Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
All Recalls
92,900 total recalls in our database
Discovery MR950, with affected software versions: 7T23.0; Nuclear Magnetic Resonance Imaging System - research use only
GE Medical Systems
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
SIGNA Artist, with affected software versions: DV26.0, DV27.1, DV28.1, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE Medical Systems
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
SIGNA Hero, with affected software versions: PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE Medical Systems
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
SIGNA Premier, with affected software versions: RX27.0 to RX27.3, RX28.0, RX29.0 to RX29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE Medical Systems
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
SIGNA MAGNUS, with affected software versions: MR29.1, RX29.1; Nuclear Magnetic Resonance Imaging System - research use only
GE Medical Systems
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
Grafts were found to have been approved, released, shipped and implanted with an invalid test results.
IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway
1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE Medical Systems
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P, REF: 169-93 containing software version 3.9.1.9
CareFusion 303
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
Luminos dRF Max- Intended to visualize anatomical structures by converting an x-ray pattern into a visible image. Model Number: 10762471
Siemens Medical Solutions USA
The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
American Contract Systems Laparoscopy Litho Std Pack convenience kit and American Contract Systems Lap Chole convenience kit
American Contract Systems
Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility
SIGNA Prime, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System
GE Medical Systems
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
American Contract Systems Special Procedure Pack convenience kit
American Contract Systems
Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility
American Contract Systems Cataract Pack convenience kit
American Contract Systems
Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
American Contract Systems Extremity convenience kit, American Contract Systems Hand Pack convenience kit, and American Contract Systems Total Knee Pack convenience kit
American Contract Systems
Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility