Alaris System PC Unit Model 8000, a modular infusion pump and monitoring system

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers. ** US Serial numbers: 2326370, 2326382, 2326385, 2326387, 3668143, 2326400, 2326401, 2326402, 2326416, 2326418, 2326419, 2326423, 2326647, 2326664, 2326819, 2326821, 2326834, 2326837, 2326842, 2326873, 2326882, 2326883, 2326926, 9300064, 9300081, 9300083, 9300117, 9300149, 9300162, 9300163, 9300164, 9300171, 9300174, 9300176, 9300284, 9300302, 9300313, 9300317, 9300333, 9300334, 9300485, 3669844, 9300531, 9300532, 9300533, 9300534, 9300535, 9300538, 3686560, 4011408, 3696229, 3000933, 4122102, 4123517, 4123611, 4123651, 4123981, 4124350, 4124764, 4124999, 4125045, 4125149, 4125888, 4125923, 4126044, 4126051, 4126072, 4126073, 4126076, 4126077, 4126092, 4126093, 4126127, 4126128, 4126131, 4126146, 4126147, 4126148, 4126149, 4126151, 4126205, 4126206, 4126208, 4126212, 4126213, 4126214, 4126285, 4126286, 4126287, 4126288, 4126290, 4126299, 4126302, 4126315, 4126318, 4126358, 4126360, 4126387, 4126390, 4126398, 4126400, 4126424, 4126427, 4126428, 4126436, 4126437, 4126438, 4126449, 4126450, 4126472, 4126473, 4126474, 4126475, 4126478, 4126504, 4126512, 4126535, 4126536, 4126537, 4126539, 4126589, 4126590, 4126591, 4126595, 4126596, 4126597, 4126651, 4126652, 4126653, 4126654, 4126655, 4126656, 4126657, 4126696, 4126697, 4126698, 4126699, 4126700, 4126703, 4126704, 4126705, 4126738, 4126740, 4126741, 4126751, 4126761, 4126765, 4126774, 4126795, 4126844, 4126845, 4126887, 4127208, 4127355, 3640679, 3640680, 3641766, 3641790, 3642182, 3642188, 3642194, 3642208, 3642221, 3642227, 3642234, 3642235, 3642251, 3642264, 3642278, 3643225, 3643235, 3643236, 3643237, 3649470, 3652550, 3653689, 3653709, 3653727, 3653732, 3653739, 3653744, 3653747, 3653755, 3653773, 3665625, 3665629, 3665642, 3665645, 3665647, 3665649, 3665652, 3665657, 3665659, 3665666, 3665698, 3665705, 3665706, 3666423, 3666452, 3666454, 3666499, 3666500, 3666502, 3666511, 3667676, 3667687, 3667692, 3667696, 3667700, 3667704, 3667725, 3667727, 3667749, 3667768, 3667770, 3668134, 3668136, 3668140, 3668142, 3668145, 3668147, 3668148, 3668149, 3668152, 3668153, 3668154, 3668155, 3668156, 3665646, 3668158, 3668161, 3668165, 3668167, 3668168, 3668169, 3668170, 3668171, 3668179, 3668198, 3668199, 3668218, 3668221, 3668916, 3668917, 3668918, 3668919, 3668920, 3668923, 3668924, 3668927, 3668929, 3668935, 3668936, 3668939, 3668940, 3668941, 3668942, 3668944, 3668945, 3668946, 3668947, 3668949, 3668950, 3668951, 3668952, 3668959, 3668963, 3668965, 3668966, 3668971, 3668972, 3668974, 3668976, 3668977, 3668978, 3668981, 3668982, 3668984, 3668989, 3668990, 3668991, 3668992, 3668993, 3668996, 3668997, 3668998, 3669000, 3669001, 3669002, 3669005, 3669006, 3669007, 3669008, 3669009, 3669010, 3669012, 3669697, 3669700, 3669701, 3669702, 3669703, 3669704, 3669015, 3669707, 3669708, 3669711, 3669713, 3669715, 3669717, 3669718, 3669720, 3669721, 3669723, 3669724, 3669725, 3669726, 3669727, 3669728, 3669729, 3669730, 3669731, 3669733, 3669741, 3669743, 3669745, 3669746, 3669747, 3669750, 3669753, 3669754, 3669755, 3669757, 3669758, 3669762, 3669764, 3669765, 3669766, 3669767, 3669769, 3669772, 3669777, 3669779, 3669780, 3669782, 3669786, 3669787, 3669789, 3669792, 3669793, 3669794, 3669795, 3669799, 3669800, 3669801, 3669803, 3669804, 3669806, 3669809, 3669815, 3669816, 3669818, 3669824, 3669825, 3669830, 3669832, 3669833, 3669834, 3669836, 3669841, 3669843, 3669845, 3672182, 3672183, 3672184, 3672185, 3672187, 3672195, 3672196, 3672197, 3672199, 3672200, 3672201, 3672204, 3672207, 3672211, 3672212, 3672213, 3672214, 3672216, 3672217, 3672218, 3672219, 3672220, 3672223, 3672225, 3672235, 3672243, 3672249, 3672255, 3672258, 3672259, 3672260, 3672262, 3672263, 3672264, 3672265, 3672266, 3672270, 3672276, 3672277, 3672280, 3673895, 3673896, 3673897, 3673898, 3673901, 3673902, 3673903, 3673904, 3673907, 3673908, 3673912, 3673914, 3673915, 3673918, 3673919, 3673920, 3673921, 3673922, 3673924, 3673926, 3673928, 3673929, 3673930, 3673932, 3673934, 3673935, 3673937, 3673938, 3673940, 3673943, 3673946, 3673947, 3673949, 3673951, 3673952, 3673954, 3673955, 3673956, 3673958, 3673959, 3673962, 3673971, 3673972, 3673973, 3673974, 3673981, 3673992, 3673993, 3673994, 3674424, 3674428, 3674429, 3674430, 3674441, 3674444, 3674445, 3674447, 3674448, 3674449, 3674450, 3674476, 3674479, 3674480, 3674483, 3674485, 3674489, 3674494, 3674495, 3674496, 3674498, 3674499, 3674501, 3674452, 3674453, 3674454, 3674455, 3674461, 3674473, 3674503, 3674504, 3674505, 3674507, 3674509, 3674510, 3674511, 3674512, 3674513, 3674517, 3674519, 3674520, 3674521, 3675314, 3675316, 3675317, 3675320, 3675323, 3675324, 3675326, 3675330, 3675334, 3675335, 3675337, 3675338, 3675339, 3675341, 3675342, 3675343, 3675344, 3675348, 3675349, 3675350, 3675352, 3675357, 3675361, 3675369, 3675373, 3675374, 3675377, 3675378, 3675379, 3675381, 3675382, 3675384, 3675386, 3675390, 3675393, 3675394, 3675395, 3675401, 3675406, 3675410, 3675411, 3676437, 3676441, 3676442, 3676445, 3676446, 3676457, 3676459, 3676463, 3676465, 3676467, 3676476, 3676478, 3676480, 3676481, 3676482, 3676485, 3676493, 3676494, 3676500, 3676501, 3676514, 3676517, 3676519, 3676525, 3676526, 3676527, 3676528, 3676529, 3676530, 3676531, 3676532, 3676533, 3676534, 3676535, 3676536, 3677593, 3677595, 3677596, 3677597, 3677598, 3677599, 3677600, 3677601, 3677602, 3677603, 3677605, 3677606, 3677607, 3677608, 3677609, 3677610, 3677611, 3677612, 3677614, 3677616, 3677617, 3677618, 3677619, 3677620, 3677621, 3677623, 3677624, 3677625, 3677626, 3677627, 3677628, 3677629, 3677630, 3677631, 3677633, 3677634, 3677635, 3677636, 3677638, 3677639, 3677640, 3677642, 3677644, 3677645, 3677649, 3677650, 3677652, 3677653, 3677654, 3677656, 3677657, 3677658, 3677659, 3677660, 3677662, 3677664, 3677665, 3677666, 3677667, 3677668, 3677669, 3677670, 3677673, 3677674, 3677676, 3677677, 3677679, 3677680, 3677681, 3677682, 3677683, 3677684, 3677685, 3677686, 3677687, 3677688, 3677689, 3677691, 3686517, 3686520, 3686521, 3686522, 3686523, 3686527, 3686528, 3686529, 3686530, 3686533, 3686535, 3686537, 3686538, 3686540, 3686541, 3686544, 3686545, 3686546, 3686547, 3686548, 3686552, 3686553, 3686554, 3686555, 3686557, 3686558, 3686566, 3686567, 3686569, 3686570, 3686571, 3686572, 3686574, 3686577, 3686580, 3686581, 3686587, 3686593, 3686596, 3686597, 3686600, 3686605, 3686610, 3687060, 3687063, 3687064, 3687065, 3687068, 3687070, 3687077, 3687082, 3687083, 3687085, 3687087, 3687089, 3687090, 3687091, 3687092, 3687093, 3687098, 3687103, 3687108, 3687109, 3687111, 3687113, 3687117, 3687120, 3687136, 3687139, 3687144, 3688474, 3688479, 3688480, 3688481, 3688482, 3688483, 3688486, 3688317, 3688490, 3688493, 3688494, 3688495, 3688496, 3688505, 3688507, 3688515, 3688518, 3688523, 3688529, 3688537, 3688547, 3688794, 3688797, 3688849, 3688856, 3688863, 3688877, 3688884, 3688900, 3688904, 3688905, 3688929, 3688939, 3688941, 3688948, 3688949, 3688950, 3688958, 3688960, 3688961, 3688966, 3688970, 3688981, 3688983, 3688985, 3689628, 3689650, 3689652, 3689653, 3689659, 3689660, 3689684, 3689685, 3689697, 3689702, 3689706, 3689713, 3689716, 3689749, 3689756, 3689766, 3689805, 3689806, 3690533, 3690534, 3690540, 3690554, 3690556, 3690564, 3690569, 3690585, 3690588, 3690589, 3690599, 3690601, 3691431, 3691443, 3691453, 3690619, 3690512, 3690521, 3691485, 3691525, 3691533, 3691567, 3691593, 3691595, 3691596, 3691599, 3692507, 3692528, 3692540, 3692552, 3692711, 3692737, 3692744, 3693072, 3692753, 3693088, 3693099, 3693107, 3693118, 3693128, 3693136, 3693145, 3693148, 3693161, 3693166, 3693173, 3693174, 3693175, 3693188, 3693190, 3693191, 3693205, 3693207, 3693224, 3693245, 3693802, 3693804, 3693811, 3693815, 3693826, 3693827, 3693834, 3693838, 3693839, 3693872, 3694258, 3694261, 3694272, 3694276, 3694279, 3694287, 3694293, 3694296, 3694301, 3693883, 3694396, 3694397, 3694399, 3694401, 3694405, 3694408, 3694410, 3694411, 3694417, 3694423, 3694435, 3694436, 3694437, 3694438, 3694440, 3694447, 3694448, 3694451, 3694832, 3694834, 3694845, 3694848, 3695674, 3695676, 3695680, 3695683, 3695686, 