CareFusion 303

167 recalls • • Categories: Other Medical Devices, Infusion Pumps, Diagnostic Equipment, Surgical Instruments, Implants & Prosthetics

Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems.

Dec 3, 2025 Other Medical Devices Nationwide View Details →

Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

Nov 20, 2025 Other Medical Devices Nationwide View Details →

If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

Oct 17, 2025 Infusion Pumps Nationwide View Details →

Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

Sep 30, 2025 Other Medical Devices Nationwide View Details →

Various models of BD Pyxis Medbank: Reference numbers: 1137-00 1145-00 1146-00 1147-00 1148-00 138902-01 138903-01 138905-01 138906-01 138907-01 138908-01 138909-01 138910-01 138911-01 138912-01 138913-01 138914-01 138915-01 138916-01 138917-01 138918-01 138919-01 138920-01 138921-01 138922-01 138923-01 138924-01 138926-01 138927-01 138928-01 138929-01 138930-01 138931-01 138932-01 138933-01 138934-01 138936-01 138937-01 138938-01 138939-01 138940-01 138941-01 138942-01 138943-01 138944-01 138945-01 138946-01 138947-01 138948-01 138949-01 138950-01 138951-01 138952-01 138953-01 138954-01 138955-01 138956-01 138957-01 138958-01 138959-01 138960-01 138961-01 138962-01 138963-01 138964-01 138965-01 138966-01 138967-01 138968-01 138969-01 138970-01 138971-01 138972-01 138973-01 138974-01 138975-01 138976-01 138977-01 138978-01 138989-01 138990-01 139000-01 139001-01 139002-01 139011-01 139013-01 139016-01 139018-01 139019-01 139020-01 139021-01 139022-01 139023-01 139028-01 139062-01 139063-01 139073-01 139082-01 139084-01 139085-01 139086-01 139091-01 139122-01 139125-01 139126-01 139127-01 139128-01 139131-01 139141-01 139142-01 139156-01 139163-01 139170-01 139171-01 139172-01 139174-01 139175-01 155081-01 155082-01 155083-01 155084-01 155085-01 155086-01 155087-01 155088-01 155089-01 155090-01 155091-01 155092-01 155093-01 155094-01 155095-01 155096-01 155097-01 155098-01 155099-01 155100-01 155101-01 155102-01 155103-01 155104-01 155105-01 155106-01 155107-01 155108-01 155109-01 155110-01 155111-01 155112-01 155113-01 155114-01 155115-01 155116-01 155117-01 155118-01 155119-01 155120-01 155121-01 155122-01 155123-01 155124-01 155125-01 155126-01 155127-01 155128-01 155129-01 155130-01 155131-01 155132-01 155133-01 155134-01 155135-01 155136-01 155137-01 155138-01 155139-01 155140-01 155141-01 155142-01 155143-01 155144-01 155145-01 155146-01 155147-01 155148-01 155149-01 155150-01 155151-01 155152-01 155153-01 155154-01 155155-01 155156-01 155157-01 155159-01 155160-01 155161-01 155163-01 155165-01 155166-01 155167-01 155168-01 155169-01 155171-01 155288-01 155293-01 155294-01 155299-01 155302-01 155303-01 155304-01 155305-01 155310-01 155311-01 169-03 169-04 169-05 169-06 169-07 169-08 169-09 169-10 169-100 169-101 169-102 169-103 169-104 169-105 169-106 169-107 169-108 169-109 169-11 169-110 169-111 169-112 169-113 169-114 169-115 169-12 169-120 169-124 169-125 169-127 169-128 169-129 169-13 169-131 169-132 169-137 169-138 169-139 169-14 169-140 169-141 169-142 169-143 169-144 169-145 169-15 169-150 169-151 169-152 169-155 169-156 169-159 169-16 169-160 169-161 169-162 169-163 169-164 169-18 169-19 169-20 169-21 169-22 169-23 169-24 169-25 169-26 169-27 169-28 169-29 169-30 169-31 169-32 169-33 169-34 169-35 169-36 169-37 169-38 169-39 169-40 169-41 169-42 169-43 169-44 169-45 169-46 169-47 169-48 169-49 169-50 169-51 169-52 169-53 169-54 169-55 169-56 169-57 169-58 169-59 169-60 169-61 169-62 169-63 169-64 169-65 169-66 169-67 169-68 169-69 169-70 169-71 169-72 169-73 169-74 169-75 169-76 169-77 169-78 169-79 169-80 169-81 169-82 169-83 169-84 169-85 169-86 169-87 169-88 169-89 169-90 169-91 169-92 169-93 169-94 169-95 169-96 169-97 169-98 169-99

