Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
All Recalls
92,900 total recalls in our database
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
25-LHP-928, HeNe Laser System, 25 mW
PACIFIC LASERTECH
These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11097543
Siemens Healthcare Diagnostics
The potential for biased quality control (QC) and patient results when using Atellica CH UCFP lot 130414 and ADVIA Chemistry UCFP lot 140 reagents.
N-LHP-928, HeNe Laser System, 35 mW
PACIFIC LASERTECH
These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
Dimension VistaΒΏ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
Siemens Healthcare Diagnostics
During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versions
Ion Beam Applications S.A.
Safety Parameters Verification Mechanisms can be deactivated in clinical runtime
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Medline convenience kits containing semi-rigid suction liners labeled as follows: MTHZ GYN LAPAROSCOPY, Pack Number DYNJ909739
MEDLINE INDUSTRIES, LP - Northfield
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
Medline Convenience kits labeled as: 1) NHP BREAST BIOPSY PACK, Pack Number DYNJ17400B; 2) BREAST BIOPSY TRAY, Pack Number DYNJ32049A ; 3) BREAST BIOPSY PACK, Pack Number DYNJ44987G ; 4) BREAST BIOPSY PACK, Pack Number DYNJ66288
MEDLINE INDUSTRIES, LP - Northfield
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Medline Convenience kits labeled as: 1) CYSTO PACK, Pack Number DYNJ42742D; 2) CYSTO PACK, Pack Number DYNJ42799
MEDLINE INDUSTRIES, LP - Northfield
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Brand Name: PROBEAT Product Name: PROBEAT-FR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-FR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. Component: N/A
Hitachi America, Ltd., Particle Therapy Division
During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.