Lactated Ringer's Injection, USP, 1000 mL, Rx only, Manufactured by Hospira, Inc., Lake Forest, IL 60045, NDC 0409-7953-09

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot 16-836-FW, Exp 04/14
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Hospira, Inc.
Reason for Recall:
CGMP Deviations: Shipment of product not approved for release.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Lactated Ringer's Injection, USP, 1000 mL, Rx only, Manufactured by Hospira, Inc., Lake Forest, IL 60045, NDC 0409-7953-09

Product Codes/Lot Numbers:

Lot 16-836-FW, Exp 04/14

Distribution:

Distributed in: NC

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: D-215-2013

Related Recalls

Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.

Feb 19, 2013 Other Medical Devices Nationwide View Details →