Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral, 36mm. Orthopedic hip implant component.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All serial numbers. UDI-DI numbers: 10885862022202, 10885862022219, 10885862022226, 10885862207081, 10885862022233, 10885862022240, 10885862022257
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Exactech, Inc.
- Reason for Recall:
- Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral, 36mm. Orthopedic hip implant component.
Product Codes/Lot Numbers:
All serial numbers. UDI-DI numbers: 10885862022202, 10885862022219, 10885862022226, 10885862207081, 10885862022233, 10885862022240, 10885862022257
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2118-2021
Related Recalls
Brand Name: Equinoxe Ergo Product Name: Ergo Modular Impactor Handle Model/Catalog Number: 321-09-05
Exactech
Impactor handle may be missing cross-pin
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.