Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00813502013252; Lot Numbers: 1727033, 1725590
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abiomed, Inc.
- Reason for Recall:
- There was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally introduced during manufacturing at the joint between the sidearm and the hub may cause the joint to detach.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457
Product Codes/Lot Numbers:
UDI-DI: 00813502013252; Lot Numbers: 1727033, 1725590
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2128-2023
Related Recalls
Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.