RANDOX LABORATORIES

7 recalls • • Categories: Infusion Pumps, Other Medical Devices, Diagnostic Equipment

Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.

Apr 5, 2024 Infusion Pumps Nationwide View Details →

Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results.

Nov 29, 2023 Other Medical Devices Nationwide View Details →

CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

Nov 20, 2023 Other Medical Devices Nationwide View Details →

There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human Assayed Multi-Sera Level 2 (HN1530) for TBIC taget value and on IFU of Human Assayed Multi-Sera Level 3 (HE 1532) for Lipase target value.

Nov 13, 2023 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.

Oct 11, 2023 Other Medical Devices Nationwide View Details →

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Sep 6, 2023 Infusion Pumps Nationwide View Details →

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Sep 6, 2023 Infusion Pumps Nationwide View Details →

there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect, which may lead to misreporting patient results. Target and range values are correct, and risk is mitigated; most clinical labs will use the target and range values in clinical practice.

Aug 3, 2023 Other Medical Devices View Details →

As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.

Jul 19, 2023 Diagnostic Equipment View Details →

(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.

Jun 21, 2023 Diagnostic Equipment Nationwide View Details →

Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.

Jun 13, 2023 Other Medical Devices Nationwide View Details →

Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range

May 4, 2023 Diagnostic Equipment View Details →

Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range

May 4, 2023 Diagnostic Equipment View Details →

Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to an elevation in Quality Control and Patient results of up to +13% for lgA and +51% for lgM respectively, which may lead to a delay in running patient samples or erroneous elevated test results

Nov 1, 2022 Other Medical Devices Nationwide View Details →

Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to an elevation in Quality Control and Patient results of up to +13% for lgA and +51% for lgM respectively, which may lead to a delay in running patient samples or erroneous elevated test results

Nov 1, 2022 Other Medical Devices Nationwide View Details →

***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

Aug 1, 2022 Infusion Pumps Nationwide View Details →

A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.

Jun 9, 2022 Infusion Pumps Nationwide View Details →

A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.

Jun 9, 2022 Infusion Pumps Nationwide View Details →

Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.

Jun 22, 2022 Infusion Pumps Nationwide View Details →

Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.

Mar 25, 2022 Other Medical Devices View Details →

An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.

Aug 11, 2020 Surgical Instruments Nationwide View Details →

An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.

Aug 11, 2020 Surgical Instruments Nationwide View Details →

The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.

Aug 9, 2019 Diagnostic Equipment Nationwide View Details →

The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.

Aug 9, 2019 Diagnostic Equipment Nationwide View Details →

Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).

Jul 12, 2019 Diagnostic Equipment Nationwide View Details →

An error in the assignment of the control range pH in Blood Gas Control

Feb 13, 2019 Diagnostic Equipment View Details →

Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials.

Oct 2, 2018 Other Medical Devices View Details →

Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials.

Oct 2, 2018 Other Medical Devices View Details →

Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials.

Oct 2, 2018 Other Medical Devices View Details →

The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.

Jun 8, 2018 Diagnostic Equipment View Details →

The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains.

May 22, 2018 Other Medical Devices View Details →

The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains.

May 22, 2018 Other Medical Devices View Details →

The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains.

May 22, 2018 Other Medical Devices View Details →

Randox has now released further steps to avoid contamination of the Lipase Assay on Rx instruments.

Mar 2, 2018 Infusion Pumps Nationwide View Details →

Randox confirmed low calibration absorbance with HbA1c, Catalog Number HA3830, Batch 414553. The reagent may not continue to meet its specific performance characteristics throughout the shelf life of the product.

Jun 7, 2017 Surgical Instruments View Details →

According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a factor of 10. Instrument specific values are correct. This does not affect the performance or stability claims of the product. As control results will be outside of the quoted range this may result in a delay in reporting the sample test results. However as the correct instrument specific means are also included in the value sheet this delay is unlikely. Customers have been referred to their Medical Director for further advice

Apr 27, 2017 Diagnostic Equipment View Details →

An internal real-time stability monitoring program revealed decreased recovery of myoglobin (-20% bias) for Randox Immunoturbidimetric and Siemens Centaur Classic methods and CKMB (-21% bias) for Roche e411, BioMerieux Vidas, and Siemens Centaur Classic methods.

Sep 16, 2016 Diagnostic Equipment View Details →

According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.

Aug 2, 2016 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

According to the firm, the IgE Calibrator Series IE2492 (lot 387-392IE) value assignment has now been aligned to the 3rd International Standard for human serum IgE 11/234. Turbidimetric (Non-IFCC Cal.) Quality Control Targets and Ranges now apply. The calibrator value sheet is to be updated with the new values.

Sep 7, 2016 Other Medical Devices View Details →