Lack of Assurance of Sterility; potential that a low level of endotoxins may be present in the diluent vials.
All Recalls
95,123 total recalls in our database
Venlafaxine Hydrochloride Tablets, 75 mg*, 100-count tablets per bottle, Rx only, Manufactured by: Cadila Healthcare Ltd., Ahmedabad, India; Distributed by: Zydus Pharmaceuticals USA Inc., Pennington, NJ 08534, NDC 68382-021-01.
Zydus Pharmaceuticals USA
Failed Tablet/Capsule Specifications: Pharmacist complaint of an excessive amount of broken and/or chipped tablets in the bottle.
CGMP Deviations: Shipment of product not approved for release.
Derma-Smoothe/FS (fluocinolone acetonide), 0.01% Topical Oil (Body Oil), Rx only, Hill Dermaceuticals, Inc., Sanford, Florida 32773 --- NDC 28105-150-04
Hill Dermaceuticals
Subpotent; 12 month stability timepoint
Labeling: Not Elsewhere Classified: This product is misbranded because the product is not sterile and the labeling is misleading in relation to sterility claims.
PHYSICIANS TOTAL CARE, TETRACYCLINE, 500 mg, 30 CAPSULES, bottle MFG. BY: TEVA PHARMACEUTICALS USA, SELLERSVILLE PA 18960 NDC 54868-0025-09
Physicians Total Care
Presence of Foreign Substance(s): There is a potential for foreign particulate matter in the API.
Labeling: Not Elsewhere Classified: This product is misbranded because the product is not sterile and the labeling is misleading in relation to sterility claims.
Labeling: Not Elsewhere Classified: This product is misbranded because the product is not sterile and the labeling is misleading in relation to sterility claims.
BS-200 Chemistry Analyzer Designed for clinical laboratory use.
Mindray DS USA, Inc. d.b.a. Mindray North America
An issue has been identified with the BS-200 where clinical chemistry results may be inaccurate.
There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.
Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system
Siemens Medical Solutions USA
The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.
The Dens Instrument Set has been designed to facilitate the anterior internal fixation of the Dens (odontoid process). There are currently no screws specifically indicated to treat Dens fractures via anterior screw fixation therefore the affected product will be removed.
The recall was initiated because Carefusion has confirmed that when the PAS 2000 drawers due to inaccessability of drawers caused by system down time or lock-up.
An issue with the plunger rod seal was identified where the seal had the potential to migrate inside the pump case top during slow rate/ high back pressure infusions.
Intraosseous Fixation System (IFS) 1.7 mm, 2.0 mm & 2.5 mm Cannulated Drill Bit. Intended to stabilize and aid in fixation of fractures.
Orthohelix Surgical Designs
The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire.
STERRAD 200 System Carriage, P/N 10208. Used as a tool to expedite loading and unloading processed items from a STERRAD 200 sterilizer.
Advanced Sterilization Products
Advanced Sterilization Products (ASP) is recalling the STERRAD 200 System Carriage because it can potentially become dislodged when an excessive load is applied.
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
The product package did not contain the correct version of the package insert listing the limitations of use and performance requirements.
Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP CAL is an in vitro diagnostic product for the calibration of the Alkaline Phosphatase method on the Dimension Vista" System.
Siemens Healthcare Diagnostics
Correlation slope bias and accuracy shift
Medtronic Archer Super Stiff Guidewire 0.035 inch (0.89mm), Model ARCSJ200W. Intended to facilitate catheter placement and exchange during procedures in the aorta.
Medtronic Inc. Cardiac Rhythm Disease Management
Medtronic has identified an issue with a specific lot of the Archer 0.035 inch (0.89mm) Super Stiff Guidewire, model ARCSJ200W, where the PTFE (Teflon) coating has the potential to separate from the wire during in-vivo use.