Allia Moveo angiographic X-ray system
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00198953052488; Serial Numbers: C4-25-001, C4-25-002, C4-26-001, C4-26-002, C4-26-004, C4-26-005
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE Medical Systems, LLC
- Reason for Recall:
- GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Allia Moveo angiographic X-ray system
Product Codes/Lot Numbers:
UDI-DI: 00198953052488; Serial Numbers: C4-25-001, C4-25-002, C4-26-001, C4-26-002, C4-26-004, C4-26-005
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2318-2026
Related Recalls
Allia IGS 3 Pulse angiographic X-ray system
GE Medical Systems
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Allia IGS Pulse angiographic X-ray system
GE Medical Systems
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Allia IGS 5 Pulse angiographic X-ray system
GE Medical Systems
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.