OEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Two populations are potentially affected by this notification. 1. Listed in Appendix A - OEC Elite Mobile C-arm Systems (GTIN: 00001234567895) and OEC 3D Mobile C-arm Systems (GTIN: 00195278096944) that potentially have an X-ray field size greater than the 4% SID limit and do not contain the acceptance limit for the X-ray field size measurement in the System Service Manual. 2. Systems NOT listed in Appendix A - OEC Elite Mobile C-arm Systems (GTIN: 00001234567895) and OEC 3D Mobile C-arm Systems (GTIN: 00195278096944) that are within the allowed SID limit. However, the acceptance limit for the X-ray field size measurement is not stated in the System Service Manual. Appendix A Serial Numbers FAHXTX00002 FAHXTX00005 FAHXTX00014 FAHXTX00018 FAHXTX00019 FAHXTX00020 FAHXTX00024 FAHXTX00028 FAHXTX00034 FAHXTX00037 FAHXTX00043 FAHXTX00052 FAHXTX00063 FAHXTX00066 FAHXTX00068 FAHXTX00070 FAHXTX00075 FAHXTX00078 FAHXTX00079 FAHXTX00080 FAHXTX00081 FAHXTX00083 FAHXTX00092 FAHXTX00093 FAHXTX00097 FAHXTX00110 FAHXTX00111 FAHXTX00112 FAHXTX00114 FAHXTX00115 FAHXTX00116 FAHXTX00117 FAHXTX00118 FAHXTX00119 FAHXTX00120 FAHXTX00121 FAHXTX00122 FAHXTX00124 FAHXTX00125 FAHXTX00135 FAHXTX00137 FAHXTX00138 FAHXTX00151 FAHXTX00152 FAHXTX00155 FAHXTX00156 FAHXTX00162 FAHXTX00163 FAHXTX00166 FAHXTX00170 FAHXTX00171 FAHXTX00177 FAHXTX00178 FAHXTX00188 FAHXTX00192 FAHXTX00193 FAHXTX00202 FAHXTX00204 FAHXTX00211 FAHXTX00212 FAHXTX00217 FAHXTX00219 FAHXTX00220 FAHXTX00221 FAHXTX00222 FAHXTX00223 FAHXTX00224 FAHXTX00226 FAHXTX00227 FAHXTX00229 FAHXTX00230 FAHXTX00236 FAHXTX00240 FAHXTX00244 FAHXTX00261 FAHXTX00265 FAHXTX00266 FAHXTX00269 FAHXTX00271 FAHXTX00272 FAHXTX00273 FAHXTX00274 FAHXTX00280 FAHXTX00283 FAHXTX00284 FAHXTX00296 FAHXTX00299 FAHXTX00300 FAHXTX00304 FAHXTX00305 FAHXTX00308 FAHXTX00312 FAHXTX00313 FAHXTX00320 FAHXTX00323 FAHXTX00325 FAHXTX00332 FAHXTX00340 FAHXTX00342 FAHXTX00343 FAHXTX00352 FAHXTX00353 FAHXTX00354 FAHXTX00357 FAHXTX00358 FAHXTX00359 FAHXTX00360 FAHXTX00363 FAHXTX00376 FAHXTX00379 FAHXTX00381 FAHXTX00382 FAHXTX00390 FAHXTX00392 FAHXTX00399 FAHXTX00400 FAHXTX00410 FAHXTX00413 FAHXTX00414 FAHXTX00415 FAHXTX00420 FAHXTX00421 FAHXTX00429 FAHXTX00439 FAHXTX00453 FAHXTX00455 FAHXTX00456 FAHXTX00463 FAHXTX00465 FAHXTX00466 FAHXTX00479 FAHXTX00481 FAHXTX00483 FAHXTX00484 FAHXTX00486 FAHXTX00488 FAHXTX00495 FAHXTX00497 FAHXTX00498 FAHXTX00507 FAHXTX00509 FAHXTX00513 FAHXTX00523 FAHXTX00530 FAHXTX00534 FAHXTX00538 FAHXTX00539 FAHXTX00540 FAHXTX00547 FAHXTX00549 FAHXTX00555 FAHXTX00557 FAHXTX00560 FAHXTX00565 FAHXTX00569 FAHXTX00570 FAHXTX00571 FAHXTX00572 FAHXTX00575 FAHXTX00576 FAHXTX00579 FAHXTX00592 FAHXTX00595 FAHXTX00610 FAHXTX00611 FAHXXX00029 FAHXXX00030 FAHXXX00038 FAHXXX00040 FAHXXX00056 FAHXXX00057 FAHXXX00059 FAHXXX00060 FAHXXX00061 FAHXXX00069 FAHXXX00082 FAHXXX00085 FAHXXX00086 FAHXXX00089 FAHXXX00091 FAHXXX00093 FAHXXX00097 FAHXXX00098 FAHXXX00099 FAHXXX00105 FAHXXX00106 FAHXXX00112 FAHXXX00113 FAHXXX00114 FAHXXX00116 FAHXXX00117 FAHXXX00119 FAHXXX00123 FAHXXX00142 FAHXXX00144 FAHXXX00145 FAHXXX00147 FAHXXX00150 FAHXXX00151 FAHXXX00153 FAHXXX00155 FAHXXX00164 FAHXXX00173 FAHXXX00176 FAHXXX00178 FAHXXX00179 FAHXXX00188 FAHXXX00190 FAHXXX00193 FAHXXX00196 FAHXXX00199 FAHXXX00200 FAHXXX00207 FAHXXX00208 FAHXXX00209 FAMHTX00001 FAMHTX00005 FAMHTX00007 FAMHTX00012 FAMHTX00014 FAMHTX00041 FAMHTX00051 FAMHTX00055 FAMHTX00064 FAMHTX00069 FAMHTX00081 FAMHTX00085 FAMHXX00003 FAMHXX00011 FAMHXX00019 FAMHXX00023 FAXXTE00003 FAXXTE00006 FAXXTE00009 FAXXTE00011 FAXXTE00013 FAXXTE00014 FAXXTE00015 FAXXTE00017 FAXXTE00021 FAXXTE00022 FAXXTE00024 FAXXTE00025 FAXXTE00028 FAXXTE00031 FAXXTE00032 FAXXTE00033 FAXXTE00034 FAXXTE00035 FAXXTE00037 FAXXTE00038 FAXXTE00039 FAXXTE00040 FAXXTE00044 FAXXTE00046 FAXXTE00047 FAXXTE00049 FAXXTE00050 FAXXTE00051 FAXXTE00052 FAXXTE00056 FAXXTE00062 FAXXTE00063 FAXXTE00064 FAXXTE00067 FAXXTE00071 FAXXTE00072 FAXXTE00073 FAXXTE00076 FAXXTE00084 FAXXTE00085 FAXXTE00091 FAXXTE00092 FAXXTE00094 FAXXTE00095 FAXXTE00099 FAXXTE00100 FAXXTE00101 FAXXTE00104 FAXXTE00105 FAXXTE00106 FAXXTE00107 FAXXTE00109 FAXXTE00110 FAXXTE00114 FAXXTE00116 FAXXTE00117 FAXXTE00120 FAXXTE00121 FAXXTE00128 FAXXTE00129 FAXXTE00134 FAXXTE00136 FAXXTE00137 FAXXTE00141 FAXXTE00144 FAXXTE00145 FAXXTE00147 FAXXTE00149 FAXXTE00163 FAXXTE00164 FAXXTE00170 FAXXTE00173 FAXXTE00175 FAXXTE00179 FAXXTE00180 FAXXTE00182 FAXXTE00184 FAXXTE00185 FAXXTE00195 FAXXTE00197 FAXXTE00205 FAXXTE00212 FAXXTE00220 FAXXTE00224 FAXXTE00225 FAXXTE00227 FAXXTE00228 FAXXTE00229 FAXXTE00230 FAXXTE00231 FAXXTE00234 FAXXTE00236 FAXXTE00237 FAXXTE00238 FAXXTE00240 FAXXTE00241 FAXXTE00243 FAXXTE00246 FAXXTE00248 FAXXTE00253 FAXXTE00256 FAXXTE00257 FAXXTE00258 FAXXTE00259 FAXXTE00263 FAXXTE00267 FAXXTE00269 FAXXTE00270 FAXXTE00271 FAXXTE00272 FAXXTE00277 FAXXTE00278 FAXXTE00279 FAXXTE00280 FAXXTE00282 FAXXTE00286 FAXXTE00291 FAXXTE00293 FAXXTE00295 FAXXTE00301 FAXXTE00317 FAXXTE00319 FAXXTE00320 FAXXTE00321 FAXXTE00323 FAXXTE00328 FAXXTE00329 FAXXTE00330 FAXXTE00331 FAXXTE00334 FAXXTE00336 FAXXTE00337 FAXXTE00338 FAXXTE00339 FAXXTE00340 FAXXTE00342 FAXXTE00344 FAXXTE00352 FAXXTE00353 FAXXTE00354 FAXXTE00357 FAXXTE00358 FAXXTE00361 FAXXTE00362 FAXXTE00363 FAXXTE00372 FAXXTE00373 FAXXTE00376 FAXXTE00379 FAXXTE00380 FAXXTE00383 FAXXTE00389 FAXXTE00392 FAXXTE00393 FAXXTE00396 FAXXTE00407 FAXXTE00413 FAXXTE00418 FAXXTE00422 FAXXTE00424 FAXXTE00425 FAXXTE00429 FAXXTE00431 FAXXTE00433 FAXXTE00434 FAXXTE00436 FAXXTE00439 FAXXTE00442 FAXXTE00443 FAXXTE00451 FAXXTE00452 FAXXTE00454 FAXXTE00455 FAXXTE00460 FAXXTE00461 FAXXTE00472 FAXXTE00475 FAXXTE00477 FAXXTE00481 FAXXTE00482 FAXXTE00485 FAXXTE00486 FAXXTE00487 FAXXTE00489 FAXXTE00490 FAXXTE00491 FAXXTE00492 FAXXTE00495 FAXXTE00499 FAXXTE00506 FAXXTE00508 FAXXTE00511 FAXXTE00515 FAXXTE00516 FAXXTE00517 FAXXTE00518 FAXXTE00519 FAXXTE00522 FAXXTE00523 FAXXTE00525 FAXXTE00526 FAXXTE00527 FAXXTE00532 FAXXTE00533 FAXXTE00536 FAXXTE00539 FAXXTE00540 FAXXTE00543 FAXXTE00544 FAXXTE00545 FAXXTE00546 FAXXTE00547 FAXXTE00549 FAXXTE00550 FAXXTE00551 FAXXTE00553 FAXXTE00555 FAXXTE00560 FAXXTE00561 FAXXTE00562 FAXXTE00566 FAXXTE00570 FAXXTE00571 FAXXTE00572 FAXXTE00573 FAXXTE00582 FAXXTE00591 FAXXTE00593 FAXXTE00594 FAXXTE00596 FAXXTE00604 FAXXTE00607 FAXXTE00608 FAXXTE00611 FAXXTE00615 FAXXTE00616 FAXXTE00617 FAXXTE00618 FAXXTE00624 FAXXTE00631 FAXXTE00638 FAXXTE00639 FAXXTE00643 FAXXTE00646 FAXXTE00650 FAXXTE00651 FAXXTE00652 FAXXTE00653 FAXXTE00657 FAXXTE00661 FAXXTE00664 FAXXTE00667 FAXXTE00669 FAXXTE00672 FAXXTE00680 FAXXTE00681 FAXXTE00684 FAXXTE00685 FAXXTE00687 FAXXTE00688 FAXXTE00701 FAXXTE00706 FAXXTE00707 FAXXTE00710 FAXXTE00712 FAXXTE00713 FAXXTE00716 FAXXTE00718 FAXXTE00720 FAXXTE00724 FAXXTE00725 FAXXTE00728 FAXXTE00731 FAXXTE00734 FAXXTE00735 FAXXTE00737 FAXXTE00738 FAXXTE00747 FAXXTE00748 FAXXTE00749 FAXXTE00752 FAXXTE00753 FAXXTE00762 FAXXTE00766 FAXXTE00772 FAXXTE00789 FAXXTE00794 FAXXTE00799 FAXXTE00801 FAXXTE00805 FAXXTE00809 FAXXTE00810 FAXXTE00811 FAXXTE00814 FAXXTE00819 FAXXTE00820 FAXXTE00829 FAXXTE00831 FAXXTE00843 FAXXTE00853 FAXXTE00854 FAXXTE00874 FAXXTE00881 FAXXTE00893 FAXXTE00894 FAXXTE00896 FAXXTE00897 FAXXTE00899 FAXXTE00902 FAXXTE00903 FAXXTE00906 FAXXTE00909 FAXXTE00914 FAXXTE00915 FAXXTE00916 FAXXTE00917 FAXXTE00918 FAXXTE00919 FAXXTE00922 FAXXTE00925 FAXXTE00930 FAXXTE00932 FAXXTE00936 FAXXTE00939 FAXXTE00940 FAXXTE00945 FAXXTE00958 FAXXTE00961 FAXXTE00964 FAXXTE00970 FAXXTE00973 FAXXTE00979 FAXXTE00980 FAXXTE00983 FAXXTE00985 FAXXTE00987 FAXXTE00990 FAXXTE00993 FAXXTE00995 FAXXTE00996 FAXXTE00999 FAXXTE01006 FAXXTE01029 FAXXTE01030 FAXXTE01048 FAXXTE01053 FAXXTE01057 FAXXTE01064 FAXXTE01068 FAXXTE01084 FAXXTE01085 FAXXTE01086 FAXXTE01088 FAXXTE01089 FAXXTE01092 FAXXTE01114 FAXXTE01124 FAXXTE01139 FAXXTE01140 FAXXTE01141 FAXXTE01145 FAXXTE01154 FAXXTE01156 FAXXTE01159 FAXXTE01163 FAXXTE01166 FAXXTE01167 FAXXTE01168 FAXXTE01169 FAXXTE01171 FAXXTE01180 FAXXTE01183 FAXXTE01191 FAXXTE01193 FAXXTE01194 FAXXTE01195 FAXXTE01204 FAXXTE01205 FAXXTE01206 FAXXTE01210 FAXXTE01212 FAXXTE01219 FAXXTE01222 FAXXTP00003 FAXXTP00004 FAXXTP00007 FAXXTP00009 FAXXTP00010 FAXXTP00011 FAXXTP00012 FAXXTP00014 