Venue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (Model Numbers H45135VG, H45135VGW), R3 Surgery for China (Model Number H45113VGSU), R3 PeriOP for China (Model Number H45113VGPO), R3 Expert for China (Model Number H45113VGEX ), R3 Pro for China (Model Number H45113VGPR), R3 Performance for China, R3 Intervention for China, R3 Surgery for China (Model Number H45113VGSU) diagnostic ultrasound systems

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: Venue Go R2 system (GTIN 00840682138840); Venue Go R3 system (GTIN 00840682190503 / GTIN 00195278358011); Venue Go R4 system (GTIN 00195278516510 / GTIN 00195278516527); Venue Go R3 Surgery for China (GTIN 00195278547866) Venue Go R3 PeriOP for China (GTIN 00195278548344) Venue Go R3 Expert for China (GTIN 00195278547262) Venue Go R3 Pro for China (GTIN 00195278547828) Venue Go R3 Performance for China (GTIN 00195278547842) Venue Go R3 Intervention for China (GTIN 00195278548061) Venue Go R5 system (GTIN 00195278450319 / GTIN 00195278451774); System ID: VENGO000382 VGB002128 VGB003021 VGD000325 VGD002313 VGD000452 VGD000488 VGD002390 VGC006411 VGB002040 VGD000354 VGC000685 VGD001444 VGC006698 VGC007776 RVGC000034 RVGC000035 VGD002256 VGD002533 VGD002535 VGD002536 VGB001162 VGC007292 VGC007119 VGC006083 VGC006841 VGB001011 VGC000147 VGD001510 RVGC000559 RVGC000562 VGD000044 VGD001371 VGD001372 VGD001375 VGD001374 VGD001373 VGD001625 VGD001638 VGC000526 RVGD002184 VGD002513 VGB002239 VGD001705 VGD002307 VGC000746 VGC004373 VGC004378 VGC006707 RVGC007218 RVGC000591 VGD001899 RVGC004147 VGC004351 VGC007098 VGC007102 VGC007105 RVGC000024 RVGC004165 RVGD001551 RVGB000480 RVGB002675 VGC007113 VGD000091 RVGD001076 RVGB000467 VENGO000545 VGC007350 VGD001731 VGB002526 VGC000227 VGD001964 VGD002225 RVGC000303 RVGB000757 RVGB000797 RVGC000302 RVGC000554 VGC000799 RVGC006372 RVGC006771 VGC007879 VGC007880 VGD000524 RVGB000791 VGD001873 VGD002221 VGD002227 VGD002467 VGD000730 VGD001872 VGC007607 VGC007611 RVGC007204 VGC006763 VGD000731 VGD000843 RVGB000694 VGD000553 VGD000554 VGD000558 VGD001336 VGC000898 VGD000200 VGC006715 VGD002460 VGD002462 VGD002427 RVGB000032 RVGC000594 RVGC000617 VGD001738 VGD001747 RVGC006758 RVGD001055 VGB000796 VGB000901 RVGD001795 VGD001400 VGD001401 VGB002539 VGB002613 VGB001125 VGD001200 VGD001257 VGD001258 VGD001258 VGD001257 VGD001261 VGD001257 VGC000386 RVGD001071 VGD002275 VGD001701 VGD001704 VENGO000317 VGB002122 VGD001687 VGD001884 VGD001885 VGD001889 VGC006031 VGC006036 VGC004366 VGC004521 RVGD001797 VGB001150 VGB000766 VGC000947 VGC000948 VGC006790 VGD000142 VGD001602 VGD000814 VGC006847 VGD001463 VGD001475 VGD001584 VGD002463 VGB000701 VGC000109 VGD001673 VGB000737 VGC000379 VGC004001 VGD001058 VGD001923 VGD001982 VGD001984 VGC007456 VGC007643 VGC007644 VGD002283 VGD002285 VGB002527 VGB002531 VGD001569 VGD002397 VGD002506 VGB000720 VGB000722 VGB000723 VGD001090 VGB002102 VGC000435 VGC000951 VGC004160 VGC007136 VGB002574 VGC007639 VGD002403 VGD002407 VGD002411 RVGB000122 VGC000866 VGC006812 VGC006814 VGC000082 VGC000084 VGC000086 VGC000085 VGC006402 VGC006406 VGC006652 VGC007049 VGC007053 VGC006409 VGD000035 VGD002484 VGD001037 VGD001150 VGD001900 VGD002237 VGB000165 VGC000943 RVGB000184 VENGO000700 RVGC000185 RVGC007163 VGC000321 RVGB000265 VGB000674 RVGC006341 VGD000399 VGD002328 VGB001415 RVGC000488 14540 VGB001140 VGB001916 VGB002688 RVGB000226 RVGB000472 VGD000932 RVGD001065 RVGD001427 VGC000671 VGC006086 VGC000156 VGC006113 VGC006127 RVGB000425 RVGB000171 VGC006710 VGD000603 VGD002018 VGC000879 RVGB000073 VGD002487 VGD002375 VGD000969 VGD000970 VGC007104 VGB001132 14617 14280 VGD002377 VENGO000703 VGB000710 VGB000707 VGB002548 VGD001005 