ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model number: 08139789, serial numbers: 5419, 5784, 5830
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- There may be an existing dark current phenomenon on ARTISTE LINAC in combination with IMRT or mARC treatments using unflat beams. Software issue.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.
Product Codes/Lot Numbers:
Model number: 08139789, serial numbers: 5419, 5784, 5830
Distribution:
Distributed in: US, UT, WI, NY
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0984-2015
Related Recalls
MAMMOMAT Inspiration;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Fusion;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Revelation;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.