FAXXTP00018 FAXXTP00019 FAXXTP00020 FAXXXE00003 FAXXXE00004 FAXXXE00005 FAXXXE00007 FAXXXE00011 FAXXXE00012 FAXXXE00013 FAXXXE00017 FAXXXE00020 FAXXXE00021 FAXXXE00023 FAXXXE00024 FAXXXE00025 FAXXXE00026 FAXXXE00030 FAXXXE00032 FAXXXE00033 FAXXXE00036 FAXXXE00037 FAXXXE00038 FAXXXE00039 FAXXXE00043 FAXXXE00044 FAXXXE00045 FAXXXE00046 FAXXXE00047 FAXXXE00049 FAXXXE00053 FAXXXE00054 FAXXXE00056 FAXXXE00062 FAXXXE00071 FAXXXE00072 FAXXXE00076 FAXXXE00081 FAXXXE00082 FAXXXE00083 FAXXXE00085 FAXXXE00087 FAXXXE00091 FAXXXE00092 FAXXXE00093 FAXXXE00095 FAXXXE00096 FAXXXE00097 FAXXXE00098 FAXXXE00100 FAXXXE00105 FAXXXE00107 FAXXXE00109 FAXXXE00117 FAXXXE00120 FAXXXE00121 FAXXXE00124 FAXXXE00141 FAXXXE00147 FAXXXE00152 FAXXXE00155 FAXXXE00159 FAXXXE00164 FAXXXE00165 FAXXXE00167 FAXXXE00172 FAXXXE00174 FAXXXE00176 FAXXXE00180 FAXXXE00181 FAXXXE00183 FAXXXE00184 FAXXXE00185 FAXXXE00188 FAXXXE00196 FAXXXE00197 FAXXXE00198 FAXXXE00199 FAXXXE00200 FAXXXE00204 FAXXXE00206 FAXXXE00208 FAXXXE00210 FAXXXE00215 FAXXXE00217 FAXXXE00218 FAXXXE00222 FAXXXE00225 FAXXXE00227 FAXXXE00228 FAXXXE00229 FAXXXE00230 FAXXXE00231 FAXXXE00232 FAXXXE00235 FAXXXE00236 FAXXXE00238 FAXXXE00240 FAXXXE00244 FAXXXE00247 FAXXXE00251 FAXXXE00253 FAXXXE00254 FAXXXE00255 FAXXXE00256 FAXXXE00257 FAXXXE00261 FAXXXE00269 FAXXXE00270 FAXXXE00273 FAXXXE00275 FAXXXE00278 FAXXXE00279 FAXXXE00280 FAXXXE00282 FAXXXE00283 FAXXXE00287 FAXXXE00288 FAXXXE00290 FAXXXE00291 FAXXXE00292 FAXXXE00296 FAXXXE00297 FAXXXE00302 FAXXXE00303 FAXXXE00306 FAXXXE00312 FAXXXE00314 FAXXXE00315 FAXXXE00318 FAXXXE00319 FAXXXE00320 FAXXXE00321 FAXXXE00324 FAXXXE00325 FAXXXE00327 FAXXXE00328 FAXXXE00329 FAXXXE00330 FAXXXE00332 FAXXXE00335 FAXXXE00336 FAXXXE00339 FAXXXE00340 FAXXXE00343 FAXXXE00345 FAXXXE00346 FAXXXE00348 FAXXXE00354 FAXXXE00356 FAXXXE00359 FAXXXX00014 FAXXXX00020 FBHVXX00037 FBHVXX00040 FBHXTX00002 FBHXTX00006 FBHXTX00007 FBHXTX00009 FBHXTX00010 FBHXTX00011 FBHXTX00013 FBHXTX00019 FBHXTX00020 FBHXTX00021 FBHXTX00022 FBHXTX00023 FBHXTX00024 FBHXTX00025 FBHXTX00026 FBHXTX00031 FBHXTX00032 FBHXTX00033 FBHXTX00035 FBHXTX00038 FBHXTX00045 FBHXTX00047 FBHXTX00048 FBHXTX00051 FBHXTX00057 FBHXTX00059 FBHXTX00062 FBHXTX00068 FBHXTX00070 FBHXTX00071 FBHXTX00074 FBHXTX00079 FBHXTX00080 FBHXTX00081 FBHXTX00095 FBHXTX00098 FBHXTX00102 FBHXTX00103 FBHXTX00104 FBHXTX00106 FBHXTX00114 FBHXTX00115 FBHXTX00125 FBHXTX00126 FBHXTX00128 FBHXTX00130 FBHXTX00137 FBHXTX00144 FBHXTX00149 FBHXTX00151 FBHXTX00154 FBHXTX00159 FBHXTX00161 FBHXTX00163 FBHXTX00172 FBHXTX00173 FBHXTX00177 FBHXTX00184 FBHXTX00198 FBHXTX00199 FBHXTX00201 FBHXTX00203 FBHXTX00204 FBHXTX00205 FBHXTX00208 FBHXTX00210 FBHXTX00218 FBHXTX00223 FBHXTX00231 FBHXTX00233 FBHXTX00234 FBHXTX00235 FBHXTX00237 FBHXTX00241 FBHXTX00261 FBHXTX00266 FBHXTX00274 FBHXTX00275 FBHXTX00276 FBHXTX00277 FBHXTX00278 FBHXTX00279 FBHXTX00281 FBHXTX00282 FBHXTX00287 FBHXTX00290 FBHXTX00295 FBHXTX00316 FBHXTX00317 FBHXTX00319 FBHXTX00320 FBHXTX00324 FBHXTX00330 FBHXTX00335 FBHXTX00336 FBHXTX00337 FBHXTX00338 FBHXTX00339 FBHXTX00343 FBHXTX00344 FBHXTX00345 FBHXTX00348 FBHXTX00349 FBHXTX00350 FBHXTX00351 FBHXTX00352 FBHXTX00353 FBHXTX00354 FBHXTX00355 FBHXTX00356 FBHXTX00357 FBHXTX00360 FBHXTX00361 FBHXTX00362 FBHXTX00365 FBHXTX00366 FBHXTX00368 FBHXTX00369 FBHXTX00371 FBHXTX00375 FBHXTX00379 FBHXTX00381 FBHXTX00411 FBHXTX00415 FBHXTX00416 FBHXTX00417 FBHXTX00429 FBHXTX00431 FBHXTX00432 FBHXTX00433 FBHXTX00436 FBHXTX00438 FBHXTX00442 FBHXTX00444 FBHXTX00447 FBHXTX00448 FBHXTX00451 FBHXTX00458 FBHXTX00462 FBHXTX00468 FBHXTX00479 FBHXTX00482 FBHXTX00490 FBHXTX00491 FBHXTX00493 FBHXTX00498 FBHXTX00499 FBHXTX00500 FBHXTX00501 FBHXTX00503 FBHXTX00504 FBHXTX00505 FBHXTX00507 FBHXTX00508 FBHXTX00509 FBHXTX00510 FBHXTX00514 FBHXTX00516 FBHXTX00520 FBHXTX00524 FBHXTX00529 FBHXTX00535 FBHXTX00537 FBHXTX00541 FBHXTX00543 FBHXTX00546 FBHXTX00549 FBHXTX00551 FBHXTX00555 FBHXTX00558 FBHXTX00561 FBHXTX00565 FBHXTX00568 FBHXTX00570 FBHXTX00571 FBHXTX00580 FBHXTX00583 FBHXTX00584 FBHXTX00586 FBHXTX00587 FBHXTX00589 FBHXTX00593 FBHXTX00597 FBHXTX00599 FBHXTX00600 FBHXTX00601 FBHXTX00605 FBHXTX00606 FBHXTX00613 FBHXTX00614 FBHXTX00616 FBHXTX00617 FBHXTX00624 FBHXTX00630 FBHXTX00641 FBHXTX00642 FBHXTX00644 FBHXTX00649 FBHXTX00650 FBHXTX00651 FBHXTX00655 FBHXTX00664 FBHXTX00665 FBHXTX00667 FBHXTX00669 FBHXTX00671 FBHXTX00672 FBHXTX00673 FBHXTX00679 FBHXTX00680 FBHXTX00682 FBHXTX00684 FBHXTX00686 FBHXTX00687 FBHXTX00689 FBHXTX00690 FBHXTX00691 FBHXTX00693 FBHXTX00698 FBHXTX00700 FBHXTX00703 FBHXTX00713 FBHXTX00719 FBHXTX00721 FBHXTX00727 FBHXTX00737 FBHXTX00745 FBHXTX00748 FBHXTX00753 FBHXTX00755 FBHXTX00757 FBHXTX00758 FBHXTX00760 FBHXTX00762 FBHXTX00764 FBHXTX00769 FBHXTX00770 FBHXTX00771 FBHXTX00774 FBHXTX00783 FBHXTX00785 FBHXTX00798 FBHXTX00800 FBHXTX00803 FBHXTX00804 FBHXTX00821 FBHXTX00823 FBHXTX00826 FBHXTX00834 FBHXTX00837 FBHXTX00839 FBHXTX00841 FBHXTX00843 FBHXTX00844 FBHXTX00847 FBHXTX00848 FBHXTX00849 FBHXTX00866 FBHXTX00867 FBHXTX00894 FBHXTX00896 FBHXTX00897 FBHXTX00898 FBHXTX00899 FBHXTX00901 FBHXTX00909 FBHXTX00912 FBHXTX00913 FBHXTX00914 FBHXTX00916 FBHXTX00920 FBHXTX00921 FBHXTX00922 FBHXTX00924 FBHXTX00925 FBHXTX00926 FBHXTX00930 FBHXTX00933 FBHXTX00934 FBHXTX00936 FBHXTX00937 FBHXTX00938 FBHXTX00942 FBHXTX00943 FBHXTX00946 FBHXTX00952 FBHXTX00954 FBHXTX00955 FBHXTX00964 FBHXTX00971 FBHXTX00972 FBHXTX00974 FBHXTX00977 FBHXTX00980 FBHXTX00982 FBHXTX00986 FBHXTX00991 FBHXTX00995 FBHXTX00996 FBHXTX01002 FBHXTX01006 FBHXTX01013 FBHXTX01014 FBHXTX01016 FBHXTX01031 FBHXTX01040 FBHXTX01045 FBHXTX01048 FBHXTX01049 FBHXTX01050 FBHXTX01062 FBHXTX01063 FBHXTX01070 FBHXTX01074 FBHXTX01075 FBHXTX01076 FBHXTX01077 FBHXTX01078 FBHXTX01079 FBHXTX01080 FBHXTX01100 FBHXTX01101 FBHXTX01102 FBHXTX01104 FBHXTX01108 FBHXTX01109 FBHXTX01110 FBHXTX01111 FBHXTX01119 FBHXTX01123 FBHXTX01126 FBHXTX01127 FBHXTX01130 FBHXTX01131 FBHXTX01132 FBHXTX01135 FBHXTX01138 FBHXTX01139 FBHXTX01140 FBHXTX01145 FBHXTX01147 FBHXTX01150 FBHXTX01153 FBHXTX01154 FBHXTX01156 FBHXTX01157 FBHXTX01158 FBHXTX01165 FBHXTX01170 FBHXTX01176 FBHXTX01180 FBHXTX01184 FBHXTX01189 FBHXTX01193 FBHXTX01194 FBHXTX01200 FBHXTX01201 FBHXTX01202 FBHXTX01217 FBHXTX01222 FBHXTX01223 FBHXTX01228 FBHXTX01229 FBHXTX01230 FBHXTX01231 FBHXTX01232 FBHXTX01233 FBHXTX01234 FBHXTX01239 FBHXTX01240 FBHXTX01241 FBHXTX01244 FBHXTX01246 FBHXTX01251 FBHXTX01252 FBHXTX01264 FBHXTX01265 FBHXTX01267 FBHXTX01268 FBHXTX01269 FBHXTX01270 FBHXTX01271 FBHXTX01273 FBHXTX01286 FBHXTX01287 FBHXTX01289 FBHXTX01291 FBHXTX01303 FBHXTX01306 FBHXTX01311 FBHXTX01319 FBHXTX01322 FBHXTX01323 FBHXTX01326 FBHXTX01330 FBHXTX01331 FBHXTX01348 FBHXTX01356 FBHXTX01368 FBHXTX01371 FBHXTX01373 FBHXTX01374 FBHXTX01381 FBHXTX01382 FBHXTX01387 FBHXTX01388 FBHXTX01389 FBHXTX01397 FBHXTX01400 FBHXTX01405 FBHXTX01410 FBHXTX01421 FBHXXX00016 FBHXXX00027 FBHXXX00032 FBHXXX00036 FBHXXX00041 FBHXXX00042 FBHXXX00045 FBHXXX00050 FBHXXX00054 FBHXXX00056 FBHXXX00060 FBHXXX00061 FBHXXX00063 FBHXXX00065 FBHXXX00067 FBHXXX00069 FBHXXX00074 FBHXXX00077 FBHXXX00078 FBHXXX00081 FBHXXX00085 FBHXXX00088 FBHXXX00091 FBHXXX00092 FBHXXX00093 FBHXXX00094 FBHXXX00097 FBHXXX00099 FBHXXX00100 FBHXXX00101 FBHXXX00102 FBHXXX00104 FBHXXX00105 FBHXXX00106 FBHXXX00107 FBHXXX00108 FBHXXX00109 FBHXXX00110 FBHXXX00111 FBHXXX00113 FBHXXX00117 FBHXXX00118 FBHXXX00119 FBHXXX00120 FBHXXX00121 FBHXXX00122 FBHXXX00126 FBHXXX00127 FBHXXX00132 FBHXXX00133 FBHXXX00134 FBHXXX00135 FBHXXX00143 FBHXXX00148 FBHXXX00156 FBHXXX00158 FBHXXX00159 FBHXXX00160 FBHXXX00166 FBHXXX00169 FBHXXX00170 FBHXXX00171 FBHXXX00179 FBHXXX00180 FBHXXX00182 FBHXXX00185 FBHXXX00186 FBHXXX00187 FBHXXX00189 FBHXXX00192 FBHXXX00196 FBHXXX00202 FBHXXX00204 FBHXXX00205 FBHXXX00206 FBHXXX00207 FBHXXX00208 FBHXXX00211 FBHXXX00213 FBHXXX00218 FBHXXX00225 FBHXXX00231 FBHXXX00238 FBHXXX00239 FBHXXX00240 FBHXXX00243 FBHXXX00246 FBHXXX00247 FBHXXX00248 FBHXXX00249 FBHXXX00251 FBHXXX00256 FBHXXX00257 FBHXXX00265 FBHXXX00266 FBHXXX00269 FBHXXX00270 FBHXXX00273 FBHXXX00275 FBHXXX00277 FBHXXX00280 FBHXXX00281 FBHXXX00285 FBHXXX00286 FBHXXX00288 FBHXXX00290 FBHXXX00291 FBHXXX00292 FBHXXX00293 FBHXXX00298 FBHXXX00300 FBHXXX00301 FBHXXX00303 FBHXXX00304 FBHXXX00305 FBHXXX00309 FBHXXX00311 