OPTETRAK Comprehensive Knee System, labeled as the following: a. OPTETRAK All-polyethylene CR Tibial Components, with Product Line Numbers: 200-11-XX, 200-12-XX, 200-13-XX, 200-14-XX, 200-15-XX, 200-16-XX; b. OPTETRAK All-polyethylene PS Tibial Components, with Product Line Numbers: 204-11-XX, 204-12-XX, 204-13-XX, 204-14-XX, 204-15-XX, 204-16-XX; c. OPTETRAK HI-FLEX PS Polyethylene Tibial Inserts, with Product Line Numbers: 244-20-XX, 244-21-XX, 244-22-XX, 244-23-XX, 244-24-XX, 244-25-XX; d. OPTETRAK CR Tibial Inserts, with Product Line Numbers: 200-21-XX, 200-22-XX, 200-23-XX, 200-24-XX, 200-25-XX, 200-26-XX, 200-50-XX, 200-51-XX; e. OPTETRAK CR Tibial Slope + Insert, with Product Line Numbers: 200-56-XX, 200-57-XX, 200-61-XX, 200-62-XX, 200-63-XX, 200-64-XX, 200-65-XX; f. OPTETRAK CR Tibial Slope ++ Insert, with Product Line Numbers: 200-57-XX, 200-71-XX, 200-72-XX, 200-73-XX, 200-74-XX, 200-75-XX; g. OPTETRAK PS Tibial Inserts, with Product Line Numbers: 204-21-XX, 204-22-XX 204-23-XX, 204-24-XX, 204-25-XX, 204-26-XX, 204-50-XX, 204-51-XX; h. OPTETRAK "MOMB" Non-Mod Molded Insert, with Product Line Numbers: 204-91-XX, 204-92-XX, 204-93-XX, 204-94-XX, 204-95-XX; i. OPTETRAK CC Tibial Insert, with Product Line Numbers: 208-21-XX, 208-22-XX, 208-23-XX, 208-24-XX, 208-25-XX, 208-51-XX; j. OPTETRAK B-Series PS Tibial Insert, with Product Line Numbers: 224-21-XX, 224-22-XX, 224-23-XX, 224-24-XX. Affected 510(k) numbers include: K932690, K933494, K932776, K011976, K082022, K933610, K030686, K954208, K010434, K033883. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.
Class I - DangerousWhat Should You Do?
- Check if you have this product: a. Item Number (Device Identifier; Serial number(s)): 200-11-11 (10885862039934; 1010599); 200-12-09 (10885862039965; 0850770, 0850771, 0850773, 0850776, 0850778, 1052587, 1052589, 1070149, 1070150, 1070151, 1070152, 1267585, 1267586, 1267588, 1318739, 1318740, 1318742, 1375833, 1375836, 1375837, 1375839, 1375844); 200-12-11 (10885862039972; 0825651, 0825653, 0825654, 0825657, 0825659, 1052802, 1052803, 1052804, 1309224); 200-12-13 (10885862039989; 0763041, 0763043, 0820424, 0820431, 1053775); 200-13-09 (10885862039996; 0765380, 0765381, 0765382, 0765383, 0789723, 0789724, 0789726, 0789727, 0789728, 0789729, 0789730, 0789731, 0789732, 0897683, 0998654, 0998655, 0998656, 0998657, 0998658, 0998659, 0998660, 0998661, 0998662, 0998663, 0998664, 0998665, 0998666, 1073908, 1073909, 1073910, 1073911, 1084202, 1084203, 1084204, 1084205, 1261436, 1261437, 1261438, 1261439, 1261440, 1316122, 1316123, 1316124, 1316125, 1316126, 1319147, 1319148, 1319149, 1319150, 1319151, 1407772, 1407774, 1407775, 1407776, 1407777, 1407778, 1407779, 1407780, 1407781, 1516714, 1516715, 1516716, 1516717, 1516718, 1976718, 1976719, 1976720, 1976721, 1976722, 2061574, 2097473, 2097474, 2097475, 2097476, 2097477, 2097478, 2097479, 2145890, 2145891, 2145892, 2145894, 2145895, 2145896, 2559776); 200-13-11 (10885862040008; 0769998, 0770002, 0770003, 0770004, 0818558, 0818559, 0818560, 0818562, 0818563, 0818564, 0818565, 1027748, 1027750, 1027752, 1027754, 1084902, 1291129, 1291130, 1291131, 1319746, 1319747, 1319748, 1319752, 1319753, 1319754, 1319756); 200-13-13 (10885862040015; 0767606, 0767607, 0767608, 0767609, 0953412, 1063120, 1307627); 200-14-09 (10885862040039; 0607230, 0607231, 0607232, 0607233, 0607234, 0607235, 0607236, 0607237, 0607238, 1033676, 1033677, 1033678, 1033679, 1040785, 1040786, 1040787, 1040788, 1065019, 1065020, 1065021, 1065022, 1065023, 1065024, 1065025, 1153592, 1153593, 1153594, 1153595, 1153596, 1153597, 1153598, 1153599, 1153600, 1215318, 1215319, 1215320, 1215322, 1281462, 1281463, 1281464, 1281465, 1321078, 1321079, 1321080, 1321081, 1321082, 1389683, 1389684, 1389685, 1400987, 1400988, 1400989, 1400990, 2012144, 2012145, 2048510, 2048511, 2207641, 2207642, 2207643, 2345327, 2345328, 2414649, 2414652, 2454327, 2454328, 2454329, 2565530, 2565533, 2565534); 200-14-11 (10885862040046; 0942385, 0942386, 0942393, 1050502, 1310607); 200-14-13 (10885862040053; 0942535, 1048228, 1048232, 1153792); 200-15-09 (10885862040077; 0822759, 0822761, 0822762, 0822764, 0822765, 0822766, 0822767, 1048609, 1048611, 1048612, 1048614, 1048615, 1048616, 1048617, 1136150, 1136151, 1136152, 1136154, 1214791, 1214792, 1214793, 1214794, 1214795, 1214796, 1283152, 1283153, 1283154, 1283155, 1439852, 1439853, 1439855, 1439856, 2146108, 2146113); 200-15-11 (10885862040084; 1065636, 1065638); 200-15-13 (10885862040091; 0954497, 0954501); 200-16-11 (10885862040121; 1040633, 1040634, 1040635, 1040636, 1073405, 1073407, 1084258); 200-16-13 (10885862040138; 0956887); b. Item Number (Device Identifier; Serial number(s)): 204-11-09 (10885862047823; 0961663, 0961665, 0961666, 0961667, 0961668, 0961669, 0961671, 0961672, 1045876, 1045877, 1045879, 1070620, 1070622, 1070623, 1106799, 1106800, 1106801, 1106802, 1106803, 1106805, 1106807, 1193435, 1193437, 1193438, 1193439, 1232147, 1232151, 1255910, 1255911, 1255912, 1255913, 1255914, 1270806, 1270807, 1270808, 1270809, 1308220, 1308221, 1308222, 1308223, 1308224, 1308225, 1377016, 1377017, 1377018, 1377019, 1377020, 1377021, 1377022, 1377023, 1377024, 1377025, 1545261, 1545262, 1545263, 1545264, 1545265, 1565700, 1565701, 1565702, 1565703, 1565704, 1582779, 1582780, 1582781, 1582782, 1582783, 1582784, 1582785, 1582786, 1582787, 1582788, 1582789, 1582790, 1582791, 1582792, 1582793, 1582794, 1582795, 1582796, 1582797, 1582798, 1659777, 1659778, 1659779, 1659780, 1659781, 1659782, 1659783, 1659784, 1659785, 1659786, 1659787, 1659788, 1659789, 1659790, 1659791, 1659792, 1659793, 1659794, 1659795, 1659796, 1659797, 1659798, 1659799, 1659800, 1659801, 1704209, 1704210, 1704211, 1704212, 1786558, 1786559, 1786560, 1786561, 1786562, 1786563, 1786564, 1786565, 1797016, 1797017, 1797018, 1797019, 1797020, 1797021, 1797022, 1797023, 1797024, 1797025, 1797026, 1797027, 1797028, 1817561, 1817562, 1817563, 1817564, 1817565, 1817566, 1817567, 1817568, 1817569, 1817570, 1817571, 1817572, 1817573, 1842319, 1842320, 1842321, 1842324, 1842325, 1901249, 1901250, 1901251, 1901252, 1901253, 1901255, 1901256, 1901259, 1901260, 1901261, 1901262, 1901263, 1901264, 1901265, 1901266, 1901267, 1901268, 1901269, 1901270, 1901271, 1901272, 1901273, 1901274, 1901275, 1901276, 2266634, 2266635, 2266636, 2266638, 2276586, 2276587, 2276588, 2276589, 2397396, 2397397, 2397398, 2397399, 2397400, 2397401, 2397402, 2397403, 2397404, 2397405, 2425639, 2425640, 2425641, 2425642, 2425643, 2425644, 2425645, 2425646, 2425647, 2425648, 2425650, 2425651, 2425652, 2425653, 2425654, 2425655, 2425656, 2430257, 2430258, 2430259, 2430260, 2430261, 2430262, 2430263, 2430264, 2430265, 2615191, 2615192, 2615193, 2615194, 2615195, 2615196, 2615197, 2615198, 2615199, 2615200, 2653102, 2653103, 2653104, 2653105, 2653106, 2653107, 2653108, 2653109, 2653110, 2653111, 2653112, 2653113, 2653114, 2653115, 2653116, 2653117, 2653118, 2653119, 2653120, 2653121, 2653123, 2653124, 2653125, 2653126, 2653127, 2653128, 2653129, 2653130, 2653131, 3545584, 3545585, 3545586, 3545587, 3545588, 3545589, 3545590, 3545591, 3545592, 3545593, 3636693, 3636694, 3636695, 3636696, 3636697, 3636698, 3636699, 3636700, 3636701, 3636702, 3654361, 3654362, 3654363, 3654364, 3654365, 3654366, 3654367, 3654368, 3654369, 3654370, 3660720, 3660721, 3660722, 3660723, 3660724, 3781241, 3781242, 3781243, 3781244, 3781245, 3792869, 3792870, 3792871, 3792872, 3792873, 3809326, 3809327, 3809328, 3809329, 3809330, 4080806, 4080807, 4080808, 4080809, 4080810, 4080811, 4080812, 4080813, 4080814, 4080815, 4149218, 4149219, 4149220, 4149221, 4149222, 4175249, 4175250, 4175251, 4175252, 4175253, 4175254, 4175255, 4175256, 4205182, 4205183, 4205184, 4205185, 4205186, 4331818, 4331819, 4331820, 4331821, 4331822, 4331823, 4331824, 4331825, 4331826, 4367437, 4367438, 4367439, 4367440, 4367441, 4367442, 4367443, 4367444, 4399568, 4399569, 4399570, 4399571, 4399572, 4399573, 4399574, 4399575, 4399576, 4399577, 4933848, 4933849, 4933851, 4933852, 4933853, 4933854, 4933855, 4933856, 4933857, 4933858, 4933859, 4933860, 4933861, 4933862, 4933863, 4933864, 4933865, 4933866, 4933867, 4933868, 4933869, 4933870, 4933871, 4933872, 4933873, 4933874, 4933875, 4933876, 4933877, 4933878, 4933879, 4933880, 4933881, 4933882, 4933883, 4933884, 4933885, 4933887, 4945297, 4945298, 4945299, 4945300, 4945301, 4945302, 4945303, 4945304, 4945305, 4945306, 4945307, 4945308, 4945309, 4945310, 4945311, 4945312, 4945313, 4945314, 4945315, 4945316, 4945317, 4945318, 4945319, 4945320, 4945321, 4945322, 4945323, 