COVIDIEN

22 recalls • • Categories: Infusion Pumps, Surgical Instruments, Diagnostic Equipment, Other Medical Devices, Patient Monitors +1 more

Due to customer complaint regarding incorrect display box labeling.

Apr 15, 2026 Other Medical Devices Nationwide View Details →

Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.

Jan 30, 2026 Diagnostic Equipment Nationwide View Details →

Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting

Jun 23, 2025 Diagnostic Equipment Nationwide View Details →

Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.

May 21, 2025 Patient Monitors Nationwide View Details →

Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.

Sep 19, 2024 Surgical Instruments Nationwide View Details →

Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway

Jun 24, 2024 Patient Monitors Nationwide View Details →

One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.

Jan 25, 2024 Surgical Instruments View Details →

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

Feb 14, 2024 Diagnostic Equipment Nationwide View Details →

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

Feb 14, 2024 Diagnostic Equipment Nationwide View Details →

Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.

Feb 21, 2024 Other Medical Devices Nationwide View Details →

Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.

Dec 21, 2023 Other Medical Devices Nationwide View Details →

Sofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 S-176 SOFSILK* 0 BLK 12X45CM PCT X24 S-182 SOFSILK* 5-0 BLK 12X45CM PCT X36 S-184 SOFSILK* 3-0 BLK 12X45CM PCT X36 S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36 S-245 SOFSILK* 2-0 BLK 12X60CM PCT X36 S-246 SOFSILK* 0 BLK 6X60CM PCT X36 S-2782K SOFSILK* 4-0 WHI 45CM SS24 DA S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36 S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36 S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36 S-316 SOFSILK* 0 BLK 12X75CM PCT X24 S-317 SOFSILK* 1 BLK 12X75CM PCT X24 S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36 S-606 SOFSILK* 0 BLK 2X150CM PCT X36 S606-12 SOFSILK* 0 BLK 2X150CM PCT X12 S-610 SOFSILK* 5 BLK 2X150CM PCT X12 SOFSILK 2-0 30 BK CV-24 GS453-2 SOFSLK 4-0 24 BK CV-15 DA VS581-2 ***Added 4/1/24*** CS-792 SOFSILK* 0 BLK 100CM GCC90 X36 S-1172 SOFSILK* 6-0 BLK 45CM C1 X36 S-1735 SOFSILK* 4-0 BLK 30CM C1 DA X12 S-1750K SOFSILK* 6-0 BLK 45CM HE1 DAX12 S-1768K SOFSILK* 7-0 BLK 45CM HE6 DAX12 S-1789K SOFSILK* 4-0 BLK 45CM HE3 X12 S-187 SOFSILK* 1 BLK 6X45CM PCT X36 S-243 SOFSILK* 4-0 BLK 12X60CM PCT X36 S-2780K SOFSILK* 6-0 BLK 45CM SS-24DA S-282 SOFSILK* 5 BLK 3X60CM PCT X36 S-318 SOFSILK* 2 BLK 12X75CM PCT X24 S-608 SOFSILK* 2 BLK 2X150CM PCT X36 VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36 VS872-2 SOFSILK 3-0 30 BK CV-23 VS872-2 SS-1694G SOFSILK 4/0 BLK 45CM P24 X12 SS-5641 SOFSILK* 4-0 BLK 45CM P13 X36 SS-5649G SOFSILK* 6-0 BLK 45CM P13 X12 SS-645 SOFSILK* 4-0 BLK 75CM C12 X36 SS-675 SOFSILK* 3-0 BLK 75CM C16 X36 SS-681 SOFSILK* 6-0 BLK 45CM C12 X36 VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36 XX-5280 SOFSILK* 0 BLK 150CM PCT X36 -- Ti-Cron Coated Braided Polyester Sutures: 3087-31 TICRON* 4-0 BLU 75CM CV-331DA 3226-31 TICRON* 4-0 BLU 60CM CV330 DA 88863026-71 TICRON* 1 BLU 75CM HOS12X36 88863070-51 TICRON* 2-0 BLU 75CM C15X36 88863087-31 TICRON* 4-0 BLU 75CM CV33DA 88863092-71 TICRON* 1 BLU 75CM KV37X36 88863147-83 TICRON 2 BLU 5X75CM KV37X12 88863185-41 TICRON* 3-0 BLU 75CM CV305 88863186-31 TICRON* 4-0 BLU 75CM CV331 88863226-31 TICRON* 4-0 BLU 60CM CV330DA 88863226-41 TICRON* 3-0 BLU 60CM CV330DA 88863229-31 TICRON* 4-0 BLU 75CM CV307DA 88863271-31 TICRON* 4-0 BLU 90CM CV331DA 88863309-71 TICRON* 1 BLU 75CM GS11X36 -- Monosof Monofilament Nylon Sutures: N-63 MONOSOF* 3-0 BLK 12X45CM PCT X36 SN-3695 MONOSOF* 5-0 BLK 45CM P10 X12 -- Steel Monofilament Stainless Steel Sutures 88862224-49 STEEL 7 4X45CM SCC X12