FBHXXX00313 FBHXXX00319 FBHXXX00320 FBHXXX00321 FBHXXX00322 FBHXXX00326 FBHXXX00327 FBHXXX00330 FBHXXX00331 FBHXXX00332 FBHXXX00333 FBHXXX00338 FBHXXX00339 FBHXXX00342 FBHXXX00352 FBHXXX00353 FBHXXX00356 FBHXXX00361 FBHXXX00362 FBHXXX00365 FBHXXX00367 FBHXXX00368 FBHXXX00371 FBHXXX00372 FBHXXX00374 FBHXXX00389 FBHXXX00390 FBHXXX00391 FBHXXX00393 FBHXXX00394 FBHXXX00395 FBHXXX00401 FBHXXX00402 FBMHDX00013 FBMHDX00015 FBMHDX00018 FBMHDX00019 FBMHDX00020 FBMHDX00021 FBMHDX00024 FBMHDX00025 FBMHDX00026 FBMHDX00028 FBMHDX00029 FBMHDX00030 FBMHDX00043 FBMHDX00044 FBMHDX00046 FBMHDX00047 FBMHDX00049 FBMHDX00054 FBMHDX00056 FBMHDX00058 FBMHDX00059 FBMHDX00061 FBMHDX00065 FBMHDX00066 FBMHDX00067 FBMHDX00071 FBMHDX00073 FBMHDX00078 FBMHDX00080 FBMHDX00081 FBMHDX00082 FBMHDX00083 FBMHDX00085 FBMHDX00086 FBMHDX00087 FBMHDX00089 FBMHDX00090 FBMHDX00092 FBMHDX00095 FBMHDX00097 FBMHDX00098 FBMHDX00100 FBMHDX00101 FBMHDX00104 FBMHDX00107 FBMHDX00108 FBMHDX00110 FBMHDX00111 FBMHDX00112 FBMHDX00118 FBMHDX00119 FBMHDX00120 FBMHDX00122 FBMHDX00124 FBMHDX00125 FBMHDX00126 FBMHDX00127 FBMHDX00129 FBMHDX00130 FBMHDX00133 FBMHDX00134 FBMHDX00136 FBMHDX00137 FBMHDX00139 FBMHDX00140 FBMHDX00141 FBMHDX00143 FBMHDX00144 FBMHDX00147 FBMHDX00151 FBMHDX00152 FBMHDX00153 FBMHDX00155 FBMHDX00157 FBMHDX00163 FBMHDX00166 FBMHDX00168 FBMHDX00169 FBMHDX00170 FBMHDX00173 FBMHDX00177 FBMHDX00180 FBMHDX00182 FBMHDX00184 FBMHDX00186 FBMHDX00187 FBMHDX00188 FBMHDX00189 FBMHDX00193 FBMHDX00195 FBMHDX00196 FBMHDX00197 FBMHDX00198 FBMHDX00200 FBMHDX00201 FBMHDX00203 FBMHDX00208 FBMHDX00210 FBMHDX00212 FBMHDX00216 FBMHDX00218 FBMHDX00220 FBMHDX00221 FBMHDX00224 FBMHDX00226 FBMHDX00227 FBMHDX00231 FBMHDX00237 FBMHDX00238 FBMHDX00239 FBMHDX00240 FBMHDX00245 FBMHDX00250 FBMHDX00252 FBMHDX00253 FBMHDX00255 FBMHDX00262 FBMHDX00264 FBMHDX00266 FBMHDX00271 FBMHDX00273 FBMHDX00282 FBMHDX00284 FBMHDX00286 FBMHDX00287 FBMHDX00290 FBMHDX00292 FBMHDX00294 FBMHDX00296 FBMHDX00297 FBMHDX00301 FBMHDX00302 FBMHDX00310 FBMHDX00313 FBMHDX00314 FBMHDX00318 FBMHDX00321 FBMHDX00326 FBMHDX00328 FBMHDX00333 FBMHDX00335 FBMHDX00346 FBMHDX00350 FBMHDX00352 FBMHDX00363 FBMHDX00366 FBMHDX00369 FBMHDX00374 FBMHDX00375 FBMHDX00379 FBMHDX00386 FBMHDX00387 FBMHDX00388 FBMHDX00389 FBMHDX00395 FBMHDX00397 FBMHDX00398 FBMHDX00399 FBMHDX00400 FBMHDX00401 FBMHDX00403 FBMHDX00406 FBMHDX00409 FBMHDX00410 FBMHDX00412 FBMHDX00413 FBMHDX00416 FBMHDX00423 FBMHDX00424 FBMHDX00425 FBMHDX00430 FBMHDX00431 FBMHDX00437 FBMHDX00439 FBMHDX00440 FBMHDX00441 FBMHDX00445 FBMHDX00449 FBMHDX00451 FBMHDX00452 FBMHDX00453 FBMHDX00458 FBMHDX00459 FBMHDX00460 FBMHDX00461 FBMHDX00462 FBMHDX00463 FBMHDX00464 FBMHDX00466 FBMHDX00468 FBMHDX00471 FBMHDX00477 FBMHDX00480 FBMHDX00487 FBMHDX00490 FBMHDX00491 FBMHDX00500 FBMHDX00509 FBMHDX00510 FBMHDX00516 FBMHDX00518 FBMHDX00521 FBMHDX00522 FBMHDX00523 FBMHDX00524 FBMHDX00525 FBMHDX00527 FBMHDX00528 FBMHDX00531 FBMHDX00532 FBMHDX00534 FBMHDX00542 FBMHDX00543 FBMHDX00544 FBMHDX00552 FBMHDX00554 FBMHTX00006 FBMHTX00007 FBMHTX00008 FBMHTX00017 FBMHTX00019 FBMHTX00020 FBMHTX00021 FBMHTX00024 FBMHTX00028 FBMHTX00030 FBMHTX00033 FBMHTX00041 FBMHTX00044 FBMHTX00046 FBMHTX00052 FBMHTX00053 FBMHTX00054 FBMHTX00057 FBMHTX00059 FBMHTX00060 FBMHTX00065 FBMHTX00066 FBMHTX00068 FBMHTX00070 FBMHTX00074 FBMHTX00077 FBMHTX00078 FBMHTX00081 FBMHTX00084 FBMHTX00088 FBMHTX00091 FBMHTX00094 FBMHTX00104 FBMHTX00106 FBMHTX00107 FBMHTX00113 FBMHTX00115 FBMHTX00116 FBMHTX00117 FBMHTX00118 FBMHTX00122 FBMHTX00126 FBMHTX00127 FBMHTX00129 FBMHTX00130 FBMHTX00133 FBMHTX00139 FBMHTX00140 FBMHTX00141 FBMHTX00143 FBMHTX00147 FBMHTX00151 FBMHTX00155 FBMHTX00161 FBMHTX00162 FBMHTX00163 FBMHTX00164 FBMHTX00166 FBMHTX00167 FBMHTX00169 FBMHTX00170 FBMHTX00171 FBMHTX00172 FBMHTX00173 FBMHTX00179 FBMHTX00180 FBMHTX00185 FBMHTX00186 FBMHTX00187 FBMHTX00188 FBMHTX00191 FBMHTX00195 FBMHTX00200 FBMHTX00203 FBMHTX00205 FBMHTX00210 FBMHTX00220 FBMHTX00221 FBMHTX00222 FBMHTX00223 FBMHTX00225 FBMHTX00230 FBMHTX00238 FBMHTX00245 FBMHTX00246 FBMHTX00250 FBMHTX00252 FBMHTX00255 FBMHTX00259 FBMHTX00275 FBMHTX00276 FBMHTX00278 FBMHTX00279 FBMHTX00280 FBMHTX00282 FBMHTX00283 FBMHTX00285 FBMHTX00290 FBMHTX00292 FBMHTX00293 FBMHTX00298 FBMHTX00306 FBMHTX00309 FBMHTX00310 FBMHTX00313 FBMHTX00329 FBMHTX00330 FBMHTX00333 FBMHTX00343 FBMHTX00344 FBMHTX00347 FBMHTX00353 FBMHTX00354 FBMHTX00358 FBMHTX00363 FBMHTX00369 FBMHTX00372 FBMHTX00383 FBMHTX00389 FBMHTX00392 FBMHTX00393 FBMHTX00399 FBMHTX00404 FBMHTX00405 FBMHTX00407 FBMHTX00408 FBMHTX00409 FBMHTX00410 FBMHTX00432 FBMHTX00433 FBMHTX00436 FBMHTX00437 FBMHTX00444 FBMHTX00446 FBMHTX00462 FBMHTX00467 FBMHTX00471 FBMHTX00473 FBMHTX00477 FBMHTX00488 FBMHTX00492 FBMHXX00011 FBMHXX00016 FBMHXX00017 FBMHXX00019 FBMHXX00024 FBMHXX00026 FBMHXX00032 FBMHXX00034 FBMHXX00037 FBMHXX00038 FBMHXX00043 FBMHXX00047 FBMHXX00049 FBMHXX00051 FBMHXX00052 FBMHXX00055 FBMHXX00057 FBMHXX00060 FBMHXX00065 FBMHXX00067 FBMHXX00072 FBMHXX00073 FBMHXX00081 FBMHXX00082 FBMHXX00083 FBMHXX00085 FBMHXX00086 FBMHXX00087 FBMHXX00089 FBMHXX00096 FBMHXX00099 FBMHXX00101 FBMHXX00102 FBMHXX00103 FBMHXX00104 FBXXTE00005 FBXXTE00010 FBXXTE00012 FBXXTE00013 FBXXTE00015 FBXXTE00016 FBXXTE00018 FBXXTE00025 FBXXTE00028 FBXXTE00029 FBXXTE00030 FBXXTE00032 FBXXTE00033 FBXXTE00035 FBXXTE00038 FBXXTE00039 FBXXTE00040 FBXXTE00043 FBXXTE00044 FBXXTE00047 FBXXTE00050 FBXXTE00051 FBXXTE00053 FBXXTE00054 FBXXTE00055 FBXXTE00056 FBXXTE00057 FBXXTE00064 FBXXTE00067 FBXXTE00075 FBXXTE00079 FBXXTE00082 FBXXTE00085 FBXXTE00088 FBXXTE00091 FBXXTE00092 FBXXTE00093 FBXXTE00098 FBXXTE00101 FBXXTE00102 FBXXTE00104 FBXXTE00105 FBXXTE00108 FBXXTE00110 FBXXTE00113 FBXXTE00114 FBXXTE00115 FBXXTE00117 FBXXTE00119 FBXXTE00120 FBXXTE00122 FBXXTE00123 FBXXTE00124 FBXXTE00125 FBXXTE00126 FBXXTE00129 FBXXTE00130 FBXXTE00131 FBXXTE00134 FBXXTE00136 FBXXTE00137 FBXXTE00138 FBXXTE00139 FBXXTE00148 FBXXTE00155 FBXXTE00158 FBXXTE00159 FBXXTE00160 FBXXTE00161 FBXXTE00163 FBXXTE00164 FBXXTE00165 FBXXTE00167 FBXXTE00172 FBXXTE00173 FBXXTE00175 FBXXTE00177 FBXXTE00180 FBXXTE00181 FBXXTE00182 FBXXTE00186 FBXXTE00187 FBXXTE00190 FBXXTE00196 FBXXTE00199 FBXXTE00203 FBXXTE00204 FBXXTE00206 FBXXTE00208 FBXXTE00209 FBXXTE00216 FBXXTE00217 FBXXTE00222 FBXXTE00223 FBXXTE00228 FBXXTE00229 FBXXTE00234 FBXXTE00237 FBXXTE00238 FBXXTE00249 FBXXTE00251 FBXXTE00252 FBXXTE00254 FBXXTE00256 FBXXTE00262 FBXXTE00263 FBXXTE00266 FBXXTE00267 FBXXTE00272 FBXXTE00279 FBXXTE00282 FBXXTE00283 FBXXTE00285 FBXXTE00288 FBXXTE00294 FBXXTE00300 FBXXTE00306 FBXXTE00316 FBXXTE00317 FBXXTE00318 FBXXTE00319 FBXXTE00321 FBXXTE00325 FBXXTE00326 FBXXTE00328 FBXXTE00329 FBXXTE00332 FBXXTE00339 FBXXTE00342 FBXXTE00346 FBXXTE00347 FBXXTE00353 FBXXTE00356 FBXXTE00357 FBXXTE00358 FBXXTE00359 FBXXTE00360 FBXXTE00384 FBXXTE00386 FBXXTE00389 FBXXTE00390 FBXXTE00391 FBXXTE00393 FBXXTE00401 FBXXTE00403 FBXXTE00406 FBXXTE00407 FBXXTE00409 FBXXTE00415 FBXXTE00420 FBXXTE00422 FBXXTE00430 FBXXTE00431 FBXXTE00432 FBXXTE00433 FBXXTE00434 FBXXTE00436 FBXXTE00443 FBXXTE00450 FBXXTE00454 FBXXTE00467 FBXXTE00480 FBXXTE00485 FBXXTE00496 FBXXTE00497 FBXXTE00502 FBXXTE00504 FBXXTE00506 FBXXTE00507 FBXXTE00509 FBXXTE00512 FBXXTE00513 FBXXTE00514 FBXXTE00515 FBXXTE00519 FBXXTE00520 FBXXTE00525 FBXXTE00528 FBXXTE00533 FBXXTE00534 FBXXTE00535 FBXXTE00537 FBXXTE00539 FBXXTE00543 FBXXTE00550 FBXXTE00553 FBXXTE00555 FBXXTE00557 FBXXTE00558 FBXXTE00559 FBXXTE00560 FBXXTE00561 FBXXTE00562 FBXXTE00563 FBXXTE00564 FBXXTE00565 FBXXTE00566 FBXXTE00567 FBXXTE00569 FBXXTE00571 FBXXTE00572 FBXXTE00575 FBXXTE00576 FBXXTE00577 FBXXTE00578 FBXXTE00579 FBXXTE00581 FBXXTE00582 FBXXTE00584 FBXXTE00585 FBXXTE00586 FBXXTE00587 FBXXTE00588 FBXXTE00589 FBXXTE00593 FBXXTE00595 FBXXTE00596 FBXXTE00597 FBXXTE00600 FBXXTE00602 FBXXTE00603 FBXXTE00613 FBXXTE00614 FBXXTE00617 FBXXTE00623 FBXXTE00625 FBXXTE00630 FBXXTE00631 FBXXTE00636 FBXXTE00638 FBXXTE00639 FBXXTE00640 FBXXTE00641 FBXXTE00642 FBXXTE00643 FBXXTE00644 FBXXTE00645 FBXXTE00646 FBXXTE00652 FBXXTE00654 FBXXTE00657 FBXXTE00659 FBXXTE00662 FBXXTE00663 FBXXTE00665 FBXXTE00668 FBXXTE00669 FBXXTE00681 FBXXTE00682 FBXXTE00684 FBXXTE00685 FBXXTE00692 FBXXTE00694 FBXXTE00700 FBXXTE00701 FBXXTE00702 FBXXTE00704 FBXXTE00708 FBXXTE00709 FBXXTE00712 