VGD002230 VGD001573 VGB001934 VGB002489 VGB002565 VGC006105 VGD001043 VGD001482 VGC000407 VGD002126 VGB001009 VGD002509 VGD000914 VGC006958 VGD000350 VGD000353 VGD002170 VGC000099 VGC000115 VGC000099 VGC000146 VGD001298 VGD001301 VGB002073 VGC000269 VGD001275 RVGB000420 VGD001933 VGB001168 VGD002234 VGC006278 VGC007636 VGD001719 VGC006661 VGB002734 VGC006294 VGC006296 VGB002469 VGD002189 VGC000718 VGD002362 VGC000521 VGB000466 VGC000324 VGC000319 VGC000322 VGC000322 VGC000324 VGC000327 RVGC000019 VGB000998 VGC000278 VGD000138 VGC006956 VGD001161 VGD001163 RVGB000437 VGB000911 VGB001876 VGD002280 RVGC000570 VGD000513 VGC006534 VGD000902 RVGC000041 VGC006995 VGD001390 RVGB002663 RVGB002674 VGC006562 VGC006563 VGC006567 VGC006569 VGC006570 VGC006574 VGD002329 RVGC000040 VGB002970 VGD002036 RVGB000116 RVGB000160 RVGB000832 VGD002466 RVGD000137 VGB003006 VGD000307 VGD002178 VGB004063 VGC004102 VGC004110 VGD001382 VGD001749 VGD001656 VGC006697 VGC007044 VGC006837 VGD001949 VGC000487 VGD002003 VGD002062 VGD000753 VGD000761 VGC006816 VGB002524 VGC000279 VGC006665 VGD002056 VGD001051 VGD001052 VGD001639 VGB000684 VENGO000686 VGB000688 VGC007135 VGB000712 VGD002516 VGB000865 VGC000130 VGC000480 VGC000703 VGC000704 VGC007107 VGD000536 VGB002772 VGC006340 VGC000346 VGD002370 VGC006702 VGD002017 RVGB000066 RVGB000793 RVGB002244 VGD001517 VGD001526 VGB001268 VGB002573 VGC000323 VGC000364 VGC000741 VGC000750 VGC000753 VGC006443 RVGC007168 VGC007421 VGC007422 VGC004000 VGC006634 VGC007326 VGD002350 VGC006672 VGD000306 VGC004205 VGC004206 VGD000176 VGD001514 VGB004017 VGB004019 VGC000081 VGC000083 VGC000090 VGC000089 VGC000092 VGC000091 VGC000087 VGC000093 VGC007032 VGC007037 VGC007032 VGC007053 VGC007037 VGC007032 VGC007048 VGC007051 VGC007052 VGD001713 VGD001718 VGC000086 VGD001897 VGC007741 VGD001194 VGD002172 VGC007455 VGD002346 VGC007023 RVGB000383 VGD000541 VGD000546XX VGD000955 VGD001377 VGD002233 VGC000094 VGD000297 RVGB000808 VGD000777 RVGD001755 VGB002547 VGD001023 VGC006506 VGC006508 VGD000938 RVGC006368 RVGC006370 VGD000901 VGB000992 VGB001020 VGB001246 VGB001999 VGB002696 VGB002281 VGB002554 VGB002555 VGB002556 VGB002557 VGB002559 VGB002561 VGC000911 VGC000970 VGD002132 VGD002134 VGD002273 RVGC006682 VGC000680 VGC006650 VGC006669 VGD000424 VGD000445 VGB000666 VGB002462 RVGB002692 VGB002781 VGC004144 VGD001694 VGD001695 VGB002098 RVGD001798 VGD002340 VGB000699 VGB002117 VGC004161 VGC004263 VGC004266 VGC007387 VGD002082 VGD002181 VGC006093 VGD000235 VGD001217 VGB000756 VGB002870 RVGC000045 VGC004458 VGD002041 VGD002051 VGD002041 VGD000247 RVGD001450 RVGB000470 VGD001645 VGD001847 VGC000760 VGC000763 VGC000785 VGC007041 VGD000391 VGC007400 VGC006723 VGD000040 RVGC000168 VGC004038 VGD000045 RVGB000427 RVGD001537 VGB002100 VGB002104 VGB002106 VGB002108 VGC006103 VGC006678 VGC007039 VGB001414 VGD002205 VGD002279 VGD002315 VGC007060 VGC004141 VGC000072 VENGO000293 VGD000310 VGD000317 VGD000310 VGC000461 VGC006070 VGC007596 VGD000985 VGC004078 RVGC000585 RVGC004182 VGD002473 RVGC000001 RVGC006764 VGB000655 VGD000343 VGB000649 VGD001830 VGC000936 VGC007431 VGC000004 VGC004463 VGC004512 VGC006575 VGD000538 VGD000555 VGC000004 VGC000098 VGC000097 VGC000106 VGC000105 VGC004513 VGC004514 VGC007064 VGC007069 VGC007073 VGD000234 VGB001004 VGB002039 VGC006675 VGD000342 VGB002170 VGD001710 VGC004497 VGC006062 VGD001857 VGD002033 VGC000742 VGC000756 VGC000903 VGC004116 VGC004118 VGC004506 VGC009011 VGD002337 VGD002338 VGC000143 RVGB001984 VGC006052 VGB002004 VGC000688 VGC000701 VGC000720 VGC007210 VGD002300 VGC000670 VGD000925 VGB001015 RVGB000252 VGC000417 VGB000692 VGB001480 VGB001491 VGB002016 VGB002052 VGB002123 VGC000533 VGC000921 VGD000715 