Philips Healthcare

57 recalls • • Categories: Infusion Pumps, Implants & Prosthetics, Other Medical Devices, Diagnostic Equipment

Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.

Aug 2, 2023 Infusion Pumps Nationwide View Details →

Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: "Press the cine pedal until the acquisition stops". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.

Feb 2, 2022 Implants & Prosthetics Nationwide View Details →

In the Azurion system, the user can add a new study to a patient by selecting the option "Add Study". The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.

May 26, 2021 Other Medical Devices Nationwide View Details →

The Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and the X-ray control certification statement [21 CFR 1010.2]. Philips has determined that some systems at customer locations did not have these labels applied.

Sep 22, 2021 Other Medical Devices Nationwide View Details →
Class I - Dangerous

The Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and the X-ray control certification statement [21 CFR 1010.2]. Philips has determined that some systems at customer locations did not have these labels applied.

Sep 22, 2021 Other Medical Devices Nationwide View Details →

3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.

Sep 24, 2021 Other Medical Devices Nationwide View Details →

The Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and the X-ray control certification statement [21 CFR 1010.2]. Philips has determined that some systems at customer locations did not have these labels applied.

Sep 22, 2021 Other Medical Devices Nationwide View Details →

During recent evaluations of the Philips DigitalDiagnost C50 system, Philips identified a missing label. This notification is intended to provide you with information regarding what the issue is, and relevant affected product, the actions you are required to take and the actions planned by Philips to correct the issue. If you need any or further information or support concerning this issue, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.

Dec 21, 2018 Other Medical Devices View Details →

The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.

Jun 11, 2018 Implants & Prosthetics Nationwide View Details →

The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.

Jun 11, 2018 Implants & Prosthetics Nationwide View Details →

The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.

Jun 11, 2018 Implants & Prosthetics Nationwide View Details →

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Sep 6, 2015 Other Medical Devices Nationwide View Details →

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Sep 6, 2015 Other Medical Devices Nationwide View Details →

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Sep 6, 2015 Other Medical Devices Nationwide View Details →

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Sep 6, 2015 Other Medical Devices Nationwide View Details →

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Sep 6, 2015 Other Medical Devices Nationwide View Details →

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Sep 6, 2015 Other Medical Devices Nationwide View Details →

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Sep 6, 2015 Other Medical Devices Nationwide View Details →

If a customer creates customized trend scales in the trend review tile and the iX or primary server reboots for any reason, the iX(s) will enter a reboot loop. If one iX reboots, the issue will impact only those patients monitored on that device. If the primary server reboots, all patients across all iX devices will be impacted.

Mar 5, 2013 Diagnostic Equipment Nationwide View Details →

Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.

Jan 13, 2014 Other Medical Devices Nationwide View Details →

The Instructions for Use (IFU) for the MobileDiagnost wDR fails to comply with a Federal standard. The IFU does not provide tolerance levels for several generator values as required by 1020.30(h)(3).

Jun 11, 2013 Diagnostic Equipment Nationwide View Details →

The MRx may deliver a non-synchronized cardioversion shock resulting in the delivery of incorrect therapy, which may induce ventricular fibrillation.

Jun 19, 2013 Diagnostic Equipment Nationwide View Details →

IntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity

Jul 8, 2013 Diagnostic Equipment Nationwide View Details →

Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.

Apr 5, 2010 Other Medical Devices Nationwide View Details →

When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.

Oct 30, 2012 Other Medical Devices Nationwide View Details →

When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.

Oct 25, 2012 Other Medical Devices Nationwide View Details →

VS2+ screen may lock up and stop monitoring during patient use

Jan 14, 2013 Diagnostic Equipment Nationwide View Details →

Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms.

Nov 9, 2012 Diagnostic Equipment Nationwide View Details →

With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "Right patient side" marker, although this can appear on the left patient side.

Oct 30, 2012 Diagnostic Equipment Nationwide View Details →

HeartStart XL Defibrillator/Monitor (M4735A) used on a patient when AC power is not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone.

Sep 11, 2012 Diagnostic Equipment Nationwide View Details →

Therapy Cable Connection wear may lead to a malfunction in detection of defibrillation pads/paddles therapy Cables

Jun 20, 2012 Diagnostic Equipment Nationwide View Details →

Philips SureSigns VSi/VS2+ patient monitors may not be properly secured and may fall from the wall mount

Aug 15, 2012 Diagnostic Equipment Nationwide View Details →

Essenta DR C-Arm may fall

Jul 25, 2012 Diagnostic Equipment View Details →

Defibrillator/monitor cannot analyze an ECG report during acquisition of 12 lead ECG

Jun 20, 2012 Diagnostic Equipment View Details →

Xcelera Connect R2.1 L 1 SP2 , incomplete information arriving from unformatted reports interface

Jun 14, 2012 Diagnostic Equipment Nationwide View Details →

HeartStart MRx Unexpected Pads/Paddles ECG Failure

Jun 20, 2012 Diagnostic Equipment Nationwide View Details →

MultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement

May 21, 2012 Implants & Prosthetics Nationwide View Details →

With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side

May 22, 2012 Other Medical Devices Nationwide View Details →

If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

May 11, 2012 Other Medical Devices View Details →

When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie

Oct 19, 2011 Other Medical Devices Nationwide View Details →