FBXXTE00713 FBXXTE00714 FBXXTE00717 FBXXTE00720 FBXXTE00723 FBXXTE00725 FBXXTE00728 FBXXTE00735 FBXXTE00736 FBXXTE00737 FBXXTE00740 FBXXTE00742 FBXXTE00744 FBXXTE00745 FBXXTE00746 FBXXTE00748 FBXXTE00749 FBXXTE00755 FBXXTE00769 FBXXTE00771 FBXXTE00772 FBXXTE00774 FBXXTE00775 FBXXTE00776 FBXXTE00777 FBXXTE00779 FBXXTE00783 FBXXTE00787 FBXXTE00790 FBXXTE00792 FBXXTE00801 FBXXTE00802 FBXXTE00805 FBXXTE00807 FBXXTE00811 FBXXTE00812 FBXXTE00813 FBXXTE00814 FBXXTE00815 FBXXTE00825 FBXXTE00828 FBXXTE00829 FBXXTE00831 FBXXTE00832 FBXXTE00833 FBXXTE00834 FBXXTE00837 FBXXTE00839 FBXXTE00841 FBXXTE00842 FBXXTE00845 FBXXTE00846 FBXXTE00851 FBXXTE00852 FBXXTE00856 FBXXTE00859 FBXXTE00860 FBXXTE00861 FBXXTE00862 FBXXTE00863 FBXXTE00864 FBXXTE00865 FBXXTE00867 FBXXTE00868 FBXXTE00877 FBXXTE00878 FBXXTE00880 FBXXTE00883 FBXXTE00886 FBXXTE00887 FBXXTE00889 FBXXTE00895 FBXXTE00896 FBXXTE00897 FBXXTE00900 FBXXTE00904 FBXXTE00915 FBXXTE00917 FBXXTE00919 FBXXTE00925 FBXXTE00926 FBXXTE00929 FBXXTE00936 FBXXTE00942 FBXXTE00952 FBXXTE00953 FBXXTE00954 FBXXTE00957 FBXXTE00958 FBXXTE00960 FBXXTE00962 FBXXTE00963 FBXXTE00975 FBXXTE00979 FBXXTE00980 FBXXTE00981 FBXXTE00985 FBXXTE00994 FBXXTE00999 FBXXTE01003 FBXXTE01004 FBXXTE01005 FBXXTE01013 FBXXTE01014 FBXXTE01015 FBXXTE01018 FBXXTE01020 FBXXTE01027 FBXXTE01030 FBXXTE01033 FBXXTE01034 FBXXTE01035 FBXXTE01037 FBXXTE01038 FBXXTE01039 FBXXTE01041 FBXXTE01047 FBXXTE01048 FBXXTE01049 FBXXTE01052 FBXXTE01054 FBXXTE01055 FBXXTE01058 FBXXTE01059 FBXXTE01060 FBXXTE01061 FBXXTE01065 FBXXTE01067 FBXXTE01068 FBXXTE01070 FBXXTE01073 FBXXTE01074 FBXXTE01075 FBXXTE01077 FBXXTE01080 FBXXTE01083 FBXXTE01084 FBXXTE01085 FBXXTE01088 FBXXTE01089 FBXXTE01093 FBXXTE01096 FBXXTE01097 FBXXTE01098 FBXXTE01100 FBXXTE01102 FBXXTE01103 FBXXTE01104 FBXXTE01106 FBXXTE01108 FBXXTE01109 FBXXTE01112 FBXXTE01113 FBXXTE01117 FBXXTE01119 FBXXTE01121 FBXXTE01123 FBXXTE01126 FBXXTE01129 FBXXTE01130 FBXXTE01131 FBXXTE01132 FBXXTE01143 FBXXTE01146 FBXXTE01148 FBXXTE01153 FBXXTE01154 FBXXTE01160 FBXXTE01165 FBXXTE01166 FBXXTE01167 FBXXTE01168 FBXXTE01169 FBXXTE01173 FBXXTE01177 FBXXTE01181 FBXXTE01183 FBXXTE01184 FBXXTE01185 FBXXTE01186 FBXXTE01190 FBXXTE01193 FBXXTE01194 FBXXTE01197 FBXXTE01198 FBXXTE01200 FBXXTE01201 FBXXTE01203 FBXXTE01207 FBXXTE01209 FBXXTE01210 FBXXTE01211 FBXXTE01216 FBXXTE01218 FBXXTE01222 FBXXTE01224 FBXXTE01226 FBXXTE01227 FBXXTE01228 FBXXTE01230 FBXXTE01231 FBXXTE01232 FBXXTE01233 FBXXTE01235 FBXXTE01238 FBXXTE01241 FBXXTE01249 FBXXTE01253 FBXXTE01257 FBXXTE01260 FBXXTE01262 FBXXTE01263 FBXXTE01267 FBXXTE01268 FBXXTE01272 FBXXTE01273 FBXXTE01274 FBXXTE01275 FBXXTE01276 FBXXTE01277 FBXXTE01278 FBXXTE01281 FBXXTE01282 FBXXTE01286 FBXXTE01287 FBXXTE01288 FBXXTE01289 FBXXTE01290 FBXXTE01291 FBXXTE01292 FBXXTE01293 FBXXTE01294 FBXXTE01296 FBXXTE01298 FBXXTE01301 FBXXTE01305 FBXXTE01306 FBXXTE01307 FBXXTE01310 FBXXTE01312 FBXXTE01313 FBXXTE01314 FBXXTE01315 FBXXTE01317 FBXXTE01320 FBXXTE01325 FBXXTE01326 FBXXTE01331 FBXXTE01333 FBXXTE01337 FBXXTE01343 FBXXTE01344 FBXXTE01349 FBXXTE01353 FBXXTE01355 FBXXTE01356 FBXXTE01363 FBXXTE01371 FBXXTE01377 FBXXTE01382 FBXXTE01384 FBXXTE01385 FBXXTE01387 FBXXTE01397 FBXXTE01399 FBXXTE01404 FBXXTE01407 FBXXTE01410 FBXXTE01411 FBXXTE01412 FBXXTE01414 FBXXTE01416 FBXXTE01417 FBXXTE01418 FBXXTE01419 FBXXTE01421 FBXXTE01422 FBXXTE01430 FBXXTE01432 FBXXTE01436 FBXXTE01437 FBXXTE01438 FBXXTE01441 FBXXTE01444 FBXXTE01446 FBXXTE01448 FBXXTE01449 FBXXTE01451 FBXXTE01452 FBXXTE01453 FBXXTE01454 FBXXTE01457 FBXXTE01458 FBXXTE01460 FBXXTE01461 FBXXTE01462 FBXXTE01463 FBXXTE01465 FBXXTE01466 FBXXTE01467 FBXXTE01469 FBXXTE01472 FBXXTE01474 FBXXTE01476 FBXXTE01477 FBXXTE01478 FBXXTE01479 FBXXTE01480 FBXXTE01482 FBXXTE01483 FBXXTE01485 FBXXTE01486 FBXXTE01487 FBXXTE01488 FBXXTE01489 FBXXTE01491 FBXXTE01494 FBXXTE01496 FBXXTE01497 FBXXTE01500 FBXXTE01501 FBXXTE01502 FBXXTE01504 FBXXTE01505 FBXXTE01507 FBXXTE01508 FBXXTE01509 FBXXTE01510 FBXXTE01512 FBXXTE01513 FBXXTE01514 FBXXTE01515 FBXXTE01516 FBXXTE01517 FBXXTE01518 FBXXTE01519 FBXXTE01520 FBXXTE01521 FBXXTE01522 FBXXTE01525 FBXXTE01526 FBXXTE01528 FBXXTE01529 FBXXTE01532 FBXXTE01535 FBXXTE01537 FBXXTE01538 FBXXTE01539 FBXXTE01540 FBXXTE01542 FBXXTE01543 FBXXTE01544 FBXXTE01546 FBXXTE01547 FBXXTE01549 FBXXTE01551 FBXXTE01553 FBXXTE01555 FBXXTE01557 FBXXTE01559 FBXXTE01563 FBXXTE01566 FBXXTE01569 FBXXTE01570 FBXXTE01571 FBXXTE01572 FBXXTE01574 FBXXTE01579 FBXXTE01585 FBXXTE01586 FBXXTE01587 FBXXTE01590 FBXXTE01593 FBXXTE01602 FBXXTE01604 FBXXTE01613 FBXXTE01617 FBXXTE01618 FBXXTE01619 FBXXTE01620 FBXXTE01630 FBXXTE01632 FBXXTE01634 FBXXTE01644 FBXXTE01650 FBXXTE01651 FBXXTE01654 FBXXTE01663 FBXXTE01667 FBXXTE01672 FBXXTE01673 FBXXTE01677 FBXXTE01678 FBXXTE01681 FBXXTE01686 FBXXTE01687 FBXXTE01688 FBXXTE01690 FBXXTE01691 FBXXTE01692 FBXXTE01702 FBXXTE01704 FBXXTE01708 FBXXTE01718 FBXXTE01719 FBXXTE01720 FBXXTE01722 FBXXTE01723 FBXXTE01725 FBXXTE01727 FBXXTE01728 FBXXTE01731 FBXXTE01732 FBXXTE01734 FBXXTE01740 FBXXTE01748 FBXXTE01750 FBXXTE01754 FBXXTE01756 FBXXTE01758 FBXXTE01760 FBXXTE01761 FBXXTE01762 FBXXTE01765 FBXXTE01766 FBXXTE01767 FBXXTE01768 FBXXTE01769 FBXXTE01770 FBXXTE01771 FBXXTE01775 FBXXTE01779 FBXXTE01786 FBXXTE01787 FBXXTE01789 FBXXTE01801 FBXXTE01803 FBXXTE01805 FBXXTE01806 FBXXTE01811 FBXXTE01812 FBXXTE01820 FBXXTE01822 FBXXTE01826 FBXXTE01829 FBXXTE01830 FBXXTE01832 FBXXTE01835 FBXXTE01837 FBXXTE01839 FBXXTE01841 FBXXTE01845 FBXXTE01846 FBXXTE01847 FBXXTE01851 FBXXTE01852 FBXXTE01853 FBXXTE01856 FBXXTE01858 FBXXTE01860 FBXXTE01861 FBXXTE01862 FBXXTE01863 FBXXTE01865 FBXXTE01874 FBXXTE01875 FBXXTE01876 FBXXTE01877 FBXXTE01881 FBXXTE01884 FBXXTE01885 FBXXTE01889 FBXXTE01890 FBXXTE01892 FBXXTE01901 FBXXTE01903 FBXXTE01906 FBXXTE01908 FBXXTE01909 FBXXTE01913 FBXXTE01919 FBXXTE01921 FBXXTE01922 FBXXTE01923 FBXXTE01926 FBXXTE01929 FBXXTE01930 FBXXTE01931 FBXXTE01932 FBXXTE01933 FBXXTE01935 FBXXTE01936 FBXXTE01937 FBXXTE01940 FBXXTE01948 FBXXTE01950 FBXXTE01951 FBXXTE01967 FBXXTE01973 FBXXTE01975 FBXXTE01976 FBXXTE01978 FBXXTE01979 FBXXTE01980 FBXXTE01981 FBXXTE01987 FBXXTE01991 FBXXTE01992 FBXXTE01993 FBXXTE01994 FBXXTE01995 FBXXTE01996 FBXXTE02003 FBXXTE02004 FBXXTE02007 FBXXTE02012 FBXXTE02013 FBXXTE02014 FBXXTE02016 FBXXTE02018 FBXXTE02020 FBXXTE02021 FBXXTE02022 FBXXTE02024 FBXXTE02025 FBXXTE02026 FBXXTE02039 FBXXTE02040 FBXXTE02045 FBXXTE02046 FBXXTE02048 FBXXTE02049 FBXXTE02053 FBXXTE02054 FBXXTE02056 FBXXTE02059 FBXXTE02061 FBXXTE02064 FBXXTE02065 FBXXTE02066 FBXXTE02067 FBXXTE02070 FBXXTE02074 FBXXTE02075 FBXXTE02085 FBXXTE02092 FBXXTE02096 FBXXTE02097 FBXXTE02104 FBXXTE02105 FBXXTE02108 FBXXTE02109 FBXXTE02113 FBXXTE02115 FBXXTE02117 FBXXTE02118 FBXXTE02119 FBXXTE02121 FBXXTE02122 FBXXTE02126 FBXXTE02133 FBXXTE02139 FBXXTE02140 FBXXTE02141 FBXXTE02142 FBXXTE02148 FBXXTE02149 FBXXTE02150 FBXXTE02151 FBXXTE02153 FBXXTE02154 FBXXTE02157 FBXXTE02162 FBXXTE02163 FBXXTE02165 FBXXTE02168 FBXXTE02179 FBXXTE02187 FBXXTE02197 FBXXTE02199 FBXXTE02200 FBXXTE02201 FBXXTE02202 FBXXTE02207 FBXXTE02208 FBXXTE02213 FBXXTE02216 FBXXTE02244 FBXXTE02250 FBXXTE02251 FBXXTE02252 FBXXTE02257 FBXXTE02262 FBXXTE02279 FBXXTE02280 FBXXTE02281 FBXXTE02295 FBXXTE02300 FBXXTE02303 FBXXTE02304 FBXXTE02311 FBXXTE02312 FBXXTE02314 FBXXTE02318 FBXXTE02320 FBXXTE02323 FBXXTE02325 FBXXTE02331 FBXXTE02332 FBXXTE02333 FBXXTE02334 FBXXTE02337 FBXXTE02338 FBXXTE02340 FBXXTE02341 FBXXTE02344 FBXXTE02345 FBXXTE02350 FBXXTE02355 FBXXTE02356 FBXXTE02357 FBXXTE02358 FBXXTE02359 FBXXTE02365 FBXXTE02374 FBXXTE02385 FBXXTE02387 FBXXTE02388 FBXXTE02389 FBXXTE02390 FBXXTE02391 FBXXTE02401 FBXXTE02414 FBXXTE02418 FBXXTE02421 FBXXTE02425 FBXXTE02426 FBXXTE02428 FBXXTE02429 FBXXTE02430 FBXXTE02431 FBXXTE02432 FBXXTE02433 FBXXTE02434 FBXXTE02435 FBXXTE02437 FBXXTE02438 FBXXTE02439 FBXXTE02442 FBXXTE02448 FBXXTE02453 FBXXTE02454 FBXXTE02455 FBXXTE02458 FBXXTE02462 FBXXTE02463 FBXXTE02464 FBXXTE02468 FBXXTE02469 FBXXTE02470 FBXXTE02471 FBXXTE02472 FBXXTE02475 FBXXTE02476 FBXXTE02477 FBXXTE02478 FBXXTE02481 FBX
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
GE Medical Systems, LLC
Reason for Recall:
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

OEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.