VGD001031 VGD001337 VGD001337 VGD001337 VGD001337 VGD001611 VGD001953 VGD001955 RVGB000354 VGC000530 VGC000200 VGC000219 VGC000213 VGC000219 VGC000219 VGC000219 VGC000219 VGC004107 VGC004490 VGC006338 VGD000790 VGD001160 VGD001782 VGB000693 RVGC006684 VGC007562 VGC006759 VGD002209 VGD000884 VGD001481 VGC000602 VGD001457 VGD001468 VGB002311 VGC000481 VGC000489 VGC006041 VGC006043 VGC006094 VGC006124 VGC006126 VGC006913 VGC007631 VGC007635 VGD002402 VGD002405 VGD002408 VGD001118 VGD002376 VGC000645 VGC007671 VGC006794 VGC006169 VGC004086 RVGC004150 VGC004403 VGD000155 VGD001737 VGD001986 VGD002005 VGD002016 VGD002025 VGD002030 VGD002034 VENGO000403 VGB000503 VGB000504 VGB000510 VGC004516 VGC006378 VGC007027 VGD002507 VENGO000339 VGB002157 VGB000805 VGC007707 VGD000917 VGB000805 VGB000671 VGB000680 VGB000685 VGC000138 VGC000140 VGC000142 VGC000151 VGC000152 VGC004518 VGC004518 VGC000103 VGD002323 VGD002326 VGD002327 VGD002392 VGB001982 VGC006720 VGD001486 VGB001917 VGC007733 VGC007233 VGD000057 RVGB000835 VGD000302 RVGB000199 VENGO000385 VGC000077 VGC000759 VGC000762 VGC006030 VGC006035 VGD000926 VGB002643 VGD002086 VGD002086 VGC004479 VGC004461 VGC004027 VENGO000459 VGB002383 VGD002013 VGD001388 VGD001918 VGC004030 RVGC007198 VGD001085 VGD001266 VGD001278 VGB001269 VGC000403 VGC004488 VGC004493 VGB002029 VGC000385 VGC006029 VGC006034 VGC006444 VGC006450 VGC006967 VGC006988 VGC000485 VGC007781 VGB002129 VGB002130 VGC000144 VGC000934 VGC000935 VGC004026 VGC004508 VGC004526 VGC007063 VGC000953 VGC004515 VGC006071 VGB000667 VGC004509 VGC000950 VGB002101 VGC000912 VGC000932 VGC000965 VGC007450 VGD002414 VGD000935 VGD002434 VGD002450 VGB002431 VGC004197 VGC004199 VGC004201 VGC004203 VGB000873 VGB001002 VGB001994 VGD001004 VGB002520 VGC000421 VGC000413 VGD002530 VGD002538 VGB002030 VGD000832 VGD001197 VGD001378 VGB000673 VGD002265 VGC000522 VGC006243 VGC006252 VGC000897 VGB002140 VGB002141 VGB002492 VGB002512 VGB002626 VGB002641 VGC000073 VGC000074 VGC000074 VGC000478 VGC000479 VGC000539 VGC000895 VGC004143 VGC006084 VGC006097 VGC006139 VGC007075 VGC007594 VGC007601 VGC007637 VGD000797 VGD001459 VGD001460 VGD001540 VGD001823 VGD002341 VGD002511 VGC004034 VGC006140 VGC007002 VGB002532 VGC004023 VGC004517 RVGC007169 VGC000463 RVGB000402 VGC000924 RVGC006848 RVGC007194 VGD000590 VGC004088 VGC007618 RVGC000066 RVGD000338 RVGD001066 RVGD001067 RVGD001068 RVGD001070 RVGD001466 RVGD001506 RVGD001550 RVGD001622 RVGD001624 RVGD001789 RVGD001801 VGD000341 VGD000362 RVGC000012 VGD000311 VGD001032 VGD001034 VGD001958 VGD001483 VGB002103 VGC004559 VGC007835 VGD002011 VGD002021 VGB001104 VGC000929 VGD002413 VGC007630 VGC000392 VGD001493 VGB001174 VGC000532 VGB002125 VGB002533 VGC006419 VGC007395 VGD001120 VGD001121 VGD001151 VGD001152 VGB002267 VENGO000493 RVGC006687 VGC004066 RVGC007160 VGC006593 RVGB002253 RVGC004124 3061702 VENGO000444 RVGC006608 RVGB000130 VGC007121 VGB002025 RVGC000592 VGD002260 VGB001183 VGB001187 VGB001189 VGB002706 VGC000916 VGC006653 VGC006664 VGD001667 VGC000515 VGD000345 VGC006818 VGB000813 VENGO000528 VGB000842 VGC000484 VGC000890 VGC004481 VGD000330 VGD002004 VGD002503 VGD002508 VGD000336 VGC000958 VGD001836 VGC004248 VGC004255 RVGC006352 RVGC006774 RVGC006775 RVGC007154 RVGC007155 RVGC007166 RVGC007192 VGD000111 VGD000113 VGD000116 VGD001429 VGD001518 VGD002136 VGD002254 VENGO000687 VENGO000695 VGD001740 VGD001741 VGD002075 VGC004467 VGC004487 VGC007725 VGD001089 VGC000308 VGC000312 VGC000314 VGD000229 RVGD000081 VGD001515 3137004 VGD001686 VGC007599 VGB000390 VGB001021 RVGC000044 RVGC000046 RVGC000047 RVGC000062 RVGC000064 RVGC000248 RVGC000043 VGD000765 VGC007006 VGD001564 VENGO000422 VENGOB000327 VGB002082 VGB002378 