The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed.
Infusion Pumps
🏥 Medical Devices • 4,262 recalls
The firm is recalling ToggleLocs and Ziploops manufactured in May and July of 2014 as they were not processed following a standard work process, which resulted in complaints of sterile packaging not being sealed.
Product is mislabeled as Double Swivel Connector with Suction Port and is actually Double Swivel Connector without Suction port. If the incorrect component is not identified prior to clinical use, the incorrect connector would be replaced during the case without risk to the patient and could result in a short delay of the procedure.
If the patient sits on either wing, not squarely on the center section of the Stretcher Lift, tipping and injury to the operator or patient could result. The device manual was updated with an additional warning statement.
GE Healthcare has become aware of a potential safety issue involving MRI systems due to software versions not being maintained properly at some sites.
Becton Dickinson received an elevated number of complaints for blood or infusate/medication leakage due to tearing of the silicone septum.
Alaris PC units, Model No. 8015. Infusion pump.
CareFusion 303
CareFusion is recalling the Alaris PC unit because of an error code. The error code may occur upon power on during the "Power-On Self Test" due to a keypad issue.
Siemens LANTIS Oncology Information System Servers; allows the radiation therapist to deliver treatment to patient using the MEVATRON and all available accessories
Siemens Medical Solutions USA
There is a potential safety risk when using LANTIS server software with operating systems with which it has not been validated or released which can lead to an incorrect treatment to the patient.
ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.
Siemens Medical Solutions USA
There may be an existing dark current phenomenon on ARTISTE LINAC in combination with IMRT or mARC treatments using unflat beams. Software issue.
Medstream 81" (206 cm) 20 drop Universal Administration Sets. Disposable IV infusion set with 3 year expiration date.
US Infusion Inc
Several complaints regarding the IV sets leaking at the Y site.
Cases of Cure Sleeve labeled with part number 4513, which fits 13mm tips, have inner boxes labeled with part number 4500 which fit 7mm and 8mm tips. There is a low risk associated with the use of the product as the sleeve may tear causing a potential for cross contamination.
KimVent Closed Suction System for Adults, Double Swivel Elbow, 12 Fr., Sterile, Distributed by Kimberly Clark Global Systems, LLC.
Halyard Health
Defect: A hole or crack was found which may cause air to leak.
TriaDyne Proventa Critical Care Therapy System. The TriaDyne Proventa Critical Care Therapy System provides a comprehensive system of pulmonary and skin care therapies for the critically ill, immobilized patient.
Arjo Hospital Equipment AB
The recalled devices labeling and instructions for use contain unapproved medical claims.
KimVent Closed Suction System for Adults, Double Swivel Elbow, 14 Fr. Sterile, Distributed by Kimberly Clark Global Systems, LLC.
Halyard Health
Defect: A hole or crack was found which may cause air to leak.
One lot of SMARTSET GHV Gentamicin Bone Cement is partially agglomerated . This may result in surgical delays while a replacement package of cement is retrieved and mixed.
Defect: A hole or crack was found which may cause air to leak.
RotoRest Delta Advanced Therapy System. The RotoRest Delta Advanced Kinetic Therapy System offers kinetic therapy for the most aggressive treatment of pulmonary complications.
Arjo Hospital Equipment AB
The recalled devices labeling and instructions for use contain unapproved medical claims.
Incorrect CT to ED when using the Monaco Image Statistics Tool.
One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.
Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.