Product Codes/Lot Numbers:

Two populations are potentially affected by this notification. 1. Listed in Appendix A - OEC Elite Mobile C-arm Systems (GTIN: 00001234567895) and OEC 3D Mobile C-arm Systems (GTIN: 00195278096944) that potentially have an X-ray field size greater than the 4% SID limit and do not contain the acceptance limit for the X-ray field size measurement in the System Service Manual. 2. Systems NOT listed in Appendix A - OEC Elite Mobile C-arm Systems (GTIN: 00001234567895) and OEC 3D Mobile C-arm Systems (GTIN: 00195278096944) that are within the allowed SID limit. However, the acceptance limit for the X-ray field size measurement is not stated in the System Service Manual. Appendix A Serial Numbers FAHXTX00002 FAHXTX00005 FAHXTX00014 FAHXTX00018 FAHXTX00019 FAHXTX00020 FAHXTX00024 FAHXTX00028 FAHXTX00034 FAHXTX00037 FAHXTX00043 FAHXTX00052 FAHXTX00063 FAHXTX00066 FAHXTX00068 FAHXTX00070 FAHXTX00075 FAHXTX00078 FAHXTX00079 FAHXTX00080 FAHXTX00081 FAHXTX00083 FAHXTX00092 FAHXTX00093 FAHXTX00097 FAHXTX00110 FAHXTX00111 FAHXTX00112 FAHXTX00114 FAHXTX00115 FAHXTX00116 FAHXTX00117 FAHXTX00118 FAHXTX00119 FAHXTX00120 FAHXTX00121 FAHXTX00122 FAHXTX00124 FAHXTX00125 FAHXTX00135 FAHXTX00137 FAHXTX00138 FAHXTX00151 FAHXTX00152 FAHXTX00155 FAHXTX00156 FAHXTX00162 FAHXTX00163 FAHXTX00166 FAHXTX00170 FAHXTX00171 FAHXTX00177 FAHXTX00178 FAHXTX00188 FAHXTX00192 FAHXTX00193 FAHXTX00202 FAHXTX00204 FAHXTX00211 FAHXTX00212 FAHXTX00217 FAHXTX00219 FAHXTX00220 FAHXTX00221 FAHXTX00222 FAHXTX00223 FAHXTX00224 FAHXTX00226 FAHXTX00227 FAHXTX00229 FAHXTX00230 FAHXTX00236 FAHXTX00240 FAHXTX00244 FAHXTX00261 FAHXTX00265 FAHXTX00266 FAHXTX00269 FAHXTX00271 FAHXTX00272 FAHXTX00273 FAHXTX00274 FAHXTX00280 FAHXTX00283 FAHXTX00284 FAHXTX00296 FAHXTX00299 FAHXTX00300 FAHXTX00304 FAHXTX00305 FAHXTX00308 FAHXTX00312 FAHXTX00313 FAHXTX00320 FAHXTX00323 FAHXTX00325 FAHXTX00332 FAHXTX00340 FAHXTX00342 FAHXTX00343 FAHXTX00352 FAHXTX00353 FAHXTX00354 FAHXTX00357 FAHXTX00358 FAHXTX00359 FAHXTX00360 FAHXTX00363 FAHXTX00376 FAHXTX00379 FAHXTX00381 FAHXTX00382 FAHXTX00390 FAHXTX00392 FAHXTX00399 FAHXTX00400 FAHXTX00410 FAHXTX00413 FAHXTX00414 FAHXTX00415 FAHXTX00420 FAHXTX00421 FAHXTX00429 FAHXTX00439 FAHXTX00453 FAHXTX00455 FAHXTX00456 FAHXTX00463 FAHXTX00465 FAHXTX00466 FAHXTX00479 FAHXTX00481 FAHXTX00483 FAHXTX00484 FAHXTX00486 FAHXTX00488 FAHXTX00495 FAHXTX00497 FAHXTX00498 FAHXTX00507 FAHXTX00509 FAHXTX00513 FAHXTX00523 FAHXTX00530 FAHXTX00534 FAHXTX00538 FAHXTX00539 FAHXTX00540 FAHXTX00547 FAHXTX00549 FAHXTX00555 FAHXTX00557 FAHXTX00560 FAHXTX00565 FAHXTX00569 FAHXTX00570 FAHXTX00571 FAHXTX00572 FAHXTX00575 FAHXTX00576 FAHXTX00579 FAHXTX00592 FAHXTX00595 FAHXTX00610 FAHXTX00611 FAHXXX00029 FAHXXX00030 FAHXXX00038 FAHXXX00040 FAHXXX00056 FAHXXX00057 FAHXXX00059 FAHXXX00060 FAHXXX00061 FAHXXX00069 FAHXXX00082 FAHXXX00085 FAHXXX00086 FAHXXX00089 FAHXXX00091 FAHXXX00093 FAHXXX00097 FAHXXX00098 FAHXXX00099 FAHXXX00105 FAHXXX00106 FAHXXX00112 FAHXXX00113 FAHXXX00114 FAHXXX00116 FAHXXX00117 FAHXXX00119 FAHXXX00123 FAHXXX00142 FAHXXX00144 FAHXXX00145 FAHXXX00147 FAHXXX00150 FAHXXX00151 FAHXXX00153 FAHXXX00155 FAHXXX00164 FAHXXX00173 FAHXXX00176 FAHXXX00178 FAHXXX00179 FAHXXX00188 FAHXXX00190 FAHXXX00193 FAHXXX00196 FAHXXX00199 FAHXXX00200 FAHXXX00207 FAHXXX00208 FAHXXX00209 FAMHTX00001 FAMHTX00005 FAMHTX00007 FAMHTX00012 FAMHTX00014 FAMHTX00041 FAMHTX00051 FAMHTX00055 FAMHTX00064 FAMHTX00069 FAMHTX00081 FAMHTX00085 FAMHXX00003 FAMHXX00011 FAMHXX00019 FAMHXX00023 FAXXTE00003 FAXXTE00006 FAXXTE00009 FAXXTE00011 FAXXTE00013 FAXXTE00014 FAXXTE00015 FAXXTE00017 FAXXTE00021 FAXXTE00022 FAXXTE00024 FAXXTE00025 FAXXTE00028 FAXXTE00031 FAXXTE00032 FAXXTE00033 FAXXTE00034 FAXXTE00035 FAXXTE00037 FAXXTE00038 FAXXTE00039 FAXXTE00040 FAXXTE00044 FAXXTE00046 FAXXTE00047 FAXXTE00049 FAXXTE00050 FAXXTE00051 FAXXTE00052 FAXXTE00056 FAXXTE00062 FAXXTE00063 FAXXTE00064 FAXXTE00067 FAXXTE00071 FAXXTE00072 FAXXTE00073 FAXXTE00076 FAXXTE00084 FAXXTE00085 FAXXTE00091 FAXXTE00092 FAXXTE00094 FAXXTE00095 FAXXTE00099 FAXXTE00100 FAXXTE00101 FAXXTE00104 FAXXTE00105 FAXXTE00106 FAXXTE00107 FAXXTE00109 FAXXTE00110 FAXXTE00114 FAXXTE00116 FAXXTE00117 FAXXTE00120 FAXXTE00121 FAXXTE00128 FAXXTE00129 FAXXTE00134 FAXXTE00136 FAXXTE00137 FAXXTE00141 FAXXTE00144 FAXXTE00145 FAXXTE00147 FAXXTE00149 FAXXTE00163 FAXXTE00164 FAXXTE00170 FAXXTE00173 FAXXTE00175 FAXXTE00179 FAXXTE00180 FAXXTE00182 FAXXTE00184 FAXXTE00185 FAXXTE00195 FAXXTE00197 FAXXTE00205 FAXXTE00212 FAXXTE00220 FAXXTE00224 FAXXTE00225 FAXXTE00227 FAXXTE00228 FAXXTE00229 FAXXTE00230 FAXXTE00231 FAXXTE00234 FAXXTE00236 FAXXTE00237 FAXXTE00238 FAXXTE00240 FAXXTE00241 FAXXTE00243 FAXXTE00246 FAXXTE00248 FAXXTE00253 FAXXTE00256 FAXXTE00257 FAXXTE00258 FAXXTE00259 FAXXTE00263 FAXXTE00267 FAXXTE00269 FAXXTE00270 FAXXTE00271 FAXXTE00272 FAXXTE00277 FAXXTE00278 FAXXTE00279 FAXXTE00280 FAXXTE00282 FAXXTE00286 FAXXTE00291 FAXXTE00293 FAXXTE00295 FAXXTE00301 FAXXTE00317 FAXXTE00319 FAXXTE00320 FAXXTE00321 FAXXTE00323 FAXXTE00328 FAXXTE00329 FAXXTE00330 FAXXTE00331 FAXXTE00334 FAXXTE00336 FAXXTE00337 FAXXTE00338 FAXXTE00339 FAXXTE00340 FAXXTE00342 FAXXTE00344 FAXXTE00352 FAXXTE00353 FAXXTE00354 FAXXTE00357 FAXXTE00358 FAXXTE00361 FAXXTE00362 FAXXTE00363 FAXXTE00372 FAXXTE00373 FAXXTE00376 FAXXTE00379 FAXXTE00380 FAXXTE00383 FAXXTE00389 FAXXTE00392 FAXXTE00393 FAXXTE00396 FAXXTE00407 FAXXTE00413 FAXXTE00418 FAXXTE00422 FAXXTE00424 FAXXTE00425 FAXXTE00429 FAXXTE00431 FAXXTE00433 FAXXTE00434 FAXXTE00436 FAXXTE00439 FAXXTE00442 FAXXTE00443 FAXXTE00451 FAXXTE00452 FAXXTE00454 FAXXTE00455 FAXXTE00460 FAXXTE00461 FAXXTE00472 FAXXTE00475 FAXXTE00477 FAXXTE00481 FAXXTE00482 FAXXTE00485 FAXXTE00486 FAXXTE00487 FAXXTE00489 FAXXTE00490 FAXXTE00491 FAXXTE00492 FAXXTE00495 FAXXTE00499 FAXXTE00506 FAXXTE00508 FAXXTE00511 FAXXTE00515 FAXXTE00516 FAXXTE00517 FAXXTE00518 FAXXTE00519 FAXXTE00522 FAXXTE00523 FAXXTE00525 FAXXTE00526 FAXXTE00527 FAXXTE00532 FAXXTE00533 FAXXTE00536 FAXXTE00539 FAXXTE00540 FAXXTE00543 FAXXTE00544 FAXXTE00545 FAXXTE00546 FAXXTE00547 FAXXTE00549 FAXXTE00550 FAXXTE00551 FAXXTE00553 FAXXTE00555 FAXXTE00560 FAXXTE00561 FAXXTE00562 FAXXTE00566 FAXXTE00570 FAXXTE00571 FAXXTE00572 FAXXTE00573 FAXXTE00582 FAXXTE00591 FAXXTE00593 FAXXTE00594 FAXXTE00596 FAXXTE00604 FAXXTE00607 FAXXTE00608 FAXXTE00611 FAXXTE00615 FAXXTE00616 FAXXTE00617 FAXXTE00618 FAXXTE00624 FAXXTE00631 FAXXTE00638 FAXXTE00639 FAXXTE00643 FAXXTE00646 FAXXTE00650 FAXXTE00651 FAXXTE00652 FAXXTE00653 FAXXTE00657 FAXXTE00661 FAXXTE00664 FAXXTE00667 FAXXTE00669 FAXXTE00672 FAXXTE00680 FAXXTE00681 FAXXTE00684 FAXXTE00685 FAXXTE00687 FAXXTE00688 FAXXTE00701 FAXXTE00706 FAXXTE00707 FAXXTE00710 FAXXTE00712 FAXXTE00713 FAXXTE00716 FAXXTE00718 FAXXTE00720 FAXXTE00724 FAXXTE00725 FAXXTE00728 FAXXTE00731 FAXXTE00734 FAXXTE00735 FAXXTE00737 FAXXTE00738 FAXXTE00747 FAXXTE00748 FAXXTE00749 FAXXTE00752 FAXXTE00753 FAXXTE00762 FAXXTE00766 FAXXTE00772 FAXXTE00789 FAXXTE00794 FAXXTE00799 FAXXTE00801 FAXXTE00805 FAXXTE00809 FAXXTE00810 FAXXTE00811 FAXXTE00814 FAXXTE00819 FAXXTE00820 FAXXTE00829 FAXXTE00831 FAXXTE00843 FAXXTE00853 FAXXTE00854 FAXXTE00874 FAXXTE00881 FAXXTE00893 FAXXTE00894 FAXXTE00896 FAXXTE00897 FAXXTE00899 FAXXTE00902 FAXXTE00903 FAXXTE00906 FAXXTE00909 FAXXTE00914 FAXXTE00915 FAXXTE00916 FAXXTE00917 FAXXTE00918 FAXXTE00919 FAXXTE00922 FAXXTE00925 FAXXTE00930 FAXXTE00932 FAXXTE00936 FAXXTE00939 FAXXTE00940 FAXXTE00945 FAXXTE00958 FAXXTE00961 FAXXTE00964 FAXXTE00970 FAXXTE00973 FAXXTE00979 FAXXTE00980 FAXXTE00983 FAXXTE00985 FAXXTE00987 FAXXTE00990 FAXXTE00993 FAXXTE00995 FAXXTE00996 FAXXTE00999 FAXXTE01006 FAXXTE01029 FAXXTE01030 FAXXTE01048 FAXXTE01053 FAXXTE01057 FAXXTE01064 FAXXTE01068 FAXXTE01084 FAXXTE01085 FAXXTE01086 FAXXTE01088 FAXXTE01089 FAXXTE01092 FAXXTE01114 FAXXTE01124 FAXXTE01139 FAXXTE01140 FAXXTE01141 FAXXTE01145 FAXXTE01154 FAXXTE01156 FAXXTE01159 FAXXTE01163 FAXXTE01166 FAXXTE01167 FAXXTE01168 FAXXTE01169 FAXXTE01171 FAXXTE01180 FAXXTE01183 FAXXTE01191 FAXXTE01193 FAXXTE01194 FAXXTE01195 FAXXTE01204 FAXXTE01205 FAXXTE01206 FAXXTE01210 FAXXTE01212 FAXXTE01219 FAXXTE01222 FAXXTP00003 FAXXTP00004 FAXXTP00007 FAXXTP00009 FAXXTP00010 FAXXTP00011 FAXXTP00012 FAXXTP00014 FAXXTP00018 FAXXTP00019 FAXXTP00020 FAXXXE00003 FAXXXE00004 