4945324, 4945325, 4945326, 4945327, 4945328, 4945329, 4945330, 4945331, 4945332, 4945333, 4945334, 5086208, 5086209, 5086210, 5086211, 5086212, 5086213, 5086214, 5086215, 5086216, 5086217, 5086218, 5086219, 5086220, 5172053, 5172054, 5172055, 5172056, 5172057, 5172058, 5172059, 5180020, 5180021, 5180022, 5180023, 5180024, 5180025, 5180026, 5188774, 5188775, 5188776, 5209513); 204-11-11 (10885862047830; 0971418, 0971419, 0971420, 0971421, 0971422, 0971423, 1015616, 1015617, 1015618, 1015619, 1047575, 1047577, 1047578, 1066792, 1066793, 1066794, 1087082, 1087083, 1087084, 1087085, 1087086, 1168686, 1168687, 1187090, 1187092, 1187093, 1187094, 1187096, 1187098, 1187099, 1255900, 1255901, 1255902, 1255903, 1255904, 1255905, 1255906, 1255907, 1255908, 1255909, 1270537, 1270540, 1294283, 1294284, 1294285, 1294287, 1294288, 1294289, 1294290, 1307222, 1307223, 1307224, 1307225, 1392229, 1392230, 1392231, 1392232, 1392233, 1576699, 1576700, 1576701, 1576702, 1576703, 1576704, 1576705, 1585601, 1585602, 1585603, 1585604, 1585605, 1585606, 1585607, 1585608, 1585609, 1585610, 1585611, 1585612, 1585613, 1585614, 1585615, 1585616, 1585617, 1585618, 1585619, 1585620, 1585621, 1585622, 1585623, 1585624, 1585625, 1585626, 1585627, 1585628, 1585629, 1585630, 1585631, 1585632, 1745600, 1745601, 1745602, 1745603, 1745604, 1774126, 1774127, 1774128, 1774129, 1774130, 1774131, 1774132, 1774133, 1774134, 1774135, 1774136, 1774137, 1786761, 1786762, 1786763, 1786764, 1786765, 1786766, 1786767, 1786768, 1810253, 1810254, 1810255, 1810256, 1810257, 1810258, 1810259, 1810260, 1810261, 1821731, 1821732, 1821733, 1821734, 1821735, 1821736, 1821737, 1821738, 1821739, 1821740, 1821741, 1865512, 1865513, 1865515, 1897369, 1897370, 1897371, 1897373, 1897374, 1897375, 1897376, 1897377, 2265900, 2265901, 2265902, 2265903, 2265904, 2306144, 2306145, 2306146, 2306147, 2306148, 2306149, 2306150, 2306151, 2306153, 2306154, 2306155, 2306156, 2306157, 2306158, 2306159, 2306160, 2306161, 2306163, 2306164, 2306165, 2306166, 2306167, 2306168, 2306169, 2306170, 2306171, 2306172, 2430406, 2430407, 2430408, 2430409, 2430410, 2430411, 2430412, 2430413, 2430414, 2430415, 2582070, 2582071, 2582072, 2582073, 2582074, 2582075, 2582076, 2628063, 2628064, 2628065, 2628066, 2628067, 2628068, 2628069, 2628070, 2628071, 2628072, 2628073, 2628074, 2628075, 3545581, 3545582, 3545583, 3616391, 3616392, 3616393, 3616394, 3616395, 3636926, 3636927, 3644006, 3654371, 3654372, 3654373, 3654374, 3654375, 3654376, 3729486, 3756690, 3775777, 3775778, 3776705, 3776706, 3776707, 3776708, 3776709, 3811782, 3811783, 3811784, 3811785, 3811786, 3811787, 3811788, 3811789, 3811790, 3811791, 4053394, 4053395, 4053396, 4053397, 4053398, 4053399, 4053400, 4053401, 4053402, 4053403, 4053404, 4053405, 4053406, 4053407, 4053408, 4205187, 4205188, 4205189, 4205190, 4205191, 4226766, 4226767, 4226768, 4226769, 4226770, 4226772, 4226773, 4226774, 4226775, 4349216, 4349217, 4349218, 4349219, 4349220, 4349221, 4349222, 4349223, 4349224, 4384105, 4384106, 4384107, 4384108, 4384109, 4384110, 4384111, 4384112, 4384113, 4384114, 4816187, 4816188, 4816189, 4846027, 4846028, 4846029, 4846030, 4846031, 4846032, 4846033, 4846034, 4846035, 4846036, 4846037, 4846038, 4846039, 4846040, 4846041, 4846042, 4846043, 4846044, 4846045, 4846046, 4846047, 4846048, 4846049, 4846050, 4846051, 4846052, 4934866, 4934867, 4934868, 4934869, 4934870, 4934871, 4934872, 4934873, 4934874, 4934875, 4934876, 4934877, 4934878, 4934879, 4934880, 4934881, 4934882, 4934883, 4934884, 4934885, 4934886, 4934887, 4934888, 4934889, 4934890, 4965321, 4965322, 4965323, 4965324, 4965325, 4965326, 4965327, 4965328, 4965329, 4965330, 4965331, 4965332, 4965333, 4965334, 4965335, 4965336, 4965337, 4965338, 4965339, 4965340, 4965341, 4965342, 4965343, 4965344, 4965345, 5009433, 5009434, 5009435, 5009436, 5009437, 5009438, 5009439, 5009440, 5009441, 5009442, 5009443, 5009444, 5009446, 5009447, 5009448, 5009449, 5009450, 5009451, 5009452, 5009453, 5009454, 5009455, 5009457, 5179063); 204-11-13 (10885862047847; 0961673, 0961674, 0961675, 0961677, 0961678, 0961680, 1021516, 1021517, 1021518, 1021519, 1031608, 1031609, 1031610, 1053779, 1053780, 1053781, 1098258, 1098259, 1201250, 1201251, 1201252, 1201254, 1201255, 1259004, 1259005, 1259006, 1259007, 1259008, 1307220, 1307221, 1328794, 1328795, 1328796, 1328797, 1328798, 1328799, 1328800, 1328801, 1328802, 1328803, 1328804, 1328805, 1328806, 1597232, 1597233, 1597234, 1597235, 1597236, 1597237, 1597239, 1597240, 1597241, 1627673, 1627674, 1627675, 1627676, 1627677, 1627678, 1627679, 1647931, 1647932, 1647933, 1647934, 1647935, 1659319, 1659320, 1659321, 1659323, 1766122, 1766123, 1766124, 1766125, 1766126, 1766127, 1766128, 1766129, 1766130, 1766131, 1766132, 1766133, 1766134, 1767874, 1767875, 1767876, 1767877, 1767878, 1767879, 1767880, 1767881, 1767882, 1767883, 1767884, 1767885, 1767886, 1852794, 1852795, 1852796, 1852797, 1852798, 1852799, 1852800, 1852801, 1852802, 1852803, 1903394, 1903395, 1903396, 1903398, 1903399, 1903401, 1903402, 1903403, 1903404, 1903405, 1903406, 1903410, 1903411, 1903412, 2259858, 2259859, 2259860, 2259861, 2259862, 2259863, 2259864, 2259865, 2259866, 2259867, 2398501, 2398502, 2398503, 2398504, 2398505, 2398506, 2398507, 2398508, 2398509, 2430396, 2430397, 2430398, 2430399, 2430400, 2430401, 2430402, 2430403, 2430404, 2430405, 2640147, 2640148, 2640149, 2640150, 2640151, 2640152, 2640153, 2640154, 2640155, 2640156, 2725662, 2725664, 2725665, 2725666, 2725667, 2725668, 2725669, 2725671, 2725672, 2725674, 2725676, 2725677, 2725678, 2725679, 2725680, 2725681, 2815981, 2926604, 2926605, 2926606, 2926607, 2926608, 2926609, 2926610, 2926611, 2926612, 3647961, 3647963, 3647964, 3647966, 4081311, 4081312, 4081313, 4081314, 4081315, 4124295, 4124298, 4124299, 4124300, 4124301, 4124302, 4124303, 4124304, 4124305, 4124307, 4124308, 4124309, 4124310, 4124311, 4124312, 4124314, 4124315, 4124316, 4124317, 4124318, 4384097, 4384098, 4384099, 4384100, 4384101, 4384102, 4384103, 4384104, 5084582, 5084583, 5084584, 5084585, 5084586); 204-12-09 (10885862047885; 0881042, 0881043, 0895287, 0895288, 0895289, 0895290, 0895291, 0895292, 0895293, 0895294, 0895295, 0957833, 0957834, 0957835, 0957836, 0957837, 1019604, 1019605, 1021520, 1021521, 1021522, 1021523, 1021524, 1045910, 1045911, 1045912, 1045913, 1045914, 1079090, 1079091, 1079092, 1079093, 1097012, 1097013, 1097014, 1097015, 1122249, 1122250, 1122251, 1122252, 1122253, 1150183, 1150185, 1173036, 1173037, 1173038, 1173039, 1173040, 1208280, 1215635, 1215636, 1215639, 1230263, 1230264, 1230265, 1230266, 1230267, 1230268, 1230269, 1230270, 1237521, 1237522, 1237523, 1237524, 1237525, 1265772, 1265773, 1265774, 1294243, 1294244, 1294245, 1300708, 1300709, 1300710, 1300711, 1300712, 1300713, 1300714, 1300715, 1324324, 1324325, 1324326, 1324327, 1324328, 1324330, 1324332, 1324333, 1326334, 1326335, 1326336, 1326337, 1326338, 1326339, 1347634, 1347635, 1347636, 1347637, 1357215, 1357216, 1357217, 1357219, 1357220, 1357221, 1357222, 1357223, 1357224, 1357225, 1357226, 1357227, 1369862, 1369863, 1369864, 1369865, 1369866, 1371266, 1371267, 1371268, 1371269, 1371270, 1371271, 1371272, 1371273, 1371274, 1371275, 1441530, 1441531, 1441532, 1441533, 1441534, 1458179, 1458180, 1458181, 1458182, 1458183, 1458184, 1458185, 1458186, 1458187, 1458188, 1552930, 1552931, 1552932, 1552933, 1552934, 1552935, 1552936, 1552937, 1552938, 1552939, 1577909, 1577910, 1577911, 1577912, 1577913, 1577914, 1577915, 1577916, 1577917, 1577918, 1584330, 1584331, 1584332, 1584333, 1584334, 1584335, 1584336, 1584337, 1584338, 1584339, 1584340, 1584341, 1584342, 1584343, 1584344, 1584345, 1584346, 1584347, 1584348, 1584349, 1584350, 1584351, 1584352, 1584353, 1584354, 1584355, 1584356, 1584357, 1584358, 1707644, 1707645, 1707646, 1707647, 1707648, 1707649, 1707650, 1707651, 1742692, 1742693, 1742694, 1742695, 1742696, 1742697, 1742698, 1742699, 1766190, 1766191, 1766192, 1766193, 1766194, 1766195, 1766196, 1766197, 1766198, 1778860, 1778861, 1778862, 1778863, 1778864, 1778865, 1778866, 1778867, 1778868, 1778869, 1778870, 1787517, 1787518, 1787519, 1787520, 1787521, 1787522, 1787523, 1787524, 1787525, 1787526, 1787527, 1787528, 1787529, 1787530, 1787531, 1787532, 1787533, 1787534, 1787535, 1787536, 1787537, 1787538, 1787539, 1787540, 1796205, 1796206, 1796207, 1796208, 1796209, 1796210, 1796211, 1796212, 1817740, 1817741, 1817742, 1817743, 1817744, 1817745, 1841280, 1841281, 1841282, 1841283, 1841284, 1841285, 1841286, 1841287, 1841288, 1841289, 2067761, 2067762, 2067763, 