3695689, 3695690, 3695692, 3695694, 3695695, 3695696, 3695697, 3694897, 3694899, 3694900, 3694901, 3694903, 3694904, 3694908, 3694909, 3694915, 3694916, 3694917, 3694919, 3694920, 3694921, 3694923, 3694924, 3694868, 3694874, 3694882, 3694884, 3694889, 3694892, 3694894, 3694896, 3695718, 3695730, 3695731, 3695751, 3696127, 3696176, 3696774, 3696792, 3696804, 3696833, 3697438, 3697476, 3697532, 3697546, 3698216, 3698978, 3699035, 3699413, 3699416, 3699442, 3699445, 3699448, 3699450, 3699487, 3699489, 3699490, 3699494, 3699499, 3699500, 3699502, 3699504, 3699505, 3699521, 3699523, 3699525, 3699528, 3699529, 3700330, 3700334, 3700336, 3700340, 3700345, 3700349, 3700350, 3700351, 3700352, 3700355, 3700357, 3700360, 3700370, 3700371, 3700372, 3700373, 3700389, 3700394, 3700415, 3700421, 3700423, 3700424, 3700430, 3700435, 3700437, 3700440, 3700442, 3700445, 3700452, 3700455, 3700466, 3700467, 3700469, 3700471, 3703189, 3703190, 3703192, 3703196, 3703200, 3703205, 3703209, 3703213, 3703214, 3703215, 3703220, 3703227, 3703238, 3703242, 3703244, 3703246, 3703249, 3703253, 3703258, 3703266, 3703274, 3703279, 3703281, 3703284, 3703286, 3703289, 3703292, 3703294, 3703295, 3703299, 3703303, 3703316, 3703317, 3703320, 3703424, 3703427, 3703430, 3703432, 3703448, 3703449, 3703454, 3703459, 3703873, 3703881, 3703883, 3703887, 3703897, 3703898, 3703899, 3703904, 3703907, 3703909, 3703910, 3703466, 3703471, 3703473, 3703489, 3703491, 3703494, 3703497, 3703504, 3703505, 3703508, 3703911, 3703913, 3703915, 3703925, 3703928, 3703929, 3703522, 3703523, 3703540, 3703542, 3703543, 3703550, 3703551, 3703556, 3703561, 3703562, 3703563, 3703566, 3704681, 3704685, 3704692, 3704695, 3704700, 3704704, 3704706, 3704709, 3704874, 3704881, 3704882, 3704885, 3704891, 3704893, 3704897, 3704898, 3704904, 3704905, 3704906, 3704916, 3000539, 3000544, 3000547, 3000550, 3000554, 3000559, 3000576, 3000614, 3000633, 3000650, 3000690, 3000715, 3000716, 3000721, 3000789, 3000866, 3000884, 3000885, 3000896, 3000908, 3000916, 3000925, 3000926, 3000930, 3000952, 3000971, 3000985, 3002240, 3002278, 3002313, 3002342, 3002405, 3004052, 3005018, 3005203, 3005211, 3005981, 3006169, 3006329, 3007684, 3008695, 3010151, 3011159, 3017712, 3020842, 3022979, 3022985, 3022986, 3022994, 3023019, 3023037, 3023050, 3023074, 3023075, 3023080, 3023091, 3023105, 3023121, 3023134, 3023147, 3023149, 3025395, 3025900, 3025909, 3025913, 3025927, 3025934, 3025935, 3025950, 3026765, 3030242, 3030247, 3030332, 3030350, 3032999, 3034297, 3050806, 3050823, 3051206, 3051584, 3051649, 3052029, 3052748, 3054182, 3055061, 3055106, 3056257, 3057073, 3057258, 3058502, 3058531, 3058543, 3058572, 3064697, 3064708, 3068367, 3068963, 3068969, 3070208, 3070210, 3070647, 3070673, 3071646, 3072144, 3077759, 3077763, 3078103, 3078123, 3078129, 3086220, 3088296, 3088320, 3088327, 3088329, 3090370, 3093141, 3092461, 3093858, 3095864, 3099862, 3099867, 3099878, 3106939, 3114600, 3114614, 3114615, 3114617, 3114622, 3116409, 3117996, 3119582, 3119601, 3119635, 3119966, 3121018, 3123169, 3123171, 3123172, 3123174, 3123198, 3123242, 3123573, 3124189, 3151681, 3151702, 3151704, 3151705, 3151711, 3151751, 3151753, 3151755, 3152421, 3152434, 3152468, 3152498, 3152499, 3152580, 3152582, 3152586, 3152593, 3152597, 3152598, 3152599, 3152603, 3152623, 3152629, 3152632, 3152972, 3153292, 3153624, 3153815, 3153840, 3153856, 3153857, 3153859, 3154161, 3154192, 3154195, 3154216, 3154227, 3154228, 3157248, 3157281, 3157349, 3157350, 3163260, 3163261, 3163265, 3163266, 3163267, 3164129, 3310371, 3323381, 3325736, 3346184, 4000044, 4000728, 4000731, 4000790, 4000994, 4001072, 4001102, 4001125, 4001275, 4001375, 4001658, 4001768, 4001854, 4002075, 4002110, 4002143, 4002181, 4002197, 4002268, 4002269, 4002285, 4002399, 4002427, 4002510, 4002517, 4002784, 4002786, 4002816, 4002830, 4002837, 4002861, 4002887, 4002899, 4002905, 4002981, 4002883, 4003168, 4003208, 4003254, 4003302, 4003319, 4003396, 4003444, 4003492, 4003548, 4003733, 4003844, 4003851, 4003857, 4003865, 4003875, 4003877, 4003905, 4003913, 4003935, 4003940, 4003952, 4003964, 4004015, 4004031, 4004048, 4004063, 4004070, 4004427, 4004469, 4004471, 4004474, 4004588, 4004622, 4005541, 4006231, 4006276, 4006496, 4006509, 4006553, 4006610, 4006658, 4006660, 4006707, 4006866, 4006885, 4006922, 4006975, 4007015, 4007189, 4007264, 4007266, 4007300, 4007356, 4007366, 4007382, 4007815, 4007890, 4007957, 4007965, 4007976, 4007983, 4008034, 4008055, 4008082, 4008097, 4008100, 4008108, 4008119, 4008126, 4008211, 4008251, 4008256, 4008301, 4008432, 4008456, 4008476, 4008491, 4008510, 4008530, 4008533, 4008565, 4008588, 4008932, 4009061, 4009126, 4009184, 4009211, 4009489, 4010164, 4010704, 4010863, 4010876, 4010913, 4010915, 4010919, 4010925, 4010930, 4010943, 4010947, 4011007, 4011013, 4011037, 4011054, 4011076, 4011084, 4011125, 4011127, 4011178, 4011179, 4011187, 4011191, 4011203, 4011212, 4011222, 4011228, 4011232, 4011237, 4011242, 4011251, 4011256, 4011268, 4011270, 4011284, 4011287, 4011300, 4011306, 4011332, 4011367, 4011374, 4011489, 4011582, 4011630, 4011635, 4011640, 4011686, 4011688, 4011732, 4011752, 4011875, 4011888, 4011899, 4011931, 4011942, 4020773, 4021028, 3640694, 3640697, 3640701, 3640717, 3640730, 3641708, 3643979, 3644032, 3644590, 3644601, 3645219, 3645225, 3645921, 3645941, 3645955, 3645964, 3647074, 3648419, 3649414, 3649423, 3651131, 3651139, 3651142, 3651183, 3651187, 3651214, 3652532, 3653720, 9856062, 9856093, 9856094, 9856097, 9856099, 9856109, 9856119, 9856126, 9856138, 9856153, 9856221, 9856237, 9856246, 9856267, 9856277, 9856284, 9856292, 9856293, 9856302, 9856317, 9856319, 9856326, 9856340, 9856354, 9856424, 9856428, 9856429, 9856433, 9856436, 9856438, 9856451, 9856456, 9856459, 9856461, 9856462, 9856464, 9856468, 9856471, 9856488, 9856491, 9856502, 9856505, 9856506, 9856513, 9856630, 9856672, 9856674, 9856680, 9856682, 9856684, 9856686, 9856694, 9856707, 9856732, 9856737, 9856748, 9856750, 9856753, 9856901, 9856936, 9856939, 9856940, 9856941, 9856946, 9856948, 9856959, 9856993, 9857043, 9857203, 9857233, 9857413, 9857414, 9857549, 9857550, 9857725, 9857845, 9857871, 9858119, 9858124, 9858136, 9858231, 9858263, 9858288, 9858327, 9858371, 9858377, 9858402, 9858403, 9858415, 9858440, 9858526, 9858627, 9858638, 9858644, 9858664, 9858683, 9858690, 9858700, 9858701, 9858703, 9858713, 9858746, 9858750, 9858764, 9858767, 9858777, 9858778, 9858780, 9858781, 9858789, 9858801, 9858815, 9858924, 9858930, 9858934, 9858946, 9858947, 9858948, 9858949, 9858950, 9858951, 9858952, 9858955, 9858956, 9858970, 9858972, 9858982, 9858984, 9858989, 9858992, 9858993, 9858994, 9858997, 9858998, 9859000, 9859007, 9859009, 9859017, 9859018, 9859031, 9859037, 9859038, 9859044, 9859045, 9859046, 9859063, 9859065, 9859074, 9859086, 9859088, 9859091, 9859202, 9859217, 9859235, 9859252, 9859260, 9859277, 9859291, 9859294, 9859310, 9859312, 9859322, 9859373, 9859479, 9859501, 9859503, 9859508, 9859515, 9859520, 9859524, 9859534, 9859541, 9859566, 9859612, 9859613, 9859616, 9859641, 9859647, 9859652, 9859666, 9859668, 9859670, 9859691, 9859696, 4016046, 9859769, 9859840, 9859895, 9859899, 9859905, 9859916, 9859917, 9859931, 9859954, 9859975, 9859976, 9860000, 9860001, 9860006, 9860056, 9860110, 9860134, 9860249, 9860260, 9860262, 9860276, 9860280, 9860282, 9860284, 9860309, 9860318, 9860323, 9860335, 9860338, 9860359, 9860404, 9860410, 9860413, 9860422, 9860533, 9860534, 9860549, 9860564, 9860668, 9860680, 9860701, 9860704, 9860727, 9860852, 9860873, 9860895, 9860901, 9860949, 9860962, 9860991, 9860995, 9861131, 9861133, 9861140, 9861141, 9861158, 9861159, 9861200, 9861217, 9861289, 9861329, 9861332, 9861374, 9861378, 9861434, 9861451, 9861452, 9861458, 9861461, 9861475, 9861478, 9861502, 9861517, 9861531, 9861534, 9861555, 9861571, 9861576, 9861586, 9861590, 9861597, 9861620, 9861625, 9861637, 9861639, 9861652, 9861654, 9861702, 9861703, 9861704, 9861705, 9861717, 9861718, 9861719, 9861767, 9861785, 9861810, 9861814, 9861823, 9861825, 9861840, 9861845, 9861851, 9861852, 9861855, 9861865, 9861866, 9861867, 9861875, 9861881, 9861889, 9861890, 9861914, 9861920, 9861921, 9861927, 