FAXXXE00005 FAXXXE00007 FAXXXE00011 FAXXXE00012 FAXXXE00013 FAXXXE00017 FAXXXE00020 FAXXXE00021 FAXXXE00023 FAXXXE00024 FAXXXE00025 FAXXXE00026 FAXXXE00030 FAXXXE00032 FAXXXE00033 FAXXXE00036 FAXXXE00037 FAXXXE00038 FAXXXE00039 FAXXXE00043 FAXXXE00044 FAXXXE00045 FAXXXE00046 FAXXXE00047 FAXXXE00049 FAXXXE00053 FAXXXE00054 FAXXXE00056 FAXXXE00062 FAXXXE00071 FAXXXE00072 FAXXXE00076 FAXXXE00081 FAXXXE00082 FAXXXE00083 FAXXXE00085 FAXXXE00087 FAXXXE00091 FAXXXE00092 FAXXXE00093 FAXXXE00095 FAXXXE00096 FAXXXE00097 FAXXXE00098 FAXXXE00100 FAXXXE00105 FAXXXE00107 FAXXXE00109 FAXXXE00117 FAXXXE00120 FAXXXE00121 FAXXXE00124 FAXXXE00141 FAXXXE00147 FAXXXE00152 FAXXXE00155 FAXXXE00159 FAXXXE00164 FAXXXE00165 FAXXXE00167 FAXXXE00172 FAXXXE00174 FAXXXE00176 FAXXXE00180 FAXXXE00181 FAXXXE00183 FAXXXE00184 FAXXXE00185 FAXXXE00188 FAXXXE00196 FAXXXE00197 FAXXXE00198 FAXXXE00199 FAXXXE00200 FAXXXE00204 FAXXXE00206 FAXXXE00208 FAXXXE00210 FAXXXE00215 FAXXXE00217 FAXXXE00218 FAXXXE00222 FAXXXE00225 FAXXXE00227 FAXXXE00228 FAXXXE00229 FAXXXE00230 FAXXXE00231 FAXXXE00232 FAXXXE00235 FAXXXE00236 FAXXXE00238 FAXXXE00240 FAXXXE00244 FAXXXE00247 FAXXXE00251 FAXXXE00253 FAXXXE00254 FAXXXE00255 FAXXXE00256 FAXXXE00257 FAXXXE00261 FAXXXE00269 FAXXXE00270 FAXXXE00273 FAXXXE00275 FAXXXE00278 FAXXXE00279 FAXXXE00280 FAXXXE00282 FAXXXE00283 FAXXXE00287 FAXXXE00288 FAXXXE00290 FAXXXE00291 FAXXXE00292 FAXXXE00296 FAXXXE00297 FAXXXE00302 FAXXXE00303 FAXXXE00306 FAXXXE00312 FAXXXE00314 FAXXXE00315 FAXXXE00318 FAXXXE00319 FAXXXE00320 FAXXXE00321 FAXXXE00324 FAXXXE00325 FAXXXE00327 FAXXXE00328 FAXXXE00329 FAXXXE00330 FAXXXE00332 FAXXXE00335 FAXXXE00336 FAXXXE00339 FAXXXE00340 FAXXXE00343 FAXXXE00345 FAXXXE00346 FAXXXE00348 FAXXXE00354 FAXXXE00356 FAXXXE00359 FAXXXX00014 FAXXXX00020 FBHVXX00037 FBHVXX00040 FBHXTX00002 FBHXTX00006 FBHXTX00007 FBHXTX00009 FBHXTX00010 FBHXTX00011 FBHXTX00013 FBHXTX00019 FBHXTX00020 FBHXTX00021 FBHXTX00022 FBHXTX00023 FBHXTX00024 FBHXTX00025 FBHXTX00026 FBHXTX00031 FBHXTX00032 FBHXTX00033 FBHXTX00035 FBHXTX00038 FBHXTX00045 FBHXTX00047 FBHXTX00048 FBHXTX00051 FBHXTX00057 FBHXTX00059 FBHXTX00062 FBHXTX00068 FBHXTX00070 FBHXTX00071 FBHXTX00074 FBHXTX00079 FBHXTX00080 FBHXTX00081 FBHXTX00095 FBHXTX00098 FBHXTX00102 FBHXTX00103 FBHXTX00104 FBHXTX00106 FBHXTX00114 FBHXTX00115 FBHXTX00125 FBHXTX00126 FBHXTX00128 FBHXTX00130 FBHXTX00137 FBHXTX00144 FBHXTX00149 FBHXTX00151 FBHXTX00154 FBHXTX00159 FBHXTX00161 FBHXTX00163 FBHXTX00172 FBHXTX00173 FBHXTX00177 FBHXTX00184 FBHXTX00198 FBHXTX00199 FBHXTX00201 FBHXTX00203 FBHXTX00204 FBHXTX00205 FBHXTX00208 FBHXTX00210 FBHXTX00218 FBHXTX00223 FBHXTX00231 FBHXTX00233 FBHXTX00234 FBHXTX00235 FBHXTX00237 FBHXTX00241 FBHXTX00261 FBHXTX00266 FBHXTX00274 FBHXTX00275 FBHXTX00276 FBHXTX00277 FBHXTX00278 FBHXTX00279 FBHXTX00281 FBHXTX00282 FBHXTX00287 FBHXTX00290 FBHXTX00295 FBHXTX00316 FBHXTX00317 FBHXTX00319 FBHXTX00320 FBHXTX00324 FBHXTX00330 FBHXTX00335 FBHXTX00336 FBHXTX00337 FBHXTX00338 FBHXTX00339 FBHXTX00343 FBHXTX00344 FBHXTX00345 FBHXTX00348 FBHXTX00349 FBHXTX00350 FBHXTX00351 FBHXTX00352 FBHXTX00353 FBHXTX00354 FBHXTX00355 FBHXTX00356 FBHXTX00357 FBHXTX00360 FBHXTX00361 FBHXTX00362 FBHXTX00365 FBHXTX00366 FBHXTX00368 FBHXTX00369 FBHXTX00371 FBHXTX00375 FBHXTX00379 FBHXTX00381 FBHXTX00411 FBHXTX00415 FBHXTX00416 FBHXTX00417 FBHXTX00429 FBHXTX00431 FBHXTX00432 FBHXTX00433 FBHXTX00436 FBHXTX00438 FBHXTX00442 FBHXTX00444 FBHXTX00447 FBHXTX00448 FBHXTX00451 FBHXTX00458 FBHXTX00462 FBHXTX00468 FBHXTX00479 FBHXTX00482 FBHXTX00490 FBHXTX00491 FBHXTX00493 FBHXTX00498 FBHXTX00499 FBHXTX00500 FBHXTX00501 FBHXTX00503 FBHXTX00504 FBHXTX00505 FBHXTX00507 FBHXTX00508 FBHXTX00509 FBHXTX00510 FBHXTX00514 FBHXTX00516 FBHXTX00520 FBHXTX00524 FBHXTX00529 FBHXTX00535 FBHXTX00537 FBHXTX00541 FBHXTX00543 FBHXTX00546 FBHXTX00549 FBHXTX00551 FBHXTX00555 FBHXTX00558 FBHXTX00561 FBHXTX00565 FBHXTX00568 FBHXTX00570 FBHXTX00571 FBHXTX00580 FBHXTX00583 FBHXTX00584 FBHXTX00586 FBHXTX00587 FBHXTX00589 FBHXTX00593 FBHXTX00597 FBHXTX00599 FBHXTX00600 FBHXTX00601 FBHXTX00605 FBHXTX00606 FBHXTX00613 FBHXTX00614 FBHXTX00616 FBHXTX00617 FBHXTX00624 FBHXTX00630 FBHXTX00641 FBHXTX00642 FBHXTX00644 FBHXTX00649 FBHXTX00650 FBHXTX00651 FBHXTX00655 FBHXTX00664 FBHXTX00665 FBHXTX00667 FBHXTX00669 FBHXTX00671 FBHXTX00672 FBHXTX00673 FBHXTX00679 FBHXTX00680 FBHXTX00682 FBHXTX00684 FBHXTX00686 FBHXTX00687 FBHXTX00689 FBHXTX00690 FBHXTX00691 FBHXTX00693 FBHXTX00698 FBHXTX00700 FBHXTX00703 FBHXTX00713 FBHXTX00719 FBHXTX00721 FBHXTX00727 FBHXTX00737 FBHXTX00745 FBHXTX00748 FBHXTX00753 FBHXTX00755 FBHXTX00757 FBHXTX00758 FBHXTX00760 FBHXTX00762 FBHXTX00764 FBHXTX00769 FBHXTX00770 FBHXTX00771 FBHXTX00774 FBHXTX00783 FBHXTX00785 FBHXTX00798 FBHXTX00800 FBHXTX00803 FBHXTX00804 FBHXTX00821 FBHXTX00823 FBHXTX00826 FBHXTX00834 FBHXTX00837 FBHXTX00839 FBHXTX00841 FBHXTX00843 FBHXTX00844 FBHXTX00847 FBHXTX00848 FBHXTX00849 FBHXTX00866 FBHXTX00867 FBHXTX00894 FBHXTX00896 FBHXTX00897 FBHXTX00898 FBHXTX00899 FBHXTX00901 FBHXTX00909 FBHXTX00912 FBHXTX00913 FBHXTX00914 FBHXTX00916 FBHXTX00920 FBHXTX00921 FBHXTX00922 FBHXTX00924 FBHXTX00925 FBHXTX00926 FBHXTX00930 FBHXTX00933 FBHXTX00934 FBHXTX00936 FBHXTX00937 FBHXTX00938 FBHXTX00942 FBHXTX00943 FBHXTX00946 FBHXTX00952 FBHXTX00954 FBHXTX00955 FBHXTX00964 FBHXTX00971 FBHXTX00972 FBHXTX00974 FBHXTX00977 FBHXTX00980 FBHXTX00982 FBHXTX00986 FBHXTX00991 FBHXTX00995 FBHXTX00996 FBHXTX01002 FBHXTX01006 FBHXTX01013 FBHXTX01014 FBHXTX01016 FBHXTX01031 FBHXTX01040 FBHXTX01045 FBHXTX01048 FBHXTX01049 FBHXTX01050 FBHXTX01062 FBHXTX01063 FBHXTX01070 FBHXTX01074 FBHXTX01075 FBHXTX01076 FBHXTX01077 FBHXTX01078 FBHXTX01079 FBHXTX01080 FBHXTX01100 FBHXTX01101 FBHXTX01102 FBHXTX01104 FBHXTX01108 FBHXTX01109 FBHXTX01110 FBHXTX01111 FBHXTX01119 FBHXTX01123 FBHXTX01126 FBHXTX01127 FBHXTX01130 FBHXTX01131 FBHXTX01132 FBHXTX01135 FBHXTX01138 FBHXTX01139 FBHXTX01140 FBHXTX01145 FBHXTX01147 FBHXTX01150 FBHXTX01153 FBHXTX01154 FBHXTX01156 FBHXTX01157 FBHXTX01158 FBHXTX01165 FBHXTX01170 FBHXTX01176 FBHXTX01180 FBHXTX01184 FBHXTX01189 FBHXTX01193 FBHXTX01194 FBHXTX01200 FBHXTX01201 FBHXTX01202 FBHXTX01217 FBHXTX01222 FBHXTX01223 FBHXTX01228 FBHXTX01229 FBHXTX01230 FBHXTX01231 FBHXTX01232 FBHXTX01233 FBHXTX01234 FBHXTX01239 FBHXTX01240 FBHXTX01241 FBHXTX01244 FBHXTX01246 FBHXTX01251 FBHXTX01252 FBHXTX01264 FBHXTX01265 FBHXTX01267 FBHXTX01268 FBHXTX01269 FBHXTX01270 FBHXTX01271 FBHXTX01273 FBHXTX01286 FBHXTX01287 FBHXTX01289 FBHXTX01291 FBHXTX01303 FBHXTX01306 FBHXTX01311 FBHXTX01319 FBHXTX01322 FBHXTX01323 FBHXTX01326 FBHXTX01330 FBHXTX01331 FBHXTX01348 FBHXTX01356 FBHXTX01368 FBHXTX01371 FBHXTX01373 FBHXTX01374 FBHXTX01381 FBHXTX01382 FBHXTX01387 FBHXTX01388 FBHXTX01389 FBHXTX01397 FBHXTX01400 FBHXTX01405 FBHXTX01410 FBHXTX01421 FBHXXX00016 FBHXXX00027 FBHXXX00032 FBHXXX00036 FBHXXX00041 FBHXXX00042 FBHXXX00045 FBHXXX00050 FBHXXX00054 FBHXXX00056 FBHXXX00060 FBHXXX00061 FBHXXX00063 FBHXXX00065 FBHXXX00067 FBHXXX00069 FBHXXX00074 FBHXXX00077 FBHXXX00078 FBHXXX00081 FBHXXX00085 FBHXXX00088 FBHXXX00091 FBHXXX00092 FBHXXX00093 FBHXXX00094 FBHXXX00097 FBHXXX00099 FBHXXX00100 FBHXXX00101 FBHXXX00102 FBHXXX00104 FBHXXX00105 FBHXXX00106 FBHXXX00107 FBHXXX00108 FBHXXX00109 FBHXXX00110 FBHXXX00111 FBHXXX00113 FBHXXX00117 FBHXXX00118 FBHXXX00119 FBHXXX00120 FBHXXX00121 FBHXXX00122 FBHXXX00126 FBHXXX00127 FBHXXX00132 FBHXXX00133 FBHXXX00134 FBHXXX00135 FBHXXX00143 FBHXXX00148 FBHXXX00156 FBHXXX00158 FBHXXX00159 FBHXXX00160 FBHXXX00166 FBHXXX00169 FBHXXX00170 FBHXXX00171 FBHXXX00179 FBHXXX00180 FBHXXX00182 FBHXXX00185 FBHXXX00186 FBHXXX00187 FBHXXX00189 FBHXXX00192 FBHXXX00196 FBHXXX00202 FBHXXX00204 FBHXXX00205 FBHXXX00206 FBHXXX00207 FBHXXX00208 FBHXXX00211 FBHXXX00213 FBHXXX00218 FBHXXX00225 FBHXXX00231 FBHXXX00238 FBHXXX00239 FBHXXX00240 FBHXXX00243 FBHXXX00246 FBHXXX00247 FBHXXX00248 FBHXXX00249 FBHXXX00251 FBHXXX00256 FBHXXX00257 FBHXXX00265 FBHXXX00266 FBHXXX00269 FBHXXX00270 FBHXXX00273 FBHXXX00275 FBHXXX00277 FBHXXX00280 FBHXXX00281 FBHXXX00285 FBHXXX00286 FBHXXX00288 FBHXXX00290 FBHXXX00291 FBHXXX00292 FBHXXX00293 FBHXXX00298 FBHXXX00300 FBHXXX00301 FBHXXX00303 FBHXXX00304 FBHXXX00305 FBHXXX00309 FBHXXX00311 FBHXXX00313 FBHXXX00319 FBHXXX00320 FBHXXX00321 FBHXXX00322 FBHXXX00326 FBHXXX00327 FBHXXX00330 FBHXXX00331 