2067764, 2067765, 2067766, 2067767, 2067768, 2067769, 2067770, 2142051, 2142052, 2142053, 2142054, 2142055, 2142056, 2142057, 2142058, 2142059, 2142060, 2142061, 2255718, 2255719, 2255720, 2255721, 2255722, 2255723, 2255724, 2255725, 2255726, 2306719, 2306720, 2306721, 2306722, 2306723, 2306724, 2306725, 2306726, 2306727, 2306728, 2306729, 2306730, 2306731, 2306732, 2306733, 2306734, 2306735, 2306736, 2306737, 2306738, 2370257, 2370258, 2370259, 2370260, 2370261, 2370262, 2370263, 2370264, 2370265, 2370266, 2370267, 2370268, 2370269, 2370270, 2370271, 2370272, 2370274, 2370275, 2415451, 2415452, 2415453, 2415454, 2415455, 2415456, 2415458, 2415459, 2415460, 2415461, 2415462, 2415463, 2415464, 2415465, 2415466, 2415467, 2415468, 2415469, 2415470, 2416268, 2416269, 2416270, 2416271, 2416272, 2416273, 2416274, 2416275, 2416276, 2416277, 2416278, 2416279, 2416280, 2416281, 2416282, 2416283, 2416284, 2416285, 2416286, 2416287, 2496195, 2496197, 2496198, 2496199, 2496200, 2496201, 2496202, 2496203, 2496204, 2591481, 2591482, 2591483, 2591484, 2591485, 2591486, 2591487, 2591488, 2591489, 2591490, 2609808, 2609809, 2609810, 2609811, 2609812, 2609813, 2609814, 2609815, 2609816, 2609817, 2609818, 2609819, 2609820, 2609821, 2609822, 2609823, 2609824, 2609825, 2609826, 2628076, 2628077, 2628078, 2628079, 2628080, 2628081, 2628082, 2628083, 2628084, 2628085, 2628086, 2628087, 2628088, 2628089, 2702396, 2702397, 2702398, 2702399, 2702400, 2702401, 2702402, 2702403, 2702404, 2702405, 2702406, 2702407, 2702408, 2702409, 2702410, 2702411, 2702412, 2702413, 2702414, 2702415, 2702416, 2702417, 2702418, 2702419, 2726285, 2726286, 2726287, 2726288, 2726289, 2726290, 2726291, 2726292, 2726293, 2726294, 2726295, 2726296, 2726297, 2726298, 2726299, 2726300, 2849419, 2849420, 2849421, 2849422, 2849423, 2849424, 2849425, 2849426, 2849427, 2849428, 2849429, 2849430, 2849431, 2849432, 2849433, 2849434, 2849435, 2849436, 2849437, 2849438, 2976096, 2976097, 2976098, 2976099, 2976100, 2976101, 2976102, 2976103, 2976105, 3546135, 3546136, 3546137, 3546138, 3546139, 3546140, 3546141, 3546142, 3546143, 3546144, 3615053, 3615054, 3615055, 3615056, 3615057, 3615058, 3615059, 3615060, 3615061, 3615062, 3654398, 3654399, 3654400, 3654401, 3654402, 3654403, 3654404, 3654405, 3654406, 3654407, 3654438, 3654439, 3654440, 3654441, 3654442, 3654443, 3654444, 3654445, 3654446, 3654447, 3786571, 3786572, 3786573, 3786574, 3786575, 3786576, 3786577, 3786578, 3786579, 3786580, 3786581, 3786582, 3786583, 3786584, 3786585, 3786586, 3786587, 4053409, 4053410, 4082563, 4082564, 4082565, 4082566, 4082567, 4082568, 4082569, 4082570, 4143349, 4143350, 4143351, 4143352, 4143353, 4145166, 4145167, 4145168, 4145169, 4145170, 4145171, 4145172, 4145173, 4145174, 4145175, 4145176, 4145177, 4145178, 4145179, 4145180, 4145181, 4145182, 4145183, 4145184, 4145185, 4208780, 4208781, 4208782, 4208783, 4208784, 4213454, 4213455, 4213456, 4213457, 4213458, 4213459, 4213460, 4213461, 4213462, 4213463, 4213464, 4213465, 4213466, 4213467, 4213468, 4215594, 4215595, 4215596, 4215597, 4215598, 4215599, 4215600, 4215601, 4215602, 4280394, 4280395, 4280396, 4280397, 4280398, 4280399, 4280400, 4280401, 4293675, 4293676, 4293677, 4293678, 4293679, 4293680, 4293682, 4293683, 4293684, 4317141, 4317142, 4317143, 4317144, 4317145, 4317146, 4317147, 4317148, 4317149, 4317150, 4367445, 4367446, 4367447, 4367448, 4367449, 4367450, 4367451, 4367452, 4367453, 4367454, 4372329, 4372330, 4372331, 4372332, 4372333, 4372334, 4372335, 4372336, 4372337, 4384115, 4384116, 4384117, 4384118, 4384119, 4384120, 4384121, 4384122, 4384123, 4384124, 4399645, 4399646, 4399647, 4399648, 4399649, 4399650, 4399651, 4399652, 4399653, 4399654, 4793183, 4793184, 4793185, 4793186, 4793187, 4793188, 4793189, 4793190, 4793191, 4793192, 4793193, 4793194, 4793195, 4793196, 4793197, 4793198, 4793199, 4793200, 4793201, 4793202, 4793203, 4793204, 4793205, 4793206, 4793207, 4793208, 4856022, 4856023, 4856024, 4863234, 4863235, 4863236, 4863237, 4863238, 4863239, 4863240, 4863241, 4863242, 4863243, 4863244, 4863245, 4863246, 4863247, 4924904, 4924905, 4924906, 4924907, 4924908, 4924909, 4924910, 4924911, 4924912, 4924913, 4924914, 4924915, 4924916, 4924917, 4924918, 4924919, 4924920, 4924921, 4924922, 4924923, 4955637, 4955638, 4955639, 4955640, 4955641, 4955642, 4955643, 4955644, 4955645, 4955646, 4955647, 4955648, 4955649, 4955650, 4955651, 4955652, 4955653, 4955654, 4955655, 4955656, 4964001, 4964002, 4964003, 4964004, 4964005, 4964006, 4964007, 4964008, 4964009, 4964010, 4964011, 4964012, 4964013, 4964014, 4964015, 4964016, 4964017, 4964018, 4964019, 4964020, 4964021, 4964022, 4964023, 4964024, 4964025, 4964026, 4964027, 4964028, 4964029, 4964030, 4964031, 4964032, 4964033, 4964034, 4964035, 4964036, 4964037, 4964038, 4964039, 4964040, 4964041, 4964042, 4964043, 4964044, 4964045, 4964046, 4964047, 4964048, 4964049, 4964050, 4965358, 4965359, 4965360, 4965361, 4965362, 4965363, 4965364, 4965365, 4965366, 4965367, 4965368, 4965369, 4965370, 4965371, 4965372, 4965373, 4965374, 4965375, 4965376, 4965377, 4965378, 4965380, 4965381, 4965382, 4965383, 4965384, 4965385, 4965386, 4965387, 4965388, 4965390, 4965391, 4965392, 4965393, 4965394, 5039055, 5039056, 5039057, 5039058, 5039059, 5039060, 5039061, 5039062, 5039063, 5039064, 5039065, 5039066, 5039067, 5039068, 5039069, 5039070, 5039071, 5039072, 5039073, 5039074, 5039075, 5039076, 5039077, 5039078, 5039079, 5039080, 5039081, 5039082, 5039083, 5039084, 5049879, 5049882, 5049885, 5049887, 5049888, 5049889, 5049892, 5049895, 5049896); 204-12-11 (10885862047892; 0856716, 0856718, 0856719, 0856720, 0856721, 0856722, 0856723, 0856724, 0856725, 0856726, 0894046, 0894047, 0894048, 0894049, 0894050, 0894051, 0894052, 0894053, 0894054, 0894056, 0894057, 0973488, 0973489, 0973490, 0973492, 0973493, 0973494, 0973495, 0973496, 0973497, 0973498, 0973499, 0973500, 0973501, 0973502, 0973504, 0973505, 0973506, 1047385, 1047386, 1047387, 1047388, 1047389, 1086399, 1086401, 1086402, 1086403, 1086404, 1086405, 1086406, 1086408, 1086410, 1086411, 1086412, 1086417, 1216096, 1216098, 1216100, 1230623, 1230624, 1230625, 1230627, 1230628, 1230629, 1230630, 1230631, 1230635, 1230636, 1230637, 1230638, 1230641, 1274067, 1274068, 1274069, 1274070, 1274071, 1274072, 1274073, 1274074, 1274075, 1274076, 1274077, 1274078, 1274079, 1274080, 1308393, 1308394, 1308395, 1308396, 1308397, 1308398, 1308399, 1308400, 1308401, 1308402, 1365381, 1365382, 1365383, 1365384, 1365385, 1365386, 1365387, 1365388, 1370851, 1370852, 1370853, 1370854, 1370855, 1406148, 1406149, 1406150, 1406151, 1406152, 1480206, 1480207, 1480208, 1480209, 1480210, 1518785, 1518786, 1518787, 1518788, 1518789, 1564109, 1564110, 1564111, 1564112, 1564113, 1564114, 1564115, 1564116, 1564117, 1564118, 1586396, 1586397, 1586398, 1586399, 1586400, 1586401, 1586402, 1586403, 1586404, 1586405, 1586407, 1586408, 1586409, 1586410, 1586411, 1586412, 1586413, 1586414, 1586415, 1586416, 1586417, 1586418, 1586419, 1586420, 1586421, 1678542, 1678543, 1678544, 1678545, 1678546, 1678547, 1678548, 1678549, 1678550, 1678551, 1678552, 1678553, 1678554, 1678555, 1742060, 1742061, 1742062, 1742063, 1742064, 1742065, 1742066, 1773352, 1773353, 1773354, 1773355, 1773356, 1773357, 1773358, 1773359, 1773360, 1773361, 1773362, 1773363, 1773364, 1773365, 1773366, 1773367, 1773368, 1773369, 1773370, 1773371, 1773372, 1773373, 1773374, 1773375, 1773376, 1773377, 1773378, 1773379, 1778871, 1778872, 1778873, 1778874, 1778875, 1778876, 1778877, 1778878, 1778879, 1778880, 1800896, 1800897, 1800899, 1800900, 1800901, 1800902, 1800903, 1800904, 1810477, 1810478, 1810479, 1810480, 1810481, 1810482, 1810483, 1810484, 1810485, 1810486, 1810487, 1817574, 1817575, 1817576, 1817577, 1817578, 1817579, 1821882, 1821883, 1821884, 1821885, 1821886, 1821887, 1890701, 1890702, 1890703, 1890704, 1890705, 2162605, 2162606, 2162607, 2264838, 2264839, 2264840, 2264841, 2264842, 2264843, 2365744, 2365745, 2365746, 2365747, 2365748, 2365749, 2365750, 2365751, 2365752, 2365753, 2365754, 2365755, 2365756, 2365757, 2365758, 2370369, 2370370, 2370371, 2370372, 2370373, 2370374, 2370375, 2370376, 2370377, 2370378, 2370379, 2370380, 2415415, 2415416, 2415417, 2415418, 2415419, 2415420, 2415421, 2415422, 2415423, 2415424, 2415425, 2415426, 2415427, 2415428, 2415429, 2415430, 2415431, 2415432, 2415433, 2415434, 2419005, 2419006, 2419007, 2419008, 2419009, 2419010, 2419011, 2419012, 2419013, 2419014, 2419015, 2419016, 2419017, 2419018, 2419019, 2419020, 2419021, 2419022, 2419023, 2419024, 2463210, 2463211, 