Immunoassay Premium Plus Controls (3x4x5mL and 12x5mL packages)
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog numbers: IA3109 (Lots: 1476EC, 1573EC, 1619EC) IA3110 (Lots 1477EDC, 1599EDC, 1621EC, 1662EC) IA3111 (Lots: 1478EC, 1508EC, 1578EC, 1647EC) IA3112 (Lots: 359199, 377412, 373302, 382147, 382679, 395668, 416979)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
â ď¸ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Randox Laboratories
- Reason for Recall:
- The ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Immunoassay Premium Plus Controls (3x4x5mL and 12x5mL packages)
Product Codes/Lot Numbers:
Catalog numbers: IA3109 (Lots: 1476EC, 1573EC, 1619EC) IA3110 (Lots 1477EDC, 1599EDC, 1621EC, 1662EC) IA3111 (Lots: 1478EC, 1508EC, 1578EC, 1647EC) IA3112 (Lots: 359199, 377412, 373302, 382147, 382679, 395668, 416979)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0970-2018
Related Recalls
RX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay on automated analysers. Catalogue Number: CYS4004
Randox Laboratories
Cystatin C Reagent marketed without a 510 (k)
Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine Catalog Number: EV4115
Randox Laboratories
Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.
Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.
Randox Laboratories
Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.