9861938, 9861943, 9861944, 9861949, 9861954, 9861972, 9861993, 9861994, 9861996, 9862014, 9862056, 9862057, 9862058, 9862086, 9862087, 9862089, 9862090, 9862107, 9862146, 9862154, 9862155, 9862156, 9862181, 9862183, 9862185, 9862195, 9862198, 9862200, 9862208, 9862209, 9862210, 9862212, 9862219, 9862227, 9862242, 9862243, 9862245, 9862251, 9862266, 9862285, 9862290, 9862292, 9862294, 9862308, 9862318, 9862331, 9862346, 9862347, 9862352, 9862361, 9862397, 9862449, 9862451, 9862454, 9862512, 9862555, 9862570, 9862579, 9862622, 9862635, 9862645, 9862646, 9862648, 9862653, 9862655, 9862673, 9862701, 9862728, 9862759, 9862804, 9862805, 9862823, 9862829, 9862836, 9862886, 9862888, 9862893, 9862918, 9862922, 9862925, 9862936, 9862948, 9862959, 9862969, 9862979, 9862991, 9862993, 9863059, 9863092, 9863123, 9863128, 9863138, 9863139, 9863186, 9863221, 9863368, 9863570, 9863624, 9863776, 9863889, 9864262, 9864380, 9864408, 9864498, 9864765, 9864858, 9864928, 9864970, 9864988, 9865009, 9865159, 9865160, 9865161, 9865183, 9865190, 9865240, 9865279, 9865300, 9865302, 9865316, 9865323, 9865331, 9865339, 9865416, 9865418, 9865474, 9865485, 9865498, 9865565, 9865601, 9865636, 9865643, 9865649, 9865653, 9865673, 9865703, 9866434, 9866490, 9866495, 9866500, 9866547, 9866556, 9866561, 9866563, 9866564, 9866948, 9866949, 9866961, 9866968, 9866970, 9866982, 9866986, 9866987, 9866991, 9866995, 9866997, 9867022, 9867026, 9867027, 9867030, 9867033, 9867037, 9867038, 9867039, 9867040, 9867041, 9867042, 9867061, 9867198, 9867200, 9867205, 9867208, 9867212, 9867225, 9867227, 9867229, 9867230, 9867232, 9867234, 9867240, 9867242, 9867243, 9867252, 9867256, 9867265, 9867271, 9867273, 9867275, 9867276, 9867278, 9867281, 9867282, 9867283, 9867292, 9867297, 9867302, 9867307, 9867308, 9867312, 9867329, 9867330, 9867332, 9867440, 9867443, 9867447, 9867449, 9867453, 9867457, 9867490, 9867496, 9867497, 9867500, 9867502, 9867503, 9867505, 9867508, 9867521, 9867534, 9867554, 9867583, 9867584, 9867591, 9867596, 9867598, 9867615, 9867623, 9867624, 9867628, 9867629, 9867630, 9867676, 9867677, 9867684, 9867689, 9867694, 9867703, 9867706, 9867707, 9867718, 9867752, 9867755, 9867802, 9867875, 9867876, 9867883, 9867933, 9867941, 9867994, 9867997, 9868003, 9868022, 9868060, 9868070, 9868111, 9868137, 9868151, 9868172, 9868195, 9868198, 9868247, 9868248, 9868251, 9868282, 9868283, 9868294, 9868309, 9868311, 9868318, 9868358, 9868379, 9868393, 9868394, 9868409, 9868428, 9868489, 9868507, 9868563, 9868569, 9868671, 9868677, 9868690, 9868754, 9868856, 9868901, 9868929, 9868936, 9868939, 9868951, 9868977, 9869033, 9869062, 9869065, 9869228, 9869255, 9869278, 9869282, 9869289, 9869299, 9869382, 9869479, 9869548, 9869563, 9869567, 9869603, 9869628, 9869650, 9869667, 9869686, 9869706, 9869870, 9869893, 9869897, 9870001, 9870011, 9870029, 9870045, 9870047, 9870084, 9870116, 9870153, 9870244, 9870254, 9870329, 9870334, 9870343, 9870361, 9870388, 9870408, 9870414, 9870423, 9870439, 9870448, 9870452, 9870465, 9870471, 9870488, 9870491, 9870582, 9870591, 9870606, 9870610, 9870635, 9870655, 9870694, 9870708, 9870709, 9870712, 9870731, 9870732, 9870733, 9870734, 9870759, 9870768, 9870781, 9870802, 9870803, 9870808, 9870809, 9870823, 9870843, 9870850, 9870853, 9870865, 9870886, 9870898, 9870900, 9870907, 9871078, 9871082, 9871092, 9871095, 9871110, 9871115, 9871123, 9871125, 9871130, 9871137, 9871138, 9871181, 9871196, 9871224, 9871228, 9871262, 9871270, 9871277, 9871369, 9871371, 9871379, 9871382, 9871422, 9871466, 9871545, 9871710, 9871764, 9871868, 9871932, 9871944, 9872342, 9872361, 9872362, 9872441, 9872512, 9872516, 9872554, 9872658, 9872664, 9872670, 9872751, 9872754, 9872770, 9872784, 9872794, 9872827, 9872882, 9872913, 9872921, 9872928, 9872936, 9872964, 9872990, 9873046, 9873054, 9873090, 9873092, 9873124, 9873216, 9873221, 9873235, 9873886, 9873893, 9873973, 9876015, 9876061, 9876078, 9876108, 9876109, 9876373, 9876481, 9876977, 9877007, 9877023, 9877054, 9877090, 9877211, 9877215, 9877219, 9877237, 9877296, 9877310, 9877311, 9877341, 9877516, 9877529, 9877538, 9877558, 9877602, 9877645, 9877676, 9877682, 9877684, 9877702, 9877720, 9877721, 9877722, 9877755, 9877798, 9877828, 9877897, 9877901, 9877903, 9877923, 9877945, 9877960, 9877991, 9878005, 9878014, 9878059, 9878078, 9878104, 9878105, 9878109, 9878120, 9878146, 9878148, 9878151, 9878166, 9878169, 9878180, 9878208, 9878209, 9878216, 9878250, 9878254, 9878281, 9878282, 9878293, 9878308, 9878314, 9878338, 9878342, 9878382, 9878383, 9878384, 9878389, 9878390, 9878392, 9878404, 9878405, 9878410, 9878412, 9878418, 9878420, 9878428, 9878430, 9878431, 9878443, 9878446, 9878455, 9878458, 9878468, 9878471, 9878486, 9878489, 9878492, 9878493, 9878500, 9878503, 9878520, 9878525, 9878531, 9878545, 9878548, 9878549, 9878550, 9878551, 9878557, 9878559, 9878565, 9878576, 9878584, 9878585, 9878587, 9878591, 9878592, 9878594, 9878625, 9878626, 9878632, 9878635, 9878636, 9878638, 9878648, 9878649, 9878650, 9878684, 9878686, 9878687, 9878689, 9878690, 9878700, 9878701, 9878706, 9878707, 9878716, 9878724, 9878727, 9878766, 9878770, 9878776, 9878787, 9878790, 9878791, 9878795, 9878817, 9878818, 9878819, 9878822, 9878828, 9878829, 9878832, 9878833, 9878834, 9878837, 9878852, 9878857, 9878865, 9878873, 9878875, 9878878, 9878880, 9878883, 9878892, 9878904, 9878910, 9878915, 9878916, 9878919, 9878922, 9878925, 9878933, 9878939, 9878941, 9878942, 9878943, 9878944, 9878945, 9878956, 9878957, 9878958, 9878965, 9878966, 9878967, 9878972, 9878973, 9878974, 9878981, 9878982, 9878988, 9878989, 9878995, 9878996, 9879012, 9879013, 9879014, 9879016, 9879018, 9879020, 9879028, 9879029, 9879037, 9879038, 9879039, 9879050, 9879051, 9879052, 9879053, 9879054, 9879090, 9879091, 9879104, 9879105, 9879110, 9879111, 9879130, 9879131, 9879163, 9879166, 9879167, 9879168, 9879175, 9879182, 9879183, 9879185, 9879189, 9879196, 9879198, 9879207, 9879210, 9879211, 9879219, 9879220, 9879246, 9879247, 9879251, 9879252, 9879253, 9879254, 9879255, 9879268, 9879269, 9879279, 9879280, 9879281, 9879284, 9879287, 9879290, 9879295, 9879300, 9879301, 9879308, 9879312, 9879320, 9879324, 9879336, 9879338, 9879339, 9879345, 9879347, 9879349, 9879350, 9879367, 9879383, 9879384, 9879386, 9879388, 9879389, 9879393, 9879396, 9879399, 9879404, 9879418, 9879422, 9879423, 9879426, 9879430, 9879431, 9879433, 9879434, 9879459, 9879460, 9879462, 9879469, 9879474, 9879478, 9879484, 9879499, 9879501, 9879503, 9879505, 9879507, 9879513, 9879534, 9879535, 9879536, 9879537, 9879540, 9879543, 9879545, 9879546, 9879547, 9879548, 9879553, 9879556, 9879565, 9879566, 9879570, 9879573, 9879587, 9879593, 9879600, 9879602, 9879608, 9879610, 9879619, 9879632, 9879636, 9879638, 9879643, 9879645, 9879651, 9879653, 9879667, 9879670, 9879680, 9879681, 9879686, 9879688, 9879696, 9879702, 9879706, 9879717, 9879722, 9879739, 9879740, 9879741, 9879742, 9879757, 9879772, 9879775, 9879776, 9879777, 9879779, 9879780, 9879790, 9879791, 9879792, 9879796, 9879798, 9879799, 9879808, 9879814, 9879816, 9879818, 9879819, 9879835, 9879840, 9879845, 9879850, 9879854, 9879856, 9879873, 9879874, 9879876, 9879877, 9879878, 9879901, 9879902, 9879903, 9879905, 9879906, 9879907, 9879924, 9879933, 9879934, 9879935, 9879936, 9879937, 9879946, 9879947, 9879959, 9879962, 9879966, 9879970, 9879973, 9879991, 9880019, 9880025, 9880034, 9880039, 9880048, 9880061, 9880064, 9880067, 9880075, 9880081, 9880092, 9880107, 9880128, 9880129, 9880131, 9880132, 9880142, 9880143, 9880144, 9880149, 9880150, 9880151, 9880152, 9880159, 9880160, 9880161, 9880164, 9880165, 9880166, 9880187, 9880188, 9880189, 9880203, 9880204, 9880205, 9880233, 9880234, 9880235, 9880236, 9880242, 9880245, 9880248, 9880249, 9880256, 9880257, 9880258, 9880259, 9880269, 9880271, 9880272, 9880279, 9880280, 9880283, 9880284, 9880286, 9880288, 9880289, 9880295, 9880296, 9880311, 9880316, 9880320, 9880324, 9880331, 9880335, 9880348, 9880357, 9880360, 9880362, 9880363, 9880369, 9880372, 9880373, 9880379, 9880385, 9880393, 9880396, 9880400, 9880404, 9880410, 9880414, 9880415, 9880417, 9880418, 9880437, 9880438, 9880457, 9880459, 9880469, 9880476, 9880481, 9880498, 9880504, 9880565, 