Customer unable to obtain measurements due to the device seeing no flow or irrelevant flow condition during dialysis procedures. The user saw a zero or a very low flow value displayed on HD03 and/or no measurements produced in other modes (recirculation; access flow; or cardiac output).

Feb 17, 2015 Diagnostic Equipment Nationwide View Details →

Potential safety issue due to improper assembly associated with the power cords of T2100 and T2000 Treadmills.

Feb 16, 2015 Diagnostic Equipment View Details →

For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.

Feb 17, 2015 Other Medical Devices Nationwide View Details →

BD Posiflush Saline Flush Syringe 10 ml inlcuded in the following device kits: Bard Poly Midline 3F Full Tray (Single-Lumen) (with Microintroducer), Bard Poly Midline 3F Maximal Barrier Tray (Single-Lumen), Bard Poly Midline 4F Maximal Barrier Tray (Single-Lumen) (with Microintroducer), Bard Poly Midline BARD MYPICC KIT (4F) (Single Lumen), Groshong NXT Catheter 5F Maximal Barrier Tray (Dual-Lumen) (with MicroIntroducer) , Groshong NXT BARD MYPICC KIT (5F) (Dual-Lumen) (45cm), PowerPICC Catheter 5F Full Tray (Single-Lumen), PowerPICC Catheter 5F Maximal Barrier Tray (Single-Lumen), Poly Per-Q-Cath Catheter 3F Single-Lumen Full Tray (with Microintroducer), Poly Per-Q-Cath Catheter 4F Single-Lumen Full Tray (with Mircrointroducer), Poly Per-Q-Cath 5F Dual-Lumen Full Tray (with Microintroducer), Poly Per-Q-Cath Catheter 3F Full Tray (Single-Lumen) (with Microintroducer) (60cm), Poly Per-Q-Cath Catheter 4F Full Tray (Single-Lumen) (with Microintroducer) (60cm), Poly Per-Q-Cath Catheter 5F Full Tray (Dual-Lumen) (with Microintroducer) (60cm), Groshong NXT ClearVue 4F Catheter Full Tray (Single-Lumen) (60cm), Groshong NXT ClearVue 4F Catheter Full Tray (Single-Lumen) (60cm) (with Microintroducer), Groshong NXT ClearVue Catheter 4F Maximal Barrier Tray (Single-Lumen), PowerPICC 5F Dual-Lumen Polyurethane Catheter with MicroEZ Microintroducer Nurse Full Tray (with Lidocaine), PowerPICC Catheter 5F Maximal Barrier Tray with Microintroducer (Dual-Lumen), PowerPICC FT Catheter 5F Maximal Barrier Tray Tray (Dual-Lumen), PowerPICC Bard MYPICC Kit 5F (Dual-Lumen)(with Nitinol Guidewire)(70cm), PowerPICC Catheter 5F Maximal Barrier Tray (Dual-Lumen) (55cm) (with Microintroducer), PowerPICC Catheter 5F Full Tray (Dual-Lumen) (70cm Nitinol Guidewire) (with Microintroducer), PowerPICC Bard MYPICC Kit 5F (Dual-Lumen) RN, PowerPICC BARD MYPICC KIT (5F) (Dual-Lumen), PowerPICC Catheter 6F Full Tray (Triple-Lumen)(with Microintroducer) (with Lidocaine), PowerPICC HF Catheter 5F Full Tray (Triple-Lumen), PowerPICC HF Catheter 5F Maximal Barrier Tray (Triple-Lumen), PowerPICC Catheter 6F (Triple-Lumen) Maximal Barrier Tray, PowerPICC SOLO HF Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray (Triple-Lumen), PowerPICC HF Catheter with Sherlock Tip Location System (TLS) Stylet 5F Maximal Barrier Tray (Triple-Lumen), PowerPICC HF Bard MYPICC Kit 5F (Triple-Lumen), PowerPICC HF BARD MYPICC KIT (5F) (Triple-Lumen), PowerPICC HF Catheter Maximal Barrier Tray (5F) (Triple-Lumen) (55cm), Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet 4F Full Tray (Single-Lumen), Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet 4F Full Tray (Single-Lumen) (with Microintroducer), Groshong NXT Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray (Dual-Lumen) (45cm), Groshong NXT Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray (Dual-Lumen) (45cm) (with Microintroducer), Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet 4F Maximal Barrier Tray (Single-Lumen), Groshong NXT Catheter with Sherlock Tip Location System (TLS) Stylet 5F Maximal Barrier Tray (Dual-Lumen) (45cm), Groshong NXT Sherlock II BARD MYPICC KIT (4F) (Single-Lumen) (TLS), Poly Per-Q-Cath Catheter with Sherlock Tip Location System (TLS) Stylet 3F Full Tray (with Microintroducer), Poly Per-Q-Cath Catheter with Sherlock Tip Location System (TLS) Stylet 4F Full Tray (with Microintroducer), Poly Per-Q-Cath Catheter with Sherlock Tip Location System (TLS) Stylet 4F Full Tray, Poly Per-Q-Cath Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray, Poly Per-Q-Cath Sherlock II Bard MYPICC KIT (4F) (Sinlge-Lumen) (TLS), Poly Per-Q-Cath Sherlock II BARD MYPICC KIT (4F) (Single-Lumen) (TLS), PowerPICC Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray (Single-Lumen) (with Microintroducer), PowerPICC 5F Dual-Lumen Polyurethane Catheter with Sherlock Tip L

