BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201; BD Pyxis CII Safe, Software: V7, REF: 107-223, 107-224, 111-191, 111-192, 111-193, 111-194, 111-195, 111-196, 111-197, 111-198; BD Pyxis CII Safe, Software: V8, 111-199-01, 111-200-01, 111-201-01, 111-202-01, 111-203-01, 111-204-01, 111-205-01, 111-206-01, 107-245-01, 107-246-01; BD Pyxis CII Safe, Software: V9, REF: 107-252-01, 111-215, 111-216, 111-219, 111-221. BD Pyxis CII Safe, Software: ES, REF: 107-254-01, 107-255-01, 1116-00, 1117-00; BD Pyxis CII Safe, Software: N/A, REF: 111-113, 111-114, 111-115, 111-160-201, 111-217, 111-218, 111-220, 111-222; Tripp Lite SUPER6TEL Surge Suppressor, REF: CN 200-025. Used with BD Pyxis CII Safe System: v7.x for WES7 User Guide, DIR 10000145747/ v8.1 User Guide, DIR 10000277635/v9.0 User Guide, DIR 10000354270/ ES User Guide v1.8, DIR 10000418517
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 10885403512520; Devices Manufactured May 04, 2006 - September 26, 2023
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- CareFusion 303, Inc.
- Reason for Recall:
- Automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which could expose users to system downtime and/or potential electric shock.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201; BD Pyxis CII Safe, Software: V7, REF: 107-223, 107-224, 111-191, 111-192, 111-193, 111-194, 111-195, 111-196, 111-197, 111-198; BD Pyxis CII Safe, Software: V8, 111-199-01, 111-200-01, 111-201-01, 111-202-01, 111-203-01, 111-204-01, 111-205-01, 111-206-01, 107-245-01, 107-246-01; BD Pyxis CII Safe, Software: V9, REF: 107-252-01, 111-215, 111-216, 111-219, 111-221. BD Pyxis CII Safe, Software: ES, REF: 107-254-01, 107-255-01, 1116-00, 1117-00; BD Pyxis CII Safe, Software: N/A, REF: 111-113, 111-114, 111-115, 111-160-201, 111-217, 111-218, 111-220, 111-222; Tripp Lite SUPER6TEL Surge Suppressor, REF: CN 200-025. Used with BD Pyxis CII Safe System: v7.x for WES7 User Guide, DIR 10000145747/ v8.1 User Guide, DIR 10000277635/v9.0 User Guide, DIR 10000354270/ ES User Guide v1.8, DIR 10000418517
Product Codes/Lot Numbers:
UDI-DI: 10885403512520; Devices Manufactured May 04, 2006 - September 26, 2023
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2094-2023
Related Recalls
BD Pyxis MedBank Tower REF: 1145-00 Medication cabinet
CareFusion 303
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
BD Pyxis Mini Main REF: 349 Medication cabinet
CareFusion 303
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
BD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinet
CareFusion 303
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.