9880566, 9880576, 9880599, 9880621, 9880622, 9880635, 9880636, 9880640, 9880652, 9880665, 9880674, 9880675, 9880676, 9880705, 9880708, 9880730, 9880773, 9880786, 9880801, 9880805, 9880821, 9880823, 9880826, 9880837, 9880879, 9880880, 9880888, 9880889, 9880922, 9880923, 9880924, 9880934, 9880935, 9880948, 9880963, 9880970, 9880971, 9880972, 9880977, 9880978, 9880979, 9880992, 9881003, 9881004, 9881005, 9881006, 9881013, 9881014, 9881015, 9881021, 9881022, 9881039, 9881040, 9881041, 9881043, 9881044, 9881045, 9881061, 9881064, 9881065, 9881074, 9881077, 9881083, 9881084, 9881085, 9881109, 9881114, 9881115, 9881116, 9881117, 9881133, 9881134, 9881136, 9881159, 9881160, 9881161, 9881174, 9881175, 9881179, 9881181, 9881183, 9881197, 9881207, 9881209, 9881222, 9881226, 9881227, 9881229, 9881233, 9881237, 9881238, 9881245, 9881247, 9881252, 9881255, 9881256, 9881263, 9881269, 9881275, 9881277, 9881279, 9881282, 9881289, 9881292, 9881294, 9881301, 9881308, 9881311, 9881314, 9881329, 9881330, 9881331, 9881367, 9881380, 9881387, 9881405, 9881458, 9881460, 9881461, 4126069, 9881734, 9881763, 9881771, 9881785, 9881799, 9881804, 9881806, 9881818, 9881823, 9881834, 9881836, 9881871, 9881887, 9881888, 9881889, 9881897, 9881898, 9881914, 9881916, 9881923, 9881932, 9881936, 9881937, 9881982, 9882214, 9882234, 9882318, 9882341, 9882348, 9882367, 9882394, 9882408, 9882424, 9882480, 9882523, 9882524, 9882563, 9882566, 9882674, 9882679, 9882680, 9882692, 9882765, 9882799, 9882857, 9882930, 9883005, 9883048, 9883056, 9883064, 9883301, 9883302, 9883303, 9883304, 9883308, 9883315, 9883318, 9883319, 9883335, 9883338, 9883345, 9883346, 9883347, 9883348, 9883354, 9883357, 9883358, 9883399, 9883400, 9883408, 9883410, 9883445, 9883575, 9883829, 9884042, 9884211, 9884293, 9884340, 9884620, 9884628, 9884695, 9884757, 9884898, 9885114, 9885115, 9885166, 9885215, 9885648, 9885649, 9885681, 9885713, 9885718, 9885771, 9885881, 9885892, 9885911, 9885912, 9885942, 9885948, 9885972, 9886129, 9886256, 9886299, 9886323, 9886482, 9886500, 9886565, 9886566, 9886586, 9886602, 9886608, 9886615, 9886642, 9886665, 9886802, 9886838, 9887211, 9887319, 9887329, 9887546, 9887654, 9887847, 9888059, 9888157, 9888239, 9888428, 9888431, 9888524, 9888526, 9888555, 9888557, 9888587, 9888604, 9888715, 9888743, 9888761, 9888811, 9888820, 9888864, 9888922, 9888942, 9889000, 9889022, 9889096, 9889200, 9889311, 9889315, 9889320, 9889331, 9889339, 9889461, 9889651, 9889778, 9889930, 9889956, 9889983, 9889994, 9890007, 9890019, 9890110, 9890144, 9890222, 9890240, 9890292, 9890294, 9890381, 9890396, 9890401, 9890484, 9890769, 9890843, 9890973, 9891059, 9891125, 9891212, 9891264, 9891326, 9891345, 9891347, 9891348, 9891360, 9891361, 9891362, 9891374, 9891376, 9891384, 9891387, 9891394, 9891412, 9891413, 9891479, 9891496, 9891526, 9891545, 9891560, 9891567, 9891568, 9891569, 9891574, 9891578, 9891593, 9891594, 9891595, 9891611, 9891614, 9891631, 9891632, 9891641, 9892106, 9892244, 9892285, 9892294, 9892306, 9892333, 9892379, 9892518, 9892556, 9892914, 9893067, 9893102, 9893122, 9893155, 9893291, 9893317, 9893374, 9893410, 9893437, 9893464, 9893467, 9893492, 9893493, 9893509, 9893510, 9893511, 9893532, 9893533, 9893537, 9893541, 9893546, 9893570, 9893578, 9893580, 9893606, 9893608, 9893611, 9893612, 9893613, 9893620, 9893623, 9893624, 9893649, 9893653, 9893682, 9893712, 9893730, 9893745, 9893752, 9893757, 9893758, 9893759, 9893761, 9893764, 9893774, 9893775, 9893790, 9893793, 9893801, 9893802, 9893803, 9893809, 9893810, 9893812, 9893831, 9893832, 9893833, 9893834, 9893847, 9893849, 9893850, 9893858, 9893859, 9893883, 9893885, 9893899, 9893909, 9893920, 9893921, 9893932, 9893939, 9893955, 9893970, 9893973, 9893982, 9893985, 9893986, 9893991, 9893995, 9894006, 9894017, 9894028, 9894040, 9894059, 9894060, 9894064, 9894073, 9894074, 9894089, 9894093, 9894097, 9894104, 9894105, 9894106, 9894107, 9894108, 9894109, 9894111, 9894112, 9894114, 9894118, 9894119, 9894120, 9894121, 9894122, 9894138, 9894139, 9894140, 9894141, 9894152, 9894153, 9894154, 9894167, 9894179, 9894180, 9894182, 9894202, 9894203, 9894205, 9894226, 9894227, 9894244, 9894245, 9894246, 9894247, 9894263, 9894265, 9894271, 9894281, 9894285, 9894286, 9894303, 9894310, 9894332, 9894345, 9894347, 9894351, 9894355, 9894356, 9894360, 9894377, 9894383, 9894384, 9894385, 9894390, 9894401, 9894404, 9894405, 9894406, 9894422, 9894423, 9894437, 9894438, 9894439, 9894448, 9894449, 9894450, 9894451, 9894453, 9894458, 9894465, 9894476, 9894477, 9894490, 9894493, 9894501, 9894503, 9894504, 9894519, 9894522, 9894524, 9894533, 9894534, 9894535, 9894548, 9894550, 9894553, 9894578, 9894579, 9894580, 9894603, 9894604, 9894606, 9894618, 9894619, 9894636, 9894638, 9894657, 9894685, 9894896, 9894917, 9894930, 9894942, 9895013, 9895049, 9895050, 9895054, 9895060, 9895081, 9895090, 9895094, 9895097, 9895098, 9895119, 9895121, 9895182, 9895197, 9895198, 9895216, 9895243, 9895247, 9895256, 9895257, 9895268, 9895280, 9895294, 9895305, 9895318, 9895343, 9895344, 9895347, 9895356, 9895366, 9895373, 9895374, 9895389, 9895405, 9895415, 9895417, 9895441, 9895442, 9895456, 9895464, 9895465, 9895467, 9895511, 9895517, 9895518, 9895520, 9895543, 9895548, 9895588, 9895589, 9895591, 9895612, 9895613, 9895614, 9895637, 9895646, 9895648, 9895695, 9895697, 9895714, 9895722, 9895723, 9895732, 9895774, 9895810, 9895831, 9895870, 9895871, 9895883, 9895892, 9895894, 9895949, 9895957, 9895961, 9895964, 9895965, 9895967, 9895970, 9895978, 9895984, 9895987, 9896005, 9896024, 9896025, 9896038, 9896041, 9896046, 9896049, 9896052, 9896061, 9896063, 9896099, 9896102, 9896103, 9896114, 9896117, 9896121, 9896125, 9896128, 9896131, 9896132, 9896137, 9896138, 9896140, 9896141, 9896142, 9896149, 9896150, 9896151, 9896163, 9896164, 9896165, 9896167, 9896180, 9896181, 9896182, 9896184, 9896196, 9896197, 9896212, 9896213, 9896214, 9896215, 9896227, 9896228, 9896229, 9896230, 9896231, 9896254, 9896255, 9896256, 9896257, 9896258, 9896276, 9896278, 9896284, 9896285, 9896295, 9896296, 9896297, 9896298, 9896299, 9896319, 9896320, 9896321, 9896332, 9896333, 9896334, 9896366, 9896368, 9896369, 9896370, 9896371, 9896375, 9896384, 9896385, 9896407, 9896410, 9896416, 9896419, 9896421, 9896434, 9896435, 9896444, 9896448, 9896473, 9896481, 9896486, 9896489, 9896499, 9896501, 9896502, 9896510, 9896511, 9896523, 9896524, 9896541, 9896542, 9896547, 9896552, 9896554, 9896558, 9896570, 9896576, 9896578, 9896582, 9896583, 9896584, 9896587, 9896588, 9896589, 9896598, 9896599, 9896600, 9896601, 9896616, 9896617, 9896618, 9896626, 9896627, 9896628, 9896629, 9896650, 9896652, 9896685, 9896701, 9896702, 9896703, 9896704, 9896726, 9896731, 9896735, 9896736, 9896737, 9896739, 9896754, 9896755, 9896756, 9896781, 9896783, 9896788, 9896801, 9896809, 9897116, 9897118, 9897126, 9897134, 9897139, 9897140, 9897142, 9897172, 9897175, 9897176, 9897207, 9897222, 9897225, 9897236, 9897237, 9897267, 9897268, 9897512, 9897514, 9897523, 9897528, 9897529, 9897530, 9897565, 9897566, 9897567, 9897591, 9897594, 9897595, 9897596, 9897619, 9897637, 9897666, 9897667, 9897668, 9897669, 9897680, 9897688, 9897689, 9897711, 9897714, 9897717, 9897732, 9897741, 9897744, 9897748, 9897764, 9897781, 9897790, 9897803, 9897819, 9897820, 9897838, 9897861, 9897871, 9897883, 9897899, 9897905, 9897936, 9897946, 9897950, 9897966, 9897991, 9898029, 9898032, 9898033, 9898034, 9898035, 9898047, 9898066, 9898067, 9898072, 9898073, 9898095, 9898096, 9898100, 9898101, 9898111, 9898127, 9898128, 9898142, 9898169, 9898180, 9898181, 9898182, 9898186, 9898191, 9898235, 9898256, 9898263, 9898280, 9898282, 9898290, 9898291, 9898292, 9898296, 9898319, 9898334, 9898340, 9898341, 9898355, 9898357, 9898360, 9898370, 9898371, 9898403, 9898430, 9898434, 9898458, 9898487, 9898488, 9898519, 9898533, 9898535, 9898538, 9898542, 9898543, 9898563, 9898581, 9898584, 9898597, 9898598, 9898605, 9898610, 9898876, 9898883, 9898884, 9898900, 9898910, 9898911, 9898942, 9898966, 9898968, 9898991, 9898992, 9898993, 9899012, 9899013, 9899055, 9899056, 9899057, 9899058, 9899059, 9899072, 9899092, 9899095, 989
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
CareFusion 303, Inc.
