BD Pyxis, Name/REF: BD PYXIS MB TOWER MAIN/ 169-153, BD PYXIS MB TOWER MAIN CR/ 169-154, BD PYXIS MB TOWER MAIN W/MP/ 169-157, BD PYXIS MB TWR MN 1FM-10HH-1FH MP/ 169-143, BD PYXIS MB TWR MN 2FH-1FM-4HH-2FM MP/ 169-145, BD PYXIS MB TWR MN 2FM-8HH-1FH MP/ 169-144, BD PYXIS MB TWR MN 4FH-4HH-1FM MP/ 169-142, BD PYXIS MB TWR RXNOW MAIN W/MED PRINTER / 169-155, BD PYXIS MB TWR RXNOW MAIN W/PAT PRINTER/ 169-156, BD PYXIS MDBANK TWR MN 7HH-1HM-3FM/ 169-132, BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow/ 169-131, BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM/ 169-23, BD PYXIS MEDBANK TWR MN 10HH-2FH/ 169-24, BD PYXIS MEDBANK TWR MN 10HH-2FM/ 169-25, BD PYXIS MEDBANK TWR MN 12HH-1FF/ 169-26, BD PYXIS MEDBANK TWR MN 12HH-1FH/ 169-27, BD PYXIS MEDBANK TWR MN 12HH-1FM/ 169-28, BD PYXIS MEDBANK TWR MN 14HH/ 169-29, BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM/ 169-138, BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF/ 169-30, BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM/ 169-31, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/ 169-32, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/ 169-33, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/ 169-34, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/ 169-35, BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM/ 169-139, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/ 169-36, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/ 169-37, BD PYXIS MEDBANK TWR MN 2HH-6FM/ 169-38, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/ 169-39, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/ 169-40, BD PYXIS MEDBANK TWR MN 2HM-6FM/ 169-41, BD PYXIS MEDBANK TWR MN 3FH-4FM/ 169-42, BD PYXIS MEDBANK TWR MN 3FH-8HH MEDPASS/ 169-140, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/ 169-43, BD PYXIS MEDBANK TWR MN 4HH-10HM/ 169-44, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/ 169-45, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/ 169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/ 169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/ 169-48, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/ 169-49, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/ 169-50, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/ 169-51, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/ 169-52, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/ 169-53, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/ 169-54, BD PYXIS MEDBANK TWR MN 4HH-5FH / 169-55, BD PYXIS MEDBANK TWR MN 4HH-5FM/ 169-56, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/ 169-57, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/ 169-58, BD PYXIS MEDBANK TWR MN 4HM-5FM/ 169-59, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/ 169-60, BD PYXIS MEDBANK TWR MN 5HH-9HM/ 169-61, BD PYXIS MEDBANK TWR MN 6FH-1FM/ 169-62, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/ 169-63, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/ 169-64, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/ 169-65, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/ 169-66, BD PYXIS MEDBANK TWR MN 6HH-4FH / 169-67, BD PYXIS MEDBANK TWR MN 6HH-4FM/ 169-68, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/ 169-69, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/ 169-70, BD PYXIS MEDBANK TWR MN 6HH-8HM/ 169-71, BD PYXIS MEDBANK TWR MN 7FH/ 169-72, BD PYXIS MEDBANK TWR MN 7FM/ 169-73, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/ 169-74, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/ 169-75, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/ 169-76, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/ 169-77, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/ 169-78, BD PYXIS MEDBANK TWR MN 8HH-3FF/ 169-79, BD PYXIS MEDBANK TWR MN 8HH-3FH/ 169-80, BD PYXIS MEDBANK TWR MN 8HH-3FM/ 169-81, BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P/ 169-82, BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P/ 169-83, BD PYXIS MEDBANK TWR MN CR-12HH-1FH/ 169-84, BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P/ 169-85, BD PYXIS MEDBANK TWR MN CR-14HH-P/ 169-86, BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM/ 169-87, BD PYXIS MEDBANK TWR MN CR-2HH-6FH/ 169-88, BD PYXIS MEDBANK TWR MN CR-2HH-6FM/ 169-89, BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P/ 169-90, BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P/ 169-91, BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P/ 169-92, BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P/ 169-93, BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P/ 169-94, BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P/ 169-95, BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P/ 169-96, BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P/ 169-97, BD

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 10885403512568. Devices Manufactured November 15, 2012 - July, 25 2023 User Guide, DIR 10000414989/DIR 10000414987/DIR 10000414990, Version 00.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
CareFusion 303, Inc.
Reason for Recall:
Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

