SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot No. 14035485 14046894 14035458 14035457
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- CareFusion 303, Inc.
- Reason for Recall:
- CareFusion is recalling the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15 micron filter and it does not provide the instruction that the set should not be used with blood administration due to the potential for occluding the 15 micron filter in the drip c
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458
Product Codes/Lot Numbers:
Lot No. 14035485 14046894 14035458 14035457
Distribution:
Distributed in: FL, NM, WA, WI, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2025-2014
Related Recalls
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
BD Pyxis Pro MedStation Main, REF: 1155-00
CareFusion 303
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.