VGB002391 VGB002681 VGB002715 VGC004472 VGC004473 VGD000998 VGD001008 VGD001008 VGD001396 VGD001541 VGD002147 VGD002168 VGD001226 VGD000899 RVGB000047 VENGO000439 VGD000117 VGD001729 VENGO000245 RVGC000015 VGC007031 VGC007035 VGC007036 RVGC006769 VGC000289 VGC000382 VGC004480 VGC004485 VGD001812 VGC007641 VGD001643 VGC004094 VGC004560 VGD000054 VGD000055 VGD000049 VGB002673 VGD002053 VGB000060 VGB000658 RVGC004135 RVGC006765 VENGO000394 VGD001972 VGD001478 VGD001479 VGD000272 VGB002174 RVGB000369 VGB001184 VGB001191 VGB001901 RVGB002261 RVGB002679 VGC000110 RVGC000616 RVGC006545 RVGD001387 VGD001806 VGD002336 VGD001967 RVGB002788 VGD002475 RVGC007157 VGD002173 VGD002175 VENGO000389 VGB001899 VGC000436 VGC000080 RVGB000765 VGC007056 VGD001576 VGC006996 VGC006999 VGB001903 VGC006358 VGC006830 VGD000038 VGD000115 VGD000149 VGD001489 VGD001501 VGD001502 VGD001505 VGD001508 VGB000817 VGB001568 VGD002259 VGD000344 VGB002568 VGC004507 VGC004520 VGD001030 VGB001897 RVGB000768 RVGC000576 VGD000561 VGB002523 VGC007443 RVGB002097 VGC000408 VGC000416 VGC006044 VGD000539 VGD000545 VGD002187 VGD000324 VGC000154 VGC000636 RVGB000743 VGD001756 VGD002135 VGC004064 VGC006641 VGD000332 VGD000333 VGD000421 VGD000436 VGD000443 VGD001934 VGD002295 VGD002330 VGD002332 VGD002334 VGC007271 VGC007275 VGC007277 VGC007271 VGC006683 VGC006727 VGD000359 VGB002525 VGB002731 VGC007663 VGD002373 VGD000741 VGC006783 VGD001536 VGC004238 VGD001745 VGB000406 VGC000277 RVGC007164 VGC007389 VGB002120 VGB002133 VGB002138 RVGB000173 VGD001808 VGD001948 VGD001608 VGD001610 VGD001613 VGB002486 VGC000376 VGD001079 VGD000752 VGD000754 VGC006668 VGC000078 RVGB002271 1000000926 VGD000997 VGD002235 VGC006696 VGC006700 VGC006704 VGC006957 1000003449 VGC006122 VGB002543 VGC006074 VGD001978 VGC000171 VGC000171 VGB003051 VGC007348 VGC007363 VGB002766 VGD000245 VGD001050 VGC000055 VGB004004 VGB004005 VGB001870 RVGB000809 VGC006114 VGD000232 RVGC000030 RVGD001666 VGC006260 VGC007132 RVGB000150 RVGC000587 VENGO000076LO VENGO0002407LO VENGO000241LO VENGO001981LO VENGO001992LO VENGO000233LO VENGO000720LO VENGO004236LO VENGO006099LO VENGO007263LO VENGO000559LO VENGO000907LO VENGO001750LO VENGO002398LO VENGO000981LO VENGO002054LO VENGO000085LO VENGO002100LO VENGO004196LO VENGO000084LO VGD001708 VGD002252 VGB000807 VGD001981 VGC000519 VGD000892 VGD002065 RVGB000818 RVGD000017 VGC006525 VGC006528 VGD000608 VGD000721 VGD000722 VGD001080 VGD001084 VGD002400 VGC007597 VGD001653 VGD002144 VGC000906 VGC004494 VGC004496 VGC006321 VGD002268 VGC004264 VGC004268 VGC007775 VGD000610 VGB002066 VGC004350 VGC000558 VGC006231 VGD001285 RVGB002260 VGC006908 VGC007114 VGD001419 RVGB000170 VGB000855 VGB002331 VGB002332 VGB002335 VGC000909 VGC000967 VGC000969 VGD001240 VGB002365 VGC004037 RVGC0000521 RVGC007213 VGD000058 VENGO000290 RVGB000177 VGB002208 VGD002048 VGC007533 RVGB002668 VGD002133 VGC007001 RVGC000586 RVGC004114 RVGC004136 RVGC004146 VENGO000381 VGD001547 VGC000491 VGC007110 VGB002118 VGB002121 VGB001885 RVGC000228 VGD001316 RVGB002273 VGC000520 RVGB002249 RVGC000028 VGC000184 VGC000187 VGD000355 VGD002420 VGD000606 VGD000607 VGD002179 VGC000358 VGD001714 RVGB002270 RVGC000226 RVGC000238 VGC006785 RVGB000178 RVGB000186 VGC000766 VGC006531 VGC006537 RVGB000108 RVGB000167 RVGB000029 RVGB000125 RVGB000137 RVGB000426 RVGB000144 VGB001288 RVGC004134 VGD002076 VGD002130 VGC004005 RVGC000249 VGD001669 VGD002289 VGD002299 VENGO000642 VGC007632 RVGB000198 RVGB000828 RVGB002258 VGD001635 VGB000871 VGB001926 VENGO000278 VENGO000281 VENGO000286 VENGO000289 VENGO000296 RVGB000117 VGD000923 VGD001392 RVGB000138 RVGB000142 RVGC006777 VGC000167 RVGB000143 VGD001439 VGD001535 VGC006147 VGC006148 VGC006149 