FBHXXX00332 FBHXXX00333 FBHXXX00338 FBHXXX00339 FBHXXX00342 FBHXXX00352 FBHXXX00353 FBHXXX00356 FBHXXX00361 FBHXXX00362 FBHXXX00365 FBHXXX00367 FBHXXX00368 FBHXXX00371 FBHXXX00372 FBHXXX00374 FBHXXX00389 FBHXXX00390 FBHXXX00391 FBHXXX00393 FBHXXX00394 FBHXXX00395 FBHXXX00401 FBHXXX00402 FBMHDX00013 FBMHDX00015 FBMHDX00018 FBMHDX00019 FBMHDX00020 FBMHDX00021 FBMHDX00024 FBMHDX00025 FBMHDX00026 FBMHDX00028 FBMHDX00029 FBMHDX00030 FBMHDX00043 FBMHDX00044 FBMHDX00046 FBMHDX00047 FBMHDX00049 FBMHDX00054 FBMHDX00056 FBMHDX00058 FBMHDX00059 FBMHDX00061 FBMHDX00065 FBMHDX00066 FBMHDX00067 FBMHDX00071 FBMHDX00073 FBMHDX00078 FBMHDX00080 FBMHDX00081 FBMHDX00082 FBMHDX00083 FBMHDX00085 FBMHDX00086 FBMHDX00087 FBMHDX00089 FBMHDX00090 FBMHDX00092 FBMHDX00095 FBMHDX00097 FBMHDX00098 FBMHDX00100 FBMHDX00101 FBMHDX00104 FBMHDX00107 FBMHDX00108 FBMHDX00110 FBMHDX00111 FBMHDX00112 FBMHDX00118 FBMHDX00119 FBMHDX00120 FBMHDX00122 FBMHDX00124 FBMHDX00125 FBMHDX00126 FBMHDX00127 FBMHDX00129 FBMHDX00130 FBMHDX00133 FBMHDX00134 FBMHDX00136 FBMHDX00137 FBMHDX00139 FBMHDX00140 FBMHDX00141 FBMHDX00143 FBMHDX00144 FBMHDX00147 FBMHDX00151 FBMHDX00152 FBMHDX00153 FBMHDX00155 FBMHDX00157 FBMHDX00163 FBMHDX00166 FBMHDX00168 FBMHDX00169 FBMHDX00170 FBMHDX00173 FBMHDX00177 FBMHDX00180 FBMHDX00182 FBMHDX00184 FBMHDX00186 FBMHDX00187 FBMHDX00188 FBMHDX00189 FBMHDX00193 FBMHDX00195 FBMHDX00196 FBMHDX00197 FBMHDX00198 FBMHDX00200 FBMHDX00201 FBMHDX00203 FBMHDX00208 FBMHDX00210 FBMHDX00212 FBMHDX00216 FBMHDX00218 FBMHDX00220 FBMHDX00221 FBMHDX00224 FBMHDX00226 FBMHDX00227 FBMHDX00231 FBMHDX00237 FBMHDX00238 FBMHDX00239 FBMHDX00240 FBMHDX00245 FBMHDX00250 FBMHDX00252 FBMHDX00253 FBMHDX00255 FBMHDX00262 FBMHDX00264 FBMHDX00266 FBMHDX00271 FBMHDX00273 FBMHDX00282 FBMHDX00284 FBMHDX00286 FBMHDX00287 FBMHDX00290 FBMHDX00292 FBMHDX00294 FBMHDX00296 FBMHDX00297 FBMHDX00301 FBMHDX00302 FBMHDX00310 FBMHDX00313 FBMHDX00314 FBMHDX00318 FBMHDX00321 FBMHDX00326 FBMHDX00328 FBMHDX00333 FBMHDX00335 FBMHDX00346 FBMHDX00350 FBMHDX00352 FBMHDX00363 FBMHDX00366 FBMHDX00369 FBMHDX00374 FBMHDX00375 FBMHDX00379 FBMHDX00386 FBMHDX00387 FBMHDX00388 FBMHDX00389 FBMHDX00395 FBMHDX00397 FBMHDX00398 FBMHDX00399 FBMHDX00400 FBMHDX00401 FBMHDX00403 FBMHDX00406 FBMHDX00409 FBMHDX00410 FBMHDX00412 FBMHDX00413 FBMHDX00416 FBMHDX00423 FBMHDX00424 FBMHDX00425 FBMHDX00430 FBMHDX00431 FBMHDX00437 FBMHDX00439 FBMHDX00440 FBMHDX00441 FBMHDX00445 FBMHDX00449 FBMHDX00451 FBMHDX00452 FBMHDX00453 FBMHDX00458 FBMHDX00459 FBMHDX00460 FBMHDX00461 FBMHDX00462 FBMHDX00463 FBMHDX00464 FBMHDX00466 FBMHDX00468 FBMHDX00471 FBMHDX00477 FBMHDX00480 FBMHDX00487 FBMHDX00490 FBMHDX00491 FBMHDX00500 FBMHDX00509 FBMHDX00510 FBMHDX00516 FBMHDX00518 FBMHDX00521 FBMHDX00522 FBMHDX00523 FBMHDX00524 FBMHDX00525 FBMHDX00527 FBMHDX00528 FBMHDX00531 FBMHDX00532 FBMHDX00534 FBMHDX00542 FBMHDX00543 FBMHDX00544 FBMHDX00552 FBMHDX00554 FBMHTX00006 FBMHTX00007 FBMHTX00008 FBMHTX00017 FBMHTX00019 FBMHTX00020 FBMHTX00021 FBMHTX00024 FBMHTX00028 FBMHTX00030 FBMHTX00033 FBMHTX00041 FBMHTX00044 FBMHTX00046 FBMHTX00052 FBMHTX00053 FBMHTX00054 FBMHTX00057 FBMHTX00059 FBMHTX00060 FBMHTX00065 FBMHTX00066 FBMHTX00068 FBMHTX00070 FBMHTX00074 FBMHTX00077 FBMHTX00078 FBMHTX00081 FBMHTX00084 FBMHTX00088 FBMHTX00091 FBMHTX00094 FBMHTX00104 FBMHTX00106 FBMHTX00107 FBMHTX00113 FBMHTX00115 FBMHTX00116 FBMHTX00117 FBMHTX00118 FBMHTX00122 FBMHTX00126 FBMHTX00127 FBMHTX00129 FBMHTX00130 FBMHTX00133 FBMHTX00139 FBMHTX00140 FBMHTX00141 FBMHTX00143 FBMHTX00147 FBMHTX00151 FBMHTX00155 FBMHTX00161 FBMHTX00162 FBMHTX00163 FBMHTX00164 FBMHTX00166 FBMHTX00167 FBMHTX00169 FBMHTX00170 FBMHTX00171 FBMHTX00172 FBMHTX00173 FBMHTX00179 FBMHTX00180 FBMHTX00185 FBMHTX00186 FBMHTX00187 FBMHTX00188 FBMHTX00191 FBMHTX00195 FBMHTX00200 FBMHTX00203 FBMHTX00205 FBMHTX00210 FBMHTX00220 FBMHTX00221 FBMHTX00222 FBMHTX00223 FBMHTX00225 FBMHTX00230 FBMHTX00238 FBMHTX00245 FBMHTX00246 FBMHTX00250 FBMHTX00252 FBMHTX00255 FBMHTX00259 FBMHTX00275 FBMHTX00276 FBMHTX00278 FBMHTX00279 FBMHTX00280 FBMHTX00282 FBMHTX00283 FBMHTX00285 FBMHTX00290 FBMHTX00292 FBMHTX00293 FBMHTX00298 FBMHTX00306 FBMHTX00309 FBMHTX00310 FBMHTX00313 FBMHTX00329 FBMHTX00330 FBMHTX00333 FBMHTX00343 FBMHTX00344 FBMHTX00347 FBMHTX00353 FBMHTX00354 FBMHTX00358 FBMHTX00363 FBMHTX00369 FBMHTX00372 FBMHTX00383 FBMHTX00389 FBMHTX00392 FBMHTX00393 FBMHTX00399 FBMHTX00404 FBMHTX00405 FBMHTX00407 FBMHTX00408 FBMHTX00409 FBMHTX00410 FBMHTX00432 FBMHTX00433 FBMHTX00436 FBMHTX00437 FBMHTX00444 FBMHTX00446 FBMHTX00462 FBMHTX00467 FBMHTX00471 FBMHTX00473 FBMHTX00477 FBMHTX00488 FBMHTX00492 FBMHXX00011 FBMHXX00016 FBMHXX00017 FBMHXX00019 FBMHXX00024 FBMHXX00026 FBMHXX00032 FBMHXX00034 FBMHXX00037 FBMHXX00038 FBMHXX00043 FBMHXX00047 FBMHXX00049 FBMHXX00051 FBMHXX00052 FBMHXX00055 FBMHXX00057 FBMHXX00060 FBMHXX00065 FBMHXX00067 FBMHXX00072 FBMHXX00073 FBMHXX00081 FBMHXX00082 FBMHXX00083 FBMHXX00085 FBMHXX00086 FBMHXX00087 FBMHXX00089 FBMHXX00096 FBMHXX00099 FBMHXX00101 FBMHXX00102 FBMHXX00103 FBMHXX00104 FBXXTE00005 FBXXTE00010 FBXXTE00012 FBXXTE00013 FBXXTE00015 FBXXTE00016 FBXXTE00018 FBXXTE00025 FBXXTE00028 FBXXTE00029 FBXXTE00030 FBXXTE00032 FBXXTE00033 FBXXTE00035 FBXXTE00038 FBXXTE00039 FBXXTE00040 FBXXTE00043 FBXXTE00044 FBXXTE00047 FBXXTE00050 FBXXTE00051 FBXXTE00053 FBXXTE00054 FBXXTE00055 FBXXTE00056 FBXXTE00057 FBXXTE00064 FBXXTE00067 FBXXTE00075 FBXXTE00079 FBXXTE00082 FBXXTE00085 FBXXTE00088 FBXXTE00091 FBXXTE00092 FBXXTE00093 FBXXTE00098 FBXXTE00101 FBXXTE00102 FBXXTE00104 FBXXTE00105 FBXXTE00108 FBXXTE00110 FBXXTE00113 FBXXTE00114 FBXXTE00115 FBXXTE00117 FBXXTE00119 FBXXTE00120 FBXXTE00122 FBXXTE00123 FBXXTE00124 FBXXTE00125 FBXXTE00126 FBXXTE00129 FBXXTE00130 FBXXTE00131 FBXXTE00134 FBXXTE00136 FBXXTE00137 FBXXTE00138 FBXXTE00139 FBXXTE00148 FBXXTE00155 FBXXTE00158 FBXXTE00159 FBXXTE00160 FBXXTE00161 FBXXTE00163 FBXXTE00164 FBXXTE00165 FBXXTE00167 FBXXTE00172 FBXXTE00173 FBXXTE00175 FBXXTE00177 FBXXTE00180 FBXXTE00181 FBXXTE00182 FBXXTE00186 FBXXTE00187 FBXXTE00190 FBXXTE00196 FBXXTE00199 FBXXTE00203 FBXXTE00204 FBXXTE00206 FBXXTE00208 FBXXTE00209 FBXXTE00216 FBXXTE00217 FBXXTE00222 FBXXTE00223 FBXXTE00228 FBXXTE00229 FBXXTE00234 FBXXTE00237 FBXXTE00238 FBXXTE00249 FBXXTE00251 FBXXTE00252 FBXXTE00254 FBXXTE00256 FBXXTE00262 FBXXTE00263 FBXXTE00266 FBXXTE00267 FBXXTE00272 FBXXTE00279 FBXXTE00282 FBXXTE00283 FBXXTE00285 FBXXTE00288 FBXXTE00294 FBXXTE00300 FBXXTE00306 FBXXTE00316 FBXXTE00317 FBXXTE00318 FBXXTE00319 FBXXTE00321 FBXXTE00325 FBXXTE00326 FBXXTE00328 FBXXTE00329 FBXXTE00332 FBXXTE00339 FBXXTE00342 FBXXTE00346 FBXXTE00347 FBXXTE00353 FBXXTE00356 FBXXTE00357 FBXXTE00358 FBXXTE00359 FBXXTE00360 FBXXTE00384 FBXXTE00386 FBXXTE00389 FBXXTE00390 FBXXTE00391 FBXXTE00393 FBXXTE00401 FBXXTE00403 FBXXTE00406 FBXXTE00407 FBXXTE00409 FBXXTE00415 FBXXTE00420 FBXXTE00422 FBXXTE00430 FBXXTE00431 FBXXTE00432 FBXXTE00433 FBXXTE00434 FBXXTE00436 FBXXTE00443 FBXXTE00450 FBXXTE00454 FBXXTE00467 FBXXTE00480 FBXXTE00485 FBXXTE00496 FBXXTE00497 FBXXTE00502 FBXXTE00504 FBXXTE00506 FBXXTE00507 FBXXTE00509 FBXXTE00512 FBXXTE00513 FBXXTE00514 FBXXTE00515 FBXXTE00519 FBXXTE00520 FBXXTE00525 FBXXTE00528 FBXXTE00533 FBXXTE00534 FBXXTE00535 FBXXTE00537 FBXXTE00539 FBXXTE00543 FBXXTE00550 FBXXTE00553 FBXXTE00555 FBXXTE00557 FBXXTE00558 FBXXTE00559 FBXXTE00560 FBXXTE00561 FBXXTE00562 FBXXTE00563 FBXXTE00564 FBXXTE00565 FBXXTE00566 FBXXTE00567 FBXXTE00569 FBXXTE00571 FBXXTE00572 FBXXTE00575 FBXXTE00576 FBXXTE00577 FBXXTE00578 FBXXTE00579 FBXXTE00581 FBXXTE00582 FBXXTE00584 FBXXTE00585 FBXXTE00586 FBXXTE00587 FBXXTE00588 FBXXTE00589 FBXXTE00593 FBXXTE00595 FBXXTE00596 FBXXTE00597 FBXXTE00600 FBXXTE00602 FBXXTE00603 FBXXTE00613 FBXXTE00614 FBXXTE00617 FBXXTE00623 FBXXTE00625 FBXXTE00630 FBXXTE00631 FBXXTE00636 FBXXTE00638 FBXXTE00639 FBXXTE00640 FBXXTE00641 FBXXTE00642 FBXXTE00643 FBXXTE00644 FBXXTE00645 FBXXTE00646 FBXXTE00652 FBXXTE00654 FBXXTE00657 FBXXTE00659 FBXXTE00662 FBXXTE00663 FBXXTE00665 FBXXTE00668 FBXXTE00669 FBXXTE00681 FBXXTE00682 FBXXTE00684 FBXXTE00685 FBXXTE00692 FBXXTE00694 FBXXTE00700 FBXXTE00701 FBXXTE00702 FBXXTE00704 FBXXTE00708 FBXXTE00709 FBXXTE00712 FBXXTE00713 FBXXTE00714 FBXXTE00717 FBXXTE00720 FBXXTE00723 FBXXTE00725 FBXXTE00728 FBXXTE00735 FBXXTE00736 FBXXTE00737 FBXXTE00740 FBXXTE00742 