Atellica CI Analyzer. Catalog Numbers: 10947347.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI Number: 630414229560; Catalog Number: 10947347. Software version 1.28.51
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Healthcare Diagnostics, Inc.
Reason for Recall:
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Atellica CI Analyzer. Catalog Numbers: 10947347.

Product Codes/Lot Numbers:

UDI-DI Number: 630414229560; Catalog Number: 10947347. Software version 1.28.51

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0597-2025

Related Recalls

Class I - Dangerous

Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results

Mar 18, 2026 Other Medical Devices Nationwide View Details →

A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.

Feb 23, 2026 Other Medical Devices Nationwide View Details →