VGC006150 VGD000540 VGD001413 VGD002476 VGD001243 VGB000732 VGB000754 RVGC000013 VGC007138 RVGC007179 RVGB002256 RVGB002275 RVGB002661 RVGB002662 VGC000281 VGC000281 VGC000422 VGC000500 VGC000499 VGC006911 VGD000473 VGD002412 VGD002421 VGC000499 VGC006337 VGB001262 RVGD000013 RVGD000016 RVGD000019 RVGC006826 VGC000764 VGC000771 VGC000774 VGC006631 VGD002384 RVGB002664 RVGC007180 VGD002180 VGC006067 VGB000647 VGC000161 VGC000161 VGD002514 VGD000072 VGB000856 VGB000868 VGB001001 RVGD001410 VGD001961 VGD001965 RVGC004130 VGD001854 VGD001041 VGD001255 VGD002393 VGC000772 VGC006660 VGC006660 VGD001297 VGD002023 VGD000885 VGC000937 VGC000393 VGC000523 RVGC006840 VGD000724 VGD000043 VGD000061 VGC000190 RVGD000130 VGD000557 VGC006538 VGD000289 VGD000394 RVGB000462 RVGB000468 VGD002430 RVGB000128 RVGB000133 RVGB000145 RVGB000152 RVGB000203 RVGB000231 RVGC006828 VGD002203 RVGB000159 VGD000766 VGC000957 RVGB000069 VGB000772 VGB000774 VGD000592 VGB001006 VGB001006 VGB002784 VGB002288 VGB001176 VGC007076 VGC007082 VGC007080 VGC007079 VGC000893 VGC000930 VGD001664 RVGB000039 VGD000047 RVGD001072 VGD001319 VGD001318 VGD002290 RVGD002537 VGD002198 VGD001696 RVGB000833 VGC007071 VGB002671 VGC006192 VGC006202 VGD000075 VGD000078 VGD000079 VGD000082 VGD000886 RVGC007186 VGC006123 VGB000755 VGD001578 RVGD000444 VGC000601 VGC007736 VGC006050 VGC006064 VGC006080 VGD000060 VGC006265 VGD001868 RVGB000482 RVGB000488 VGD000908 VGD000915 VGD000922 VGD000928 VGD000929 RVGD001527 VGD001630 VGB004012 VGC007480 VGD000062 VGC004379 VGC007068 VGD002363 RVGC000571 VGD002019 VGC000163 VGC000896 VGC007541 VGC006850 VGD002143 VGD002312 VGC000390 VGB002425 VGC000258 VGC007115 VGC000765 VGD000357 RVGC004140 VGC006861 VGC006863 VGD001790 VGB002310 VGC004455 VGC007025 VGC000945 VGC006673 VGB001197 VGB001951 VGB002139 VGB002538 VGD000986 VENGO000646 RVGB002124 VGD001671 RVGB002657 VGD002009 VGB002546 VGB000999 RVGD000241 VGB000690 RVGC006548 VGD000204 VGC007434 VGB002224 VGB002230 VGD000323 VGC000889 VGC000891 VGC006676 VGC006679 VGC007600 VGC000514 VGC006693 RVGD001406 RVGB000428 VGD002357 VGB001998 VGB002041 VGC000401 VGD002495 VGD002182 RVGB000080 RVGB000110 VGB002597 VGC006566 VGD001511 RVGD000153 VGC000116 VGC000117 VGC000119 VGD001484 RVGB000798 RVGB002220 VGD001628 RVGC000020 VGD001647 VGC000702 VGD000221 VENGO000239 VGD000150 VGD002154 RVGB000135 RVGB000277 RVGB000831 VGB002250 VGC000952 VGC004055 RVGB000230 VGC006383 RVGC000573 RVGB002450 VGC000976 VGC004010 VGC006796 VGC006799 VGC006912 VGC006621 VGC006699 VGD002417 VGC000053 VGC000057 VGC000441 VGC007500 VGC007502 VGC007503 VGC007591 VGD000820 VGD001585 VGD002468 VGD002470 VGD002472 VENGO000536 VGB000672 RVGB000799 VGC000173 VGC000352 VGD001748 VGD001954 VGC006807 RVGC000501 VGD001286 VGD001519 VGB002698 VGC000527 VGC006335 VGC007546 VGD001881 VGD000143 RVGC007167 VGD001938 VGD002345 RVGD001440 VENGO000434 RVGC000229 RVGC000307 RVGD001391 VGD001492 VGC004142 VGC007623 VGD002429 RVGD000015 RVGD000135 RVGD001432 RVGB000773 VGC006659 VGD002152 VGC007409 VGC000207 VGC007837 VGD001902 VGD002064 VGB003022 VGD002499 VGD002497 VGD002498 VGD002490 VGC007805 RVGB000263 RVGC000584 VGD000304 VGD000304 VGD000320 VGC007061 VGB000664 VGB000981 VGB000982 VGB001017 VGB001019 VGC004498 VGC006032 RVGC000425 VGC004008 RVGC000578 VGD001204 RVGC000310 RVGB000164 VGD002293 VGD002415 VGB001564 VGC007806 VGC006401 RVGC000006 RVGC000016 RVGC000026 RVGC000021 RVGC000025 RVGC000027 RVGC000029 RVGB000779 VGC007134 RVGB000096 RVGB000124 RVGB000126 RVGB000136 VGB001932 VGB000691 VGD000063 RVGC004133 VGC009042 VGC009044 VGC009045 VGB000702 VGC006501 VGC007783 VGD001256 VGD001264 VGC000885 VENGO000811 VGD001919 VGD001921 VGD001925 