G228MM 140-32-51 NV EHXL NTRL LNR G132MM 140-32-52 NV EHXL NTRL LNR G232MM 140-32-53 NV EHXL NTRL LNR G332MM 140-32-54 NV EHXLNTRL LNR G4 32MM 140-32-55 NV EHXLNTRL LNR G5 32MM 140-36-52 NV EHXLNTRL LNR G2 36MM 140-36-53 NV EHXL NTRL LNR G336MM 140-36-54 NV EHXLNTRL LNR G4 36MM 140-36-55 NV EHXLNTRL LNR G5 36MM 140-40-53 NV EHXLNTRL LNR G3 40MM 140-40-54 NV EHXLNTRL LNR G4 40MM 140-40-55 NV EHXLNTRL LNR G5 40MM 142-22-68 NV EHXL ALPLNR G00 22MM 142-22-70 NV EHXL LIPLNR G00 22MM 142-28-50 NV EHXL LIP LNR G0 28MM 142-28-51 NV EHXL LIP LNR G1 28MM 142-28-52 NV EHXL LIP LNR G2 28MM 142-28-60 NV EHXL ALIPLNR G0 28MM 142-28-61 NV EHXL ALIP LNR G1 28MM 142-28-62 NV EHXL ALIP LNR G2 28MM 142-32-51 NV EHXL LIP LNR G1 32MM 142-32-52 NV EHXL LIP LNR G2 32MM 142-32-53 NV EHXL LIP LNR G3 32MM 142-32-54 NV EHXL LIP LNR G4 32MM 142-32-55 NV EHXL LIP LNR G5 32MM 142-32-61 NV EHXL ALIP LNR G1 32MM 142-32-62 NV EHXL ALIP LNR G2 32MM 142-32-63 NV EHXL ALIP LNR G3 32MM 142-32-64 NV EHXL ALIP LNR G4 32MM 142-32-65 NV EHXL ALIP LNR G5 32MM 142-36-52 NV EHXL LIP LNR G2 36MM 142-36-53 NV EHXL LIP LNR G3 36MM 142-36-54 NV EHXL LIP LNR G4 36MM 142-36-55 NV EHXL LIP LNR G5 36MM 142-36-62 NV EHXL ALIP LNR G2 36MM 142-36-63 NV EHXL ALIP LNR G3 36MM 142-36-64 NV EHXL ALIP LNR G4 36MM 142-36-65 NV EHXL ALIP LNR G5 36MM 142-40-53 NV EHXL LIP LNR G3 40MM 142-40-54 NV EHXL LIP LNR G4 40MM 142-40-55 NV EHXL LIP LNR G5 40MM 142-40-63 NV EHXL ALIP LNR G3 40MM 142-40-64 NV EHXL ALIP LNR G4 40MM 142-40-65 NV EHXL ALIP LNR G5 40MM 146-22-70 NV EHXL LATLNR G00 22MM 146-28-51 NV EHXL LAT LNR G1 28MM 146-32-51 NV EHXL LAT LNR G1 32MM 146-32-52 NV EHXL LAT LNR G2 32MM 146-32-53 NV EHXL LAT LNR G3 32MM 146-32-54 NV EHXL LAT LNR G4 32MM 146-32-55 NV EHXL LAT LNR G5 32MM 146-36-52 NV EHXL LAT LNR G2 36MM 146-36-53 NV EHXL LAT LNR G3 36MM 146-36-54 NV EHXL LAT LNR G4 36MM 146-36-55 NV EHXL LAT LNR G5 36MM 146-40-53 NV EHXL LAT LNR G3 40MM 146-40-54 NV EHXL LAT LNR G4 40MM 146-40-55 NV EHXL LAT LNR G5 40MM 148-32-51 NV EHXL 10¿ LNR G1 32MM 148-36-52 NV EHXL 10¿ LNR G2 36MM 148-36-53 NV EHXL 10¿ LNR G3 36MM 148-36-54 NV EHXL 10¿ LNR G4 36MM 148-36-55 NV EHXL 10¿ LNR G5 36MM 148-40-53 NV EHXL 10¿ LNR G3 40MM 148-40-54 NV EHXL 10¿ LNR G4 40MM 148-40-55 NV EHXL 10¿ LNR G5 40MM All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment.