Class I - Dangerous

Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

Sep 30, 2025 Other Medical Devices Nationwide View Details →

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

Sep 11, 2025 Infusion Pumps Nationwide View Details →

BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main / 10885403512629 / 303 BD Pyxis" MedStation" ES Tower / 10885403512674 / 352 BD Pyxis" Anesthesia Station ES / 10885403477836 / 327 BD Pyxis" MedStation" ES (Main) / 10885403512667 / 323 BD Pyxis" CII Safe ES / 10885403512605 / 1116-00 BD Pyxis" Enterprise Server / 10885403518348 and 10885403520341 / 1115-00 BD Pyxis" CIISafe, V9.X, SGL IM, BIO, / 10885403512520 / 111-221 BD Pyxis" CIISafe, V7.X, DBL INTG MAIN, BIO, SOLID DO / N/A / 111-194 BD Pyxis" MedBank Tower / 10885403512568 / 169-29 BD Pyxis" CIISafe, V8.X Desktop PC, Bio / N/A / 107-245-01 Supply Server / N/A / N/A CCE Enterprise SW Site License (2-5) / N/A / N/A BD Pyxis" Logistics / N/A / 806 BD Pyxis" SupplyStation" V9 / N/A / 317 Security Module / N/A / 139698-01 Supply itParActive RF V11 / N/A / 138566-01 BD Knowledge Portal for Medication Technologies / N/A / 136607-01 STOCKSTN V10.X WRD/WL SCANNERS / N/A / 132-53-02 PRS STOCKSTN V10.X WRD/WL SCANNERS / N/A / 132-53-01 STOCKSTN V10.X WIRED SCANNER / N/A / 132-52-02 PRS STOCKSTN V10.X WIRED SCANNER / N/A / 132-52-01 BD Pyxis" CII Safe ES, Desktop PC / N/A / 1156-00 BD Pyxis" SupplyStation" Panel PC V11 / N/A / 1132-00 Pyxis" SupplyRoller V11 / N/A / 1127-00 Dell 640 Server / N/A / 1128-00 BD Pyxis" SupplyStation" System V11 / N/A / 1123-00 BD Pyxis" SupplyRoller V10 / N/A / 1101-00 BD Pyxis" SupplyStation" System V10 / **** / 347 BD Pyxis CII Safe,V9.X DESKTOP PC / **** / 107-252-01 BD Pyxis CII Safe,V7.X,DBL INTG MAIN,BIO,CLEAR DO 111-191 BD Pyxis CII Safe, V9.X, DBL IM, BIO, SLD DOORS / **** / 111-215 BD Care Coordination Engine (CCE) / **** / 134056-01 The Pyxis devices are automated medication dispensing cabinet that is intended to securely store and dispense medications to a qualified and authorized healthcare provider.

Class I - Dangerous

Antivirus software was not consistently installed on impacted devices during the implementation process.

May 8, 2025 Diagnostic Equipment Nationwide View Details →

Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS MEDBANK MINI 3HH-1HM/169-111, BD PYXIS MEDBANK MINI 3HH-1FG/169-110, BD PYXIS MEDBANK MINI 2HM-1FM/169-109, BD PYXIS MEDBANK MINI 2HH-2HM/169-108, BD PYXIS MEDBANK MINI 2HH-1FM/169-107, BD PYXIS MEDBANK MINI 2HH-1FH/169-106, BD PYXIS MEDBANK MINI 2FM/169-105, BD PYXIS MEDBANK MINI 2FH/169-104, BD PYXIS MEDBANK MINI 1HH-3HM/169-103, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/169-102, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/169-101 , BD PYXIS MEDBANK MINI 1FH-2HH/155311-01, BD PYXIS MEDBANK MINI 4HH/155310-01, BD PYXIS MEDBANK MINI 1FH-1FM/155288-01, BD PYXIS MEDBANK MINI 4HM/155171-01, BD PYXIS MEDBANK MINI 3HH-1HM/155169-01, BD PYXIS MEDBANK MINI 3HH-1FG/155168-01, BD PYXIS MEDBANK MINI 2HM-1FM/155167-01, BD PYXIS MEDBANK MINI 2HH-2HM/155166-01, BD PYXIS MEDBANK MINI 2HH-1FM/155165-01, BD PYXIS MEDBANK MINI 2FM/155163-01, BD PYXIS MEDBANK MINI 1HH-3HM/155161-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/155160-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/155159-01, BD PYXIS MEDBANK MINI 2HM-1FM/139131-01, BD PYXIS MEDBANK MINI 1HH-1HM-1FM/139122-01, BD PYXIS MEDBANK MINI 3HH-1FG/139091-01, BD PYXIS MEDBANK MINI 1HH-1FH-1HM/138921-01, BD PYXIS MEDBANK MINI 4HM/138915-01, BD PYXIS MEDBANK MINI 1HH-3HM/138912-01, BD PYXIS MEDBANK MINI 4HH/138910-01, BD PYXIS MEDBANK MINI 3HH-1HM/138909-01, BD PYXIS MEDBANK MINI 2HH-1FH/138908-01, BD PYXIS MEDBANK MINI 2HH-2HM/138907-01, BD PYXIS MEDBANK MINI 2HH-1FM/138906-01, BD PYXIS MEDBANK MINI 2FM/138903-01, BD PYXIS MEDBANK MINI 2FH/138902-01, MEDBANK MINI MAIN/1147-00, MEDBANK MAIN/1137-00, BD PYXIS MEDBANK TWR MN 8HH-3FM/169-81, BD PYXIS MEDBANK TWR MN 8HH-3FH/169-80, BD PYXIS MEDBANK TWR MN 8HH-3FF/169-79, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/169-78, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/169-77, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/169-76, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/169-75, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/169-74, BD PYXIS MEDBANK TWR MN 7FM/169-73, BD PYXIS MEDBANK TWR MN 7FH/169-72, BD PYXIS MEDBANK TWR MN 6HH-8HM/169-71, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/169-70, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/169-69, BD PYXIS MEDBANK TWR MN 6HH-4FM/169-68, BD PYXIS MEDBANK TWR MN 6HH-4FH /169-67, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/169-66, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/169-65, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/169-64, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/169-63, BD PYXIS MEDBANK TWR MN 6FH-1FM/169-62, BD PYXIS MEDBANK TWR MN 5HH-9HM/169-61, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/169-60, BD PYXIS MEDBANK TWR MN 4HM-5FM/169-59, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/169-58, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/169-57, BD PYXIS MEDBANK TWR MN 4HH-5FM/169-56, BD PYXIS MEDBANK TWR MN 4HH-5FH /169-55, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/169-54, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/169-53, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/169-52, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/169-51, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/169-50, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/169-49, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/169-48, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/169-45, BD PYXIS MEDBANK TWR MN 4HH-10HM/169-44, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/169-43, BD PYXIS MEDBANK TWR MN 3FH-4FM/169-42, BD PYXIS MEDBANK TWR MN 2HM-6FM/169-41, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/169-40, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/169-39, BD PYXIS MEDBANK TWR MN 2HH-6FM/169-38, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/169-37, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/169-36, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/169-35, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/169-34, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/169-33, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/169-32, BD PY

Class I - Dangerous

Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.

