Medtronic Sofamor Danek USA

85 recalls • • Categories: Implants & Prosthetics, Other Medical Devices, Infusion Pumps, Surgical Instruments

Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11MM, Model Number: 6068113. 6. SPACER 6068116 CATALYFT PL LONG 11MM, Model Number: 6068116. 7. SPACER 6069073 CATALYFT PL40 SHORT 7MM, Model Number: 6069073. 8. SPACER 6069076 CATALYFT PL40 LONG 7MM, Model Number: 6069076. 9. SPACER 6069093 CATALYFT PL40 SHORT 9MM, Model Number: 6069093. 10. SPACER 6069096 CATALYFT PL40 LONG 9MM, Model Number: 6069096. 11. SPACER 6069113 CATALYFT PL40 SHORT 11MM, Model Number: 6069113. 12. SPACER 6069116 CATALYFT PL40 LONG 11MM, Model Number: 6069116. 13. CATALYFT PL IMPLANT SET, Model Number: SPS03143. 14. CATALYFT PL 40 IMPLANT SET, Model Number: SPS03144. 15. KIT CATALYFT PL IMPLANTS, Model Number: SPS04012. 16. KIT CATALYFT PL 40 IMPLANTS, Model Number: SPS04013. 17. CATALYFT PL 40 IMP KIT, Model Number: SPSJEP40. 18. CATALYFT PL 40 HIGH IMP KIT, Model Number: SPSJEP4H. 19. CATALYFT PL 40 SHORT IMP KIT, Model Number: SPSJEP4S. 20. CATALYFT PL 40 SHORT HIGH IMP KIT, Model Number: SPSJEP4T. 21. CATALYFT PL STRAIGHT HIGH IMP KIT, Model Number: SPSJEPSH. 22. CATALYFT PL STRAIGHT IMP KIT, Model Number: SPSJEPSI. These systems utilize the following affected Instructions for Use (IFU), Document M333023W048E, and Surgical Technique Guide, Document M333023W217.

Class I - Dangerous

Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.

Sep 16, 2025 Implants & Prosthetics Nationwide View Details →

Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.

Sep 26, 2023 Other Medical Devices View Details →

There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.

Jun 6, 2023 Implants & Prosthetics Nationwide View Details →

There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.

Jun 6, 2023 Implants & Prosthetics Nationwide View Details →

There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.

Jun 6, 2023 Implants & Prosthetics Nationwide View Details →

There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.

Jun 6, 2023 Implants & Prosthetics Nationwide View Details →

There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.

Jun 6, 2023 Implants & Prosthetics Nationwide View Details →

There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.

Jun 6, 2023 Implants & Prosthetics Nationwide View Details →

There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.

Jun 6, 2023 Implants & Prosthetics Nationwide View Details →

Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

Feb 16, 2023 Other Medical Devices Nationwide View Details →

Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

Feb 16, 2023 Other Medical Devices Nationwide View Details →

Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

Feb 16, 2023 Other Medical Devices Nationwide View Details →

Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

Feb 16, 2023 Other Medical Devices Nationwide View Details →

Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

Feb 16, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.

Feb 8, 2023 Other Medical Devices Nationwide View Details →

The shelf box identifies a small device actually contains a large device, and the shelf box identifying a large device actually contains a small device.

Mar 17, 2021 Other Medical Devices Nationwide View Details →

The shelf box identifies a small device actually contains a large device, and the shelf box identifying a large device actually contains a small device.

Mar 17, 2021 Other Medical Devices Nationwide View Details →

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Jun 8, 2020 Infusion Pumps Nationwide View Details →

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Jun 8, 2020 Infusion Pumps Nationwide View Details →

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Jun 8, 2020 Infusion Pumps Nationwide View Details →

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Jun 8, 2020 Infusion Pumps Nationwide View Details →

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Jun 8, 2020 Infusion Pumps Nationwide View Details →

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Jun 8, 2020 Infusion Pumps Nationwide View Details →

Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.

Apr 24, 2020 Other Medical Devices Nationwide View Details →

Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.

Apr 24, 2020 Other Medical Devices Nationwide View Details →

Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.

Apr 24, 2020 Other Medical Devices Nationwide View Details →

This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly indicate a 5.5 mm diameter screw when the actual screw diameter aligns with 5.0 diameter screw size requirements.

Jul 9, 2018 Other Medical Devices View Details →

The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.

Mar 13, 2018 Implants & Prosthetics Nationwide View Details →

The dovetails of poly trials, Beta 2.0, were observed as either cracked or broken.

Feb 20, 2017 Implants & Prosthetics View Details →

The dovetails of poly trials, Beta 2.0, were observed as either cracked or broken.

Feb 20, 2017 Implants & Prosthetics View Details →

The dovetails of poly trials, Beta 2.0, were observed as either cracked or broken.

Feb 20, 2017 Implants & Prosthetics View Details →

Poly Component Trial, MPCS, packaged in the following sizes and configurations: a) Poly Component Trial - MPCS 3x10L, REF 90-SRK-192310 b) Poly Component Trial - MPCS 3x10R, REF 90-SRK-191310 c) Poly Component Trial - MPCS 3x12L, REF 90-SRK-192312 d) Poly Component Trial - MPCS 3x12R, REF 90-SRK-191312 e) Poly Component Trial - MPCS 3x14L, REF 90-SRK-192314 f) Poly Component Trial - MPCS 3x14R, REF 90-SRK-191314 g) Poly Component Trial - MPCS 4x10L, REF 90-SRK-192410 h) Poly Component Trial - MPCS 4x10R, REF 90-SRK-191410 i) Poly Component Trial - MPCS 4x12L, REF 90-SRK-192412 j) Poly Component Trial - MPCS 4x12R, REF 90-SRK-191412 k) Poly Component Trial - MPCS 4x14L, REF 90-SRK-192414 l) Poly Component Trial - MPCS 4x14R, REF 90-SRK-191414 m) Poly Component Trial - MPCS 5x10L, REF 90-SRK-192510 n) Poly Component Trial - MPCS 5x10R, REF 90-SRK-191510 o) Poly Component Trial - MPCS 5x12L, REF 90-SRK-192512 p) Poly Component Trial - MPCS 5x12R, REF 90-SRK-191512 q) Poly Component Trial - MPCS 5x14L, REF 90-SRK-192514 r) Poly Component Trial - MPCS 5x14R, REF 90-SRK-191514 s) Poly Component Trial - MPCS 6x10L, REF 90-SRK-192610 t) Poly Component Trial - MPCS 6x10R, REF 90-SRK-191610 u) Poly Component Trial - MPCS 6x12L, REF 90-SRK-192612 v) Poly Component Trial - MPCS 6x12R, REF 90-SRK-191612 w) Poly Component Trial - MPCS 6x14L, REF 90-SRK-192614 x) Poly Component Trial - MPCS 6x14R, REF 90-SRK-191614 The Responsive Orthopedics (RO) Total Knee Arthroplasty (TKA) System is intended to restore alignment, stability, range of motion, and alleviate pain by replacing the articulating surfaces of the knee joint.

Class I - Dangerous

The dovetails of poly trials, Beta 2.0, were observed as either cracked or broken.

Feb 20, 2017 Implants & Prosthetics View Details →

One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.

Aug 12, 2013 Implants & Prosthetics Nationwide View Details →

Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.

Aug 30, 2012 Surgical Instruments Nationwide View Details →

Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.

Aug 30, 2012 Surgical Instruments Nationwide View Details →

Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.

Aug 30, 2012 Surgical Instruments Nationwide View Details →