Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
All Recalls
95,089 total recalls in our database
Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
RUSCH, Pocket PAC IC Intermittent Cath Kit, 8 FR, 12 FR, 14 FR, 10 FR and 16 FR, Rx Only, Sterile, Intermittent self-catheterization
Teleflex Medical
Labeling Inconsistency: The products have been labeled with the incorrect expiration date which exceeds the actual, validated 35-month shelf life.
DELFIA hFSH kit. In Vitro Diagnostic for the determination of hFSH in serum
PerkinElmer Health Sciences
A DELFIA hFSH kit lot failed real time stability testing. The real time stability data demonstrated a decrease in hFSH concentrations between -4.8% to -11.3% for control sera. The -11.3% bias was observed at an hFSH target concentration of 101 IU/L. The change observed in the serum samples within the range 1.76 67.0 IU/L hFSH averaged -2.9%. The reference range of the assay spans 0.60 to 98.0 IU/L.
Ultroid Sterile Disposable Probes sold individually
Ultroid Technologies
Products are not designed, qualified, manufactured, and/or managed under a state of control per internal quality system procedures and external laws, regulations, directives, standards, and/or guidance (such as 21 CFR 820 Quality System Regulation)
Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
Siemens Healthcare Diagnostics has confirmed that the IMMULITE/ IMMULITE 1000 Systems PSA assays are not meeting the current High-dose Hook Effect expectation in the Instructions for Use.
Medtronic is initiating a voluntary product recall for specific lot numbers of the Affinity FusionΒΏ Oxygenators with BalanceΒΏ1 Biosurface. These are distributed as stand-alone devices, or as a combination unit with the Affinity Fusion Cardiotomy/Venous Reservoir, or as part of tubing packs. Medtronic has identified an out-of-specification condition exhibiting additional plastic (flash) in the arterial sampling port adjacent to the arterial outlet port of the oxygenator.
A limited number of product pouches may not have been sealed during packaging.
Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
Possibility that a patient or user may be exposed to a very small level of touch current.
GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma GmbH, Prof.-Kuntscher-Strasse 1-5, Schonkirchen, Germany 24232; T2 K-Wire, Sterile R Intended for the temporary stabilization of bone segments or fragments
Stryker Howmedica Osteonics
A review of packaging revealed the seal integrity of the pouch may be compromised. More specifically, there is a potential that the sterile pouch is not sealed at one end due to a manufacturing error.
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
Ansell Healthcare Products is initiating a voluntary recall involving 15 cases of Gammex Non Latex Sensitive Surgical Gloves because the packaged contents incorrectly included a natural rubber Latex surgical glove.
Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication.
Nuvectra is conducting a recall due to two clinical risks that are associated with certain percutaneous leads currently being used in the Algovita Spinal Cord Stimulation System. It has been reported that in some percutaneous leads, the stylet protruded beyond the distal lead tip.
A limited number of product pouches may not have been sealed during packaging.
A limited number of product pouches may not have been sealed during packaging.
Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.