Jul 10, 2025 Other Medical Devices Nationwide View Details →

BD Alaris Pump Infusion Sets REF: 10013186, BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D DEHP FREE 3SS CV; REF: 2452-0007, BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 10D 3SS 2CV; REF: 2204-0007, BD Alaris Pump Infusion SetAS LVP 20D; REF: 24001-0007, BD Alaris Pump Infusion Set, Back Check Valve, SmartSite Y-siteAS LVP 20D SS CV; REF: 10015012, BD Alaris Pump Infusion Burette Set, 0.2 Micron Filter, Smallbore Tubing, SmartSite Port (Burette), SmartSite Y-siteAS LVP BUR 20D PE-Lined SMBORE SS 0.2M; REF: 2432-0007, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D 3SS 0.2M CV; REF: 2434-0007, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, SmartSite Y-siteAS LVP 20D SS 0.2M CV; REF: 2202-0007, BD Alaris Pump Infusion Set, 1.2 Micron FilterAS LVP 20D 1.2M; REF: 11426965, BD Alaris Pump Infusion Set, Back Check Valve, 5 SmartSite Y-sitesAS LVP 20D CV; REF: 10561554, BD Alaris Pump Infusion Set, 15 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D DEHP FREE 3SS 15M CV; REF: 10863358, BD Alaris Pump Infusion Set, 15 Micron FilterAS LVP 20D DEHP FREE 15M; REF: 11171447, BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 20D 3SS 2CV; REF: 11532269, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, PE Lined Tubing, 2 SmartSite Y-sitesAS LVP 20D PE-LINED 2SS 0.2M CV; REF: 10010454, BD Alaris Pump Infusion Set, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-siteAS LVP 20D PE-LINED TUBING SS 0.2M; REF: 2465-0007, BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, 0.2 Micron Filter, Low Sorbing Tubing (PE Lined), SmartSite Y-siteAS LVP SS BAG 20D PE-Lined SS 0.2M; REF: 10010453, ASBD Alaris Pump Infusion Set, 1.2 Micron Filter, PE Lined Tubing, SmartSite Y-site LVP 20D PE-LINED TUBING SS 1.2M REF: 11419365, BD Alaris" Pump Infusion Set, Back Check Valve, Manifold, 3-Way Stopcock, 5 SmartSite" Y-sites REF: 10813621, BD Alaris" Pump Infusion Set, Back Check Valve, 3 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 2423-0007, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 10015896, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 3 SmartSite" Y-sites REF: 24010-0007T, BD Alaris" Pump Infusion Set, Bonded Texium", Closed Male Luer with Priming Cap, Back Check Valve, 3 SmartSite" Y-sites REF: 10010483, BD Alaris" Pump Infusion Set, Vented Syringe Adapter, Smallbore Tubing REF: 2403-0007, BD Alaris" Pump Infusion Set, Half Set, SmartSite" Y-site REF: 24301-0007T, BD Alaris" Pump Infusion Set, SmartSite" Bag Access Non-Vented, 0.2 Micron Filter, PE Lined Tubing, Bonded Texium" Closed Male Luer with Priming Cap, SmartSite" Y-site

Class I - Dangerous

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

Jul 8, 2025 Surgical Instruments Nationwide View Details →

During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may lead to delay in access to medications/supplies, and delivery/replenishment of patient specific orders; which may lead to delay in patient care.

