Elekta Precise Treatment Table
Class I - DangerousWhat Should You Do?
- Check if you have this product: Precise Treatment Table Serial Numbers: (124001-125803, 125807-880, 125883-899, 125902-917, 125920-949, 125952-963, 125966-1256453, 126158-170, 126174-208, 126212-230, 126232-261, 126269-281, 126293-309,123318-330, 126338-126373, 126375-397, 126400-133999, 213000+)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Elekta, Inc.
- Reason for Recall:
- It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Elekta Precise Treatment Table
Product Codes/Lot Numbers:
Precise Treatment Table Serial Numbers: (124001-125803, 125807-880, 125883-899, 125902-917, 125920-949, 125952-963, 125966-1256453, 126158-170, 126174-208, 126212-230, 126232-261, 126269-281, 126293-309,123318-330, 126338-126373, 126375-397, 126400-133999, 213000+)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1238-2015
Related Recalls
Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
Due to adapter not being properly aligned or locked in place.