Baxter healthcare is recalling specific product codes of Vascu-Guard Peripheral Vascular Patch due to complaints received for difficulty in distinguishing the smooth from rough surface.
All Recalls
95,114 total recalls in our database
The Auxiliary Serial Input/ Output Modes on the Advisor do not function as expected, resulting in data formatting issues or incorrect data being transmitted in or out of the serial port into a data collection system.
Coreva Health Science, LLC. is recalling ActCel Hemostatic Dressing due to customer literature and web sites promoting that it can help control bleeding from open wounds and in body cavities. ActCel is indicated as a prescription device for management of topical wounds and to temporarily control external surface bleeding.
The product failed to meet the internal established statistical test requirement for heparin activity.
affected parts and lots of the Cannulated Drill Bits have the potential to break during use
The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.
Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin with plasma activator for use in the determination of the activated partial thromboplastin time (APTT) and other coagulation procedures requiring an activated partial thromboplastin reagent.
Siemens Healthcare Diagnostics
Recovery for assayed controls are out of assigned ranges. A drift >3 seconds in the normal APTT range and/or >15% in the pathological range has been observed. Patient samples with recoveries close to the medical decision points could exhibit a deviation up to 4 seconds for the normal range and up to 33% for the pathological range.
The supplier of the Devon Light Glove, a component of the Alcon Custom PAK, recalled the gloves due to the potential for splits and holes on the glove. Should the user be unaware that the Light Glove is torn/split, surgical site infections are possible.
A potential exists for order transactions from interfaced HIS (Hospital Information System) systems to be lost during a restart of interface processes when using the Order Batching feature, leading to diagnosis delay.
Brilliance iCT Computed Tomography X-ray system
Philips Medical Systems (Cleveland)
A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.
Factor deficient plasma for the Siemens BCS¿ / BCS¿ XP System. In vitro diagnostic reagent for the determination of the activity of coagulation factor IX in human plasma by coagulometric methods.
Siemens Healthcare Diagnostics
Discrepancies for Siemens BCS / BCS XP with dilution studies; erroneously high Factor IX results at the lower end of the measuring range.
The Immunodiagnostic Products Signal Reagent, Lot 3082 as the reagent packs may contain blank labels.
Incorrect expiration date on label
PRI Femoral Impactor Block for knee prosthesis surgery. Orthopedic Manual Surgical Instrument
Zimmer
Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response.
Multi-stage treatment option for the EC-5000 Operator's Manual was not approved by the FDA.
Incorrect dose calculation for Regions of Interest (ROIs) defined on a secondary image series.
Syngo.via and Syngo.x ; a software solution intended to be used for viewing, manipulation, communication, and storage of medical images.
Siemens Medical Solutions USA
measurements drawn on the 2nd and subsequent images of the series are not visible on printouts when the series is sent to print.
Zimmer CAS has determined that the potential exists for the spikes on the iASSIST Tibial Alignment Guide to bend or break during insertion or extraction from the tibia during use.
Dimension Vista Systems LOCI Thyroid Stimulating Hormone; the TSH method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone in human serum and plasma on the Dimension Vista System.
Siemens Healthcare Diagnostics
Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a Flex well.
Children's furniture
Associate Yang Enterprise Co. Ltd., of Vietnam
The plastic restraint strap used to attach armoires, combos, dressers and hutches to a wall can break and allow the unit to tip over. Falling furniture can result in a wide range of injuries to young children, from soft tissue bruising to broken bones, head injuries and death by suffocation when a child is pinned under a heavy piece of furniture.