May 8, 2025 Diagnostic Equipment Nationwide View Details →

BD PYXIS MEDBANK (Name, REF): MEDPASS MODULE, 139088-01; MINI 1FH-1FM, 169-137; MINI 1HH-1FH-1HM, 169-101; MINI 1HH-1HM-1FM, 169-102; MINI 2FH, 138902-01; MINI 2FH, 169-104; MINI 2FM, 169-105; MINI 2HH-1FH, 169-106; MINI 2HH-1FH, 138908-01; MINI 2HH-1FM, 169-107; MINI 2HH-1FM, 138906-01; MINI 3HH-1FG, 169-110; MINI 3HH-1FG, 139091-01; MINI 3HH-1HM, 169-111; MINI 4HH, 138910-01; MINI 4HH, 169-112; MINI 4HM, 169-113; MINI 4HM, 138915-01; TWR MN 10HH-1FH-1FM, 169-23; TWR MN 10HH-2FH, 169-24; TWR MN 10HH-2FH, 138955-01; TWR MN 10HH-2FM, 169-25; TWR MN 10HH-2FM, 138924-01; TWR MN 12HH-1FF, 169-26; TWR MN 12HH-1FF, 138960-01; TWR MN 12HH-1FH, 169-27; TWR MN 12HH-1FH, 138961-01; TWR MN 12HH-1FM, 169-28; TWR MN 12HH-1FM, 138926-01; TWR MN 14HH, 138927-01; TWR MN 14HH, 169-29; TWR MN 1HH-1HM-2FM-4FF, 169-30; TWR MN 2FH-2HM-4FM, 169-31; TWR MN 2HH-10HM-1FM, 169-32; TWR MN 2HH-1FH-5FM, 169-33; TWR MN 2HH-2FH-4FM, 169-34; TWR MN 2HH-4FH-2FM, 169-139; TWR MN 2HH-4HM-4FM, 139125-01; TWR MN 2HH-6FM, 169-38; TWR MN 2HH-6FM, 138919-01; TWR MN 2HH-6HM-3FM, 169-39; TWR MN 2HH-8HM-2FM, 169-40; TWR MN 2HM-6FM, 169-41; TWR MN 3FH-4FM, 169-42; TWR MN 3FH-4FM, 138933-01; TWR MN 3FH-8HH MEDPASS, 169-140; TWR MN 3HH-7HM-2FM, 169-43; TWR MN 4HH-10HM, 169-44; TWR MN 4HH-10HM, 138959-01; TWR MN 4HH-1FH-2HM-3FM, 169-45; TWR MN 4HH-1FH-4FM, 169-46; TWR MN 4HH-1FH-4HM-2FM, 169-47; TWR MN 4HH-2FH-2HM-2FM, 169-49; TWR MN 4HH-2FH-3FM, 169-50; TWR MN 4HH-2FH-3FM, 138931-01; TWR MN 4HH-3FH-2FM, 169-52; TWR MN 4HH-4FH-1FM, 169-53; TWR MN 4HH-4FH-1FM, 138940-01; TWR MN 4HH-4HM-3FM, 169-54; TWR MN 4HH-4HM-3FM, 138952-01; TWR MN 4HH-5FH, 169-55; TWR MN 4HH-5FM, 169-56; TWR MN 4HH-5FM, 138918-01; TWR MN 4HH-6HM-2FM, 169-57; TWR MN 4HM-5FM, 169-59; TWR MN 5HH-3HM-3FM, 169-60; TWR MN 5HH-9HM, 138914-01; TWR MN 5HH-9HM, 169-61; TWR MN 6FH-1FM, 169-62; TWR MN 6HH-1FH-2HM-2FM, 169-63; TWR MN 6HH-1FH-3FM, 169-64; TWR MN 6HH-2FH-2FM, 169-65; TWR MN 6HH-4FH, 169-67; TWR MN 6HH-4FH, 138957-01; TWR MN 6HH-4FM, 169-68; TWR MN 6HH-4FM, 138922-01; TWR MN 6HH-4HM-2FM, 169-69; TWR MN 6HH-8HM, 169-71; TWR MN 6HH-8HM, 138958-01; TWR MN 7FH, 169-72; TWR MN 7FH, 138920-01; TWR MN 7FM, 169-73; TWR MN 7FM, 138937-01; TWR MN 7HH-1HM-3FM, 139086-01; TWR MN 7HH-1HM-3FM, 169-74; TWR MN 8HH-1FH-2FM, 169-75; TWR MN 8HH-2FH-1FM, 169-77; TWR MN 8HH-2HM-2FM, 169-78; TWR MN 8HH-2HM-2FM, 138939-01; TWR MN 8HH-3FF, 169-79; TWR MN 8HH-3FH, 169-80; TWR MN 8HH-3FH, 138956-01; TWR MN 8HH-3FM, 169-81; TWR MN 8HH-3FM, 138923-01. BD PYXIS MEDFLEX MN (Name, REF): 1000 12HH, 139055-01; 1000 2HH-2FM, 139041-01; 1000 2HH-5FM, 139046-01; 1000 4HH-4FM, 139049-01; 1000 6HH-3FM, 139051-01; 1000 8HH-2FM, 139052-01; 2000 8HH-2FH, 139165-01; 2000 8HH-2FM, 139069-01. MEDBANK MAIN, 1137-00; MEDBANK MINI MAIN, 1147-00; MEDBANK TOWER MAIN, 1145-00; MEDFLEX, 1119-00; MEDFLEX 2.0, 1139-00

Class I - Dangerous

Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment.

Mar 11, 2025 Other Medical Devices Nationwide View Details →

MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software, REF: 139090-01; BD PYXS MEDBNK TWR MN 2HH1FH2HM2FM2FG1FF, REF: 169-100; BD PYXIS MEDBANK MINI 1HH-1FH- 1HM, REF: 169-101; BD PYXIS MEDBANK MINI 1HH-1HM- 1FM, REF: 169-102; BD PYXIS MEDBANK MINI 1HH-3HM, REF: 169-103; BD PYXIS MEDBANK MINI 2FH, REF: 169-104; BD PYXIS MEDBANK MINI 2FM, REF: 169-105; BD PYXIS MEDBANK MINI 2HH-1FH, REF: 169-106; BD PYXIS MEDBANK MINI 2HH-1FM , REF: 169-107; BD PYXIS MEDBANK MINI 2HH-2HM, REF: 169-108; BD PYXIS MEDBANK MINI 2HM-1FM, REF: 169-109; BD PYXIS MEDBANK MINI 3HH-1FG, REF: 169-110; BD PYXIS MEDBANK MINI 3HH-1HM, REF: 169-111; BD PYXIS MEDBANK MINI 4HH, REF: 169-112; BD PYXIS MEDBANK MINI 4HM, REF: 169-113; BD PYXIS MDBANK TWR MN 7HH-1HM-3FM RxNow, REF: 169-132; BD PYXIS MEDBANK MINI 1FH-1FM, REF: 169-137; BD PYXIS MEDBANK TWR MN 1FH- 10HM-1FM, REF: 169-138; BD PYXIS MEDBANK TWR MN 2HH- 4FH-2FM, REF: 169-139; OBS BD PYXIS MB TOWER MAIN, REF: 169-153; OBS BD PYXIS MB TWR RXNOW MAIN W/MED PRI, REF: 169-155; OBS BD PYXIS MB TWR RXNOW MAIN W/PAT PRI, REF: 169-156; OBS BD PYXIS MB MINI MAIN W/CUBIE DRWS, REF: 169-159; OBS BD PYXIS MB MINI MAIN W/MATRIX DRWS, REF: 169-160; OBS BD PYXIS MB MINI MAIN W/HH FALSE DRW, REF: 169-161; OBS BD PYXIS MB MINI MAIN W/FH FALSE DRW, REF: 169-162; BD PYXIS MEDBANK TWR MN 10HH- 1FH-1FM, REF: 169-23; BD PYXIS MEDBANK TWR MN 10HH- 2FH, REF: 169-24; BD PYXIS MEDBANK TWR MN 10HH- 2FM, REF: 169-25; BD PYXIS MEDBANK TWR MN 12HH- 1FF, REF: 169-26; BD PYXIS MEDBANK TWR MN 12HH- 1FH, REF: 169-27; BD PYXIS MEDBANK TWR MN 12HH- 1FM, REF: 169-28; BD PYXIS MEDBANK TWR MN 14HH, REF: 169-29; BD PYXIS MEDBANK TWR MN 1HH- 1HM-2FM-4FF, REF: 169-30; BD PYXIS MEDBANK TWR MN 2FH- 2HM-4FM, REF: 169-31; BD PYXIS MEDBANK TWR MN 2HH- 10HM-1FM, REF: 169-32; BD PYXIS MEDBANK TWR MN 2HH- 1FH-5FM, REF: 169-33; BD PYXIS MEDBANK TWR MN 2HH- 2FH-4FM, REF: 169-34; BD PYXIS MEDBANK TWR MN 2HH- 2FM-4FF, REF: 169-35; BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FF, REF: 169-36; BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FM, REF: 169-37; BD PYXIS MEDBANK TWR MN 2HH- 6FM, REF: 169-38; BD PYXIS MEDBANK TWR MN 2HH- 6HM-3FM, REF: 169-39; BD PYXIS MEDBANK TWR MN 2HH- 8HM-2FM, REF: 169-40; BD PYXIS MEDBANK TWR MN 2HM- 6FM, REF: 169-41; BD PYXIS MEDBANK TWR MN 3FH-4FM , REF: 169-42; BD PYXIS MEDBANK TWR MN 3HH- 7HM-2FM, REF: 169-43; BD PYXIS MEDBANK TWR MN 4HH- 10HM, REF: 169-44; BD PYXIS MEDBANK TWR MN 4HH- 1FH-2HM-3FM, REF: 169-45; BD PYXIS MEDBANK TWR MN 4HH- 1FH-4FM, REF: 169-46; BD PYXIS MEDBANK TWR MN 4HH- 1FH-4HM-2FM, REF: 169-47; BD PYXIS MEDBANK TWR MN 4HH- 1FH-6HM-1FM, REF: 169-48; BD PYXIS MEDBANK TWR MN 4HH- 2FH-2HM-2FM, REF: 169-49; BD PYXIS MEDBANK TWR MN 4HH- 2FH-3FM, REF: 169-50; BD PYXIS MEDBANK TWR MN 4HH- 2HM-4FM, REF: 169-51; BD PYXIS MEDBANK TWR MN 4HH- 3FH-2FM, REF: 169-52; BD PYXIS MEDBANK TWR MN 4HH- 4FH-1FM, REF: 169-53; BD PYXIS MEDBANK TWR MN 4HH- 4HM-3FM, REF: 169-54; BD PYXIS MEDBANK TWR MN 4HH-5FH, REF: 169-55; BD PYXIS MEDBANK TWR MN 4HH- 5FM, REF: 169-56; BD PYXIS MEDBANK TWR MN 4HH- 6HM-2FM, REF: 169-57; BD PYXIS MEDBANK TWR MN 4HH- 8HM-1FM, REF: 169-58; BD PYXIS MEDBANK TWR MN 4HM- 5FM, REF: 169-59; BD PYXIS MEDBANK TWR MN 5HH- 3HM-3FM, REF: 169-60; BD PYXIS MEDBANK TWR MN 5HH- 9HM, REF: 169-61; BD PYXIS MEDBANK TWR MN 6FH-1FM, REF: 169-62; BD PYXIS MEDBANK TWR MN 6HH- 1FH-2HM-2FM, REF: 169-63; BD PYXIS MEDBANK TWR MN 6HH- 1FH-3FM, REF: 169-64; BD PYXIS MEDBANK TWR MN 6HH- 2FH-2FM, REF: 169-65; BD PYXIS MEDBANK TWR MN 6HH- 2FH-2HM-1FM, REF: 169-66; BD PYXIS MEDBANK TWR MN 6HH-4FH, REF: 169-67; BD PYXIS MEDBANK TWR MN 6HH- 4FM, REF: 169-68; BD PYXIS MEDBANK TWR MN 6HH- 4HM-2FM, REF: 169-69; BD PYXIS MEDBANK TWR MN 6HH- 6HM-1FM, REF: 169-70; BD PYXIS MEDBANK TWR MN 6HH- 8HM, REF: 169-71; BD PYXIS MEDBANK TWR MN 7FH, REF: 169-72; BD PYXIS M

Class I - Dangerous

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

Oct 5, 2023 Other Medical Devices Nationwide View Details →

Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the CUBIE opening more times than needed for the transaction to be completed, which may result in a patient receiving more than the intended dose of medication.