2463212, 2463213, 2463214, 2599653, 2599654, 2599655, 2599656, 2599657, 2640117, 2640118, 2640119, 2640120, 2640121, 2640122, 2640123, 2640124, 2640125, 2640126, 2668355, 2668356, 2668357, 2668358, 2668359, 2668360, 2668361, 2668362, 2668363, 2668364, 2668365, 2668366, 2668367, 2668369, 2668370, 2668371, 2668372, 2668373, 2668374, 2726301, 2726302, 2726303, 2726304, 2726305, 2726306, 2726307, 2726310, 2849439, 2849440, 2849441, 2849442, 2849443, 2849444, 2849445, 2849446, 2849447, 2849448, 3552579, 3552580, 3552581, 3552582, 3552583, 3552584, 3552585, 3552586, 3552587, 3552588, 3782921, 3782922, 3782923, 3782924, 3782925, 3782926, 3782927, 3782928, 3782929, 3782930, 4016867, 4053339, 4053340, 4053341, 4053342, 4053343, 4053344, 4053345, 4053346, 4053347, 4082555, 4082556, 4082557, 4082558, 4082559, 4082560, 4082561, 4082562, 4143354, 4143355, 4143356, 4143357, 4143358, 4175257, 4175258, 4175259, 4175260, 4175261, 4175262, 4175263, 4175264, 4199377, 4199378, 4199379, 4199380, 4199381, 4199382, 4213442, 4213443, 4213444, 4213445, 4213446, 4213447, 4213448, 4213449, 4213450, 4213451, 4213452, 4213453, 4237448, 4237449, 4237450, 4237451, 4237452, 4237453, 4237454, 4237455, 4237456, 4237457, 4320163, 4320164, 4320165, 4320166, 4320167, 4320168, 4320169, 4320170, 4320171, 4320172, 4367455, 4367456, 4367457, 4367458, 4367459, 4367460, 4367461, 4367462, 4367463, 4367464, 4382090, 4382091, 4382092, 4382093, 4382094, 4382095, 4382096, 4382097, 4382098, 4382099, 4399635, 4399636, 4399637, 4399638, 4399639, 4399640, 4399641, 4399642, 4399643, 4399644, 4728943, 4728944, 4728945, 4734555, 4734556, 4734557, 4734558, 4734559, 4734560, 4734561, 4802599, 4802600, 4802601, 4802602, 4802603, 4802604, 4802605, 4802606, 4802607, 4802608, 4802609, 4802610, 4802611, 4802612, 4802613, 4846019, 4846020, 4846021, 4846022, 4846023, 4846024, 4846025, 4846026, 4856025, 4856026, 4856027, 4856028, 4856029, 4856030, 4856031, 4856032, 4856033, 4856034, 4856035, 4856036, 4856037, 4856038, 4856039, 4856040, 4856041, 4856042, 4856043, 4856044, 4876881, 4917323, 4917324, 4917326, 4917327, 4917328, 4917329, 4917330, 4917331, 4917332, 4917333, 4938613, 4938615, 4938617, 4938618, 4938619, 4938620, 4938621, 4938622, 4938623, 4938624, 4938625, 4938628, 4938629, 4938631, 4938632, 4938635, 4938636, 4938637, 4938639, 4948123, 4948124, 4948125, 4948126, 4948127, 4948128, 4948129, 4948130, 4948131, 4948132, 4948133, 4948134, 4948135, 4948136, 4948137, 4948138, 4948139, 4948140, 4948141, 4948142, 4948143, 4948144, 4948145, 4948146, 4948147, 4948148, 4948149, 4948150, 4948151, 4948152, 4948153, 4948154, 4948155, 4948156, 4948157, 4948158, 4948159, 4948160, 4948161, 4948162, 4952615, 4952616, 4952617, 4952618, 4952619, 4952620, 4952621, 4952622, 4952623, 4952624, 4952625, 4952626, 4952627, 4952628, 4952629, 4952630, 4952631, 4952632, 4952633, 4952634); 204-12-13 (10885862047908; 0786075, 0786076, 0786077, 0786078, 0786079, 0786080, 0856858, 0856859, 0856860, 0856861, 0856862, 0856863, 0856866, 0856867, 0856868, 0856869, 0928734, 0928735, 0928736, 0965858, 0965859, 0965860, 0965861, 0965862, 0965863, 0965864, 0965865, 0965867, 1056465, 1056466, 1056468, 1056469, 1095852, 1095853, 1095855, 1147284, 1147285, 1147286, 1147287, 1221001, 1221002, 1221003, 1221004, 1221005, 1221006, 1221007, 1221008, 1221009, 1221010, 1272679, 1324214, 1324215, 1324216, 1324217, 1324218, 1324219, 1324220, 1324221, 1324222, 1324223, 1324224, 1324225, 1324226, 1324227, 1371305, 1371306, 1371307, 1371308, 1371309, 1371310, 1371311, 1371312, 1371313, 1434119, 1434120, 1434121, 1434122, 1434123, 1439857, 1439858, 1439859, 1439860, 1439861, 1567694, 1567695, 1567696, 1567697, 1567699, 1567700, 1567701, 1567702, 1595186, 1595187, 1595188, 1595189, 1595190, 1595191, 1595192, 1595193, 1595194, 1595195, 1654384, 1654385, 1654386, 1654387, 1654388, 1654390, 1654391, 1654392, 1654393, 1654394, 1654395, 1654396, 1654397, 1654398, 1654399, 1654400, 1654401, 1654402, 1742763, 1742764, 1742765, 1742766, 1742767, 1742768, 1742769, 1799279, 1799280, 1799281, 1799282, 1799283, 1799284, 1799285, 1810263, 1810264, 1810265, 1810266, 1810267, 1810268, 1810269, 1810270, 1810271, 1822723, 1822724, 1822725, 1822726, 1822727, 1822728, 1865358, 1865359, 1865360, 1865361, 1865362, 1865364, 2159166, 2159167, 2259363, 2259364, 2259365, 2259366, 2259367, 2259368, 2370381, 2370382, 2370383, 2370384, 2370385, 2370387, 2370388, 2370389, 2370390, 2370391, 2370392, 2370393, 2370394, 2370395, 2370396, 2370397, 2370398, 2370399, 2370400, 2414611, 2414612, 2414613, 2414614, 2414615, 2414616, 2414617, 2414618, 2414619, 2414620, 2414621, 2414622, 2414623, 2414624, 2414625, 2414626, 2414627, 2414628, 2414629, 2414630, 2415648, 2415649, 2415650, 2415652, 2415653, 2415654, 2415655, 2415656, 2415657, 2415658, 2415659, 2415660, 2415661, 2415662, 2415663, 2415664, 2415666, 2708659, 2708660, 2708661, 2708662, 2708663, 2708664, 2708665, 2708666, 2708667, 2708668, 2708669, 2708670, 2708671, 2708672, 2708673, 2708674, 2708675, 2708676, 2781285, 4084294, 4084295, 4084296, 4084297, 4084298, 4084299, 4084300, 4084301, 4084302, 4084303, 4145161, 4145162, 4145163, 4145164, 4145165, 4158336, 4158337, 4158338, 4158339, 4158340, 4158341, 4158342, 4158343, 4158344, 4158345, 4158346, 4158347, 4158348, 4158349, 4234201, 4234202, 4234203, 4234204, 4234205, 4234206, 4384046, 4384047, 4384048, 4384049, 4384050, 4384051, 4384052, 4384053, 4384054, 4384055, 4910357, 4910358, 4910359, 4945388, 4945389, 5005391, 5005392, 5005393, 5005395, 5005396); 204-13-09 (10885862047946; 0738085, 0738086, 0738087, 0738088, 0738089, 0738090, 0738091, 0873178, 0873179, 0880267, 0880268, 0880269, 0880277, 0888706, 0888707, 0888708, 0888709, 0888710, 0903244, 0903245, 0903246, 0903247, 0923881, 0923883, 0923884, 0923885, 0957090, 0957091, 0957092, 0957093, 0957094, 0957095, 0957096, 0957097, 0957098, 0957099, 1023430, 1023431, 1023432, 1023433, 1023434, 1023435, 1023436, 1023437, 1023438, 1023439, 1045915, 1045916, 1045917, 1045918, 1045919, 1080068, 1080069, 1080070, 1080071, 1080072, 1084267, 1084268, 1084269, 1084270, 1084271, 1121617, 1121618, 1121619, 1121620, 1121621, 1124480, 1124481, 1124482, 1124483, 1212143, 1212144, 1212145, 1212916, 1212917, 1212918, 1212919, 1212920, 1220986, 1220987, 1220988, 1220989, 1220990, 1229011, 1229012, 1229013, 1229014, 1229015, 1229016, 1229017, 1229018, 1229019, 1255862, 1255863, 1255864, 1255865, 1255866, 1255867, 1255868, 1255869, 1255870, 1255871, 1302619, 1302620, 1302621, 1302622, 1302623, 1302624, 1302625, 1302626, 1302627, 1302628, 1302629, 1302630, 1302631, 1302632, 1302633, 1302634, 1302635, 1302636, 1302637, 1302638, 1353923, 1353924, 1353925, 1353926, 1353927, 1366615, 1366616, 1366617, 1366618, 1366619, 1366620, 1366621, 1366622, 1366623, 1366624, 1406356, 1406357, 1406358, 1406359, 1406360, 1480795, 1480796, 1480797, 1480798, 1522625, 1522626, 1534797, 1534798, 1534799, 1534800, 1576758, 1576759, 1576760, 1576761, 1576762, 1576763, 1576764, 1576765, 1576766, 1576767, 1619204, 1619205, 1619206, 1619207, 1619208, 1619209, 1619210, 1619211, 1619212, 1619213, 1661516, 1661517, 1661518, 1661519, 1661520, 1661521, 1661522, 1661523, 1661524, 1661525, 1661526, 1661527, 1661528, 1661529, 1661530, 1673843, 1673844, 1673845, 1673846, 1673847, 1673848, 1673849, 1673850, 1673851, 1673852, 1693776, 1693777, 1693778, 1693779, 1693780, 1693781, 1693782, 1693783, 1693784, 1693785, 1791221, 1791222, 1791223, 1791224, 1791225, 1791226, 1791227, 1791228, 1791229, 1799118, 1799119, 1799120, 1799121, 1799122, 1799123, 1799124, 1799125, 1799126, 1799127, 1811412, 1811413, 1811414, 1811415, 1811416, 1811417, 1811418, 1811419, 1817798, 1817799, 1817800, 1817801, 1817802, 1817803, 1831528, 1831529, 1831530, 1831531, 1831533, 1831534, 1831535, 1831536, 1899925, 1899926, 1899927, 1899928, 1899929, 2255018, 2255019, 2255020, 2255021, 2255971, 2255972, 2255973, 2255974, 2255975, 2255976, 2306080, 2306081, 2306082, 2306083, 2306084, 2306085, 2306086, 2306087, 2306088, 2306089, 2306090, 2306091, 2306092, 2306093, 2370401, 2370402, 2370403, 2370404, 2370405, 2370406, 2370407, 2370408, 2370409, 2370410, 2398805, 2398806, 2398807, 2398808, 2398809, 2398810, 2398811, 2398812, 2398813, 2398814, 2398815, 2398816, 2398817, 2398818, 2398819, 2398820, 2398821, 2415553, 2415555, 2415556, 2415557, 2415558, 2415559, 2415560, 2415561, 2415562, 2454408, 2454409, 2454410, 2454411, 2454412, 2454413, 2454414, 2454415, 2454416, 2454417, 2509357, 2509358, 2509359, 2509360, 2551485, 2551486, 2551487, 2551488, 25
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Exactech, Inc.
