All Recalls

95,075 total recalls in our database

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →
🏥 Medical Devices Class I - Dangerous

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →
🏥 Medical Devices Class I - Dangerous

Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.

Nov 4, 2024 Other Medical Devices Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Other Medical Devices Nationwide View Details →

Due to defects on primary packaging pouch that compromise the device's ability to maintain a sterile barrier.

Oct 1, 2024 Other Medical Devices Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →

MEDLINE convenience kits labeled as: 1) GYN LAPAROSCOPY-LF, REF CDS760059I; 2) GYN LAPAROSCOPY CDS #34-RF, REF CDS920099S; 3) GYN MAJOR, REF CDS980998L; 4) ROBOTIC PROCEDURE, REF CDS982128P; 5) ROBOTIC, REF CDS984262N; 6) GYN LAPAROSCOPY, REF CDS984900M; 7) GYN LAPAROSCOPY, REF CDS984900N; 8) DAVINCI PACK-LF, REF DYNJ0101304G; 9) PROSTATE ROBOTIC PACK, REF DYNJ0347372J; 10) URO GYN PACK-LF, REF DYNJ17454N; 11) DB DAVINCI ROBOT PACK-LF, REF DYNJ33592I; 12) ROBOTIC UROLOGY-RF, REF DYNJ38844N; 13) TLH PACK, REF DYNJ42792D; 14) SVMMC GEN ROBOTIC PACK, REF DYNJ43770L; 15) SVMMC ROBOTIC KIDNEY PACK, REF DYNJ43778L; 16) SVMMC GYN ROBOTIC PACK, REF DYNJ43782K; 17) SVMMC ROBOTIC PROSTATE PACK, REF DYNJ43862J; 18) DA VINCI HYSTERECTOMY, REF DYNJ44863O; 19) DA VINCI PROSTATE/COLPOPEXY, REF DYNJ44864S; 20) LAP-GYN PACK, REF DYNJ44876L; 21) LAP-GYN PACK, REF DYNJ44876M; 22) ROBOT PACK, REF DYNJ45184F; 23) GYN ROBOTIC PACK, REF DYNJ47766Q; 24) ST. ANNE'S MULTI PORT PACK, REF DYNJ48638G; 25) RBTC COLON/KIDNEY PK RFID-LF, REF DYNJ52397M; 26) GYN LAP PACK, REF DYNJ52524C; 27) ROBOTIC PACK, REF DYNJ53650D; 28) LAPAROTOMY PACK, REF DYNJ59079D; 29) ROBOTIC GYN PACK, REF DYNJ62107B; 30) GYN UROL BASIC LAPAROSCOPY, REF DYNJ63326D; 31) FAIRFAX DAVINCI TRAY, REF DYNJ63606C; 32) GYN LAP PACK, REF DYNJ64234B; 33) ROBOT GYN SHARED, REF DYNJ64793A; 34) ROBOT GYN SHARED, REF DYNJ64793B; 35) ROBOT GYN SHARED, REF DYNJ64793C; 36) GYN DAVINCI PACK, REF DYNJ65373A; 37) ROBOT PACK, REF DYNJ67190C; 38) TUBAL LIGATION PACK, REF DYNJ68563A; 39) TUBAL LIGATION PACK, REF DYNJ68563B; 40) DAVINCI GYN PACK, REF DYNJ69169A; 41) GYN LAPAROSCOPY PACK, REF DYNJ69490C; 42) ROBOTIC PACK, REF DYNJ84776A; 43) ROBOTIC LAVH, REF DYNJ901281G; 44) ROBOTIC LAP CHOLE, REF DYNJ901282J; 45) GYN LAPAROTOMY, REF DYNJ902498J; 46) WMC ROBOTIC BASE-LF, REF DYNJ902583M; 47) HHI ROBOTIC CDS, REF DYNJ902953G; 48) HHI ROBOTIC CDS, REF DYNJ902953I; 49) HHI ROBOTIC CDS, REF DYNJ902953J; 50) HHI ROBOTIC CDS, REF DYNJ902953K; 51) ROBOTIC GYN & GU, REF DYNJ903459F; 52) SM DAVINCI, REF DYNJ903539G; 53) WEST GYN LAPAROSCOPY, REF DYNJ904490I; 54) ROBOTIC GYN, REF DYNJ904792C; 55) ROBOTIC GENERAL, REF DYNJ904793B; 56) ROBOT, REF DYNJ906112B; 57) ROBOTIC URO LH, REF DYNJ906445A; 58) ROBOTIC URO LH, REF DYNJ906445B; 59) ROBOTIC URO LH, REF DYNJ906445C; 60) WMC THORACIC ROBOTIC, REF DYNJ907153C; 61) DAVINCI, REF DYNJ907392A; 62) DAVINCI, REF DYNJ907392B; 63) DAVINCI ROBOTICS-HERNIA, REF DYNJ907396B; 64) DAVINCI ROBOTICS-HERNIA, REF DYNJ907396C; 65) GYN LAPAROSCOPY - NTSC, REF DYNJ907407A; 66) GYN LAPAROSCOPY - NTSC, REF DYNJ907407B; 67) ROBOTIC-LF, REF DYNJ907910; 68) ROBOTIC-LF, REF DYNJ907910A; 69) GU DAVINCI KIT, REF DYNJ908783; 70) GU DAVINCI KIT, REF DYNJ908783A; 71) GU DAVINCI PACK - LF, REF DYNJ909398A; 72) GYN LAPAROSCOPY, REF DYNJ910084; 73) MHPB GYN ROBOTIC PACK, REF DYNJ910125;