Class I - Dangerous

Due to the lack of the Ethylene Vinyl Alcohol (EVOH) barrier layer.

Dec 31, 2024 Implants & Prosthetics Nationwide View Details →

The AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), Nylon, and EVOH, or 2) LDPE and Nylon without EVOH.

Apr 26, 2024 Implants & Prosthetics View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Mar 6, 2024 Other Medical Devices Nationwide View Details →

Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.

Nov 21, 2023 Implants & Prosthetics View Details →

GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH GXL 0 DEGREE LINER 28MM SZ E, REF 130-28-25 b. ACUMATCH GXL 0 DEGREE LINER 28MM SZ F, REF 130-28-26 c. ACUMATCH GXL 0 DEGREE LINER 28MM SZ G, REF 130-28-27 d. ACUMATCH GXL 0 DEGREE LINER 28MM SZ H, REF 130-28-28 e. ACUMATCH GXL 0 DEGREE LINER 28MM SZ J, REF 130-28-29 f. ACUMATCH GXL 0 DEGREE LINER 32MM SZ G, REF 130-32-27 g. ACUMATCH GXL 0 DEGREE LINER 32MM SZ H, REF 130-32-28 f. ACUMATCH GXL 0 DEGREE LINER 32MM SZ J, REF 130-32-29 g. ACUMATCH GXL 0 DEGREE LINER 36MM SZ H, REF 130-36-28 h. ACUMATCH GXL 0 DEGREE LINER 36MM SZ J, REF 130-36-29 i. ACUMATCH GXL 0 DEGREE LINER 36MM SZ K, REF 130-36-30 j. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ E, REF 138-28-25 k. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ F, REF 138-28-26 l. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ G, REF 138-28-27 m. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ H, REF 138-28-28 n. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ J, REF 138-28-29 o. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ H, REF 138-36-28 p. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ J, REF 138-36-29 q. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ K, REF 138-36-30 r. ACUMATCH GXL 15 DEG +5 LAT LINER SZG 38MM, REF 138-36-27 s. ACUMATCH GXL 15DEG LINER 28MM SZ D, REF 132-28-24 t. ACUMATCH GXL 15DEG LINER 28MM SZ E, REF 132-28-25 u. ACUMATCH GXL 15DEG LINER 28MM SZ F, REF 132-28-26 v. ACUMATCH GXL 15DEG LINER 28MM SZ G, REF 132-28-27 w. ACUMATCH GXL 15DEG LINER 28MM SZ H, REF 132-28-28 x. ACUMATCH GXL 15DEG LINER 28MM SZ J, REF 132-28-29 y. ACUMATCH GXL 15DEG LINER 28MM SZ K, REF 132-28-30 z. ACUMATCH GXL 15DEG LINER 32MM SZ G, REF 132-32-27 aa. ACUMATCH GXL 15DEG LINER 32MM SZ H, REF 132-32-28 bb. ACUMATCH GXL 15DEG LINER 32MM SZ J, REF 132-32-29 cc. ACUMATCH GXL 15DEG LINER 32MM SZ K, REF 132-32-30 dd.ACUMATCH GXL 15DEG LINER 36MM SZ H, REF 132-36-28 ee. ACUMATCH GXL 15DEG LINER 36MM SZ J, REF 132-36-29 ff. ACUMATCH GXL 15DEG LINER 36MM SZ K, REF 132-36-30 gg. ACUMATCH GXL EXT COV LINER 28MM SZ E, REF 134-28-25 hh. ACUMATCH GXL EXT COV LINER 28MM SZ F, REF 134-28-26 ii. ACUMATCH GXL EXT COV LINER 28MM SZ G, REF 134-28-27 jj. ACUMATCH GXL EXT COV LINER 28MM SZ H, REF 134-28-28 kk. ACUMATCH GXL EXT COV LINER 28MM SZ J, REF 134-28-29 ll. ACUMATCH GXL EXT COV LINER 28MM SZ K, REF 134-28-30 mm. MCS GXL LINER 5/15 DEG 4448X28, REF 104-28-43 nn. MCS GXL LINER 5/15 DEG 5052X28, REF 104-28-44 oo. MCS GXL LINER 5/15 DEG 5458X28, REF 104-28-45 pp. MCS GXL LINER 5/15 DEG 5458X32, REF 104-32-45 qq. MCS GXL LINER 5/15 DEG 6064X32, REF 104-32-46 rr. MCS GXL LINER 5/15 DEG 6064X36, REF 104-36-46 ss. MCS GXL LINER 5/15 DEG 6670X36, REF 104-36-47