Reason for Recall:
Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Alaris System PC Unit Model 8000, a modular infusion pump and monitoring system

Product Codes/Lot Numbers:

During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers. ** US Serial numbers: 2326370, 2326382, 2326385, 2326387, 3668143, 2326400, 2326401, 2326402, 2326416, 2326418, 2326419, 2326423, 2326647, 2326664, 2326819, 2326821, 2326834, 2326837, 2326842, 2326873, 2326882, 2326883, 2326926, 9300064, 9300081, 9300083, 9300117, 9300149, 9300162, 9300163, 9300164, 9300171, 9300174, 9300176, 9300284, 9300302, 9300313, 9300317, 9300333, 9300334, 9300485, 3669844, 9300531, 9300532, 9300533, 9300534, 9300535, 9300538, 3686560, 4011408, 3696229, 3000933, 4122102, 4123517, 4123611, 4123651, 4123981, 4124350, 4124764, 4124999, 4125045, 4125149, 4125888, 4125923, 4126044, 4126051, 4126072, 4126073, 4126076, 4126077, 4126092, 4126093, 4126127, 4126128, 4126131, 4126146, 4126147, 4126148, 4126149, 4126151, 4126205, 4126206, 4126208, 4126212, 4126213, 4126214, 4126285, 4126286, 4126287, 4126288, 4126290, 4126299, 4126302, 4126315, 4126318, 4126358, 4126360, 4126387, 4126390, 4126398, 4126400, 4126424, 4126427, 4126428, 4126436, 4126437, 4126438, 4126449, 4126450, 4126472, 4126473, 4126474, 4126475, 4126478, 4126504, 4126512, 4126535, 4126536, 4126537, 4126539, 4126589, 4126590, 4126591, 4126595, 4126596, 4126597, 4126651, 4126652, 4126653, 4126654, 4126655, 4126656, 4126657, 4126696, 4126697, 4126698, 4126699, 4126700, 4126703, 4126704, 4126705, 4126738, 4126740, 4126741, 4126751, 4126761, 4126765, 4126774, 4126795, 4126844, 4126845, 4126887, 4127208, 4127355, 3640679, 3640680, 3641766, 3641790, 3642182, 3642188, 3642194, 3642208, 3642221, 3642227, 3642234, 3642235, 3642251, 3642264, 3642278, 3643225, 3643235, 3643236, 3643237, 3649470, 3652550, 3653689, 3653709, 3653727, 3653732, 3653739, 3653744, 3653747, 3653755, 3653773, 3665625, 3665629, 3665642, 3665645, 3665647, 3665649, 3665652, 3665657, 3665659, 3665666, 3665698, 3665705, 3665706, 3666423, 3666452, 3666454, 3666499, 3666500, 3666502, 3666511, 3667676, 3667687, 3667692, 3667696, 3667700, 3667704, 3667725, 3667727, 3667749, 3667768, 3667770, 3668134, 3668136, 3668140, 3668142, 3668145, 3668147, 3668148, 3668149, 3668152, 3668153, 3668154, 3668155, 3668156, 3665646, 3668158, 3668161, 3668165, 3668167, 3668168, 3668169, 3668170, 3668171, 3668179, 3668198, 3668199, 3668218, 3668221, 3668916, 3668917, 3668918, 3668919, 3668920, 3668923, 3668924, 3668927, 3668929, 3668935, 3668936, 3668939, 3668940, 3668941, 3668942, 3668944, 3668945, 3668946, 3668947, 3668949, 3668950, 3668951, 3668952, 3668959, 3668963, 3668965, 3668966, 3668971, 3668972, 3668974, 3668976, 3668977, 3668978, 3668981, 3668982, 3668984, 3668989, 3668990, 3668991, 3668992, 3668993, 3668996, 3668997, 3668998, 3669000, 3669001, 3669002, 3669005, 3669006, 3669007, 3669008, 3669009, 3669010, 3669012, 3669697, 3669700, 3669701, 3669702, 3669703, 3669704, 3669015, 3669707, 3669708, 3669711, 3669713, 3669715, 3669717, 3669718, 3669720, 3669721, 3669723, 3669724, 3669725, 3669726, 3669727, 3669728, 3669729, 3669730, 3669731, 3669733, 3669741, 3669743, 3669745, 3669746, 3669747, 3669750, 3669753, 3669754, 3669755, 3669757, 3669758, 3669762, 3669764, 3669765, 3669766, 3669767, 3669769, 3669772, 3669777, 3669779, 3669780, 3669782, 3669786, 3669787, 3669789, 3669792, 3669793, 3669794, 3669795, 3669799, 3669800, 3669801, 3669803, 3669804, 3669806, 3669809, 3669815, 3669816, 3669818, 3669824, 3669825, 3669830, 3669832, 3669833, 3669834, 3669836, 3669841, 3669843, 3669845, 3672182, 3672183, 3672184, 3672185, 3672187, 3672195, 3672196, 3672197, 3672199, 3672200, 3672201, 3672204, 3672207, 3672211, 3672212, 3672213, 3672214, 3672216, 3672217, 3672218, 3672219, 3672220, 3672223, 3672225, 3672235, 3672243, 3672249, 3672255, 3672258, 3672259, 3672260, 3672262, 3672263, 3672264, 3672265, 3672266, 3672270, 3672276, 3672277, 3672280, 3673895, 3673896, 3673897, 3673898, 3673901, 3673902, 3673903, 3673904, 3673907, 3673908, 3673912, 3673914, 3673915, 3673918, 3673919, 3673920, 3673921, 3673922, 3673924, 3673926, 3673928, 3673929, 3673930, 3673932, 3673934, 3673935, 3673937, 3673938, 3673940, 3673943, 3673946, 3673947, 3673949, 3673951, 3673952, 3673954, 3673955, 3673956, 3673958, 3673959, 3673962, 3673971, 3673972, 3673973, 3673974, 3673981, 3673992, 3673993, 3673994, 3674424, 3674428, 3674429, 3674430, 3674441, 3674444, 3674445, 3674447, 3674448, 3674449, 3674450, 3674476, 3674479, 3674480, 3674483, 3674485, 3674489, 3674494, 3674495, 3674496, 3674498, 3674499, 3674501, 3674452, 3674453, 3674454, 3674455, 3674461, 3674473, 3674503, 3674504, 3674505, 3674507, 3674509, 3674510, 3674511, 3674512, 3674513, 3674517, 3674519, 3674520, 3674521, 3675314, 3675316, 3675317, 3675320, 3675323, 3675324, 3675326, 3675330, 3675334, 3675335, 3675337, 3675338, 3675339, 3675341, 3675342, 3675343, 3675344, 3675348, 3675349, 3675350, 3675352, 3675357, 3675361, 3675369, 3675373, 3675374, 3675377, 3675378, 3675379, 3675381, 3675382, 3675384, 3675386, 3675390, 3675393, 3675394, 3675395, 3675401, 3675406, 3675410, 3675411, 3676437, 3676441, 3676442, 3676445, 3676446, 3676457, 3676459, 3676463, 3676465, 3676467, 3676476, 3676478, 3676480, 3676481, 3676482, 3676485, 3676493, 3676494, 3676500, 3676501, 3676514, 3676517, 3676519, 3676525, 3676526, 3676527, 3676528, 3676529, 3676530, 3676531, 3676532, 3676533, 3676534, 3676535, 3676536, 3677593, 3677595, 3677596, 3677597, 3677598, 3677599, 3677600, 3677601, 3677602, 3677603, 3677605, 3677606, 3677607, 3677608, 3677609, 3677610, 3677611, 3677612, 3677614, 3677616, 3677617, 3677618, 3677619, 3677620, 3677621, 3677623, 3677624, 3677625, 3677626, 3677627, 3677628, 3677629, 3677630, 3677631, 3677633, 3677634, 3677635, 3677636, 3677638, 3677639, 3677640, 3677642, 3677644, 3677645, 3677649, 3677650, 3677652, 3677653, 3677654, 3677656, 3677657, 3677658, 3677659, 3677660, 3677662, 3677664, 3677665, 3677666, 3677667, 3677668, 3677669, 3677670, 3677673, 3677674, 3677676, 3677677, 3677679, 3677680, 3677681, 3677682, 3677683, 3677684, 3677685, 3677686, 3677687, 3677688, 3677689, 3677691, 3686517, 3686520, 3686521, 3686522, 3686523, 3686527, 3686528, 3686529, 3686530, 3686533, 3686535, 3686537, 3686538, 3686540, 3686541, 3686544, 3686545, 3686546, 3686547, 3686548, 3686552, 3686553, 3686554, 3686555, 3686557, 3686558, 3686566, 3686567, 3686569, 3686570, 3686571, 3686572, 3686574, 3686577, 3686580, 3686581, 3686587, 3686593, 3686596, 3686597, 3686600, 3686605, 3686610, 3687060, 3687063, 3687064, 3687065, 3687068, 3687070, 3687077, 3687082, 3687083, 3687085, 3687087, 3687089, 3687090, 3687091, 3687092, 3687093, 3687098, 3687103, 3687108, 3687109, 3687111, 3687113, 3687117, 3687120, 3687136, 3687139, 3687144, 3688474, 3688479, 3688480, 3688481, 3688482, 3688483, 3688486, 3688317, 3688490, 3688493, 3688494, 3688495, 3688496, 3688505, 3688507, 3688515, 3688518, 3688523, 3688529, 3688537, 3688547, 3688794, 3688797, 3688849, 3688856, 3688863, 3688877, 3688884, 3688900, 3688904, 3688905, 3688929, 3688939, 3688941, 3688948, 3688949, 3688950, 3688958, 3688960, 3688961, 3688966, 3688970, 3688981, 3688983, 3688985, 3689628, 3689650, 3689652, 3689653, 3689659, 3689660, 3689684, 3689685, 3689697, 3689702, 3689706, 3689713, 3689716, 3689749, 3689756, 3689766, 3689805, 3689806, 3690533, 3690534, 3690540, 3690554, 3690556, 3690564, 3690569, 3690585, 3690588, 3690589, 3690599, 3690601, 3691431, 3691443, 3691453, 3690619, 3690512, 3690521, 3691485, 3691525, 3691533, 3691567, 3691593, 3691595, 3691596, 3691599, 3692507, 3692528, 3692540, 3692552, 3692711, 3692737, 3692744, 3693072, 3692753, 3693088, 3693099, 3693107, 3693118, 3693128, 3693136, 3693145, 3693148, 3693161, 3693166, 3693173, 3693174, 3693175, 3693188, 3693190, 3693191, 3693205, 3693207, 3693224, 3693245, 3693802, 3693804, 3693811, 3693815, 3693826, 3693827, 3693834, 3693838, 3693839, 3693872, 3694258, 3694261, 3694272, 3694276, 3694279, 3694287, 3694293, 3694296, 3694301, 3693883, 3694396, 3694397, 3694399, 3694401, 3694405, 3694408, 3694410, 3694411, 3694417, 3694423, 3694435, 3694436, 3694437, 3694438, 3694440, 3694447, 3694448, 3694451, 3694832, 3694834, 3694845, 3694848, 3695674, 3695676, 