VGD001929 VGD001930 VGD001951 VGC006038 VGD001996 VGD000789 VGC006624 VGD001343 VGD001344 VGD001345 VGD001346 VGD001347 VGD001349 VGD001352 VENGO000324 VGB002294 VGB000834 VGB000866 VGD001179 VGD001187 VGD001190 VGD001191 VGD002081 VGD002091 VGD001189 RVGC000070 VGC007687 VGC007687 VENGO000717 VENGO000718 VENGO000719 VENGO000721 VGC000137 VENGO000282 VENGO000283 VENGO000287 VENGO000288 VGB002356 VGD002459 VGD002433 VGC000177 VGC000178 VGC000177 VGC004002 VGC004002 VGD000817 VGD000819 VGC007126 VGD000738 VGD000727 VGB002725 VGC007638 RVGC006344 VGD002320 VGD002322 VGD002494 VGC004019 VGD001730 VGC006987 VGC006805 VGD001366 VGD002319 RVGC000398 VGB002362 VGB002519 VGC004348 VGC006620 VGC006671 VGC006817 VGD000896 VGD001716 VGD001891 VGB002038 VGB002042 VGD002032 VGB002303 VGB002303 VGB000713 VGB000753 RVGC000577 VGD002387 VGD002389 VGD001828 VGD002202 VGC009058 RVGC000023 VGD002007 VGC007429 VGD001332 VGD001783 VGC006039 VGD001294 VGC007211 VENGO000393 VENGO000404 RVGB000075 VENGO000386 VGC000483 VGC006511 VGC006757 VGD001454 VGD002059 VGC006435 RVGB000049 RVGB001900 RVGD001528 RVGB000031 VGC006512 VGB000689 VGB000714 VGC000715 VGC007660 VGC007670 VGD001906 VGD001909 VGD001910 VGD001914 VGC007669 VGB001441 VGB001570 RVGB000424 VGC009030 RVGC000049 RVGD000014 RVGD000124 VGB000854 VGC000315 VGC000925 VENGO000285 VGB002773 VGC007000 VGC007059 VGC000974 VGD002453 VGB002172 VGC000412 VGC000513 VGC000513 VGC000528 VGC000902 VGD001487 VGC000949 RVGB002272 RVGB002262 RVGC000556 VGC007122 RVGB000824 VGC007481 VGB002001 RVGB000284 RVGC006680 VGC006142 VGD001011 VGD000052 VGD002523 VGD000596 VGB000677 VGB000678 VGB002521 VGC000767 VGC004183 VGC004344 VGC000108 VGC007721 VGB002158 VGC004251 VGC004257 VGD002291 VGD002304 VGD002305 RVGB000786 RVGB000821 VGB001986 VGC009013 VGC004438 VGD002368 VGD001467 VGB001962 VGB000683 VGD001183 VGD002047 RVGC000575 VGC000325 VGC000326 VGD002258 VGC006129 VGC007452 VGC006712 VGC006138 VGC004080 VGC006106 VGC006136 VGC006146 VGB000665 VGB000670 VGD000046 VGC000265 RVGD001420 VGB000668 VENGO000733 VGB002359 VGB002361 VGD001621 RVGD001600 RVGB000041 VGC006786 VGD002491 RVGB000792 RVGB002269 VGC000972 VGD000199 VGD000393 VGC000420 VGC006918 VGD001874 VGD001875 VGC000170 VGC006824 VGC006825 VGD001959 VGC007442 VGC007810 VGB001991 VGB001997 VGB002490 VGC000113 VGC000357 VGC000357 VGC007123 VGD001047 VGB001192 VGD001640 VGC000328 VGC007077 VGC007077 VGC007100 VGC007100 VGD001650 VENGO000295 RVGB002264 VGC004245 VGC004249 VGC004525 VGC004527 VGD000909 VGD000910 VENGOB000337 VGB001166 RVGC007158 VGD001042 RVGB000156 RVGB000421 VGB000452 VGB000456 VGB000474 VGC000359 VGC004016 VGC007070 VGC007140 VGC007330 VGC007338 VGC007339 VGC007340 VGC007745 VGC007764 VGC007769 VGD000944 VGD001678 VGD001980 VGC007074 VGC007081 VGC007336 VGD001173 VGD001198 RVGB000123 RVGB000193 VGC000939 RVGD001056 RVGD001525 RVGD001543 VGD000801 VGD001581 VGD001768 VGD001353 VGD001016 RVGC000036 VGC006045 VGC006058 VGC006057 VGC006058 VGD001123 VGD001534 VGD002114 VGD001751 VGD002115 VGD002108 VGD002109 VGD002115 VGD002109 VGD002106 VGD002114 VGD002115 VGD000846 VGD001832 VGC000975 VGC004180 VGC006834 VGD000583 VGD000584 VGD000888 VGD000903 VGD000988 VGD001835 VGD001837 VGB001118 VGD000795 VGD001572 VGC006061 VGD001876 RVGC000071 VGC006557 VGC006565 VGC006568 VGC006584 VGC009006 VGC009005 VGC009008 VGC009004 VGD002084 VGC000928 VGC000923 VGC000910 VGC000918 VGC007150 VGC007152 VGD000159 VGD000882 VGD000883 VGD000937 VGD001102 VGD001108 VGD001815 VGD001817 VGD001818 VGD001819 VGD001826 VGD001826 VGD001819 VGD001831 VGC007072 VGD002372 VGD002378 VGC004542 VGC006477 VGC006477 VGC006481 VGC006498 VGD000879 VGD000880 VGD000881 VGD000912 VGC004541 