- Reason for Recall:
- Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
OPTETRAK Comprehensive Knee System, labeled as the following: a. OPTETRAK All-polyethylene CR Tibial Components, with Product Line Numbers: 200-11-XX, 200-12-XX, 200-13-XX, 200-14-XX, 200-15-XX, 200-16-XX; b. OPTETRAK All-polyethylene PS Tibial Components, with Product Line Numbers: 204-11-XX, 204-12-XX, 204-13-XX, 204-14-XX, 204-15-XX, 204-16-XX; c. OPTETRAK HI-FLEX PS Polyethylene Tibial Inserts, with Product Line Numbers: 244-20-XX, 244-21-XX, 244-22-XX, 244-23-XX, 244-24-XX, 244-25-XX; d. OPTETRAK CR Tibial Inserts, with Product Line Numbers: 200-21-XX, 200-22-XX, 200-23-XX, 200-24-XX, 200-25-XX, 200-26-XX, 200-50-XX, 200-51-XX; e. OPTETRAK CR Tibial Slope + Insert, with Product Line Numbers: 200-56-XX, 200-57-XX, 200-61-XX, 200-62-XX, 200-63-XX, 200-64-XX, 200-65-XX; f. OPTETRAK CR Tibial Slope ++ Insert, with Product Line Numbers: 200-57-XX, 200-71-XX, 200-72-XX, 200-73-XX, 200-74-XX, 200-75-XX; g. OPTETRAK PS Tibial Inserts, with Product Line Numbers: 204-21-XX, 204-22-XX 204-23-XX, 204-24-XX, 204-25-XX, 204-26-XX, 204-50-XX, 204-51-XX; h. OPTETRAK "MOMB" Non-Mod Molded Insert, with Product Line Numbers: 204-91-XX, 204-92-XX, 204-93-XX, 204-94-XX, 204-95-XX; i. OPTETRAK CC Tibial Insert, with Product Line Numbers: 208-21-XX, 208-22-XX, 208-23-XX, 208-24-XX, 208-25-XX, 208-51-XX; j. OPTETRAK B-Series PS Tibial Insert, with Product Line Numbers: 224-21-XX, 224-22-XX, 224-23-XX, 224-24-XX. Affected 510(k) numbers include: K932690, K933494, K932776, K011976, K082022, K933610, K030686, K954208, K010434, K033883. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.
Product Codes/Lot Numbers:
a. Item Number (Device Identifier; Serial number(s)): 200-11-11 (10885862039934; 1010599); 200-12-09 (10885862039965; 0850770, 0850771, 0850773, 0850776, 0850778, 1052587, 1052589, 1070149, 1070150, 1070151, 1070152, 1267585, 1267586, 1267588, 1318739, 1318740, 1318742, 1375833, 1375836, 1375837, 1375839, 1375844); 200-12-11 (10885862039972; 0825651, 0825653, 0825654, 0825657, 0825659, 1052802, 1052803, 1052804, 1309224); 200-12-13 (10885862039989; 0763041, 0763043, 0820424, 0820431, 1053775); 200-13-09 (10885862039996; 0765380, 0765381, 0765382, 0765383, 0789723, 0789724, 0789726, 0789727, 0789728, 0789729, 0789730, 0789731, 0789732, 0897683, 0998654, 0998655, 0998656, 0998657, 0998658, 0998659, 0998660, 0998661, 0998662, 0998663, 0998664, 0998665, 0998666, 1073908, 1073909, 1073910, 1073911, 1084202, 1084203, 1084204, 1084205, 1261436, 1261437, 1261438, 1261439, 1261440, 1316122, 1316123, 1316124, 1316125, 1316126, 1319147, 1319148, 1319149, 1319150, 1319151, 1407772, 1407774, 1407775, 1407776, 1407777, 1407778, 1407779, 1407780, 1407781, 1516714, 1516715, 1516716, 1516717, 1516718, 1976718, 1976719, 1976720, 1976721, 1976722, 2061574, 2097473, 2097474, 2097475, 2097476, 2097477, 2097478, 2097479, 2145890, 2145891, 2145892, 2145894, 2145895, 2145896, 2559776); 200-13-11 (10885862040008; 0769998, 0770002, 0770003, 0770004, 0818558, 0818559, 0818560, 0818562, 0818563, 0818564, 0818565, 1027748, 1027750, 1027752, 1027754, 1084902, 1291129, 1291130, 1291131, 1319746, 1319747, 1319748, 1319752, 1319753, 1319754, 1319756); 200-13-13 (10885862040015; 0767606, 0767607, 0767608, 0767609, 0953412, 1063120, 1307627); 200-14-09 (10885862040039; 0607230, 0607231, 0607232, 0607233, 0607234, 0607235, 0607236, 0607237, 0607238, 1033676, 1033677, 1033678, 1033679, 1040785, 1040786, 1040787, 1040788, 1065019, 1065020, 1065021, 1065022, 1065023, 1065024, 1065025, 1153592, 1153593, 1153594, 1153595, 1153596, 1153597, 1153598, 1153599, 1153600, 1215318, 1215319, 1215320, 1215322, 1281462, 1281463, 1281464, 1281465, 1321078, 1321079, 1321080, 1321081, 1321082, 1389683, 1389684, 1389685, 1400987, 1400988, 1400989, 1400990, 2012144, 2012145, 2048510, 2048511, 2207641, 2207642, 2207643, 2345327, 2345328, 2414649, 2414652, 2454327, 2454328, 2454329, 2565530, 2565533, 2565534); 200-14-11 (10885862040046; 0942385, 0942386, 0942393, 1050502, 1310607); 200-14-13 (10885862040053; 0942535, 1048228, 1048232, 1153792); 200-15-09 (10885862040077; 0822759, 0822761, 0822762, 0822764, 0822765, 0822766, 0822767, 1048609, 1048611, 1048612, 1048614, 1048615, 1048616, 1048617, 1136150, 1136151, 1136152, 1136154, 1214791, 1214792, 1214793, 1214794, 1214795, 1214796, 1283152, 1283153, 1283154, 1283155, 1439852, 1439853, 1439855, 1439856, 2146108, 2146113); 200-15-11 (10885862040084; 1065636, 1065638); 200-15-13 (10885862040091; 0954497, 0954501); 200-16-11 (10885862040121; 1040633, 1040634, 1040635, 1040636, 1073405, 1073407, 1084258); 200-16-13 (10885862040138; 0956887); b. Item Number (Device Identifier; Serial number(s)): 204-11-09 (10885862047823; 0961663, 0961665, 0961666, 0961667, 0961668, 0961669, 0961671, 0961672, 1045876, 1045877, 1045879, 1070620, 1070622, 1070623, 1106799, 1106800, 1106801, 1106802, 1106803, 1106805, 1106807, 1193435, 1193437, 1193438, 1193439, 1232147, 1232151, 1255910, 1255911, 1255912, 1255913, 1255914, 1270806, 1270807, 1270808, 1270809, 1308220, 1308221, 1308222, 1308223, 1308224, 1308225, 1377016, 1377017, 1377018, 1377019, 1377020, 1377021, 1377022, 1377023, 1377024, 1377025, 1545261, 1545262, 1545263, 1545264, 1545265, 1565700, 1565701, 1565702, 1565703, 1565704, 1582779, 1582780, 1582781, 1582782, 1582783, 1582784, 1582785, 1582786, 1582787, 1582788, 1582789, 1582790, 1582791, 1582792, 1582793, 1582794, 1582795, 1582796, 1582797, 1582798, 1659777, 1659778, 1659779, 1659780, 1659781, 1659782, 1659783, 1659784, 1659785, 1659786, 1659787, 1659788, 1659789, 1659790, 1659791, 1659792, 1659793, 1659794, 1659795, 1659796, 1659797, 1659798, 1659799, 1659800, 1659801, 1704209, 1704210, 1704211, 1704212, 1786558, 1786559, 1786560, 1786561, 1786562, 1786563, 1786564, 1786565, 1797016, 1797017, 1797018, 1797019, 1797020, 1797021, 1797022, 1797023, 1797024, 1797025, 1797026, 1797027, 1797028, 1817561, 1817562, 1817563, 1817564, 1817565, 1817566, 1817567, 1817568, 1817569, 1817570, 1817571, 1817572, 1817573, 1842319, 1842320, 1842321, 1842324, 1842325, 1901249, 1901250, 1901251, 1901252, 1901253, 1901255, 1901256, 1901259, 1901260, 1901261, 1901262, 1901263, 1901264, 1901265, 1901266, 1901267, 1901268, 1901269, 1901270, 1901271, 1901272, 1901273, 1901274, 1901275, 1901276, 2266634, 2266635, 2266636, 2266638, 2276586, 2276587, 2276588, 2276589, 2397396, 2397397, 2397398, 2397399, 2397400, 2397401, 2397402, 2397403, 2397404, 2397405, 2425639, 2425640, 2425641, 2425642, 2425643, 2425644, 2425645, 2425646, 2425647, 2425648, 2425650, 2425651, 2425652, 2425653, 2425654, 2425655, 2425656, 2430257, 2430258, 2430259, 2430260, 2430261, 2430262, 2430263, 2430264, 2430265, 2615191, 2615192, 2615193, 2615194, 2615195, 2615196, 2615197, 2615198, 2615199, 2615200, 2653102, 2653103, 2653104, 2653105, 2653106, 2653107, 2653108, 2653109, 2653110, 2653111, 2653112, 2653113, 2653114, 2653115, 2653116, 2653117, 2653118, 2653119, 2653120, 2653121, 2653123, 2653124, 2653125, 2653126, 2653127, 2653128, 2653129, 2653130, 2653131, 3545584, 3545585, 3545586, 3545587, 3545588, 3545589, 3545590, 3545591, 3545592, 3545593, 3636693, 3636694, 3636695, 3636696, 3636697, 3636698, 3636699, 3636700, 3636701, 3636702, 3654361, 3654362, 3654363, 3654364, 3654365, 3654366, 3654367, 3654368, 3654369, 3654370, 3660720, 3660721, 3660722, 3660723, 3660724, 3781241, 3781242, 3781243, 3781244, 3781245, 3792869, 3792870, 3792871, 3792872, 3792873, 3809326, 3809327, 3809328, 3809329, 3809330, 4080806, 4080807, 4080808, 4080809, 4080810, 4080811, 4080812, 4080813, 4080814, 4080815, 4149218, 4149219, 4149220, 4149221, 4149222, 4175249, 4175250, 4175251, 4175252, 4175253, 4175254, 4175255, 4175256, 4205182, 4205183, 4205184, 4205185, 4205186, 4331818, 4331819, 4331820, 4331821, 4331822, 4331823, 4331824, 4331825, 4331826, 4367437, 4367438, 4367439, 4367440, 4367441, 4367442, 4367443, 4367444, 4399568, 4399569, 4399570, 4399571, 4399572, 4399573, 4399574, 4399575, 4399576, 4399577, 4933848, 4933849, 4933851, 4933852, 4933853, 4933854, 4933855, 4933856, 4933857, 4933858, 4933859, 4933860, 4933861, 4933862, 4933863, 4933864, 4933865, 4933866, 4933867, 4933868, 4933869, 4933870, 4933871, 4933872, 4933873, 4933874, 4933875, 4933876, 4933877, 4933878, 4933879, 4933880, 4933881, 4933882, 4933883, 4933884, 4933885, 4933887, 4945297, 4945298, 4945299, 4945300, 4945301, 4945302, 4945303, 4945304, 4945305, 4945306, 4945307, 4945308, 4945309, 4945310, 4945311, 4945312, 4945313, 4945314, 4945315, 4945316, 4945317, 4945318, 4945319, 4945320, 4945321, 4945322, 4945323, 4945324, 4945325, 4945326, 