Class I - Dangerous

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION GXL LINER, NEUTRAL, 28MM ID, GROUP 1 CUPS, REF 130-28-51 b. NOVATION GXL LINER, NEUTRAL, 32MM ID, GROUP 2 CUPS, REF 130-32-52 c. NOVATION GXL LINER, NEUTRAL, 36MM ID, GROUP 3 CUPS, REF 130-36-53 d. NOVATION GXL LINER, NEUTRAL, 36MM ID, GROUP 4 CUPS, REF 130-36-54 e. NOVATION GXL LINER, NEUTRAL, 36MM ID, GROUP 5 CUPS, REF 130-36-55 f. NOVATION GXL LINER, +5 LAT 32MM ID, GROUP 3 CUPS, REF 136-32-53 g. NOVATION GXL LINER, +5 LAT 32MM ID, GROUP 4 CUPS, REF 136-32-54 h. NOVATION GXL LINER, +5 LAT 32MM ID, GROUP 5 CUPS, REF 136-32-55 i. NOVATION GXL LINER, +5LAT, 28MM G1-48/50MM CUPS, REF 136-28-51 j. NOVATION GXL LINER, +5LAT, 32MM G1-48/50MM CUPS, REF 136-32-51 k. NOVATION GXL LINER, +5LAT, 32MM G2-52/54MM CUPS, REF 136-32-52 l. NOVATION GXL LINER, +5LAT, 36MM G2-52/54MM CUPS, REF 136-36-52 m. NOVATION GXL LINER, +5LAT, 36MM G3-56/58MM CUPS, REF 136-36-53 n. NOVATION GXL LINER, +5LAT, 36MM G4-60/62MM CUPS, REF 136-36-54 o. NOVATION GXL LINER, +5LAT, 36MM G5-64/66/68MM CUPS, REF 136-36-55 p. NOVATION GXL LINER, 10 DEG FACE, 32MM ID, GRP 1 CUP, REF 138-32-51 q. NOVATION GXL LINER, 10 DEG FACE, 32MM ID, GRP 3 CUP, REF 138-32-53 r. NOVATION GXL LINER, 10 DEG FACE, 36MM ID, GRP 2 CUP, REF 138-36-52 s. NOVATION GXL LINER, 10 DEG FACE, 36MM ID, GRP 3 CUP, REF 138-36-53 t. NOVATION GXL LINER, 10 DEG FACE, 36MM ID, GRP 4 CUP, REF 138-36-54 u. NOVATION GXL LINER, 10 DEG FACE, 36MM ID, GRP 5 CUP, REF 138-36-55 v. NOVATION GXL LINER, 10 DEG FACE, REPLACED BY 138-36-52, REF 138-32-52 w. NOVATION GXL LINER, LIPPED 32MM ID GROUP 2 CUPS, REF 132-32-52 x. NOVATION GXL LINER, LIPPED ANT, 32MM ID, GROUP 1 CUPS, REF 132-32-61 y. NOVATION GXL LINER, LIPPED ANT, 36MM ID, GROUP 2 CUPS, REF 132-36-62 z. NOVATION GXL LINER, LIPPED ANT, 36MM ID, GROUP 3 CUPS, REF 132-36-63 aa. NOVATION GXL LINER, LIPPED ANT, 36MM ID, GROUP 4 CUPS, REF 132-36-64 bb. NOVATION GXL LINER, LIPPED, 36MM ID, GROUP 3 CUPS, REF 132-36-53 cc. NOVATION GXL LINER, LIPPED, 28MM ID, GROUP 1 CUPS, REF 132-28-51 dd. NOVATION GXL LINER, LIPPED, 28MM ID, GROUP 2 CUPS, REF 132-28-52 ee. NOVATION GXL LINER, LIPPED, 32MM ID GROUP 4 CUPS, REF 132-32-54 ff. NOVATION GXL LINER, LIPPED, 32MM ID, GROUP 3 CUPS, REF 132-32-53 gg. NOVATION GXL LINER, LIPPED, 32MM ID, GROUP 5 CUPS, REF 132-32-55 hh. NOVATION GXL LINER, LIPPED, 36MM ID GROUP 4 CUPS, REF 132-36-54 ii. NOVATION GXL LINER, LIPPED, 36MM ID, GROUP 5 CUPS, REF 132-36-55 jj. NOVATION GXL LINER, NEUTRAL, 28MM ID, GROUP 2 CUPS, REF 130-28-52 kk. NOVATION GXL LINER, NEUTRAL, 32MM ID, GROUP 3 CUPS, REF 130-32-53 ll. NOVATION GXL LINER, NEUTRAL, 32MM ID, GROUP 4 CUPS, REF 130-32-54 mm. NOVATION GXL LINER, NEUTRAL, 32MM ID, GROUP 5 CUPS, REF 130-32-55