VGD000351 VGC004408 VGD000591 VGC006130 VGC006141 VGC006448 VGC006452 VGC006452 VGC006791 VGC006795 VGD000126 VGD000127 VGD000126 VGD000818 VGD000821 VGD000859 VGD000861 VGD000862 VGD000865 VGD000990 VGD001081 VGD001101 VGD001125 VGD001339 VGD001341 VGD001341 VGD002212 VGD002218 VGD002219 VGD002212 VGD002232 VGD002218 VGD002232 VGD002218 VGB001902 RVGC000581 VGC006654 RVGC000305 VGB001005 VGB002142 VGB002147 VGC004454 VGC004511 VGC006717 VGD001878 RVGC006835 RVGC006369 RVGC006677 VGC006651 VGD000887 VGD000889 VGD000889 VGD001003 VGD001279 VGD001283 VGD002211 VGD002216 VGD002211 VGD002211 VGC007765 VGD000594 VGC006657 VGD001376 VGD001469 VGD001470 VGD002303 VGD002324 VGD001851 VGC007272 VGC007814 VGD001252 VGC006947 VGD002122 VGD002123 RVGB000036 RVGB000046 VGB000654 VGC007127 VGC007231 VGD000601 VGD000609 VGD001549 VGD001632 VGD001800 VGB002530 VGD000890 VGD001024 VGD001317 VGD001335 VGD001869 RVGB000129 VGC007603 VGD000595 VGD000597 VGD000598 VGD000599 VGD000602 VGD001883 VGD000597 VGD002171 RVGB000155 VGD000600 VGC006120 VGC006125 VGC006133 VGC006135 VGD002145 VGD000957 VGD000960 VGD000961 VGD000964 VGB001022 VGC006096 VGC006109 VGC006112 VGC006131 VGC006360 VGC006361 VGB002534 VGC006527 VGD001082 VGD000292 VGD000314 VGD001009 VGD001015 VGD001098 VGD001099 VGD001103 VGD001106 VGD001110 VGD001111 VGD001119 VGD001122 VGD001271 VGD001288 VGD001289 VGD001293 VGD001342 VGD002022 VGD002057 VGD002093 VGD002094 VGD002096 VGD002097 VGD002098 VGD002101 VGD002102 VGD002103 VGD002104 VGD001879 VGD001882 VGD001888 VGD001890 VGD001894 RVGB000146 RVGB000147 RVGB000148 RVGB000166 RVGB000172 RVGB000151 RVGC006345 VGD001155 VGD001615 RVGB000481 RVGC000022 VGD001464 VGC000744 VGC000745 VGC000748 VGC000749 VGC000752 VGC000754 VGC007449 VGD002185 VGD001557 VGC007619 VGC007629 VENGO000297 RVGC006773 RVGC007212 VENGO000266 RVGC000418 RVGC006689 RVGD001195 RVGC000039 VGD000329 RVGB000140 RVGC007224 VGD002015 RVGB002665 RVGB002669 VGC000913 VGC006709 VGC007029 VGC007030 VGC007033 VGC007034 VGC007038 VGC007040 VGC007045 VGD000924 VGD001046 VGD001054 VGD001046 VGD001231 VGD001296 VGD001303 VGD001310 VGD001307 VGD001304 VGD001313 RVGB002268 VGC004462 VGD002479 VGD000231 VGD000231 VGD001892 VGD002186 VGD002190 VGD000965 RVGB000027 RVGB000054 RVGB000056 RVGB000059 RVGB000062 VGB000067 RVGB000154 RVGB000183 RVGB000229 VGB000514 VGB000661 RVGB000750 RVGC000051 RVGC000056 RVGC000059 RVGD000348 RVGD001078 RVGD001178 RVGD001399 RVGD001415 RVGD001425 RVGD001430 RVGD001431 RVGD001438 RVGD001441 RVGD001442 RVGD001521 RVGD001522 RVGD001532 RVGD001788 RVGD001792 RVGD001794 RVGD001799 RVGD001803 RVGD001810 RVGD002192 RVGD002194 RVGB000157 VGB000715 VENGO000770 RVGB000051 VGD000782 VGB000679 RVGB000068 VGC007768 RVGD001520 VGD001636 VGC000509 VGC000770 RVGC006829 VGD002382 VGD002383 VGD002382 RVGB000781 RVGB000781 RVGC000242 VGC004386 VGC006326 VGB000566 VGB000675 VGB001950 RVGB000052 RVGB000153 VGB002131 VGB002137 RVGC000593 RVGC000598 RVGC000611 VGC000946 VGC006662 RVGC006770 VGC004369 VGD000833 VGD000528 RVGC000239 VGB000711 RVGC000607 VGD002269 VGD001253 VGD002239 RVGC004121 VGC007062 VGD002469 VENGO000345 RVGC006968 VGC006299 RVGC0000541 VGC006069 VGC006072 VGC006073 VGC006085 VGC006076 VGC006078 VGC006079 VGC006076 VGC006075 VGC006098 VGC006100 VGC006101 VGC006110 VGC006639 VGC007129 VGC007130 VGC007521 VGC007538 VGC007539 VGC007540 VGD000313 VGD000315 VGD000319 VGD000905 VGD000418 RVGD001074 RVGC000017 VGB000104 VGC007197 VGC006821 RVGB000175 RVGB000763 RVGC000003 VGD000904 VGD002210 VENGO000508 RVGB000376 VGC000220 RVGB000127 VGB001185 VGB002360 VGC007228 VGC007388 RVGD001732 RVGC000608 RVGC004115 VGD002505 VGC004185 RVGB000174 RVGB000761 RVGB000761 VENGO000262LO