4945327, 4945328, 4945329, 4945330, 4945331, 4945332, 4945333, 4945334, 5086208, 5086209, 5086210, 5086211, 5086212, 5086213, 5086214, 5086215, 5086216, 5086217, 5086218, 5086219, 5086220, 5172053, 5172054, 5172055, 5172056, 5172057, 5172058, 5172059, 5180020, 5180021, 5180022, 5180023, 5180024, 5180025, 5180026, 5188774, 5188775, 5188776, 5209513); 204-11-11 (10885862047830; 0971418, 0971419, 0971420, 0971421, 0971422, 0971423, 1015616, 1015617, 1015618, 1015619, 1047575, 1047577, 1047578, 1066792, 1066793, 1066794, 1087082, 1087083, 1087084, 1087085, 1087086, 1168686, 1168687, 1187090, 1187092, 1187093, 1187094, 1187096, 1187098, 1187099, 1255900, 1255901, 1255902, 1255903, 1255904, 1255905, 1255906, 1255907, 1255908, 1255909, 1270537, 1270540, 1294283, 1294284, 1294285, 1294287, 1294288, 1294289, 1294290, 1307222, 1307223, 1307224, 1307225, 1392229, 1392230, 1392231, 1392232, 1392233, 1576699, 1576700, 1576701, 1576702, 1576703, 1576704, 1576705, 1585601, 1585602, 1585603, 1585604, 1585605, 1585606, 1585607, 1585608, 1585609, 1585610, 1585611, 1585612, 1585613, 1585614, 1585615, 1585616, 1585617, 1585618, 1585619, 1585620, 1585621, 1585622, 1585623, 1585624, 1585625, 1585626, 1585627, 1585628, 1585629, 1585630, 1585631, 1585632, 1745600, 1745601, 1745602, 1745603, 1745604, 1774126, 1774127, 1774128, 1774129, 1774130, 1774131, 1774132, 1774133, 1774134, 1774135, 1774136, 1774137, 1786761, 1786762, 1786763, 1786764, 1786765, 1786766, 1786767, 1786768, 1810253, 1810254, 1810255, 1810256, 1810257, 1810258, 1810259, 1810260, 1810261, 1821731, 1821732, 1821733, 1821734, 1821735, 1821736, 1821737, 1821738, 1821739, 1821740, 1821741, 1865512, 1865513, 1865515, 1897369, 1897370, 1897371, 1897373, 1897374, 1897375, 1897376, 1897377, 2265900, 2265901, 2265902, 2265903, 2265904, 2306144, 2306145, 2306146, 2306147, 2306148, 2306149, 2306150, 2306151, 2306153, 2306154, 2306155, 2306156, 2306157, 2306158, 2306159, 2306160, 2306161, 2306163, 2306164, 2306165, 2306166, 2306167, 2306168, 2306169, 2306170, 2306171, 2306172, 2430406, 2430407, 2430408, 2430409, 2430410, 2430411, 2430412, 2430413, 2430414, 2430415, 2582070, 2582071, 2582072, 2582073, 2582074, 2582075, 2582076, 2628063, 2628064, 2628065, 2628066, 2628067, 2628068, 2628069, 2628070, 2628071, 2628072, 2628073, 2628074, 2628075, 3545581, 3545582, 3545583, 3616391, 3616392, 3616393, 3616394, 3616395, 3636926, 3636927, 3644006, 3654371, 3654372, 3654373, 3654374, 3654375, 3654376, 3729486, 3756690, 3775777, 3775778, 3776705, 3776706, 3776707, 3776708, 3776709, 3811782, 3811783, 3811784, 3811785, 3811786, 3811787, 3811788, 3811789, 3811790, 3811791, 4053394, 4053395, 4053396, 4053397, 4053398, 4053399, 4053400, 4053401, 4053402, 4053403, 4053404, 4053405, 4053406, 4053407, 4053408, 4205187, 4205188, 4205189, 4205190, 4205191, 4226766, 4226767, 4226768, 4226769, 4226770, 4226772, 4226773, 4226774, 4226775, 4349216, 4349217, 4349218, 4349219, 4349220, 4349221, 4349222, 4349223, 4349224, 4384105, 4384106, 4384107, 4384108, 4384109, 4384110, 4384111, 4384112, 4384113, 4384114, 4816187, 4816188, 4816189, 4846027, 4846028, 4846029, 4846030, 4846031, 4846032, 4846033, 4846034, 4846035, 4846036, 4846037, 4846038, 4846039, 4846040, 4846041, 4846042, 4846043, 4846044, 4846045, 4846046, 4846047, 4846048, 4846049, 4846050, 4846051, 4846052, 4934866, 4934867, 4934868, 4934869, 4934870, 4934871, 4934872, 4934873, 4934874, 4934875, 4934876, 4934877, 4934878, 4934879, 4934880, 4934881, 4934882, 4934883, 4934884, 4934885, 4934886, 4934887, 4934888, 4934889, 4934890, 4965321, 4965322, 4965323, 4965324, 4965325, 4965326, 4965327, 4965328, 4965329, 4965330, 4965331, 4965332, 4965333, 4965334, 4965335, 4965336, 4965337, 4965338, 4965339, 4965340, 4965341, 4965342, 4965343, 4965344, 4965345, 5009433, 5009434, 5009435, 5009436, 5009437, 5009438, 5009439, 5009440, 5009441, 5009442, 5009443, 5009444, 5009446, 5009447, 5009448, 5009449, 5009450, 5009451, 5009452, 5009453, 5009454, 5009455, 5009457, 5179063); 204-11-13 (10885862047847; 0961673, 0961674, 0961675, 0961677, 0961678, 0961680, 1021516, 1021517, 1021518, 1021519, 1031608, 1031609, 1031610, 1053779, 1053780, 1053781, 1098258, 1098259, 1201250, 1201251, 1201252, 1201254, 1201255, 1259004, 1259005, 1259006, 1259007, 1259008, 1307220, 1307221, 1328794, 1328795, 1328796, 1328797, 1328798, 1328799, 1328800, 1328801, 1328802, 1328803, 1328804, 1328805, 1328806, 1597232, 1597233, 1597234, 1597235, 1597236, 1597237, 1597239, 1597240, 1597241, 1627673, 1627674, 1627675, 1627676, 1627677, 1627678, 1627679, 1647931, 1647932, 1647933, 1647934, 1647935, 1659319, 1659320, 1659321, 1659323, 1766122, 1766123, 1766124, 1766125, 1766126, 1766127, 1766128, 1766129, 1766130, 1766131, 1766132, 1766133, 1766134, 1767874, 1767875, 1767876, 1767877, 1767878, 1767879, 1767880, 1767881, 1767882, 1767883, 1767884, 1767885, 1767886, 1852794, 1852795, 1852796, 1852797, 1852798, 1852799, 1852800, 1852801, 1852802, 1852803, 1903394, 1903395, 1903396, 1903398, 1903399, 1903401, 1903402, 1903403, 1903404, 1903405, 1903406, 1903410, 1903411, 1903412, 2259858, 2259859, 2259860, 2259861, 2259862, 2259863, 2259864, 2259865, 2259866, 2259867, 2398501, 2398502, 2398503, 2398504, 2398505, 2398506, 2398507, 2398508, 2398509, 2430396, 2430397, 2430398, 2430399, 2430400, 2430401, 2430402, 2430403, 2430404, 2430405, 2640147, 2640148, 2640149, 2640150, 2640151, 2640152, 2640153, 2640154, 2640155, 2640156, 2725662, 2725664, 2725665, 2725666, 2725667, 2725668, 2725669, 2725671, 2725672, 2725674, 2725676, 2725677, 2725678, 2725679, 2725680, 2725681, 2815981, 2926604, 2926605, 2926606, 2926607, 2926608, 2926609, 2926610, 2926611, 2926612, 3647961, 3647963, 3647964, 3647966, 4081311, 4081312, 4081313, 4081314, 4081315, 4124295, 4124298, 4124299, 4124300, 4124301, 4124302, 4124303, 4124304, 4124305, 4124307, 4124308, 4124309, 4124310, 4124311, 4124312, 4124314, 4124315, 4124316, 4124317, 4124318, 4384097, 4384098, 4384099, 4384100, 4384101, 4384102, 4384103, 4384104, 5084582, 5084583, 5084584, 5084585, 5084586); 204-12-09 (10885862047885; 0881042, 0881043, 0895287, 0895288, 0895289, 0895290, 0895291, 0895292, 0895293, 0895294, 0895295, 0957833, 0957834, 0957835, 0957836, 0957837, 1019604, 1019605, 1021520, 1021521, 1021522, 1021523, 1021524, 1045910, 1045911, 1045912, 1045913, 1045914, 1079090, 1079091, 1079092, 1079093, 1097012, 1097013, 1097014, 1097015, 1122249, 1122250, 1122251, 1122252, 1122253, 1150183, 1150185, 1173036, 1173037, 1173038, 1173039, 1173040, 1208280, 1215635, 1215636, 1215639, 1230263, 1230264, 1230265, 1230266, 1230267, 1230268, 1230269, 1230270, 1237521, 1237522, 1237523, 1237524, 1237525, 1265772, 1265773, 1265774, 1294243, 1294244, 1294245, 1300708, 1300709, 1300710, 1300711, 1300712, 1300713, 1300714, 1300715, 1324324, 1324325, 1324326, 1324327, 1324328, 1324330, 1324332, 1324333, 1326334, 1326335, 1326336, 1326337, 1326338, 1326339, 1347634, 1347635, 1347636, 1347637, 1357215, 1357216, 1357217, 1357219, 1357220, 1357221, 1357222, 1357223, 1357224, 1357225, 1357226, 1357227, 1369862, 1369863, 1369864, 1369865, 1369866, 1371266, 1371267, 1371268, 1371269, 1371270, 1371271, 1371272, 1371273, 1371274, 1371275, 1441530, 1441531, 1441532, 1441533, 1441534, 1458179, 1458180, 1458181, 1458182, 1458183, 1458184, 1458185, 1458186, 1458187, 1458188, 1552930, 1552931, 1552932, 1552933, 1552934, 1552935, 1552936, 1552937, 1552938, 1552939, 1577909, 1577910, 1577911, 1577912, 1577913, 1577914, 1577915, 1577916, 1577917, 1577918, 1584330, 1584331, 1584332, 1584333, 1584334, 1584335, 1584336, 1584337, 1584338, 1584339, 1584340, 1584341, 1584342, 1584343, 1584344, 1584345, 1584346, 1584347, 1584348, 1584349, 1584350, 1584351, 1584352, 1584353, 1584354, 1584355, 1584356, 1584357, 1584358, 1707644, 1707645, 1707646, 1707647, 1707648, 1707649, 1707650, 1707651, 1742692, 1742693, 1742694, 1742695, 1742696, 1742697, 1742698, 1742699, 1766190, 1766191, 1766192, 1766193, 1766194, 1766195, 1766196, 1766197, 1766198, 1778860, 1778861, 1778862, 1778863, 1778864, 1778865, 1778866, 1778867, 1778868, 1778869, 1778870, 1787517, 1787518, 1787519, 1787520, 1787521, 1787522, 1787523, 1787524, 1787525, 1787526, 1787527, 1787528, 1787529, 1787530, 1787531, 1787532, 1787533, 1787534, 1787535, 1787536, 1787537, 1787538, 1787539, 1787540, 1796205, 1796206, 1796207, 1796208, 1796209, 1796210, 1796211, 1796212, 1817740, 1817741, 1817742, 1817743, 1817744, 1817745, 1841280, 1841281, 1841282, 1841283, 1841284, 1841285, 1841286, 1841287, 1841288, 1841289, 2067761, 2067762, 2067763, 2067764, 2067765, 2067766, 2067767, 2067768, 2067769, 2067770, 2142051, 2142052, 2142053, 