Class I - Dangerous

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 0 DEGREE LINER 28MM 46MM, REF 130-28-05; b. ACUMATCH 0 DEGREE LINER 28MM 48MM, REF 130-28-06; c. ACUMATCH 0 DEGREE LINER 28MM 50/52MM, REF 130-28-07; d. ACUMATCH 0 DEGREE LINER 28MM 54/56MM, REF 130-28-08; e. ACUMATCH 0 DEGREE LINER 28MM 58/64MM, REF 130-28-09; f. ACUMATCH 15 DEGREE LINER 22MM SZ D, REF 132-22-04; g. ACUMATCH 15 DEGREE LINER 22MM SZ E, REF 132-22-05; h. ACUMATCH 15 DEGREE LINER 22MM SZ F, REF 132-22-06; i. ACUMATCH 15 DEGREE LINER 22MM SZ G, REF 132-22-07; j. ACUMATCH 15 DEGREE LINER 22MM SZ H, REF 132-22-08; k. ACUMATCH 15 DEGREE LINER 28MM SZ E, REF 132-28-05; l. ACUMATCH 15 DEGREE LINER 28MM SZ F, REF 132-28-06; m. ACUMATCH 15 DEGREE LINER 28MM SZ G, REF 132-28-07; n. ACUMATCH 15 DEGREE LINER 28MM SZ H, REF 132-28-08; o. ACUMATCH 15 DEGREE LINER 28MM SZ J, REF 132-28-09; p. ACUMATCH 15 DEGREE LINER 28MM SZ K, REF 132-28-10; q. ACUMATCH 15 DEGREE LINER 32MM SZ G, REF 132-32-07; r. ACUMATCH 15 DEGREE LINER 32MM SZ H, REF 132-32-08; s. ACUMATCH 15 DEGREE LINER 32MM SZ J, REF 132-32-09; t. ACUMATCH 15 DEGREE LINER 32MM SZ K, REF 132-32-10; u. ACUMATCH EXT COV LINER 28MM 46MM, REF 134-28-05; v. ACUMATCH EXT COV LINER 28MM 48MM, REF 134-28-06; w. ACUMATCH EXT COV LINER 28MM 50/52MM, REF 134-28-07; x. ACUMATCH EXT COV LINER 28MM 54/56MM, REF 134-28-08; y. ACUMATCH EXT COV LINER 28MM 58/64MM, REF 134-28-09; z. ACUMATCH EXT COV LINER 28MM 66/70MM, REF 134-28-10; aa. ACUMATCH LAT LINER 28MM 46MM, REF 138-28-05; bb. ACUMATCH LAT LINER 28MM 48MM, REF 138-28-06; cc. ACUMATCH LAT LINER 28MM 50/52MM, REF 138-28-07; dd. ACUMATCH LAT LINER 28MM 54/56MM, REF 138-28-08; ee. ACUMATCH LAT LINER 28MM 58/64MM, REF 138-28-09; ff. ACUMATCH 15 DEG LINER, SZ F, 28MM ID, REF CSK-1663; gg. ACUMATCH 15 DEG LINER, SZ G, 28MM ID, REF CSK-1664; hh. ACUMATCH 15 DEG LINER, SZ H, 28MM ID, REF CSK-1665

