Alaris System PC Unit Model 8000 modular infusion pump and monitoring system
Class I - DangerousWhat Should You Do?
- Check if you have this product: All serial numbers
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- CareFusion 303, Inc.
- Reason for Recall:
- If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Alaris System PC Unit Model 8000 modular infusion pump and monitoring system
Product Codes/Lot Numbers:
All serial numbers
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2740-2020
Related Recalls
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
BD Pyxis Pro MedStation Main, REF: 1155-00
CareFusion 303
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.