BD Pyxis, Name/REF: BD PYXIS MB TOWER MAIN/ 169-153, BD PYXIS MB TOWER MAIN CR/ 169-154, BD PYXIS MB TOWER MAIN W/MP/ 169-157, BD PYXIS MB TWR MN 1FM-10HH-1FH MP/ 169-143, BD PYXIS MB TWR MN 2FH-1FM-4HH-2FM MP/ 169-145, BD PYXIS MB TWR MN 2FM-8HH-1FH MP/ 169-144, BD PYXIS MB TWR MN 4FH-4HH-1FM MP/ 169-142, BD PYXIS MB TWR RXNOW MAIN W/MED PRINTER / 169-155, BD PYXIS MB TWR RXNOW MAIN W/PAT PRINTER/ 169-156, BD PYXIS MDBANK TWR MN 7HH-1HM-3FM/ 169-132, BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow/ 169-131, BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM/ 169-23, BD PYXIS MEDBANK TWR MN 10HH-2FH/ 169-24, BD PYXIS MEDBANK TWR MN 10HH-2FM/ 169-25, BD PYXIS MEDBANK TWR MN 12HH-1FF/ 169-26, BD PYXIS MEDBANK TWR MN 12HH-1FH/ 169-27, BD PYXIS MEDBANK TWR MN 12HH-1FM/ 169-28, BD PYXIS MEDBANK TWR MN 14HH/ 169-29, BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM/ 169-138, BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF/ 169-30, BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM/ 169-31, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/ 169-32, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/ 169-33, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/ 169-34, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/ 169-35, BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM/ 169-139, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/ 169-36, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/ 169-37, BD PYXIS MEDBANK TWR MN 2HH-6FM/ 169-38, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/ 169-39, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/ 169-40, BD PYXIS MEDBANK TWR MN 2HM-6FM/ 169-41, BD PYXIS MEDBANK TWR MN 3FH-4FM/ 169-42, BD PYXIS MEDBANK TWR MN 3FH-8HH MEDPASS/ 169-140, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/ 169-43, BD PYXIS MEDBANK TWR MN 4HH-10HM/ 169-44, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/ 169-45, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/ 169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/ 169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/ 169-48, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/ 169-49, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/ 169-50, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/ 169-51, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/ 169-52, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/ 169-53, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/ 169-54, BD PYXIS MEDBANK TWR MN 4HH-5FH / 169-55, BD PYXIS MEDBANK TWR MN 4HH-5FM/ 169-56, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/ 169-57, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/ 169-58, BD PYXIS MEDBANK TWR MN 4HM-5FM/ 169-59, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/ 169-60, BD PYXIS MEDBANK TWR MN 5HH-9HM/ 169-61, BD PYXIS MEDBANK TWR MN 6FH-1FM/ 169-62, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/ 169-63, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/ 169-64, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/ 169-65, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/ 169-66, BD PYXIS MEDBANK TWR MN 6HH-4FH / 169-67, BD PYXIS MEDBANK TWR MN 6HH-4FM/ 169-68, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/ 169-69, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/ 169-70, BD PYXIS MEDBANK TWR MN 6HH-8HM/ 169-71, BD PYXIS MEDBANK TWR MN 7FH/ 169-72, BD PYXIS MEDBANK TWR MN 7FM/ 169-73, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/ 169-74, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/ 169-75, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/ 169-76, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/ 169-77, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/ 169-78, BD PYXIS MEDBANK TWR MN 8HH-3FF/ 169-79, BD PYXIS MEDBANK TWR MN 8HH-3FH/ 169-80, BD PYXIS MEDBANK TWR MN 8HH-3FM/ 169-81, BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P/ 169-82, BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P/ 169-83, BD PYXIS MEDBANK TWR MN CR-12HH-1FH/ 169-84, BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P/ 169-85, BD PYXIS MEDBANK TWR MN CR-14HH-P/ 169-86, BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM/ 169-87, BD PYXIS MEDBANK TWR MN CR-2HH-6FH/ 169-88, BD PYXIS MEDBANK TWR MN CR-2HH-6FM/ 169-89, BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P/ 169-90, BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P/ 169-91, BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P/ 169-92, BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P/ 169-93, BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P/ 169-94, BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P/ 169-95, BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P/ 169-96, BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P/ 169-97, BD

Product Codes/Lot Numbers:

UDI-DI: 10885403512568. Devices Manufactured November 15, 2012 - July, 25 2023 User Guide, DIR 10000414989/DIR 10000414987/DIR 10000414990, Version 00.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2256-2024

Related Recalls