FBXXTE00744 FBXXTE00745 FBXXTE00746 FBXXTE00748 FBXXTE00749 FBXXTE00755 FBXXTE00769 FBXXTE00771 FBXXTE00772 FBXXTE00774 FBXXTE00775 FBXXTE00776 FBXXTE00777 FBXXTE00779 FBXXTE00783 FBXXTE00787 FBXXTE00790 FBXXTE00792 FBXXTE00801 FBXXTE00802 FBXXTE00805 FBXXTE00807 FBXXTE00811 FBXXTE00812 FBXXTE00813 FBXXTE00814 FBXXTE00815 FBXXTE00825 FBXXTE00828 FBXXTE00829 FBXXTE00831 FBXXTE00832 FBXXTE00833 FBXXTE00834 FBXXTE00837 FBXXTE00839 FBXXTE00841 FBXXTE00842 FBXXTE00845 FBXXTE00846 FBXXTE00851 FBXXTE00852 FBXXTE00856 FBXXTE00859 FBXXTE00860 FBXXTE00861 FBXXTE00862 FBXXTE00863 FBXXTE00864 FBXXTE00865 FBXXTE00867 FBXXTE00868 FBXXTE00877 FBXXTE00878 FBXXTE00880 FBXXTE00883 FBXXTE00886 FBXXTE00887 FBXXTE00889 FBXXTE00895 FBXXTE00896 FBXXTE00897 FBXXTE00900 FBXXTE00904 FBXXTE00915 FBXXTE00917 FBXXTE00919 FBXXTE00925 FBXXTE00926 FBXXTE00929 FBXXTE00936 FBXXTE00942 FBXXTE00952 FBXXTE00953 FBXXTE00954 FBXXTE00957 FBXXTE00958 FBXXTE00960 FBXXTE00962 FBXXTE00963 FBXXTE00975 FBXXTE00979 FBXXTE00980 FBXXTE00981 FBXXTE00985 FBXXTE00994 FBXXTE00999 FBXXTE01003 FBXXTE01004 FBXXTE01005 FBXXTE01013 FBXXTE01014 FBXXTE01015 FBXXTE01018 FBXXTE01020 FBXXTE01027 FBXXTE01030 FBXXTE01033 FBXXTE01034 FBXXTE01035 FBXXTE01037 FBXXTE01038 FBXXTE01039 FBXXTE01041 FBXXTE01047 FBXXTE01048 FBXXTE01049 FBXXTE01052 FBXXTE01054 FBXXTE01055 FBXXTE01058 FBXXTE01059 FBXXTE01060 FBXXTE01061 FBXXTE01065 FBXXTE01067 FBXXTE01068 FBXXTE01070 FBXXTE01073 FBXXTE01074 FBXXTE01075 FBXXTE01077 FBXXTE01080 FBXXTE01083 FBXXTE01084 FBXXTE01085 FBXXTE01088 FBXXTE01089 FBXXTE01093 FBXXTE01096 FBXXTE01097 FBXXTE01098 FBXXTE01100 FBXXTE01102 FBXXTE01103 FBXXTE01104 FBXXTE01106 FBXXTE01108 FBXXTE01109 FBXXTE01112 FBXXTE01113 FBXXTE01117 FBXXTE01119 FBXXTE01121 FBXXTE01123 FBXXTE01126 FBXXTE01129 FBXXTE01130 FBXXTE01131 FBXXTE01132 FBXXTE01143 FBXXTE01146 FBXXTE01148 FBXXTE01153 FBXXTE01154 FBXXTE01160 FBXXTE01165 FBXXTE01166 FBXXTE01167 FBXXTE01168 FBXXTE01169 FBXXTE01173 FBXXTE01177 FBXXTE01181 FBXXTE01183 FBXXTE01184 FBXXTE01185 FBXXTE01186 FBXXTE01190 FBXXTE01193 FBXXTE01194 FBXXTE01197 FBXXTE01198 FBXXTE01200 FBXXTE01201 FBXXTE01203 FBXXTE01207 FBXXTE01209 FBXXTE01210 FBXXTE01211 FBXXTE01216 FBXXTE01218 FBXXTE01222 FBXXTE01224 FBXXTE01226 FBXXTE01227 FBXXTE01228 FBXXTE01230 FBXXTE01231 FBXXTE01232 FBXXTE01233 FBXXTE01235 FBXXTE01238 FBXXTE01241 FBXXTE01249 FBXXTE01253 FBXXTE01257 FBXXTE01260 FBXXTE01262 FBXXTE01263 FBXXTE01267 FBXXTE01268 FBXXTE01272 FBXXTE01273 FBXXTE01274 FBXXTE01275 FBXXTE01276 FBXXTE01277 FBXXTE01278 FBXXTE01281 FBXXTE01282 FBXXTE01286 FBXXTE01287 FBXXTE01288 FBXXTE01289 FBXXTE01290 FBXXTE01291 FBXXTE01292 FBXXTE01293 FBXXTE01294 FBXXTE01296 FBXXTE01298 FBXXTE01301 FBXXTE01305 FBXXTE01306 FBXXTE01307 FBXXTE01310 FBXXTE01312 FBXXTE01313 FBXXTE01314 FBXXTE01315 FBXXTE01317 FBXXTE01320 FBXXTE01325 FBXXTE01326 FBXXTE01331 FBXXTE01333 FBXXTE01337 FBXXTE01343 FBXXTE01344 FBXXTE01349 FBXXTE01353 FBXXTE01355 FBXXTE01356 FBXXTE01363 FBXXTE01371 FBXXTE01377 FBXXTE01382 FBXXTE01384 FBXXTE01385 FBXXTE01387 FBXXTE01397 FBXXTE01399 FBXXTE01404 FBXXTE01407 FBXXTE01410 FBXXTE01411 FBXXTE01412 FBXXTE01414 FBXXTE01416 FBXXTE01417 FBXXTE01418 FBXXTE01419 FBXXTE01421 FBXXTE01422 FBXXTE01430 FBXXTE01432 FBXXTE01436 FBXXTE01437 FBXXTE01438 FBXXTE01441 FBXXTE01444 FBXXTE01446 FBXXTE01448 FBXXTE01449 FBXXTE01451 FBXXTE01452 FBXXTE01453 FBXXTE01454 FBXXTE01457 FBXXTE01458 FBXXTE01460 FBXXTE01461 FBXXTE01462 FBXXTE01463 FBXXTE01465 FBXXTE01466 FBXXTE01467 FBXXTE01469 FBXXTE01472 FBXXTE01474 FBXXTE01476 FBXXTE01477 FBXXTE01478 FBXXTE01479 FBXXTE01480 FBXXTE01482 FBXXTE01483 FBXXTE01485 FBXXTE01486 FBXXTE01487 FBXXTE01488 FBXXTE01489 FBXXTE01491 FBXXTE01494 FBXXTE01496 FBXXTE01497 FBXXTE01500 FBXXTE01501 FBXXTE01502 FBXXTE01504 FBXXTE01505 FBXXTE01507 FBXXTE01508 FBXXTE01509 FBXXTE01510 FBXXTE01512 FBXXTE01513 FBXXTE01514 FBXXTE01515 FBXXTE01516 FBXXTE01517 FBXXTE01518 FBXXTE01519 FBXXTE01520 FBXXTE01521 FBXXTE01522 FBXXTE01525 FBXXTE01526 FBXXTE01528 FBXXTE01529 FBXXTE01532 FBXXTE01535 FBXXTE01537 FBXXTE01538 FBXXTE01539 FBXXTE01540 FBXXTE01542 FBXXTE01543 FBXXTE01544 FBXXTE01546 FBXXTE01547 FBXXTE01549 FBXXTE01551 FBXXTE01553 FBXXTE01555 FBXXTE01557 FBXXTE01559 FBXXTE01563 FBXXTE01566 FBXXTE01569 FBXXTE01570 FBXXTE01571 FBXXTE01572 FBXXTE01574 FBXXTE01579 FBXXTE01585 FBXXTE01586 FBXXTE01587 FBXXTE01590 FBXXTE01593 FBXXTE01602 FBXXTE01604 FBXXTE01613 FBXXTE01617 FBXXTE01618 FBXXTE01619 FBXXTE01620 FBXXTE01630 FBXXTE01632 FBXXTE01634 FBXXTE01644 FBXXTE01650 FBXXTE01651 FBXXTE01654 FBXXTE01663 FBXXTE01667 FBXXTE01672 FBXXTE01673 FBXXTE01677 FBXXTE01678 FBXXTE01681 FBXXTE01686 FBXXTE01687 FBXXTE01688 FBXXTE01690 FBXXTE01691 FBXXTE01692 FBXXTE01702 FBXXTE01704 FBXXTE01708 FBXXTE01718 FBXXTE01719 FBXXTE01720 FBXXTE01722 FBXXTE01723 FBXXTE01725 FBXXTE01727 FBXXTE01728 FBXXTE01731 FBXXTE01732 FBXXTE01734 FBXXTE01740 FBXXTE01748 FBXXTE01750 FBXXTE01754 FBXXTE01756 FBXXTE01758 FBXXTE01760 FBXXTE01761 FBXXTE01762 FBXXTE01765 FBXXTE01766 FBXXTE01767 FBXXTE01768 FBXXTE01769 FBXXTE01770 FBXXTE01771 FBXXTE01775 FBXXTE01779 FBXXTE01786 FBXXTE01787 FBXXTE01789 FBXXTE01801 FBXXTE01803 FBXXTE01805 FBXXTE01806 FBXXTE01811 FBXXTE01812 FBXXTE01820 FBXXTE01822 FBXXTE01826 FBXXTE01829 FBXXTE01830 FBXXTE01832 FBXXTE01835 FBXXTE01837 FBXXTE01839 FBXXTE01841 FBXXTE01845 FBXXTE01846 FBXXTE01847 FBXXTE01851 FBXXTE01852 FBXXTE01853 FBXXTE01856 FBXXTE01858 FBXXTE01860 FBXXTE01861 FBXXTE01862 FBXXTE01863 FBXXTE01865 FBXXTE01874 FBXXTE01875 FBXXTE01876 FBXXTE01877 FBXXTE01881 FBXXTE01884 FBXXTE01885 FBXXTE01889 FBXXTE01890 FBXXTE01892 FBXXTE01901 FBXXTE01903 FBXXTE01906 FBXXTE01908 FBXXTE01909 FBXXTE01913 FBXXTE01919 FBXXTE01921 FBXXTE01922 FBXXTE01923 FBXXTE01926 FBXXTE01929 FBXXTE01930 FBXXTE01931 FBXXTE01932 FBXXTE01933 FBXXTE01935 FBXXTE01936 FBXXTE01937 FBXXTE01940 FBXXTE01948 FBXXTE01950 FBXXTE01951 FBXXTE01967 FBXXTE01973 FBXXTE01975 FBXXTE01976 FBXXTE01978 FBXXTE01979 FBXXTE01980 FBXXTE01981 FBXXTE01987 FBXXTE01991 FBXXTE01992 FBXXTE01993 FBXXTE01994 FBXXTE01995 FBXXTE01996 FBXXTE02003 FBXXTE02004 FBXXTE02007 FBXXTE02012 FBXXTE02013 FBXXTE02014 FBXXTE02016 FBXXTE02018 FBXXTE02020 FBXXTE02021 FBXXTE02022 FBXXTE02024 FBXXTE02025 FBXXTE02026 FBXXTE02039 FBXXTE02040 FBXXTE02045 FBXXTE02046 FBXXTE02048 FBXXTE02049 FBXXTE02053 FBXXTE02054 FBXXTE02056 FBXXTE02059 FBXXTE02061 FBXXTE02064 FBXXTE02065 FBXXTE02066 FBXXTE02067 FBXXTE02070 FBXXTE02074 FBXXTE02075 FBXXTE02085 FBXXTE02092 FBXXTE02096 FBXXTE02097 FBXXTE02104 FBXXTE02105 FBXXTE02108 FBXXTE02109 FBXXTE02113 FBXXTE02115 FBXXTE02117 FBXXTE02118 FBXXTE02119 FBXXTE02121 FBXXTE02122 FBXXTE02126 FBXXTE02133 FBXXTE02139 FBXXTE02140 FBXXTE02141 FBXXTE02142 FBXXTE02148 FBXXTE02149 FBXXTE02150 FBXXTE02151 FBXXTE02153 FBXXTE02154 FBXXTE02157 FBXXTE02162 FBXXTE02163 FBXXTE02165 FBXXTE02168 FBXXTE02179 FBXXTE02187 FBXXTE02197 FBXXTE02199 FBXXTE02200 FBXXTE02201 FBXXTE02202 FBXXTE02207 FBXXTE02208 FBXXTE02213 FBXXTE02216 FBXXTE02244 FBXXTE02250 FBXXTE02251 FBXXTE02252 FBXXTE02257 FBXXTE02262 FBXXTE02279 FBXXTE02280 FBXXTE02281 FBXXTE02295 FBXXTE02300 FBXXTE02303 FBXXTE02304 FBXXTE02311 FBXXTE02312 FBXXTE02314 FBXXTE02318 FBXXTE02320 FBXXTE02323 FBXXTE02325 FBXXTE02331 FBXXTE02332 FBXXTE02333 FBXXTE02334 FBXXTE02337 FBXXTE02338 FBXXTE02340 FBXXTE02341 FBXXTE02344 FBXXTE02345 FBXXTE02350 FBXXTE02355 FBXXTE02356 FBXXTE02357 FBXXTE02358 FBXXTE02359 FBXXTE02365 FBXXTE02374 FBXXTE02385 FBXXTE02387 FBXXTE02388 FBXXTE02389 FBXXTE02390 FBXXTE02391 FBXXTE02401 FBXXTE02414 FBXXTE02418 FBXXTE02421 FBXXTE02425 FBXXTE02426 FBXXTE02428 FBXXTE02429 FBXXTE02430 FBXXTE02431 FBXXTE02432 FBXXTE02433 FBXXTE02434 FBXXTE02435 FBXXTE02437 FBXXTE02438 FBXXTE02439 FBXXTE02442 FBXXTE02448 FBXXTE02453 FBXXTE02454 FBXXTE02455 FBXXTE02458 FBXXTE02462 FBXXTE02463 FBXXTE02464 FBXXTE02468 FBXXTE02469 FBXXTE02470 FBXXTE02471 FBXXTE02472 FBXXTE02475 FBXXTE02476 FBXXTE02477 FBXXTE02478 FBXXTE02481 FBX

Distribution:

Distributed in: US

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1819-2025

Related Recalls

Class I - Dangerous

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Apr 20, 2026 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Apr 20, 2026 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Apr 20, 2026 Other Medical Devices Nationwide View Details β†’