Centron Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Centron 1.0.10 Centron 1.0.10.1 Centron 1.0.10.5
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Healthcare
Reason for Recall:
The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Centron Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures.

Product Codes/Lot Numbers:

Centron 1.0.10 Centron 1.0.10.1 Centron 1.0.10.5

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2500-2018

Related Recalls

Class I - Dangerous

Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.

Aug 2, 2023 Infusion Pumps Nationwide View Details β†’

Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: "Press the cine pedal until the acquisition stops". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.

Feb 2, 2022 Implants & Prosthetics Nationwide View Details β†’
Class I - Dangerous

In the Azurion system, the user can add a new study to a patient by selecting the option "Add Study". The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.

May 26, 2021 Other Medical Devices Nationwide View Details β†’