3695680, 3695683, 3695686, 3695689, 3695690, 3695692, 3695694, 3695695, 3695696, 3695697, 3694897, 3694899, 3694900, 3694901, 3694903, 3694904, 3694908, 3694909, 3694915, 3694916, 3694917, 3694919, 3694920, 3694921, 3694923, 3694924, 3694868, 3694874, 3694882, 3694884, 3694889, 3694892, 3694894, 3694896, 3695718, 3695730, 3695731, 3695751, 3696127, 3696176, 3696774, 3696792, 3696804, 3696833, 3697438, 3697476, 3697532, 3697546, 3698216, 3698978, 3699035, 3699413, 3699416, 3699442, 3699445, 3699448, 3699450, 3699487, 3699489, 3699490, 3699494, 3699499, 3699500, 3699502, 3699504, 3699505, 3699521, 3699523, 3699525, 3699528, 3699529, 3700330, 3700334, 3700336, 3700340, 3700345, 3700349, 3700350, 3700351, 3700352, 3700355, 3700357, 3700360, 3700370, 3700371, 3700372, 3700373, 3700389, 3700394, 3700415, 3700421, 3700423, 3700424, 3700430, 3700435, 3700437, 3700440, 3700442, 3700445, 3700452, 3700455, 3700466, 3700467, 3700469, 3700471, 3703189, 3703190, 3703192, 3703196, 3703200, 3703205, 3703209, 3703213, 3703214, 3703215, 3703220, 3703227, 3703238, 3703242, 3703244, 3703246, 3703249, 3703253, 3703258, 3703266, 3703274, 3703279, 3703281, 3703284, 3703286, 3703289, 3703292, 3703294, 3703295, 3703299, 3703303, 3703316, 3703317, 3703320, 3703424, 3703427, 3703430, 3703432, 3703448, 3703449, 3703454, 3703459, 3703873, 3703881, 3703883, 3703887, 3703897, 3703898, 3703899, 3703904, 3703907, 3703909, 3703910, 3703466, 3703471, 3703473, 3703489, 3703491, 3703494, 3703497, 3703504, 3703505, 3703508, 3703911, 3703913, 3703915, 3703925, 3703928, 3703929, 3703522, 3703523, 3703540, 3703542, 3703543, 3703550, 3703551, 3703556, 3703561, 3703562, 3703563, 3703566, 3704681, 3704685, 3704692, 3704695, 3704700, 3704704, 3704706, 3704709, 3704874, 3704881, 3704882, 3704885, 3704891, 3704893, 3704897, 3704898, 3704904, 3704905, 3704906, 3704916, 3000539, 3000544, 3000547, 3000550, 3000554, 3000559, 3000576, 3000614, 3000633, 3000650, 3000690, 3000715, 3000716, 3000721, 3000789, 3000866, 3000884, 3000885, 3000896, 3000908, 3000916, 3000925, 3000926, 3000930, 3000952, 3000971, 3000985, 3002240, 3002278, 3002313, 3002342, 3002405, 3004052, 3005018, 3005203, 3005211, 3005981, 3006169, 3006329, 3007684, 3008695, 3010151, 3011159, 3017712, 3020842, 3022979, 3022985, 3022986, 3022994, 3023019, 3023037, 3023050, 3023074, 3023075, 3023080, 3023091, 3023105, 3023121, 3023134, 3023147, 3023149, 3025395, 3025900, 3025909, 3025913, 3025927, 3025934, 3025935, 3025950, 3026765, 3030242, 3030247, 3030332, 3030350, 3032999, 3034297, 3050806, 3050823, 3051206, 3051584, 3051649, 3052029, 3052748, 3054182, 3055061, 3055106, 3056257, 3057073, 3057258, 3058502, 3058531, 3058543, 3058572, 3064697, 3064708, 3068367, 3068963, 3068969, 3070208, 3070210, 3070647, 3070673, 3071646, 3072144, 3077759, 3077763, 3078103, 3078123, 3078129, 3086220, 3088296, 3088320, 3088327, 3088329, 3090370, 3093141, 3092461, 3093858, 3095864, 3099862, 3099867, 3099878, 3106939, 3114600, 3114614, 3114615, 3114617, 3114622, 3116409, 3117996, 3119582, 3119601, 3119635, 3119966, 3121018, 3123169, 3123171, 3123172, 3123174, 3123198, 3123242, 3123573, 3124189, 3151681, 3151702, 3151704, 3151705, 3151711, 3151751, 3151753, 3151755, 3152421, 3152434, 3152468, 3152498, 3152499, 3152580, 3152582, 3152586, 3152593, 3152597, 3152598, 3152599, 3152603, 3152623, 3152629, 3152632, 3152972, 3153292, 3153624, 3153815, 3153840, 3153856, 3153857, 3153859, 3154161, 3154192, 3154195, 3154216, 3154227, 3154228, 3157248, 3157281, 3157349, 3157350, 3163260, 3163261, 3163265, 3163266, 3163267, 3164129, 3310371, 3323381, 3325736, 3346184, 4000044, 4000728, 4000731, 4000790, 4000994, 4001072, 4001102, 4001125, 4001275, 4001375, 4001658, 4001768, 4001854, 4002075, 4002110, 4002143, 4002181, 4002197, 4002268, 4002269, 4002285, 4002399, 4002427, 4002510, 4002517, 4002784, 4002786, 4002816, 4002830, 4002837, 4002861, 4002887, 4002899, 4002905, 4002981, 4002883, 4003168, 4003208, 4003254, 4003302, 4003319, 4003396, 4003444, 4003492, 4003548, 4003733, 4003844, 4003851, 4003857, 4003865, 4003875, 4003877, 4003905, 4003913, 4003935, 4003940, 4003952, 4003964, 4004015, 4004031, 4004048, 4004063, 4004070, 4004427, 4004469, 4004471, 4004474, 4004588, 4004622, 4005541, 4006231, 4006276, 4006496, 4006509, 4006553, 4006610, 4006658, 4006660, 4006707, 4006866, 4006885, 4006922, 4006975, 4007015, 4007189, 4007264, 4007266, 4007300, 4007356, 4007366, 4007382, 4007815, 4007890, 4007957, 4007965, 4007976, 4007983, 4008034, 4008055, 4008082, 4008097, 4008100, 4008108, 4008119, 4008126, 4008211, 4008251, 4008256, 4008301, 4008432, 4008456, 4008476, 4008491, 4008510, 4008530, 4008533, 4008565, 4008588, 4008932, 4009061, 4009126, 4009184, 4009211, 4009489, 4010164, 4010704, 4010863, 4010876, 4010913, 4010915, 4010919, 4010925, 4010930, 4010943, 4010947, 4011007, 4011013, 4011037, 4011054, 4011076, 4011084, 4011125, 4011127, 4011178, 4011179, 4011187, 4011191, 4011203, 4011212, 4011222, 4011228, 4011232, 4011237, 4011242, 4011251, 4011256, 4011268, 4011270, 4011284, 4011287, 4011300, 4011306, 4011332, 4011367, 4011374, 4011489, 4011582, 4011630, 4011635, 4011640, 4011686, 4011688, 4011732, 4011752, 4011875, 4011888, 4011899, 4011931, 4011942, 4020773, 4021028, 3640694, 3640697, 3640701, 3640717, 3640730, 3641708, 3643979, 3644032, 3644590, 3644601, 3645219, 3645225, 3645921, 3645941, 3645955, 3645964, 3647074, 3648419, 3649414, 3649423, 3651131, 3651139, 3651142, 3651183, 3651187, 3651214, 3652532, 3653720, 9856062, 9856093, 9856094, 9856097, 9856099, 9856109, 9856119, 9856126, 9856138, 9856153, 9856221, 9856237, 9856246, 9856267, 9856277, 9856284, 9856292, 9856293, 9856302, 9856317, 9856319, 9856326, 9856340, 9856354, 9856424, 9856428, 9856429, 9856433, 9856436, 9856438, 9856451, 9856456, 9856459, 9856461, 9856462, 9856464, 9856468, 9856471, 9856488, 9856491, 9856502, 9856505, 9856506, 9856513, 9856630, 9856672, 9856674, 9856680, 9856682, 9856684, 9856686, 9856694, 9856707, 9856732, 9856737, 9856748, 9856750, 9856753, 9856901, 9856936, 9856939, 9856940, 9856941, 9856946, 9856948, 9856959, 9856993, 9857043, 9857203, 9857233, 9857413, 9857414, 9857549, 9857550, 9857725, 9857845, 9857871, 9858119, 9858124, 9858136, 9858231, 9858263, 9858288, 9858327, 9858371, 9858377, 9858402, 9858403, 9858415, 9858440, 9858526, 9858627, 9858638, 9858644, 9858664, 9858683, 9858690, 9858700, 9858701, 9858703, 9858713, 9858746, 9858750, 9858764, 9858767, 9858777, 9858778, 9858780, 9858781, 9858789, 9858801, 9858815, 9858924, 9858930, 9858934, 9858946, 9858947, 9858948, 9858949, 9858950, 9858951, 9858952, 9858955, 9858956, 9858970, 9858972, 9858982, 9858984, 9858989, 9858992, 9858993, 9858994, 9858997, 9858998, 9859000, 9859007, 9859009, 9859017, 9859018, 9859031, 9859037, 9859038, 9859044, 9859045, 9859046, 9859063, 9859065, 9859074, 9859086, 9859088, 9859091, 9859202, 9859217, 9859235, 9859252, 9859260, 9859277, 9859291, 9859294, 9859310, 9859312, 9859322, 9859373, 9859479, 9859501, 9859503, 9859508, 9859515, 9859520, 9859524, 9859534, 9859541, 9859566, 9859612, 9859613, 9859616, 9859641, 9859647, 9859652, 9859666, 9859668, 9859670, 9859691, 9859696, 4016046, 9859769, 9859840, 9859895, 9859899, 9859905, 9859916, 9859917, 9859931, 9859954, 9859975, 9859976, 9860000, 9860001, 9860006, 9860056, 9860110, 9860134, 9860249, 9860260, 9860262, 9860276, 9860280, 9860282, 9860284, 9860309, 9860318, 9860323, 9860335, 9860338, 9860359, 9860404, 9860410, 9860413, 9860422, 9860533, 9860534, 9860549, 9860564, 9860668, 9860680, 9860701, 9860704, 9860727, 9860852, 9860873, 9860895, 9860901, 9860949, 9860962, 9860991, 9860995, 9861131, 9861133, 9861140, 9861141, 9861158, 9861159, 9861200, 9861217, 9861289, 9861329, 9861332, 9861374, 9861378, 9861434, 9861451, 9861452, 9861458, 9861461, 9861475, 9861478, 9861502, 9861517, 9861531, 9861534, 9861555, 9861571, 9861576, 9861586, 9861590, 9861597, 9861620, 9861625, 9861637, 9861639, 9861652, 9861654, 9861702, 9861703, 9861704, 9861705, 9861717, 9861718, 9861719, 9861767, 9861785, 9861810, 9861814, 9861823, 9861825, 9861840, 9861845, 9861851, 9861852, 9861855, 9861865, 9861866, 9861867, 9861875, 9861881, 9861889, 9861890, 9861914, 9861920, 9861921, 9861927, 9861938, 9861943, 9861944, 9861949, 9861954, 9861972, 9861993, 9861994, 