All Recalls

95,123 total recalls in our database

The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.

Jun 19, 2012 Surgical Instruments Nationwide View Details →

Drager PT 4000 Phototherapy System

Draeger Medical Systems

🏥 Medical Devices Class I - Dangerous

A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip promptly in the event of an internal short-circuit. The possible consequence is the wall outlet circuit breaker would trip first on the device may severely overheat. There have not b

Jul 2, 2012 Other Medical Devices Nationwide View Details →

Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.

Sep 8, 2011 Diagnostic Equipment Nationwide View Details →

In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.

May 22, 2012 Diagnostic Equipment Nationwide View Details →

Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob.

Jun 22, 2012 Diagnostic Equipment Nationwide View Details →

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. On October 11, 2006, GE Healthcare recalled GE OEC InstaTrak 3500 Plus System with Software version 5.2, the surgical Navigation and Visualization Application due to software related issues.

Sep 15, 2006 Surgical Instruments Nationwide View Details →

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

May 22, 2012 Infusion Pumps Nationwide View Details →

superDimension, Inc., is conducting a recall of certain superDimension Systems operating Software Version 4.0 - 4.9. The affected systems are sold under the product code AAS00016-xx. The products being recalled have been found to malfunction when connected to the superDimension Locatable Guide (LG). Specifically, the Systems using Software Version 4.0-4.9 will not recognize the Locatable Guide

Jun 18, 2012 Surgical Instruments Nationwide View Details →

Some of the kits may exhibit a potential defect with the pouch seal. The pouch issue may compromise the sterile barrier.

May 4, 2012 Other Medical Devices View Details →

The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.

May 7, 2012 Surgical Instruments View Details →

ArjoHuntleigh has received a limited number of customer reports concerning injuries related to an exposed metal support bracket when the proactive knee pad assembly is removed.

Jun 29, 2012 Implants & Prosthetics Nationwide View Details →

Terumo CVS has received reports of the Central Control Monitor (CCM) for Terumo Advanced Perfusion System 1 losing partial or full functionality. Symptoms that have been associated with a malfunctioning CCM include: full or partial loss of display, loss of control functions; inability to distinguish various status, alert, alarm conditions, the inability to operate the touch screen, or the CCM

Jun 22, 2012 Other Medical Devices Nationwide View Details →

Edwards is recalling Suction Wand models SPC2081 and S033 because the plastic tip attached to the metal suction wand may become detached during routine use.

Jun 6, 2012 Surgical Instruments Nationwide View Details →

Terumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery time remaining if not subjected to regular use, such as would occur with a backup system. This may result in unexpected system shutdown. Prolonged battery depletion may cause permanent battery damage.

Jun 22, 2012 Other Medical Devices Nationwide View Details →

Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.

Sep 8, 2011 Diagnostic Equipment Nationwide View Details →

The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.

May 7, 2012 Surgical Instruments View Details →

The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.

May 7, 2012 Surgical Instruments View Details →

The pouch label of catalog NL850-1330 indicates an incorrect model and closing pressure of the device. The incorrect label reads: "Model: 12 mm-Medium Pressure" instead of "Model: 16 mm-Low Pressure".

Apr 4, 2012 Other Medical Devices View Details →

ArjoHuntleigh has received a limited number of customer reports concerning injuries related to an exposed metal support bracket when the proactive knee pad assembly is removed.

Jun 29, 2012 Other Medical Devices Nationwide View Details →