RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Randox Laboratories Ltd.
Reason for Recall:
When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Product Codes/Lot Numbers:

Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0328-2022

Related Recalls

Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.

Apr 5, 2024 Infusion Pumps Nationwide View Details β†’