2142054, 2142055, 2142056, 2142057, 2142058, 2142059, 2142060, 2142061, 2255718, 2255719, 2255720, 2255721, 2255722, 2255723, 2255724, 2255725, 2255726, 2306719, 2306720, 2306721, 2306722, 2306723, 2306724, 2306725, 2306726, 2306727, 2306728, 2306729, 2306730, 2306731, 2306732, 2306733, 2306734, 2306735, 2306736, 2306737, 2306738, 2370257, 2370258, 2370259, 2370260, 2370261, 2370262, 2370263, 2370264, 2370265, 2370266, 2370267, 2370268, 2370269, 2370270, 2370271, 2370272, 2370274, 2370275, 2415451, 2415452, 2415453, 2415454, 2415455, 2415456, 2415458, 2415459, 2415460, 2415461, 2415462, 2415463, 2415464, 2415465, 2415466, 2415467, 2415468, 2415469, 2415470, 2416268, 2416269, 2416270, 2416271, 2416272, 2416273, 2416274, 2416275, 2416276, 2416277, 2416278, 2416279, 2416280, 2416281, 2416282, 2416283, 2416284, 2416285, 2416286, 2416287, 2496195, 2496197, 2496198, 2496199, 2496200, 2496201, 2496202, 2496203, 2496204, 2591481, 2591482, 2591483, 2591484, 2591485, 2591486, 2591487, 2591488, 2591489, 2591490, 2609808, 2609809, 2609810, 2609811, 2609812, 2609813, 2609814, 2609815, 2609816, 2609817, 2609818, 2609819, 2609820, 2609821, 2609822, 2609823, 2609824, 2609825, 2609826, 2628076, 2628077, 2628078, 2628079, 2628080, 2628081, 2628082, 2628083, 2628084, 2628085, 2628086, 2628087, 2628088, 2628089, 2702396, 2702397, 2702398, 2702399, 2702400, 2702401, 2702402, 2702403, 2702404, 2702405, 2702406, 2702407, 2702408, 2702409, 2702410, 2702411, 2702412, 2702413, 2702414, 2702415, 2702416, 2702417, 2702418, 2702419, 2726285, 2726286, 2726287, 2726288, 2726289, 2726290, 2726291, 2726292, 2726293, 2726294, 2726295, 2726296, 2726297, 2726298, 2726299, 2726300, 2849419, 2849420, 2849421, 2849422, 2849423, 2849424, 2849425, 2849426, 2849427, 2849428, 2849429, 2849430, 2849431, 2849432, 2849433, 2849434, 2849435, 2849436, 2849437, 2849438, 2976096, 2976097, 2976098, 2976099, 2976100, 2976101, 2976102, 2976103, 2976105, 3546135, 3546136, 3546137, 3546138, 3546139, 3546140, 3546141, 3546142, 3546143, 3546144, 3615053, 3615054, 3615055, 3615056, 3615057, 3615058, 3615059, 3615060, 3615061, 3615062, 3654398, 3654399, 3654400, 3654401, 3654402, 3654403, 3654404, 3654405, 3654406, 3654407, 3654438, 3654439, 3654440, 3654441, 3654442, 3654443, 3654444, 3654445, 3654446, 3654447, 3786571, 3786572, 3786573, 3786574, 3786575, 3786576, 3786577, 3786578, 3786579, 3786580, 3786581, 3786582, 3786583, 3786584, 3786585, 3786586, 3786587, 4053409, 4053410, 4082563, 4082564, 4082565, 4082566, 4082567, 4082568, 4082569, 4082570, 4143349, 4143350, 4143351, 4143352, 4143353, 4145166, 4145167, 4145168, 4145169, 4145170, 4145171, 4145172, 4145173, 4145174, 4145175, 4145176, 4145177, 4145178, 4145179, 4145180, 4145181, 4145182, 4145183, 4145184, 4145185, 4208780, 4208781, 4208782, 4208783, 4208784, 4213454, 4213455, 4213456, 4213457, 4213458, 4213459, 4213460, 4213461, 4213462, 4213463, 4213464, 4213465, 4213466, 4213467, 4213468, 4215594, 4215595, 4215596, 4215597, 4215598, 4215599, 4215600, 4215601, 4215602, 4280394, 4280395, 4280396, 4280397, 4280398, 4280399, 4280400, 4280401, 4293675, 4293676, 4293677, 4293678, 4293679, 4293680, 4293682, 4293683, 4293684, 4317141, 4317142, 4317143, 4317144, 4317145, 4317146, 4317147, 4317148, 4317149, 4317150, 4367445, 4367446, 4367447, 4367448, 4367449, 4367450, 4367451, 4367452, 4367453, 4367454, 4372329, 4372330, 4372331, 4372332, 4372333, 4372334, 4372335, 4372336, 4372337, 4384115, 4384116, 4384117, 4384118, 4384119, 4384120, 4384121, 4384122, 4384123, 4384124, 4399645, 4399646, 4399647, 4399648, 4399649, 4399650, 4399651, 4399652, 4399653, 4399654, 4793183, 4793184, 4793185, 4793186, 4793187, 4793188, 4793189, 4793190, 4793191, 4793192, 4793193, 4793194, 4793195, 4793196, 4793197, 4793198, 4793199, 4793200, 4793201, 4793202, 4793203, 4793204, 4793205, 4793206, 4793207, 4793208, 4856022, 4856023, 4856024, 4863234, 4863235, 4863236, 4863237, 4863238, 4863239, 4863240, 4863241, 4863242, 4863243, 4863244, 4863245, 4863246, 4863247, 4924904, 4924905, 4924906, 4924907, 4924908, 4924909, 4924910, 4924911, 4924912, 4924913, 4924914, 4924915, 4924916, 4924917, 4924918, 4924919, 4924920, 4924921, 4924922, 4924923, 4955637, 4955638, 4955639, 4955640, 4955641, 4955642, 4955643, 4955644, 4955645, 4955646, 4955647, 4955648, 4955649, 4955650, 4955651, 4955652, 4955653, 4955654, 4955655, 4955656, 4964001, 4964002, 4964003, 4964004, 4964005, 4964006, 4964007, 4964008, 4964009, 4964010, 4964011, 4964012, 4964013, 4964014, 4964015, 4964016, 4964017, 4964018, 4964019, 4964020, 4964021, 4964022, 4964023, 4964024, 4964025, 4964026, 4964027, 4964028, 4964029, 4964030, 4964031, 4964032, 4964033, 4964034, 4964035, 4964036, 4964037, 4964038, 4964039, 4964040, 4964041, 4964042, 4964043, 4964044, 4964045, 4964046, 4964047, 4964048, 4964049, 4964050, 4965358, 4965359, 4965360, 4965361, 4965362, 4965363, 4965364, 4965365, 4965366, 4965367, 4965368, 4965369, 4965370, 4965371, 4965372, 4965373, 4965374, 4965375, 4965376, 4965377, 4965378, 4965380, 4965381, 4965382, 4965383, 4965384, 4965385, 4965386, 4965387, 4965388, 4965390, 4965391, 4965392, 4965393, 4965394, 5039055, 5039056, 5039057, 5039058, 5039059, 5039060, 5039061, 5039062, 5039063, 5039064, 5039065, 5039066, 5039067, 5039068, 5039069, 5039070, 5039071, 5039072, 5039073, 5039074, 5039075, 5039076, 5039077, 5039078, 5039079, 5039080, 5039081, 5039082, 5039083, 5039084, 5049879, 5049882, 5049885, 5049887, 5049888, 5049889, 5049892, 5049895, 5049896); 204-12-11 (10885862047892; 0856716, 0856718, 0856719, 0856720, 0856721, 0856722, 0856723, 0856724, 0856725, 0856726, 0894046, 0894047, 0894048, 0894049, 0894050, 0894051, 0894052, 0894053, 0894054, 0894056, 0894057, 0973488, 0973489, 0973490, 0973492, 0973493, 0973494, 0973495, 0973496, 0973497, 0973498, 0973499, 0973500, 0973501, 0973502, 0973504, 0973505, 0973506, 1047385, 1047386, 1047387, 1047388, 1047389, 1086399, 1086401, 1086402, 1086403, 1086404, 1086405, 1086406, 1086408, 1086410, 1086411, 1086412, 1086417, 1216096, 1216098, 1216100, 1230623, 1230624, 1230625, 1230627, 1230628, 1230629, 1230630, 1230631, 1230635, 1230636, 1230637, 1230638, 1230641, 1274067, 1274068, 1274069, 1274070, 1274071, 1274072, 1274073, 1274074, 1274075, 1274076, 1274077, 1274078, 1274079, 1274080, 1308393, 1308394, 1308395, 1308396, 1308397, 1308398, 1308399, 1308400, 1308401, 1308402, 1365381, 1365382, 1365383, 1365384, 1365385, 1365386, 1365387, 1365388, 1370851, 1370852, 1370853, 1370854, 1370855, 1406148, 1406149, 1406150, 1406151, 1406152, 1480206, 1480207, 1480208, 1480209, 1480210, 1518785, 1518786, 1518787, 1518788, 1518789, 1564109, 1564110, 1564111, 1564112, 1564113, 1564114, 1564115, 1564116, 1564117, 1564118, 1586396, 1586397, 1586398, 1586399, 1586400, 1586401, 1586402, 1586403, 1586404, 1586405, 1586407, 1586408, 1586409, 1586410, 1586411, 1586412, 1586413, 1586414, 1586415, 1586416, 1586417, 1586418, 1586419, 1586420, 1586421, 1678542, 1678543, 1678544, 1678545, 1678546, 1678547, 1678548, 1678549, 1678550, 1678551, 1678552, 1678553, 1678554, 1678555, 1742060, 1742061, 1742062, 1742063, 1742064, 1742065, 1742066, 1773352, 1773353, 1773354, 1773355, 1773356, 1773357, 1773358, 1773359, 1773360, 1773361, 1773362, 1773363, 1773364, 1773365, 1773366, 1773367, 1773368, 1773369, 1773370, 1773371, 1773372, 1773373, 1773374, 1773375, 1773376, 1773377, 1773378, 1773379, 1778871, 1778872, 1778873, 1778874, 1778875, 1778876, 1778877, 1778878, 1778879, 1778880, 1800896, 1800897, 1800899, 1800900, 1800901, 1800902, 1800903, 1800904, 1810477, 1810478, 1810479, 1810480, 1810481, 1810482, 1810483, 1810484, 1810485, 1810486, 1810487, 1817574, 1817575, 1817576, 1817577, 1817578, 1817579, 1821882, 1821883, 1821884, 1821885, 1821886, 1821887, 1890701, 1890702, 1890703, 1890704, 1890705, 2162605, 2162606, 2162607, 2264838, 2264839, 2264840, 2264841, 2264842, 2264843, 2365744, 2365745, 2365746, 2365747, 2365748, 2365749, 2365750, 2365751, 2365752, 2365753, 2365754, 2365755, 2365756, 2365757, 2365758, 2370369, 2370370, 2370371, 2370372, 2370373, 2370374, 2370375, 2370376, 2370377, 2370378, 2370379, 2370380, 2415415, 2415416, 2415417, 2415418, 2415419, 2415420, 2415421, 2415422, 2415423, 2415424, 2415425, 2415426, 2415427, 2415428, 2415429, 2415430, 2415431, 2415432, 2415433, 2415434, 2419005, 2419006, 2419007, 2419008, 2419009, 2419010, 2419011, 2419012, 2419013, 2419014, 2419015, 2419016, 2419017, 2419018, 2419019, 2419020, 2419021, 2419022, 2419023, 2419024, 2463210, 2463211, 2463212, 2463213, 2463214, 2599653, 2599654, 2599655, 2599656, 2599657, 2640117, 2640118, 2640119, 2640120, 2640121, 2640122, 2640123, 2640124, 2640125, 2640126, 