Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots. Part numbers: 00-2253-020-45 INTERLOCKING IM SCREW 4.5MM DIA X 20MM 00-2253-022-45 INTERLOCKING IM SCREW 4.5MM DIA X 22.5MM 00-2253-025-45 INTERLOCKING IM SCREW 4.5MM DIA X 25MM 00-2253-027-45 M/DN Intramedullary Fixation 4.5mm Dia. Cortical Screw 00-2253-030-45 INTERLOCKING IM SCREW 4.5MM DIA X 30MM 00-2253-032-45 INTERLOCKING IM SCREW 4.5MM DIA X 32.5MM 00-2253-035-45 INTERLOCKING IM SCREW 4.5MM DIA X 35MM 00-2253-037-45 INTERLOCKING IM SCREW 4.5MM DIA X 37.5MM 00-2253-040-45 INTERLOCKING IM SCREW 4.5MM DIA X 40MM 00-2253-042-45 INTERLOCKING IM SCREW 4.5MM DIA X 42.5MM 00-2253-045-45 INTERLOCKING IM SCREW 4.5MM DIA X 45MM 00-2253-047-45 INTERLOCKING IM SCREW 4.5MM DIA X 47.5MM 00-2253-050-45 INTERLOCKING IM SCREW 4.5MM DIA X 50MM 00-2253-052-45 INTERLOCKING IM SCREW 4.5MM DIA X 52.5MM 00-2253-055-45 INTERLOCKING IM SCREW 4.5MM DIA X 55MM 00-2253-057-45 INTERLOCKING IM SCREW 4.5MM DIA X 57.5MM 00-2253-060-45 INTERLOCKING IM SCREW 4.5MM DIA X 60MM 00-2253-062-45 INTERLOCKING IM SCREW 4.5MM DIA X 62.5MM 00-2253-065-45 INTERLOCKING IM SCREW 4.5MM DIA X 65MM 00-2253-067-45 INTERLOCKING IM SCREW 4.5MM DIA X 67.5MM 00-2253-070-45 INTERLOCKING IM SCREW 4.5MM DIA X 70MM 00-2253-072-45 INTERLOCKING IM SCREW 4.5MM DIA X 72.5MM 00-2253-075-45 INTERLOCKING IM SCREW 4.5MM DIA X 75MM 00-2253-077-45 INTERLOCKING IM SCREW 4.5MM DIA X 77.5MM 00-2253-080-45 INTERLOCKING IM SCREW 4.5MM DIA X 80MM 00-2253-082-45 INTERLOCKING IM SCREW 4.5MM DIA X 82.5MM 00-2253-085-45 INTERLOCKING IM SCREW 4.5MM DIA X 85MM 00-2253-087-45 INTERLOCKING IM SCREW 4.5MM DIA X 87.5MM 00-2253-090-45 INTERLOCKING IM SCREW 4.5MM DIA X 90MM
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Zimmer, Inc.
- Reason for Recall:
- The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.
Product Codes/Lot Numbers:
All lots. Part numbers: 00-2253-020-45 INTERLOCKING IM SCREW 4.5MM DIA X 20MM 00-2253-022-45 INTERLOCKING IM SCREW 4.5MM DIA X 22.5MM 00-2253-025-45 INTERLOCKING IM SCREW 4.5MM DIA X 25MM 00-2253-027-45 M/DN Intramedullary Fixation 4.5mm Dia. Cortical Screw 00-2253-030-45 INTERLOCKING IM SCREW 4.5MM DIA X 30MM 00-2253-032-45 INTERLOCKING IM SCREW 4.5MM DIA X 32.5MM 00-2253-035-45 INTERLOCKING IM SCREW 4.5MM DIA X 35MM 00-2253-037-45 INTERLOCKING IM SCREW 4.5MM DIA X 37.5MM 00-2253-040-45 INTERLOCKING IM SCREW 4.5MM DIA X 40MM 00-2253-042-45 INTERLOCKING IM SCREW 4.5MM DIA X 42.5MM 00-2253-045-45 INTERLOCKING IM SCREW 4.5MM DIA X 45MM 00-2253-047-45 INTERLOCKING IM SCREW 4.5MM DIA X 47.5MM 00-2253-050-45 INTERLOCKING IM SCREW 4.5MM DIA X 50MM 00-2253-052-45 INTERLOCKING IM SCREW 4.5MM DIA X 52.5MM 00-2253-055-45 INTERLOCKING IM SCREW 4.5MM DIA X 55MM 00-2253-057-45 INTERLOCKING IM SCREW 4.5MM DIA X 57.5MM 00-2253-060-45 INTERLOCKING IM SCREW 4.5MM DIA X 60MM 00-2253-062-45 INTERLOCKING IM SCREW 4.5MM DIA X 62.5MM 00-2253-065-45 INTERLOCKING IM SCREW 4.5MM DIA X 65MM 00-2253-067-45 INTERLOCKING IM SCREW 4.5MM DIA X 67.5MM 00-2253-070-45 INTERLOCKING IM SCREW 4.5MM DIA X 70MM 00-2253-072-45 INTERLOCKING IM SCREW 4.5MM DIA X 72.5MM 00-2253-075-45 INTERLOCKING IM SCREW 4.5MM DIA X 75MM 00-2253-077-45 INTERLOCKING IM SCREW 4.5MM DIA X 77.5MM 00-2253-080-45 INTERLOCKING IM SCREW 4.5MM DIA X 80MM 00-2253-082-45 INTERLOCKING IM SCREW 4.5MM DIA X 82.5MM 00-2253-085-45 INTERLOCKING IM SCREW 4.5MM DIA X 85MM 00-2253-087-45 INTERLOCKING IM SCREW 4.5MM DIA X 87.5MM 00-2253-090-45 INTERLOCKING IM SCREW 4.5MM DIA X 90MM
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1694-2015
Related Recalls
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.