Class I - Dangerous

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

Aug 11, 2022 Implants & Prosthetics Nationwide View Details →

Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

Aug 30, 2021 Other Medical Devices Nationwide View Details →

Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

Aug 30, 2021 Implants & Prosthetics Nationwide View Details →

Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

Aug 30, 2021 Implants & Prosthetics Nationwide View Details →

OPTETRAK Comprehensive Knee System, labeled as the following: a. OPTETRAK All-polyethylene CR Tibial Components, with Product Line Numbers: 200-11-XX, 200-12-XX, 200-13-XX, 200-14-XX, 200-15-XX, 200-16-XX; b. OPTETRAK All-polyethylene PS Tibial Components, with Product Line Numbers: 204-11-XX, 204-12-XX, 204-13-XX, 204-14-XX, 204-15-XX, 204-16-XX; c. OPTETRAK HI-FLEX PS Polyethylene Tibial Inserts, with Product Line Numbers: 244-20-XX, 244-21-XX, 244-22-XX, 244-23-XX, 244-24-XX, 244-25-XX; d. OPTETRAK CR Tibial Inserts, with Product Line Numbers: 200-21-XX, 200-22-XX, 200-23-XX, 200-24-XX, 200-25-XX, 200-26-XX, 200-50-XX, 200-51-XX; e. OPTETRAK CR Tibial Slope + Insert, with Product Line Numbers: 200-56-XX, 200-57-XX, 200-61-XX, 200-62-XX, 200-63-XX, 200-64-XX, 200-65-XX; f. OPTETRAK CR Tibial Slope ++ Insert, with Product Line Numbers: 200-57-XX, 200-71-XX, 200-72-XX, 200-73-XX, 200-74-XX, 200-75-XX; g. OPTETRAK PS Tibial Inserts, with Product Line Numbers: 204-21-XX, 204-22-XX 204-23-XX, 204-24-XX, 204-25-XX, 204-26-XX, 204-50-XX, 204-51-XX; h. OPTETRAK "MOMB" Non-Mod Molded Insert, with Product Line Numbers: 204-91-XX, 204-92-XX, 204-93-XX, 204-94-XX, 204-95-XX; i. OPTETRAK CC Tibial Insert, with Product Line Numbers: 208-21-XX, 208-22-XX, 208-23-XX, 208-24-XX, 208-25-XX, 208-51-XX; j. OPTETRAK B-Series PS Tibial Insert, with Product Line Numbers: 224-21-XX, 224-22-XX, 224-23-XX, 224-24-XX. Affected 510(k) numbers include: K932690, K933494, K932776, K011976, K082022, K933610, K030686, K954208, K010434, K033883. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.

Class I - Dangerous

Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

Aug 30, 2021 Implants & Prosthetics Nationwide View Details →

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Jun 29, 2021 Implants & Prosthetics Nationwide View Details →

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Jun 29, 2021 Implants & Prosthetics Nationwide View Details →

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Jun 29, 2021 Implants & Prosthetics Nationwide View Details →

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Jun 29, 2021 Implants & Prosthetics Nationwide View Details →

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Jun 29, 2021 Implants & Prosthetics Nationwide View Details →

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Jun 29, 2021 Implants & Prosthetics Nationwide View Details →

The HAT trial contains a retaining ring that has shown the potential to disengage during use.

May 11, 2021 Implants & Prosthetics Nationwide View Details →