Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI (GTIN): 20884521543598 Lots: N2D0002Y, N2D0004Y, N2D0195Y
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Covidien, LP
- Reason for Recall:
- Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT
Product Codes/Lot Numbers:
UDI-DI (GTIN): 20884521543598 Lots: N2D0002Y, N2D0004Y, N2D0195Y
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1228-2023
Related Recalls
Class I - Dangerous
Temperature probe devices lack FDA clearance.
Due to customer complaint regarding incorrect display box labeling.
Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.