Exactech Ergo Impactor Handle, Catalog #321-09-05.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog #321-09-05 - Lot numbers: 147544002, 275151004, 275151005, 275151006, 314934001, 400107002, and 400798002; UDI-DI 10885862557674. Updated data - 01/31/2025: Additional Lot Numbers 401819005 401819007 401819008 404552002 405492002 405253003 406041001
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Exactech, Inc.
- Reason for Recall:
- The ball and spring housed within the locking mechanism of the handle can dislodge during disassembly or during impaction.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Exactech Ergo Impactor Handle, Catalog #321-09-05.
Product Codes/Lot Numbers:
Catalog #321-09-05 - Lot numbers: 147544002, 275151004, 275151005, 275151006, 314934001, 400107002, and 400798002; UDI-DI 10885862557674. Updated data - 01/31/2025: Additional Lot Numbers 401819005 401819007 401819008 404552002 405492002 405253003 406041001
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2051-2023
Related Recalls
Brand Name: Equinoxe Ergo Product Name: Ergo Modular Impactor Handle Model/Catalog Number: 321-09-05
Exactech
Impactor handle may be missing cross-pin
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.