Bard Access Systems

🏥 Medical Devices Class I - Dangerous

Some unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.

Aug 8, 2014 Surgical Instruments Nationwide View Details →

A configuration update needs to be done on TactiSys to appropriately recognize all TactiCath catheters.

Nov 25, 2014 Surgical Instruments Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.

Jan 29, 2015 Other Medical Devices Nationwide View Details →

BD Posiflush Saline Flush Syringe 10 ml inlcuded in the following device kits: SafeStep Port Access Kit 22G x 0.75in , SafeStep Port Access Kit 22G x 0.75in with Y-Site, SafeStep Port Access Kit 22G x 1 in , SafeStep Port Access Kit 20G x 0.75in , SafeStep Port Access Kit 20G x 0.75in with Y-Site, SafeStep Port Access Kit 20G x 1 in , SafeStep Port Access Kit 20G x 1 in with Y-Site, SafeStep Port Access Kit 19G x 0.75in , SafeStep Port Access Kit 19G x 0.75in with Y-Site, SafeStep Port Access Kit 19G x 1 in , SafeStep Port Access Kit 19G x 1 in with Y-Site, SafeStep Port Access Kit 20G x 1.5in , SafeStep Port Access Kit 20G x 1.5in with Y-Site, MiniLoc Port Access Kit 20G x 1 in, MiniLoc Port Access Kit 20G x 0.75in, MiniLoc Port Access Kit 22G x 1 in, MiniLoc Port Access Kit 22G x 0.75in, MiniLoc Port Access Kit 19G x 1 in with Y-Site, MiniLoc Port Access Kit 20G x 1 in with Y-Site, MiniLoc Port Access Kit 20G x 0.75in with Y-Site, MiniLoc Port Access Kit 22G x 1.5in with Y-Site, MiniLoc Port Access Kit 22G x 0.75in with Y-Site, PowerLoc Max Port Access Kit 19G x 1.0 in with Y-Site, PowerLoc Max Port Access Kit 19G x 0.75in with Y-Site, PowerLoc Max Port Access Kit 20G x 0.75in with Y-Site, PowerLoc Max Port Access Kit 19G x 1.0 in , PowerLoc Max Port Access Kit 20G x 1.0 in , PowerLoc Max Port Access Kit 20G x 0.75in , PowerLoc Max Port Access Kit 22G x 1.0 in , PowerLoc Max Port Access Kit 22G x 0.75in The syringe is used to pre-flush the catheter prior to placement and may also be used to verify patency post-placement.

Bard Access Systems

🏥 Medical Devices Class I - Dangerous

Some unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.

Aug 8, 2014 Surgical Instruments Nationwide View Details →

Complaints have been confirmed on failed proficiency testing due to serum samples recovering too low when using some N Latex IgM reagent kit lots 44018, 44019 and 44042.

Jan 15, 2015 Diagnostic Equipment View Details →

GE Healthcare has recently become aware of a potential safety issue due to a collimator falling off a VG system during a Quality Control check, while the detectors were rotating. No injuries have been reported to date

Mar 5, 2015 Diagnostic Equipment Nationwide View Details →