KYPHON(R) Directional Bone Filler Device, Size: 3, REF F04C, Rx Only, STERILE R.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Lots
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Sofamor Danek USA Inc
- Reason for Recall:
- The directional arrow at the proximal end of the instrument may not correctly align with the cutout opening on the distal end of the instrument. Using affected bone void filler may result in the injected cement being placed in a direction unintended by the surgeon.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
KYPHON(R) Directional Bone Filler Device, Size: 3, REF F04C, Rx Only, STERILE R.
Product Codes/Lot Numbers:
All Lots
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2247-2017
Related Recalls
Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
Grafton DBM (Flex), Model Numbers: T42100, Demineralized Bone Matrix
Medtronic Sofamor Danek USA
Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.
Grafton DMB Matrix, Model Numbers: a) T42200, b) T42210; Demineralized Bone Matrix
Medtronic Sofamor Danek USA
Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.