Class I - Dangerous

Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.

Dec 21, 2023 Other Medical Devices Nationwide View Details →

Defective video laryngoscopes that did not pass quality tests established for the product, that were designated for destruction, were stolen, and subsequently distributed for sale. Defective device use may result in failed laryngoscopy and/or intubation.

Sep 28, 2023 Other Medical Devices View Details →

Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.

Aug 22, 2023 Other Medical Devices View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Jun 28, 2023 Surgical Instruments Nationwide View Details →

Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.

Jun 2, 2023 Other Medical Devices Nationwide View Details →

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY CUFFED TRACH CAN; 10CN10R, 10.0MM ADT FLEX TRACH W TG CUFF; 10UN10A, 10.0MM UNCUFF TRACH TUBE X1; 10UN10H, 10.0MM SHILEY CUFFLESS TRACH C; 10UN10R, 10.0MM ADT FLEX TRACH CUFFLESS; 4CN65A, 6.5MM TRACH TUBE W TG CUFF X1; 4CN65H, 6.5MM SHILEY CUFFED TRACH CANN; 4CN65R, 6.5MM ADT FLEX TRACH W TG CUFF X1; 4UN65A, 6.5MM UNCUFF TRACH TUBE X1; 4UN65H, 6.5MM SHILEY CUFFLESS TRACH CA; 4UN65R, 6.5MM ADT FLEX TRACH CUFFLESSX1; 5CN70A, 7.0MM TRACH TUBE W TG CUFF X1; 5CN70H, 7.0MM SHILEY CUFFED TRACH CANN; 5CN70R, 7.0MM ADT FLEX TRACH W TG CUFF X1; 5UN70A, 7.0MM UNCUFF TRACH TUBE X1; 5UN70H, 7.0MM SHILEY CUFFLESS TRACH CA; 5UN70R, 7.0MM ADT FLEX TRACH CUFFLESSX1; 6CN75A, 7.5MM TRACH TUBE W TG CUFF X1; 6CN75H, 7.5MM SHILEY CUFFED TRACH CANN; 6CN75R, 7.5MM ADT FLEX TRACH W TG CUFF X1; 6UN75H, 7.5MM SHILEY CUFFLESS TRACH CA; 6UN75R, 7.5MM ADT FLEX TRACH CUFFLESSX1; 7CN80A, 8.0MM TRACH TUBE W TG CUFF X1; 7CN80H, 8.0MM SHILEY CUFFED TRACH CANN; 7CN80R, 8.0MM ADT FLEX TRACH W TG CUFF X1; 7UN80A, 8.0MM UNCUFF TRACH TUBE X1; 7UN80H, 8.0MM SHILEY CUFFLESS TRACH CA; 7UN80R, 8.0MM ADT FLEX TRACH CUFFLESSX1; 8CN85A, 8.5MM TRACH TUBE W TG CUFF X1; 8CN85H, 8.5MM SHILEY CUFFED TRACH CANN; 8CN85R, 8.5MM ADT FLEX TRACH W TG CUFF X1; 8UN85A, 8.5MM UNCUFF TRACH TUBE X1; 8UN85H, 8.5MM SHILEY CUFFLESS TRACH CA; 8UN85R, 8.5MM ADT FLEX TRACH CUFFLESSX1; 9CN90H, 9.0MM SHILEY CUFFED TRACH CANN; 9CN90R, 9.0MM ADT FLEX TRACH W TG CUFF X1; 9UN90H, 9.0MM SHILEY CUFFLESS TRACH CA; 9UN90R, 9.0MM ADT FLEX TRACH CUFFLESSX1

Class I - Dangerous

A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

Mar 24, 2023 Other Medical Devices Nationwide View Details →

Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.

