The Tyvek pouch label for this device states: 'with holster'; however, this product does not come with a holster. Firm officials met with FDA on 5/17/2012. At which time, they reported various labeling errors with their Button Switch Pen Electrode Electrosurgical /Extendable Pencils.

Jun 5, 2012 Surgical Instruments Nationwide View Details β†’

The firm has received 112 complaints of the Eon Mini IPGs that lost the ability to communicate or recharge due to a workmanship issue resulting in loss of pain relief and subsequent explant. The firm has determined the cause of these reports to be related to process variances in the positioning of the internal battery and printed circuit board, causing a short, and therefore, prematurely depletin

Dec 19, 2011 Other Medical Devices Nationwide View Details β†’

Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump occlusion malfunctions that could result in the inability to use the pump and would then require replacement with back-up equipment. Occlusion malfunctions include stiff/difficult to adjust occlusion, loss of occlusion setting, unequal roller-to-roller occlusion and inability to adjust the occlusion. If it is

Jun 25, 2012 Other Medical Devices Nationwide View Details β†’

Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu

Jun 25, 2012 Other Medical Devices Nationwide View Details β†’

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

May 22, 2012 Other Medical Devices Nationwide View Details β†’

Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.

Sep 8, 2011 Diagnostic Equipment Nationwide View Details β†’

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

May 22, 2012 Infusion Pumps Nationwide View Details β†’

Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.

Jun 5, 2012 Surgical Instruments Nationwide View Details β†’

Firm officials reported to CIN-DO that their Electrosurgical Monopolar Reusable Bayonet Forceps are misbranded in that the label states that the tip size is: (0.5mm) when the tip size actually measures (1.0mm).

Jun 1, 2012 Surgical Instruments Nationwide View Details β†’

HOTLINE 3 Blood and Fluid Warmer( Reorder No. HL-390 ) disposable is not correctly installed , the microswitch may become damaged and unusable

Jun 27, 2012 Infusion Pumps Nationwide View Details β†’

A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.

Feb 20, 2012 Other Medical Devices Nationwide View Details β†’

Stryker Orthopaedics has become aware of the potential for cracking of the PMMA coating on their PM COCR Hip Stem product.

Dec 29, 2011 Implants & Prosthetics Nationwide View Details β†’

Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu

Jun 25, 2012 Other Medical Devices Nationwide View Details β†’

The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.

Feb 20, 2012 Infusion Pumps Nationwide View Details β†’

Firm officials reported to CIN-DO that the outer carton label of this sterile, single use device is misbranded in that it lists an incorrect expiration date. The outer carton label lists the sterile expiration date as: 2012-02-(28) [the manufacture date] ; however the device pouch label lists the sterile date as: 2017-02-(28). The pouch label lists the correct sterile expiration date.

Jun 1, 2012 Surgical Instruments Nationwide View Details β†’

Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob.

Jun 22, 2012 Diagnostic Equipment Nationwide View Details β†’

Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump occlusion malfunctions that could result in the inability to use the pump and would then require replacement with back-up equipment. Occlusion malfunctions include stiff/difficult to adjust occlusion, loss of occlusion setting, unequal roller-to-roller occlusion and inability to adjust the occlusion. If it is

Jun 25, 2012 Other Medical Devices Nationwide View Details β†’

Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.

Sep 8, 2011 Diagnostic Equipment Nationwide View Details β†’