9861996, 9862014, 9862056, 9862057, 9862058, 9862086, 9862087, 9862089, 9862090, 9862107, 9862146, 9862154, 9862155, 9862156, 9862181, 9862183, 9862185, 9862195, 9862198, 9862200, 9862208, 9862209, 9862210, 9862212, 9862219, 9862227, 9862242, 9862243, 9862245, 9862251, 9862266, 9862285, 9862290, 9862292, 9862294, 9862308, 9862318, 9862331, 9862346, 9862347, 9862352, 9862361, 9862397, 9862449, 9862451, 9862454, 9862512, 9862555, 9862570, 9862579, 9862622, 9862635, 9862645, 9862646, 9862648, 9862653, 9862655, 9862673, 9862701, 9862728, 9862759, 9862804, 9862805, 9862823, 9862829, 9862836, 9862886, 9862888, 9862893, 9862918, 9862922, 9862925, 9862936, 9862948, 9862959, 9862969, 9862979, 9862991, 9862993, 9863059, 9863092, 9863123, 9863128, 9863138, 9863139, 9863186, 9863221, 9863368, 9863570, 9863624, 9863776, 9863889, 9864262, 9864380, 9864408, 9864498, 9864765, 9864858, 9864928, 9864970, 9864988, 9865009, 9865159, 9865160, 9865161, 9865183, 9865190, 9865240, 9865279, 9865300, 9865302, 9865316, 9865323, 9865331, 9865339, 9865416, 9865418, 9865474, 9865485, 9865498, 9865565, 9865601, 9865636, 9865643, 9865649, 9865653, 9865673, 9865703, 9866434, 9866490, 9866495, 9866500, 9866547, 9866556, 9866561, 9866563, 9866564, 9866948, 9866949, 9866961, 9866968, 9866970, 9866982, 9866986, 9866987, 9866991, 9866995, 9866997, 9867022, 9867026, 9867027, 9867030, 9867033, 9867037, 9867038, 9867039, 9867040, 9867041, 9867042, 9867061, 9867198, 9867200, 9867205, 9867208, 9867212, 9867225, 9867227, 9867229, 9867230, 9867232, 9867234, 9867240, 9867242, 9867243, 9867252, 9867256, 9867265, 9867271, 9867273, 9867275, 9867276, 9867278, 9867281, 9867282, 9867283, 9867292, 9867297, 9867302, 9867307, 9867308, 9867312, 9867329, 9867330, 9867332, 9867440, 9867443, 9867447, 9867449, 9867453, 9867457, 9867490, 9867496, 9867497, 9867500, 9867502, 9867503, 9867505, 9867508, 9867521, 9867534, 9867554, 9867583, 9867584, 9867591, 9867596, 9867598, 9867615, 9867623, 9867624, 9867628, 9867629, 9867630, 9867676, 9867677, 9867684, 9867689, 9867694, 9867703, 9867706, 9867707, 9867718, 9867752, 9867755, 9867802, 9867875, 9867876, 9867883, 9867933, 9867941, 9867994, 9867997, 9868003, 9868022, 9868060, 9868070, 9868111, 9868137, 9868151, 9868172, 9868195, 9868198, 9868247, 9868248, 9868251, 9868282, 9868283, 9868294, 9868309, 9868311, 9868318, 9868358, 9868379, 9868393, 9868394, 9868409, 9868428, 9868489, 9868507, 9868563, 9868569, 9868671, 9868677, 9868690, 9868754, 9868856, 9868901, 9868929, 9868936, 9868939, 9868951, 9868977, 9869033, 9869062, 9869065, 9869228, 9869255, 9869278, 9869282, 9869289, 9869299, 9869382, 9869479, 9869548, 9869563, 9869567, 9869603, 9869628, 9869650, 9869667, 9869686, 9869706, 9869870, 9869893, 9869897, 9870001, 9870011, 9870029, 9870045, 9870047, 9870084, 9870116, 9870153, 9870244, 9870254, 9870329, 9870334, 9870343, 9870361, 9870388, 9870408, 9870414, 9870423, 9870439, 9870448, 9870452, 9870465, 9870471, 9870488, 9870491, 9870582, 9870591, 9870606, 9870610, 9870635, 9870655, 9870694, 9870708, 9870709, 9870712, 9870731, 9870732, 9870733, 9870734, 9870759, 9870768, 9870781, 9870802, 9870803, 9870808, 9870809, 9870823, 9870843, 9870850, 9870853, 9870865, 9870886, 9870898, 9870900, 9870907, 9871078, 9871082, 9871092, 9871095, 9871110, 9871115, 9871123, 9871125, 9871130, 9871137, 9871138, 9871181, 9871196, 9871224, 9871228, 9871262, 9871270, 9871277, 9871369, 9871371, 9871379, 9871382, 9871422, 9871466, 9871545, 9871710, 9871764, 9871868, 9871932, 9871944, 9872342, 9872361, 9872362, 9872441, 9872512, 9872516, 9872554, 9872658, 9872664, 9872670, 9872751, 9872754, 9872770, 9872784, 9872794, 9872827, 9872882, 9872913, 9872921, 9872928, 9872936, 9872964, 9872990, 9873046, 9873054, 9873090, 9873092, 9873124, 9873216, 9873221, 9873235, 9873886, 9873893, 9873973, 9876015, 9876061, 9876078, 9876108, 9876109, 9876373, 9876481, 9876977, 9877007, 9877023, 9877054, 9877090, 9877211, 9877215, 9877219, 9877237, 9877296, 9877310, 9877311, 9877341, 9877516, 9877529, 9877538, 9877558, 9877602, 9877645, 9877676, 9877682, 9877684, 9877702, 9877720, 9877721, 9877722, 9877755, 9877798, 9877828, 9877897, 9877901, 9877903, 9877923, 9877945, 9877960, 9877991, 9878005, 9878014, 9878059, 9878078, 9878104, 9878105, 9878109, 9878120, 9878146, 9878148, 9878151, 9878166, 9878169, 9878180, 9878208, 9878209, 9878216, 9878250, 9878254, 9878281, 9878282, 9878293, 9878308, 9878314, 9878338, 9878342, 9878382, 9878383, 9878384, 9878389, 9878390, 9878392, 9878404, 9878405, 9878410, 9878412, 9878418, 9878420, 9878428, 9878430, 9878431, 9878443, 9878446, 9878455, 9878458, 9878468, 9878471, 9878486, 9878489, 9878492, 9878493, 9878500, 9878503, 9878520, 9878525, 9878531, 9878545, 9878548, 9878549, 9878550, 9878551, 9878557, 9878559, 9878565, 9878576, 9878584, 9878585, 9878587, 9878591, 9878592, 9878594, 9878625, 9878626, 9878632, 9878635, 9878636, 9878638, 9878648, 9878649, 9878650, 9878684, 9878686, 9878687, 9878689, 9878690, 9878700, 9878701, 9878706, 9878707, 9878716, 9878724, 9878727, 9878766, 9878770, 9878776, 9878787, 9878790, 9878791, 9878795, 9878817, 9878818, 9878819, 9878822, 9878828, 9878829, 9878832, 9878833, 9878834, 9878837, 9878852, 9878857, 9878865, 9878873, 9878875, 9878878, 9878880, 9878883, 9878892, 9878904, 9878910, 9878915, 9878916, 9878919, 9878922, 9878925, 9878933, 9878939, 9878941, 9878942, 9878943, 9878944, 9878945, 9878956, 9878957, 9878958, 9878965, 9878966, 9878967, 9878972, 9878973, 9878974, 9878981, 9878982, 9878988, 9878989, 9878995, 9878996, 9879012, 9879013, 9879014, 9879016, 9879018, 9879020, 9879028, 9879029, 9879037, 9879038, 9879039, 9879050, 9879051, 9879052, 9879053, 9879054, 9879090, 9879091, 9879104, 9879105, 9879110, 9879111, 9879130, 9879131, 9879163, 9879166, 9879167, 9879168, 9879175, 9879182, 9879183, 9879185, 9879189, 9879196, 9879198, 9879207, 9879210, 9879211, 9879219, 9879220, 9879246, 9879247, 9879251, 9879252, 9879253, 9879254, 9879255, 9879268, 9879269, 9879279, 9879280, 9879281, 9879284, 9879287, 9879290, 9879295, 9879300, 9879301, 9879308, 9879312, 9879320, 9879324, 9879336, 9879338, 9879339, 9879345, 9879347, 9879349, 9879350, 9879367, 9879383, 9879384, 9879386, 9879388, 9879389, 9879393, 9879396, 9879399, 9879404, 9879418, 9879422, 9879423, 9879426, 9879430, 9879431, 9879433, 9879434, 9879459, 9879460, 9879462, 9879469, 9879474, 9879478, 9879484, 9879499, 9879501, 9879503, 9879505, 9879507, 9879513, 9879534, 9879535, 9879536, 9879537, 9879540, 9879543, 9879545, 9879546, 9879547, 9879548, 9879553, 9879556, 9879565, 9879566, 9879570, 9879573, 9879587, 9879593, 9879600, 9879602, 9879608, 9879610, 9879619, 9879632, 9879636, 9879638, 9879643, 9879645, 9879651, 9879653, 9879667, 9879670, 9879680, 9879681, 9879686, 9879688, 9879696, 9879702, 9879706, 9879717, 9879722, 9879739, 9879740, 9879741, 9879742, 9879757, 9879772, 9879775, 9879776, 9879777, 9879779, 9879780, 9879790, 9879791, 9879792, 9879796, 9879798, 9879799, 9879808, 9879814, 9879816, 9879818, 9879819, 9879835, 9879840, 9879845, 9879850, 9879854, 9879856, 9879873, 9879874, 9879876, 9879877, 9879878, 9879901, 9879902, 9879903, 9879905, 9879906, 9879907, 9879924, 9879933, 9879934, 9879935, 9879936, 9879937, 9879946, 9879947, 9879959, 9879962, 9879966, 9879970, 9879973, 9879991, 9880019, 9880025, 9880034, 9880039, 9880048, 9880061, 9880064, 9880067, 9880075, 9880081, 9880092, 9880107, 9880128, 9880129, 9880131, 9880132, 9880142, 9880143, 9880144, 9880149, 9880150, 9880151, 9880152, 9880159, 9880160, 9880161, 9880164, 9880165, 9880166, 9880187, 9880188, 9880189, 9880203, 9880204, 9880205, 9880233, 9880234, 9880235, 9880236, 9880242, 9880245, 9880248, 9880249, 9880256, 9880257, 9880258, 9880259, 9880269, 9880271, 9880272, 9880279, 9880280, 9880283, 9880284, 9880286, 9880288, 9880289, 9880295, 9880296, 9880311, 9880316, 9880320, 9880324, 9880331, 9880335, 9880348, 9880357, 9880360, 9880362, 9880363, 9880369, 9880372, 9880373, 9880379, 9880385, 9880393, 9880396, 9880400, 9880404, 9880410, 9880414, 9880415, 9880417, 9880418, 9880437, 9880438, 9880457, 9880459, 9880469, 9880476, 9880481, 9880498, 9880504, 9880565, 9880566, 9880576, 9880599, 9880621, 9880622, 9880635, 9880636, 9880640, 9880652, 9880665, 9880674, 9880675, 9880676, 9880705, 9880708, 9880730, 9880773, 