Mar 1, 2023 Other Medical Devices Nationwide View Details →

Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.

Jan 25, 2023 Other Medical Devices Nationwide View Details →

Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

Dec 2, 2022 Other Medical Devices Nationwide View Details →

Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

Dec 2, 2022 Other Medical Devices Nationwide View Details →

Sofsilk Coated Braided Silk Suture Product Description CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 CS211 CS-490 SOFSILK* 2-0 BLK 75CM GS11 X36 CS490 CS-562 SOFSILK* 2-0 BLK 75CM GS23 X36 CS562 GS-823 SOFSILK* 2-0 BLK 75CM V30 X36 GS823 GS-835 SOFSILK* 1 BLK 75CM V20 X36 GS835 S-1172 SOFSILK* 6-0 BLK 45CM C1 X36 S1172 S-1173 SOFSILK* 5-0 BLK 45CM C1 X36 S1173 S-1174 SOFSILK* 4-0 BLK 45CM C1 X36 S1174 S-1750K SOFSILK* 6-0 BLK 45CM HE1 DAX12 S1750K S-176 SOFSILK* 0 BLK 12X45CM PCT X24 S176 S-176 SOFSILK* 0 BLK 12X45CM PCT X24 S176 S-1768K SOFSILK* 7-0 BLK 45CM HE6 DAX12 S1768K S-182 SOFSILK* 5-0 BLK 12X45CM PCT X36 S182 S-183 SOFSILK* 4-0 BLK 12X45CM PCT X36 S183 S-184 SOFSILK* 3-0 BLK 12X45CM PCT X36 S184 S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36 S185 S-187 SOFSILK* 1 BLK 6X45CM PCT X36 S187 S-206 SOFSILK* 2 BLK 100CM STDLTH X36 S206 S-243 SOFSILK* 4-0 BLK 12X60CM PCT X36 S243 S-244 SOFSILK* 3-0 BLK 12X60CM PCT X36 S244 S-245 SOFSILK* 2-0 BLK 12X60CM PCT X36 S245 S-246 SOFSILK* 0 BLK 6X60CM PCT X36 S246 S-2782K SOFSILK* 4-0 WHI 45CM SS24 DA S2782K S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36 S303 S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36 S304 S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36 S305 S-316 SOFSILK* 0 BLK 12X75CM PCT X24 S316 S-317 SOFSILK* 1 BLK 12X75CM PCT X24 S317 S-318 SOFSILK* 2 BLK 12X75CM PCT X24 S318 S-346 SOFSILK* 0 BLK 12X60CM PCT X24 S346 S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36 S605 S-607 SOFSILK* 1 BLK 2X150CM PCT X36 S607 S-608 SOFSILK* 2 BLK 2X150CM PCT X36 S608 SS-523 SOFSILK* 2-0 BLK 75CM SC1 X36 SS523 SS-525 SOFSILK* 1 BLK 75CM SC1 X36 SS525 SS-5641 SOFSILK* 4-0 BLK 45CM P13 X36 SS5641 SS-5641G SOFSILK* 4-0 BLK 45CM P13 X12 SS5641G SS-5678 SOFSILK* 5-0 BLK 45CM P11 X36 SS5678 SS-5684 SOFSILK* 3-0 BLK 45CM P14 X36 SS5684 SS-623 SOFSILK* 2-0 BLK 75CM SC2 X36 SS623 SS-645 SOFSILK* 4-0 BLK 75CM C12 X36 SS645 SS-653 SOFSILK* 4-0 BLK 75CM C13 X36 SS653 SS-673 SOFSILK* 5-0 BLK 75CM C12 X36 SS673 SS-683G SOFSILK* 4-0 BLK 45CM C13 X12 SS683G SS-685G SOFSILK* 2-0 BLK 45CM C15 X12 SS685G SS-686 SOFSILK* 4-0 BLK 45CM C12 X12 SS686 SS-689 SOFSILK* 1 BLK 75CM C16 X36 SS689 SS-745 SOFSILK* 3-0 BLK 75CM C12 X36 SS745 SS-786 SOFSILK* 0 BLK 75CM C14 X36 SS786 SS-787 SOFSILK* 1 BLK 75CM C14 X36 SS787 VS-533 SOFSILK* 2-0 BLK 90CM V20 DAX36 VS533 VS-845 SOFSILK* 1 BLK 75CM V26 X36 VS845 VS-870 SOFSILK* 5-0 BLK 75CM CV23 X36 VS870 VS-871 SOFSILK* 4-0 BLK 75CM CV23 X36 VS871 SOFSILK 4-0 30 BK CV-23 VS871-2 VS8712 VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36 VS872 SOFSILK 3-0 30 BK CV-23 VS872-2 VS8722 VS-882 SOFSILK* 3-0 BLK 75CM CV15 X36 VS882 SOFSILK 5-0 30 BK CV-11 VS890-2 VS8902 XX-5278 SOFSILK* 3-0 BLK 150CM PCT X36 XX5278 XX-5280 SOFSILK* 0 BLK 150CM PCT X36 XX5280 XX-5281 SOFSILK* 1 BLK 150CM PCT X36 XX5281 XX-5284 SOFSILK* 4 BLK 150CM PCT X36 XX5284

