Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI 15099590201937 Lots 338921 339071 372223 439138 439205 439206 439207 439650 439801 439889 440026 440090 472012 472092 472136 472154
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter Inc.
- Reason for Recall:
- The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.
Product Codes/Lot Numbers:
UDI-DI 15099590201937 Lots 338921 339071 372223 439138 439205 439206 439207 439650 439801 439889 440026 440090 472012 472092 472136 472154
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0673-2025
Related Recalls
Access Total T4 Calibrator, Catalog No. 33805
Beckman Coulter
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
MicroScan Neg MIC 3J REF C54814
Beckman Coulter
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics