Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
Class I - DangerousWhat Should You Do?
- Check if you have this product: Siemens Material Number (SMN): 10445146; UDI-DI: 00842768015717; Lot Number: 23206BA;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
Product Codes/Lot Numbers:
Siemens Material Number (SMN): 10445146; UDI-DI: 00842768015717; Lot Number: 23206BA;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-3065-2024
Related Recalls
Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
Siemens Healthcare Diagnostics
A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Siemens Healthcare Diagnostics
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
Siemens Healthcare Diagnostics
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.