Mar 11, 2025 Other Medical Devices Nationwide View Details →

BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 155140-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 155143-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 155144-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 155145-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 155157-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P / 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 155153-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 155155-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 155147-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 155149-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 155150-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 169-114 BD PYXIS MEDBANK MINI CR-4HH-P / 169-115 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 169-83 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 169-85 BD PYXIS MEDBANK TWR MN CR-14HH-P / 169-86 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 169-89 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 169-91 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /169-93 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 169-95 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 169-99 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 138913-01 BD PYXIS MEDBANK MINI CR-4HH-P / 138911-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P / 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P / 139002-01 BD PYXIS MEDBANK MN 200 CR-4HH-P / 139000-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P / 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P / 139018-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P / 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P / 139021-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P / 139022-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 138975-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 138976-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 138977-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 139085-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 138964-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 138965-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 138970-01 BD Pyxis QFill Replenishment Station / 138904-01 KIT BD PYXIS QFILL REPLENISHMENT STATION / 155197-01 MEDBANK UPOS211 AIO WIN10 QFILL 1.5 / 155215-05 OPT,MEDBANK,HD,WIN10,QFILL,CR / 131944-04 The BD Pyxis" MedBank" CUBIE" Replenishment Station is an Automated Dispensing Cabinet (ADC) that is intended to be used by a qualified and authorized user to replenish medications and supplies to a BD Pyxis" MedBank" cabinet.

Class I - Dangerous

Due to a software issue, when attempting to restock the automated dispensing cabinet, user have received the banner message "Stocking window has passed, or item has been removed from purchase order. Please contact Pharmacy for assistance". When this occurs, the medication cannot be restocked and can potentially lead to delays in accessing medication for patients.

Mar 11, 2025 Other Medical Devices Nationwide View Details →

Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.

Feb 6, 2025 Other Medical Devices Nationwide View Details →

Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.

Jan 23, 2025 Diagnostic Equipment Nationwide View Details →

138913-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 169-114 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 138911-01 BD PYXIS MEDBANK MINI CR-4HH-P 169-115 BD PYXIS MEDBANK MINI CR-4HH-P 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P 139002-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P 139000-01 BD PYXIS MEDBANK MN 200 CR-4HH-P 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P 139018-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P 139021-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P 139022-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 155140-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 138975-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 169-83 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 138976-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 155143-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 169-85 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 138977-01 BD PYXIS MEDBANK TWR MN CR-14HH-P 155144-01 BD PYXIS MEDBANK TWR MN CR-14HH-P 169-86 BD PYXIS MEDBANK TWR MN CR-14HH-P 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 155145-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 139085-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 155147-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 169-89 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 138964-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 155149-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 169-91 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 155150-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 169-93 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 138965-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 155153-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 169-95 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 155155-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 138970-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 155157-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 169-99 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P