MEDLINE INDUSTRIES, LP - Northfield

🏥 Medical Devices Class I - Dangerous

Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.

Oct 9, 2024 Other Medical Devices Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →

Due to defects on primary packaging pouch that compromise the device's ability to maintain a sterile barrier.

Oct 1, 2024 Surgical Instruments Nationwide View Details →

Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.

Oct 9, 2024 Other Medical Devices Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →

MEDLINE convenience kits labeled as: 1) MAJOR LITHOTOMY CDS-LF, REF CDS760056G; 2) LAVH CDS, REF CDS983411J; 3) MAJOR VAGINAL, REF CDS983497L; 4) DAY SURG LARGE VAG CDS, REF CDS984995N; 5) GYN LAPAROTOMY HARPER PACK-LF, REF DYNJ0160684F; 6) D&C/HYSTEROSCOPY PACK, REF DYNJ01921M; 7) HYSTERECTOMY PACK-LF, REF DYNJ0488471K; 8) VAG HYST PACK-LF, REF DYNJ0532873M; 9) VAG HYST PACK-LF, REF DYNJ0532873N; 10) VAGINAL PROCEDURES PACK-LF, REF DYNJ0750737Q; 11) MINOR VAGINAL #76-RF, REF DYNJ27434R; 12) DB VAG HYSTERECTOMY-LF, REF DYNJ33590I; 13) VAG HYST PACK, REF DYNJ36628L; 14) VAG HYST PACK, REF DYNJ36628M; 15) ROBOTIC HYSTO URO PACK, REF DYNJ40623F; 16) ABDOMINAL HYST PACK, REF DYNJ44848L; 17) DA VINCI HYSTERECTOMY, REF DYNJ44863Q; 18) VAGINAL HYSTERECTOMY PACK, REF DYNJ44900O; 19) VAGINAL HYSTERECTOMY PACK, REF DYNJ44900P; 20) URO GYN, REF DYNJ50047G; 21) LAVH NEW BRAUNFELS, REF DYNJ61136D; 22) ABDOMINAL PERINEAL, REF DYNJ63371D; 23) ABDOMINAL HYST FFX, REF DYNJ63631A; 24) VAGINAL HYSTERECTOMY PACK, REF DYNJ67194B; 25) LAVH PACK, REF DYNJ67210A; 26) VAG MINOR PACK, REF DYNJ82216; 27) VAG MINOR PACK, REF DYNJ82216B; 28) ABDOMINAL HYSTERECTOMY, REF DYNJ900091F; 29) VAGINAL, REF DYNJ900292K; 30) GYN LAPAROTOMY, REF DYNJ900296J; 31) LAVH, REF DYNJ900298J; 32) LAVH, REF DYNJ900298K; 33) LITHOTOMY-LF, REF DYNJ900721C; 34) MAJOR VAGINAL-LF, REF DYNJ901807I; 35) ABDOMINAL HYSTERECTOMY, REF DYNJ906478F; 36) VAGINAL HYST, REF DYNJ907220A; 37) PRENTICE GYNE ONCOLOGY KIT, REF DYNJ908792; 38) PRENTICE TOTAL ABD HYST KIT, REF DYNJ908795; 39) PRENTICE TOTAL ABD HYST KIT, REF DYNJ908795B; 40) PRENTICE VAG HYST KIT, REF DYNJ908796; 41) PRENTICE VAG HYST KIT, REF DYNJ908796B; 42) KIT LAVH CENTRAL DUPAGE, REF DYNJ908825; 43) KIT PFANNENSTIEL GYN, REF DYNJ908826; 44) KIT VAGINAL HYSTERECTOMY, REF DYNJ908837

MEDLINE INDUSTRIES, LP - Northfield

🏥 Medical Devices Class I - Dangerous

Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.

Oct 9, 2024 Other Medical Devices Nationwide View Details →