Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 10884521782211; Serial Number: 518503
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Covidien, LLC
Reason for Recall:
Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

Product Codes/Lot Numbers:

UDI-DI: 10884521782211; Serial Number: 518503

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2008-2023

Related Recalls

Due to customer complaint regarding incorrect display box labeling.

Apr 15, 2026 Other Medical Devices Nationwide View Details →

Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.

Jan 30, 2026 Diagnostic Equipment Nationwide View Details →