Class I - Dangerous

Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

Jan 8, 2025 Other Medical Devices Nationwide View Details →

Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

Jan 8, 2025 Other Medical Devices Nationwide View Details →

155288-01 BD PYXIS MEDBANK MINI 1FH-1FM 169-137 BD PYXIS MEDBANK MINI 1FH-1FM 155311-01 BD PYXIS MEDBANK MINI 1FH-2HH 138921-01 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 155159-01 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 169-101 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 139122-01 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 155160-01 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 169-102 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 138912-01 BD PYXIS MEDBANK MINI 1HH-3HM 155161-01 BD PYXIS MEDBANK MINI 1HH-3HM 169-103 BD PYXIS MEDBANK MINI 1HH-3HM 138902-01 BD PYXIS MEDBANK MINI 2FH 169-104 BD PYXIS MEDBANK MINI 2FH 138903-01 BD PYXIS MEDBANK MINI 2FM 155163-01 BD PYXIS MEDBANK MINI 2FM 169-105 BD PYXIS MEDBANK MINI 2FM 138908-01 BD PYXIS MEDBANK MINI 2HH-1FH 169-106 BD PYXIS MEDBANK MINI 2HH-1FH 138906-01 BD PYXIS MEDBANK MINI 2HH-1FM 155165-01 BD PYXIS MEDBANK MINI 2HH-1FM 169-107 BD PYXIS MEDBANK MINI 2HH-1FM 138907-01 BD PYXIS MEDBANK MINI 2HH-2HM 155166-01 BD PYXIS MEDBANK MINI 2HH-2HM 169-108 BD PYXIS MEDBANK MINI 2HH-2HM 139131-01 BD PYXIS MEDBANK MINI 2HM-1FM 155167-01 BD PYXIS MEDBANK MINI 2HM-1FM 169-109 BD PYXIS MEDBANK MINI 2HM-1FM 139091-01 BD PYXIS MEDBANK MINI 3HH-1FG 155168-01 BD PYXIS MEDBANK MINI 3HH-1FG 169-110 BD PYXIS MEDBANK MINI 3HH-1FG 138909-01 BD PYXIS MEDBANK MINI 3HH-1HM 155169-01 BD PYXIS MEDBANK MINI 3HH-1HM 169-111 BD PYXIS MEDBANK MINI 3HH-1HM 138910-01 BD PYXIS MEDBANK MINI 4HH 155310-01 BD PYXIS MEDBANK MINI 4HH 169-112 BD PYXIS MEDBANK MINI 4HH 138915-01 BD PYXIS MEDBANK MINI 4HM 155171-01 BD PYXIS MEDBANK MINI 4HM 169-113 BD PYXIS MEDBANK MINI 4HM 1147-00 MEDBANK MINI MAIN 138943-01 BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 155081-01 BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 169-23 BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 138955-01 BD PYXIS MEDBANK TWR MN 10HH-2FH 155082-01 BD PYXIS MEDBANK TWR MN 10HH-2FH 169-24 BD PYXIS MEDBANK TWR MN 10HH-2FH 138924-01 BD PYXIS MEDBANK TWR MN 10HH-2FM 155083-01 BD PYXIS MEDBANK TWR MN 10HH-2FM 169-25 BD PYXIS MEDBANK TWR MN 10HH-2FM 138960-01 BD PYXIS MEDBANK TWR MN 12HH-1FF 155084-01 BD PYXIS MEDBANK TWR MN 12HH-1FF 169-26 BD PYXIS MEDBANK TWR MN 12HH-1FF 138961-01 BD PYXIS MEDBANK TWR MN 12HH-1FH 155085-01 BD PYXIS MEDBANK TWR MN 12HH-1FH 169-27 BD PYXIS MEDBANK TWR MN 12HH-1FH 138926-01 BD PYXIS MEDBANK TWR MN 12HH-1FM 155086-01 BD PYXIS MEDBANK TWR MN 12HH-1FM 169-28 BD PYXIS MEDBANK TWR MN 12HH-1FM 138927-01 BD PYXIS MEDBANK TWR MN 14HH 155087-01 BD PYXIS MEDBANK TWR MN 14HH 169-29 BD PYXIS MEDBANK TWR MN 14HH 155293-01 BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM 169-138 BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM 155303-01 BD PYXIS MEDBANK TWR MN 1FM-10HH-1FH 139082-01 BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 155088-01 BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 169-30 BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 155305-01 BD PYXIS MEDBANK TWR MN 2FH-1FM-4HH-2FM 139056-01 BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 155089-01 BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 169-31 BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 155304-01 BD PYXIS MEDBANK TWR MN 2FM-8HH-1FH 139045-01 BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 155090-01 BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 169-32 BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 138932-01 BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 155091-01 BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 169-33 BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 139084-01 BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 155092-01 BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 169-34 BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 138963-01 BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 155093-01 BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 169-35 BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 155294-01 BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM 169-139 BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM 138938-01 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 155094-01 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 169-36 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 139125-01 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 155095-01 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 169-37 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 138919-01 BD PYXIS MEDBANK TWR MN 2HH-6FM 155096-01 BD PYXIS MEDBANK TWR MN 2HH-6FM 169-38 BD PYXIS MEDBANK TWR MN 2HH-6

Class I - Dangerous

Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

Jan 8, 2025 Other Medical Devices Nationwide View Details →

Automated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential delays in retrieving contents/ accessing medications that may result from power loss or potential automated dispensing device malfunctions.

Jan 10, 2025 Other Medical Devices Nationwide View Details →

Labeling is insufficient regarding delays in access to medication. The firm is strengthening the label as an additional risk mitigation.

Jan 8, 2025 Other Medical Devices Nationwide View Details →

Due to an increase in complaints of door latch assembly failures that may cause the doors to be left in an open state potentially causing inventory discrepancies, a delay in access to medication and/or diversion of medication.

Feb 14, 2023 Other Medical Devices Nationwide View Details →

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

Feb 14, 2023 Other Medical Devices View Details →

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

Feb 14, 2023 Other Medical Devices View Details →

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

Feb 14, 2023 Other Medical Devices View Details →

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

Feb 14, 2023 Other Medical Devices View Details →

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

Feb 14, 2023 Other Medical Devices View Details →

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

Feb 14, 2023 Other Medical Devices View Details →

Due to an automated dispensing cabinet software issue, when a medication single order contains more than one component and at least one is unavailable for a multi-item or combination order, the order may display as grayed out; alternate location tooltip will display "Server connection failed," not listing other area devices where unavailable item(s) may be found, which may lead to delayed therapy.

Apr 5, 2023 Other Medical Devices Nationwide View Details →

Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.

Nov 14, 2024 Other Medical Devices Nationwide View Details →

Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.

Nov 14, 2024 Other Medical Devices Nationwide View Details →

Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.

Nov 14, 2024 Other Medical Devices Nationwide View Details →

If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.

May 2, 2024 Diagnostic Equipment Nationwide View Details →

Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.