Class I - Dangerous

Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

Dec 2, 2022 Other Medical Devices Nationwide View Details →

Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

Dec 2, 2022 Other Medical Devices Nationwide View Details →

Surgipro II Monofilament Polypropylene Sutures Product Description Model Number VP-421X SURGIPRO*II 4-0 90CM V20 DA7P VP421X VP-435-X SURGIPRO II 4-0 90CM KV5 DA7P VP435X VP-541X SURGIPRO*II 4-0 BLU 120CM V20DA VP541X VP-542-X SURGIPRO II 3-0 BLU 120CM V20DA VP542X VP-735-X SURGIPRO*II 7-0 60CM MV1758DA VP735X VP735X-12 SURGIPRO*II 7-0 60CM MV1758 DA VP735X12 VP-738-X SURGIPRO*II 8-0BLU 60CM CV351DA VP738X VP-744-X SURGIPRO*II 8-0 45CM MV1758DA VP744X VP-745-X SURGIPRO*II 8-0 60CM MV1758DA VP745X VP-747-X SURGIPRO*II 7-0 75CM MV1758DA VP747X VP-853 SURGIPRO* 2-0 BLU 120CM V26 DA VP853 VP-875-X SURGIPRO*II 7-0 75CM MV1759DA VP875X VP-900-X SURGIPRO*II 8-0 45CM MV1355DA VP900X VP-902-X SURGIPRO*II 8-0 60CM MV1355DA VP902X VP-421X SURGIPRO*II 4-0 90CM V20 DA7P VP421X VP-435-X SURGIPRO II 4-0 90CM KV5 DA7P VP435X VP-541X SURGIPRO*II 4-0 BLU 120CM V20DA VP541X VP-542-X SURGIPRO II 3-0 BLU 120CM V20DA VP542X VP-735-X SURGIPRO*II 7-0 60CM MV1758DA VP735X VP735X-12 SURGIPRO*II 7-0 60CM MV1758 DA VP735X12 VP-738-X SURGIPRO*II 8-0BLU 60CM CV351DA VP738X VP-744-X SURGIPRO*II 8-0 45CM MV1758DA VP744X VP-745-X SURGIPRO*II 8-0 60CM MV1758DA VP745X VP-747-X SURGIPRO*II 7-0 75CM MV1758DA VP747X VP-853 SURGIPRO* 2-0 BLU 120CM V26 DA VP853 VP-875-X SURGIPRO*II 7-0 75CM MV1759DA VP875X VP-900-X SURGIPRO*II 8-0 45CM MV1355DA VP900X VP-902-X SURGIPRO*II 8-0 60CM MV1355DA VP902X12 VP-904-X SURGIPRO*II 7-0 BLU 60CM MV1355 VP904X VPF-735-X SURGIPRO*II 7-0 60CM MVF1758DA VPF735X XX-3300 SURGIPRO* II 5-0 BLU 60CM CV1DA XX3300 XX-5224 SURGIPRO*II 5-0 BLU 90CM CV25DA XX5224

Class I - Dangerous

Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

Dec 2, 2022 Other Medical Devices Nationwide View Details →

Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

Dec 2, 2022 Other Medical Devices Nationwide View Details →

Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

Dec 2, 2022 Diagnostic Equipment Nationwide View Details →

There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

Dec 7, 2022 Surgical Instruments Nationwide View Details →

Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios.

Sep 28, 2022 Other Medical Devices Nationwide View Details →

Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios.

Sep 28, 2022 Other Medical Devices Nationwide View Details →