9880786, 9880801, 9880805, 9880821, 9880823, 9880826, 9880837, 9880879, 9880880, 9880888, 9880889, 9880922, 9880923, 9880924, 9880934, 9880935, 9880948, 9880963, 9880970, 9880971, 9880972, 9880977, 9880978, 9880979, 9880992, 9881003, 9881004, 9881005, 9881006, 9881013, 9881014, 9881015, 9881021, 9881022, 9881039, 9881040, 9881041, 9881043, 9881044, 9881045, 9881061, 9881064, 9881065, 9881074, 9881077, 9881083, 9881084, 9881085, 9881109, 9881114, 9881115, 9881116, 9881117, 9881133, 9881134, 9881136, 9881159, 9881160, 9881161, 9881174, 9881175, 9881179, 9881181, 9881183, 9881197, 9881207, 9881209, 9881222, 9881226, 9881227, 9881229, 9881233, 9881237, 9881238, 9881245, 9881247, 9881252, 9881255, 9881256, 9881263, 9881269, 9881275, 9881277, 9881279, 9881282, 9881289, 9881292, 9881294, 9881301, 9881308, 9881311, 9881314, 9881329, 9881330, 9881331, 9881367, 9881380, 9881387, 9881405, 9881458, 9881460, 9881461, 4126069, 9881734, 9881763, 9881771, 9881785, 9881799, 9881804, 9881806, 9881818, 9881823, 9881834, 9881836, 9881871, 9881887, 9881888, 9881889, 9881897, 9881898, 9881914, 9881916, 9881923, 9881932, 9881936, 9881937, 9881982, 9882214, 9882234, 9882318, 9882341, 9882348, 9882367, 9882394, 9882408, 9882424, 9882480, 9882523, 9882524, 9882563, 9882566, 9882674, 9882679, 9882680, 9882692, 9882765, 9882799, 9882857, 9882930, 9883005, 9883048, 9883056, 9883064, 9883301, 9883302, 9883303, 9883304, 9883308, 9883315, 9883318, 9883319, 9883335, 9883338, 9883345, 9883346, 9883347, 9883348, 9883354, 9883357, 9883358, 9883399, 9883400, 9883408, 9883410, 9883445, 9883575, 9883829, 9884042, 9884211, 9884293, 9884340, 9884620, 9884628, 9884695, 9884757, 9884898, 9885114, 9885115, 9885166, 9885215, 9885648, 9885649, 9885681, 9885713, 9885718, 9885771, 9885881, 9885892, 9885911, 9885912, 9885942, 9885948, 9885972, 9886129, 9886256, 9886299, 9886323, 9886482, 9886500, 9886565, 9886566, 9886586, 9886602, 9886608, 9886615, 9886642, 9886665, 9886802, 9886838, 9887211, 9887319, 9887329, 9887546, 9887654, 9887847, 9888059, 9888157, 9888239, 9888428, 9888431, 9888524, 9888526, 9888555, 9888557, 9888587, 9888604, 9888715, 9888743, 9888761, 9888811, 9888820, 9888864, 9888922, 9888942, 9889000, 9889022, 9889096, 9889200, 9889311, 9889315, 9889320, 9889331, 9889339, 9889461, 9889651, 9889778, 9889930, 9889956, 9889983, 9889994, 9890007, 9890019, 9890110, 9890144, 9890222, 9890240, 9890292, 9890294, 9890381, 9890396, 9890401, 9890484, 9890769, 9890843, 9890973, 9891059, 9891125, 9891212, 9891264, 9891326, 9891345, 9891347, 9891348, 9891360, 9891361, 9891362, 9891374, 9891376, 9891384, 9891387, 9891394, 9891412, 9891413, 9891479, 9891496, 9891526, 9891545, 9891560, 9891567, 9891568, 9891569, 9891574, 9891578, 9891593, 9891594, 9891595, 9891611, 9891614, 9891631, 9891632, 9891641, 9892106, 9892244, 9892285, 9892294, 9892306, 9892333, 9892379, 9892518, 9892556, 9892914, 9893067, 9893102, 9893122, 9893155, 9893291, 9893317, 9893374, 9893410, 9893437, 9893464, 9893467, 9893492, 9893493, 9893509, 9893510, 9893511, 9893532, 9893533, 9893537, 9893541, 9893546, 9893570, 9893578, 9893580, 9893606, 9893608, 9893611, 9893612, 9893613, 9893620, 9893623, 9893624, 9893649, 9893653, 9893682, 9893712, 9893730, 9893745, 9893752, 9893757, 9893758, 9893759, 9893761, 9893764, 9893774, 9893775, 9893790, 9893793, 9893801, 9893802, 9893803, 9893809, 9893810, 9893812, 9893831, 9893832, 9893833, 9893834, 9893847, 9893849, 9893850, 9893858, 9893859, 9893883, 9893885, 9893899, 9893909, 9893920, 9893921, 9893932, 9893939, 9893955, 9893970, 9893973, 9893982, 9893985, 9893986, 9893991, 9893995, 9894006, 9894017, 9894028, 9894040, 9894059, 9894060, 9894064, 9894073, 9894074, 9894089, 9894093, 9894097, 9894104, 9894105, 9894106, 9894107, 9894108, 9894109, 9894111, 9894112, 9894114, 9894118, 9894119, 9894120, 9894121, 9894122, 9894138, 9894139, 9894140, 9894141, 9894152, 9894153, 9894154, 9894167, 9894179, 9894180, 9894182, 9894202, 9894203, 9894205, 9894226, 9894227, 9894244, 9894245, 9894246, 9894247, 9894263, 9894265, 9894271, 9894281, 9894285, 9894286, 9894303, 9894310, 9894332, 9894345, 9894347, 9894351, 9894355, 9894356, 9894360, 9894377, 9894383, 9894384, 9894385, 9894390, 9894401, 9894404, 9894405, 9894406, 9894422, 9894423, 9894437, 9894438, 9894439, 9894448, 9894449, 9894450, 9894451, 9894453, 9894458, 9894465, 9894476, 9894477, 9894490, 9894493, 9894501, 9894503, 9894504, 9894519, 9894522, 9894524, 9894533, 9894534, 9894535, 9894548, 9894550, 9894553, 9894578, 9894579, 9894580, 9894603, 9894604, 9894606, 9894618, 9894619, 9894636, 9894638, 9894657, 9894685, 9894896, 9894917, 9894930, 9894942, 9895013, 9895049, 9895050, 9895054, 9895060, 9895081, 9895090, 9895094, 9895097, 9895098, 9895119, 9895121, 9895182, 9895197, 9895198, 9895216, 9895243, 9895247, 9895256, 9895257, 9895268, 9895280, 9895294, 9895305, 9895318, 9895343, 9895344, 9895347, 9895356, 9895366, 9895373, 9895374, 9895389, 9895405, 9895415, 9895417, 9895441, 9895442, 9895456, 9895464, 9895465, 9895467, 9895511, 9895517, 9895518, 9895520, 9895543, 9895548, 9895588, 9895589, 9895591, 9895612, 9895613, 9895614, 9895637, 9895646, 9895648, 9895695, 9895697, 9895714, 9895722, 9895723, 9895732, 9895774, 9895810, 9895831, 9895870, 9895871, 9895883, 9895892, 9895894, 9895949, 9895957, 9895961, 9895964, 9895965, 9895967, 9895970, 9895978, 9895984, 9895987, 9896005, 9896024, 9896025, 9896038, 9896041, 9896046, 9896049, 9896052, 9896061, 9896063, 9896099, 9896102, 9896103, 9896114, 9896117, 9896121, 9896125, 9896128, 9896131, 9896132, 9896137, 9896138, 9896140, 9896141, 9896142, 9896149, 9896150, 9896151, 9896163, 9896164, 9896165, 9896167, 9896180, 9896181, 9896182, 9896184, 9896196, 9896197, 9896212, 9896213, 9896214, 9896215, 9896227, 9896228, 9896229, 9896230, 9896231, 9896254, 9896255, 9896256, 9896257, 9896258, 9896276, 9896278, 9896284, 9896285, 9896295, 9896296, 9896297, 9896298, 9896299, 9896319, 9896320, 9896321, 9896332, 9896333, 9896334, 9896366, 9896368, 9896369, 9896370, 9896371, 9896375, 9896384, 9896385, 9896407, 9896410, 9896416, 9896419, 9896421, 9896434, 9896435, 9896444, 9896448, 9896473, 9896481, 9896486, 9896489, 9896499, 9896501, 9896502, 9896510, 9896511, 9896523, 9896524, 9896541, 9896542, 9896547, 9896552, 9896554, 9896558, 9896570, 9896576, 9896578, 9896582, 9896583, 9896584, 9896587, 9896588, 9896589, 9896598, 9896599, 9896600, 9896601, 9896616, 9896617, 9896618, 9896626, 9896627, 9896628, 9896629, 9896650, 9896652, 9896685, 9896701, 9896702, 9896703, 9896704, 9896726, 9896731, 9896735, 9896736, 9896737, 9896739, 9896754, 9896755, 9896756, 9896781, 9896783, 9896788, 9896801, 9896809, 9897116, 9897118, 9897126, 9897134, 9897139, 9897140, 9897142, 9897172, 9897175, 9897176, 9897207, 9897222, 9897225, 9897236, 9897237, 9897267, 9897268, 9897512, 9897514, 9897523, 9897528, 9897529, 9897530, 9897565, 9897566, 9897567, 9897591, 9897594, 9897595, 9897596, 9897619, 9897637, 9897666, 9897667, 9897668, 9897669, 9897680, 9897688, 9897689, 9897711, 9897714, 9897717, 9897732, 9897741, 9897744, 9897748, 9897764, 9897781, 9897790, 9897803, 9897819, 9897820, 9897838, 9897861, 9897871, 9897883, 9897899, 9897905, 9897936, 9897946, 9897950, 9897966, 9897991, 9898029, 9898032, 9898033, 9898034, 9898035, 9898047, 9898066, 9898067, 9898072, 9898073, 9898095, 9898096, 9898100, 9898101, 9898111, 9898127, 9898128, 9898142, 9898169, 9898180, 9898181, 9898182, 9898186, 9898191, 9898235, 9898256, 9898263, 9898280, 9898282, 9898290, 9898291, 9898292, 9898296, 9898319, 9898334, 9898340, 9898341, 9898355, 9898357, 9898360, 9898370, 9898371, 9898403, 9898430, 9898434, 9898458, 9898487, 9898488, 9898519, 9898533, 9898535, 9898538, 9898542, 9898543, 9898563, 9898581, 9898584, 9898597, 9898598, 9898605, 9898610, 9898876, 9898883, 9898884, 9898900, 9898910, 9898911, 9898942, 9898966, 9898968, 9898991, 9898992, 9898993, 9899012, 9899013, 9899055, 9899056, 9899057, 9899058, 9899059, 9899072, 9899092, 9899095, 989

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2716-2020

Related Recalls

If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

Oct 17, 2025 Infusion Pumps Nationwide View Details →

Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

Sep 30, 2025 Other Medical Devices Nationwide View Details →