May 13, 2024 Other Medical Devices Nationwide View Details →

BD Pyxis, Name/REF: BD PYXIS MB TOWER MAIN/ 169-153, BD PYXIS MB TOWER MAIN CR/ 169-154, BD PYXIS MB TOWER MAIN W/MP/ 169-157, BD PYXIS MB TWR MN 1FM-10HH-1FH MP/ 169-143, BD PYXIS MB TWR MN 2FH-1FM-4HH-2FM MP/ 169-145, BD PYXIS MB TWR MN 2FM-8HH-1FH MP/ 169-144, BD PYXIS MB TWR MN 4FH-4HH-1FM MP/ 169-142, BD PYXIS MB TWR RXNOW MAIN W/MED PRINTER / 169-155, BD PYXIS MB TWR RXNOW MAIN W/PAT PRINTER/ 169-156, BD PYXIS MDBANK TWR MN 7HH-1HM-3FM/ 169-132, BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow/ 169-131, BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM/ 169-23, BD PYXIS MEDBANK TWR MN 10HH-2FH/ 169-24, BD PYXIS MEDBANK TWR MN 10HH-2FM/ 169-25, BD PYXIS MEDBANK TWR MN 12HH-1FF/ 169-26, BD PYXIS MEDBANK TWR MN 12HH-1FH/ 169-27, BD PYXIS MEDBANK TWR MN 12HH-1FM/ 169-28, BD PYXIS MEDBANK TWR MN 14HH/ 169-29, BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM/ 169-138, BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF/ 169-30, BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM/ 169-31, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/ 169-32, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/ 169-33, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/ 169-34, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/ 169-35, BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM/ 169-139, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/ 169-36, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/ 169-37, BD PYXIS MEDBANK TWR MN 2HH-6FM/ 169-38, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/ 169-39, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/ 169-40, BD PYXIS MEDBANK TWR MN 2HM-6FM/ 169-41, BD PYXIS MEDBANK TWR MN 3FH-4FM/ 169-42, BD PYXIS MEDBANK TWR MN 3FH-8HH MEDPASS/ 169-140, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/ 169-43, BD PYXIS MEDBANK TWR MN 4HH-10HM/ 169-44, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/ 169-45, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/ 169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/ 169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/ 169-48, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/ 169-49, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/ 169-50, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/ 169-51, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/ 169-52, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/ 169-53, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/ 169-54, BD PYXIS MEDBANK TWR MN 4HH-5FH / 169-55, BD PYXIS MEDBANK TWR MN 4HH-5FM/ 169-56, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/ 169-57, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/ 169-58, BD PYXIS MEDBANK TWR MN 4HM-5FM/ 169-59, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/ 169-60, BD PYXIS MEDBANK TWR MN 5HH-9HM/ 169-61, BD PYXIS MEDBANK TWR MN 6FH-1FM/ 169-62, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/ 169-63, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/ 169-64, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/ 169-65, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/ 169-66, BD PYXIS MEDBANK TWR MN 6HH-4FH / 169-67, BD PYXIS MEDBANK TWR MN 6HH-4FM/ 169-68, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/ 169-69, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/ 169-70, BD PYXIS MEDBANK TWR MN 6HH-8HM/ 169-71, BD PYXIS MEDBANK TWR MN 7FH/ 169-72, BD PYXIS MEDBANK TWR MN 7FM/ 169-73, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/ 169-74, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/ 169-75, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/ 169-76, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/ 169-77, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/ 169-78, BD PYXIS MEDBANK TWR MN 8HH-3FF/ 169-79, BD PYXIS MEDBANK TWR MN 8HH-3FH/ 169-80, BD PYXIS MEDBANK TWR MN 8HH-3FM/ 169-81, BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P/ 169-82, BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P/ 169-83, BD PYXIS MEDBANK TWR MN CR-12HH-1FH/ 169-84, BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P/ 169-85, BD PYXIS MEDBANK TWR MN CR-14HH-P/ 169-86, BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM/ 169-87, BD PYXIS MEDBANK TWR MN CR-2HH-6FH/ 169-88, BD PYXIS MEDBANK TWR MN CR-2HH-6FM/ 169-89, BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P/ 169-90, BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P/ 169-91, BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P/ 169-92, BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P/ 169-93, BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P/ 169-94, BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P/ 169-95, BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P/ 169-96, BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P/ 169-97, BD

Class I - Dangerous

Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.

May 13, 2024 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

Sep 15, 2023 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

Sep 15, 2023 Other Medical Devices Nationwide View Details →

When global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty), 4) CII Safe Stock Out Notice are, without user warning, overwritten to the value NO , which could result in diversion of controlled substances from the automated dispensing cabinet.

Sep 6, 2023 Other Medical Devices Nationwide View Details →

After dispensing a timed dose medication for a patient, the automated dispensing cabinet software may continue to show the already dispensed medication dose is still available for dispense, which could lead to duplicate medication administration.

Aug 18, 2023 Other Medical Devices View Details →

Automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which could expose users to system downtime and/or potential electric shock.

May 22, 2023 Other Medical Devices Nationwide View Details →

Automated dispensing cabinet software is experiencing: 1) ES device download failure resulting in partial patient information loss that leads to 2) partial loss of patient transaction data, seen as a server upload processing error. Outdated/incorrect information display may contribute to the removal of medication to which the patient is allergic, incorrect dose amount, or discontinued medication.

Apr 5, 2023 Other Medical Devices Nationwide View Details →

Automated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device installations or device replacement upgrades, which may lead to an increased frequency of drawer failures, which could increase the likelihood of patients experiencing delays in receiving medications.

Feb 14, 2023 Other Medical Devices Nationwide View Details →

Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed

Jan 24, 2023 Other Medical Devices Nationwide View Details →

Due to customer complaints related to the MedBank software which may indicate that a different medication is loaded in a cubie (drawer) pocket than what is physically loaded and listed on the pocket label. As a result of this issue, a patient could receive the wrong medication or wrong dose of medication that could lead to patient harm.

Dec 12, 2022 Other Medical Devices Nationwide View Details →

The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop opening in the MedBank Application) leading to delay in access to medications or supplies needed in residential or procedural care facilities.

May 24, 2022 Other Medical Devices Nationwide View Details →

Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion start delay/inability to titrate medication. 2) Stuck keys: PC unit alarms, module exhibits Channel Error, may result in infusion interruption/start delay

Mar 3, 2021 Infusion Pumps Nationwide View Details →

Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.

Aug 4, 2020 Infusion Pumps Nationwide View Details →