2668355, 2668356, 2668357, 2668358, 2668359, 2668360, 2668361, 2668362, 2668363, 2668364, 2668365, 2668366, 2668367, 2668369, 2668370, 2668371, 2668372, 2668373, 2668374, 2726301, 2726302, 2726303, 2726304, 2726305, 2726306, 2726307, 2726310, 2849439, 2849440, 2849441, 2849442, 2849443, 2849444, 2849445, 2849446, 2849447, 2849448, 3552579, 3552580, 3552581, 3552582, 3552583, 3552584, 3552585, 3552586, 3552587, 3552588, 3782921, 3782922, 3782923, 3782924, 3782925, 3782926, 3782927, 3782928, 3782929, 3782930, 4016867, 4053339, 4053340, 4053341, 4053342, 4053343, 4053344, 4053345, 4053346, 4053347, 4082555, 4082556, 4082557, 4082558, 4082559, 4082560, 4082561, 4082562, 4143354, 4143355, 4143356, 4143357, 4143358, 4175257, 4175258, 4175259, 4175260, 4175261, 4175262, 4175263, 4175264, 4199377, 4199378, 4199379, 4199380, 4199381, 4199382, 4213442, 4213443, 4213444, 4213445, 4213446, 4213447, 4213448, 4213449, 4213450, 4213451, 4213452, 4213453, 4237448, 4237449, 4237450, 4237451, 4237452, 4237453, 4237454, 4237455, 4237456, 4237457, 4320163, 4320164, 4320165, 4320166, 4320167, 4320168, 4320169, 4320170, 4320171, 4320172, 4367455, 4367456, 4367457, 4367458, 4367459, 4367460, 4367461, 4367462, 4367463, 4367464, 4382090, 4382091, 4382092, 4382093, 4382094, 4382095, 4382096, 4382097, 4382098, 4382099, 4399635, 4399636, 4399637, 4399638, 4399639, 4399640, 4399641, 4399642, 4399643, 4399644, 4728943, 4728944, 4728945, 4734555, 4734556, 4734557, 4734558, 4734559, 4734560, 4734561, 4802599, 4802600, 4802601, 4802602, 4802603, 4802604, 4802605, 4802606, 4802607, 4802608, 4802609, 4802610, 4802611, 4802612, 4802613, 4846019, 4846020, 4846021, 4846022, 4846023, 4846024, 4846025, 4846026, 4856025, 4856026, 4856027, 4856028, 4856029, 4856030, 4856031, 4856032, 4856033, 4856034, 4856035, 4856036, 4856037, 4856038, 4856039, 4856040, 4856041, 4856042, 4856043, 4856044, 4876881, 4917323, 4917324, 4917326, 4917327, 4917328, 4917329, 4917330, 4917331, 4917332, 4917333, 4938613, 4938615, 4938617, 4938618, 4938619, 4938620, 4938621, 4938622, 4938623, 4938624, 4938625, 4938628, 4938629, 4938631, 4938632, 4938635, 4938636, 4938637, 4938639, 4948123, 4948124, 4948125, 4948126, 4948127, 4948128, 4948129, 4948130, 4948131, 4948132, 4948133, 4948134, 4948135, 4948136, 4948137, 4948138, 4948139, 4948140, 4948141, 4948142, 4948143, 4948144, 4948145, 4948146, 4948147, 4948148, 4948149, 4948150, 4948151, 4948152, 4948153, 4948154, 4948155, 4948156, 4948157, 4948158, 4948159, 4948160, 4948161, 4948162, 4952615, 4952616, 4952617, 4952618, 4952619, 4952620, 4952621, 4952622, 4952623, 4952624, 4952625, 4952626, 4952627, 4952628, 4952629, 4952630, 4952631, 4952632, 4952633, 4952634); 204-12-13 (10885862047908; 0786075, 0786076, 0786077, 0786078, 0786079, 0786080, 0856858, 0856859, 0856860, 0856861, 0856862, 0856863, 0856866, 0856867, 0856868, 0856869, 0928734, 0928735, 0928736, 0965858, 0965859, 0965860, 0965861, 0965862, 0965863, 0965864, 0965865, 0965867, 1056465, 1056466, 1056468, 1056469, 1095852, 1095853, 1095855, 1147284, 1147285, 1147286, 1147287, 1221001, 1221002, 1221003, 1221004, 1221005, 1221006, 1221007, 1221008, 1221009, 1221010, 1272679, 1324214, 1324215, 1324216, 1324217, 1324218, 1324219, 1324220, 1324221, 1324222, 1324223, 1324224, 1324225, 1324226, 1324227, 1371305, 1371306, 1371307, 1371308, 1371309, 1371310, 1371311, 1371312, 1371313, 1434119, 1434120, 1434121, 1434122, 1434123, 1439857, 1439858, 1439859, 1439860, 1439861, 1567694, 1567695, 1567696, 1567697, 1567699, 1567700, 1567701, 1567702, 1595186, 1595187, 1595188, 1595189, 1595190, 1595191, 1595192, 1595193, 1595194, 1595195, 1654384, 1654385, 1654386, 1654387, 1654388, 1654390, 1654391, 1654392, 1654393, 1654394, 1654395, 1654396, 1654397, 1654398, 1654399, 1654400, 1654401, 1654402, 1742763, 1742764, 1742765, 1742766, 1742767, 1742768, 1742769, 1799279, 1799280, 1799281, 1799282, 1799283, 1799284, 1799285, 1810263, 1810264, 1810265, 1810266, 1810267, 1810268, 1810269, 1810270, 1810271, 1822723, 1822724, 1822725, 1822726, 1822727, 1822728, 1865358, 1865359, 1865360, 1865361, 1865362, 1865364, 2159166, 2159167, 2259363, 2259364, 2259365, 2259366, 2259367, 2259368, 2370381, 2370382, 2370383, 2370384, 2370385, 2370387, 2370388, 2370389, 2370390, 2370391, 2370392, 2370393, 2370394, 2370395, 2370396, 2370397, 2370398, 2370399, 2370400, 2414611, 2414612, 2414613, 2414614, 2414615, 2414616, 2414617, 2414618, 2414619, 2414620, 2414621, 2414622, 2414623, 2414624, 2414625, 2414626, 2414627, 2414628, 2414629, 2414630, 2415648, 2415649, 2415650, 2415652, 2415653, 2415654, 2415655, 2415656, 2415657, 2415658, 2415659, 2415660, 2415661, 2415662, 2415663, 2415664, 2415666, 2708659, 2708660, 2708661, 2708662, 2708663, 2708664, 2708665, 2708666, 2708667, 2708668, 2708669, 2708670, 2708671, 2708672, 2708673, 2708674, 2708675, 2708676, 2781285, 4084294, 4084295, 4084296, 4084297, 4084298, 4084299, 4084300, 4084301, 4084302, 4084303, 4145161, 4145162, 4145163, 4145164, 4145165, 4158336, 4158337, 4158338, 4158339, 4158340, 4158341, 4158342, 4158343, 4158344, 4158345, 4158346, 4158347, 4158348, 4158349, 4234201, 4234202, 4234203, 4234204, 4234205, 4234206, 4384046, 4384047, 4384048, 4384049, 4384050, 4384051, 4384052, 4384053, 4384054, 4384055, 4910357, 4910358, 4910359, 4945388, 4945389, 5005391, 5005392, 5005393, 5005395, 5005396); 204-13-09 (10885862047946; 0738085, 0738086, 0738087, 0738088, 0738089, 0738090, 0738091, 0873178, 0873179, 0880267, 0880268, 0880269, 0880277, 0888706, 0888707, 0888708, 0888709, 0888710, 0903244, 0903245, 0903246, 0903247, 0923881, 0923883, 0923884, 0923885, 0957090, 0957091, 0957092, 0957093, 0957094, 0957095, 0957096, 0957097, 0957098, 0957099, 1023430, 1023431, 1023432, 1023433, 1023434, 1023435, 1023436, 1023437, 1023438, 1023439, 1045915, 1045916, 1045917, 1045918, 1045919, 1080068, 1080069, 1080070, 1080071, 1080072, 1084267, 1084268, 1084269, 1084270, 1084271, 1121617, 1121618, 1121619, 1121620, 1121621, 1124480, 1124481, 1124482, 1124483, 1212143, 1212144, 1212145, 1212916, 1212917, 1212918, 1212919, 1212920, 1220986, 1220987, 1220988, 1220989, 1220990, 1229011, 1229012, 1229013, 1229014, 1229015, 1229016, 1229017, 1229018, 1229019, 1255862, 1255863, 1255864, 1255865, 1255866, 1255867, 1255868, 1255869, 1255870, 1255871, 1302619, 1302620, 1302621, 1302622, 1302623, 1302624, 1302625, 1302626, 1302627, 1302628, 1302629, 1302630, 1302631, 1302632, 1302633, 1302634, 1302635, 1302636, 1302637, 1302638, 1353923, 1353924, 1353925, 1353926, 1353927, 1366615, 1366616, 1366617, 1366618, 1366619, 1366620, 1366621, 1366622, 1366623, 1366624, 1406356, 1406357, 1406358, 1406359, 1406360, 1480795, 1480796, 1480797, 1480798, 1522625, 1522626, 1534797, 1534798, 1534799, 1534800, 1576758, 1576759, 1576760, 1576761, 1576762, 1576763, 1576764, 1576765, 1576766, 1576767, 1619204, 1619205, 1619206, 1619207, 1619208, 1619209, 1619210, 1619211, 1619212, 1619213, 1661516, 1661517, 1661518, 1661519, 1661520, 1661521, 1661522, 1661523, 1661524, 1661525, 1661526, 1661527, 1661528, 1661529, 1661530, 1673843, 1673844, 1673845, 1673846, 1673847, 1673848, 1673849, 1673850, 1673851, 1673852, 1693776, 1693777, 1693778, 1693779, 1693780, 1693781, 1693782, 1693783, 1693784, 1693785, 1791221, 1791222, 1791223, 1791224, 1791225, 1791226, 1791227, 1791228, 1791229, 1799118, 1799119, 1799120, 1799121, 1799122, 1799123, 1799124, 1799125, 1799126, 1799127, 1811412, 1811413, 1811414, 1811415, 1811416, 1811417, 1811418, 1811419, 1817798, 1817799, 1817800, 1817801, 1817802, 1817803, 1831528, 1831529, 1831530, 1831531, 1831533, 1831534, 1831535, 1831536, 1899925, 1899926, 1899927, 1899928, 1899929, 2255018, 2255019, 2255020, 2255021, 2255971, 2255972, 2255973, 2255974, 2255975, 2255976, 2306080, 2306081, 2306082, 2306083, 2306084, 2306085, 2306086, 2306087, 2306088, 2306089, 2306090, 2306091, 2306092, 2306093, 2370401, 2370402, 2370403, 2370404, 2370405, 2370406, 2370407, 2370408, 2370409, 2370410, 2398805, 2398806, 2398807, 2398808, 2398809, 2398810, 2398811, 2398812, 2398813, 2398814, 2398815, 2398816, 2398817, 2398818, 2398819, 2398820, 2398821, 2415553, 2415555, 2415556, 2415557, 2415558, 2415559, 2415560, 2415561, 2415562, 2454408, 2454409, 2454410, 2454411, 2454412, 2454413, 2454414, 2454415, 2454416, 2454417, 2509357, 2509358, 2509359, 2509360, 2551485, 2551486, 2551487, 2551488, 25
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0019-2022
Related Recalls
Brand Name: Equinoxe Ergo Product Name: Ergo Modular Impactor Handle Model/Catalog Number: 321-09-05
Exactech
Impactor handle may be missing cross-pin
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.