VENGO000446 RVGC007172 VGD002471 VGC006572 VGC006718 VGC007278 VGC007808 VGC007809 VGB002062 VGB002064 RVGD001416 VGC006648 VGC004553 VGD000321 VGB002544 VGB002583 VGB002672 VGB002960 VGC004014 VGD001320 VGD002451 VGB002126 VGB002854 VGC006077 VGD001735 VGD001833 VENGO000228 VGD000823 VGD001654 VGD001674 VGD001676 VGD001677 VGD001781 VGC006820 RVGD001421 RVGB000457 VGC000400 RVGB000822 VGD002072 VGB001877 VGC007374 VGD001616 VGC000507 VGC006729 VGD000831 VGD001471 RVGD001796 VGC000920 VGB002542 VGB001113 VGC000268 RVGB000120 VGD001370 RVGB000826 VGB001178 VGD001991 VENGO000380 RVGC000564 VGD001988 RVGC006365 RVGB000196 VGD001273 VGD001603 VGB002055 VGD000331 VGD000337 VGD001985 VGD001993 RVGC006367 RVGC007165 RVGC007171 VGD001230 VGD001232 VGD001233 VGD001235 VGD001239 VGD001233 VGD001244 VGD002519 RVGC007184 VGD002464 VGD002517 VGC006509 VGD000449 VGD000587 VGD000588 VGD000605 RVGD001791 VGD001895 VGC006091 VGC004240 VGC004241 VGC004156 RVGB000070 RVGB000072 RVGB002670 VGB001270 VGB000657 VGD001631 VGD000023 VGD002481 VGD002481 VENGO000461 RVGC006741 VGD002014 VGC004009 VGB001890 VGC000329 VGD002354 VGB002145 VGB002146 VGD002454 VGD001963 RVGC009064 VGB002511 VGD000546 VGC004452 VGB002536 VGB002541 VGB004102 RVGB000819 VGC006355 VGB000708 VGC004331 VGC006688 VGD001029 VENGO000407 VGC007141 VGC007139 VGC007131 VGC007143 VGC007145 VGC007146 VGC007151 VGC007779 VGD000074 VGD000076 VGD000919 VGD000920 VGD000927 VGD000943 VGD001292 VGD001689 VGD001689 VGD001707 VGD001709 VGD000039 VGD000911 VGD001451 VGC000397 VGD000982 VGC000919 VGD000589 VGD001952 VGD002174 VGD002406 RVGC007195 VGC007288 VGC006815 VGC006980 VGB001919 VGB002299 VGB002367 VGB002370 VGB002371 VGC006629 VGD000571 VGB002373 RVGC000008 RVGC000009 VGC000658 VGC000658 VGC000661 VGC000662 VGC006111 VGC000506 VGD001461 VGB002560 VGB002563 VGC000747 VGC000955 VGC006092 VGD001966 VGD001968 VGD001969 VGD001968 VGD001990 VGD001992 VGD001994 VGD001995 VGD001966 VGD001969 VGC000188 VGC004370 VGC004375 VGC004435 VGD002366 VGD002416 VGB002591 RVGC004152 VENGO000541 VGC007042 VGD000048 VGD001839 VGD001840 RVGB000827 VGD001793 RVGD001646 VGC007133 VGC000561 RVGC004173 VGD001935 VGC007007 VGC007009 VGD000895 VGD000897 VGD000895 VGD001094 VGD001104 VGOVGC007882 VGOVGB002883 VGB000077 RVGB000025 VGC000095 VGC006618 VGB000506 RVGC007156 RVGC007161 VGD001725 VGB000659 RVGB000131 VGB001154 VGD001380 VGC006089 VENGO000233 VGC004464 VGC004495 VGD001485 VGD001488 RVGC006633 VGD001774 VGC006792 RVGB000163 RVGB000168 VGC006623 VGC006922 VGD000298 VGD002226 VGC004020 VGC007423 VGC006716 VENGO000048 VGC004105 VGC000668 VGD002243 VGB000975 VGC000595 VGC007393 VGB002372 VGC004261 VGC004466 VGC004470 VGC006137 VGC006946 VGD000404 VGC000894 VGD002396 VGD002214 VGC000695 VGD001597 VGD000767 VGB001018 VGC000971 VGC006417 VGC006674 VGC006761 VGC007604 VGD000776 VGD000780 VGD000784 VGD000785 VGD002127 VGD002125 VGD002127 VGB002399 VGD000284 VENG0000311 VENGO000325 VGC000768 VGC006063 VGC006068 VGD002215 VGD002253 VGD001225 VGD000269 VGD002482 VGC006691 RVGC007149 RVGC007182 VGD001049 VGD000303 VGD000305 VGD002318 VGB001995 VGD002465 VGD000768 VGD000770 VGC006760 VGD000544 VGD002297 VGC004006 VGD001105 VGC000334 VGC000331 RVGC006772 RVGC007181 VGC006128 VGC004387 VGB000663 VENGO000476 VGC000531 VGC000536 VGC000537 VGC000538 VGC000541 VGC000529 VGB000812 RVGB000200 RVGB002263 VENGO000492 VGD001648 VENGO000400 VENGO000405 VGC000096 VGC007112 RVGC000580 VGC006961 VGC000381 VGD002213 VGD001064 VGD001649 VGD002301 VGC006810 VGD000403 VGD001559 VGB001271 VGB001012 VGB001560 VGB002112 VGB002252 VGB002279 VGB002283 VGB002285 VGB002291 VGC000155 VGC000157 VGC000158 VGC000160 VGC000162 VGC004082 VGC000159 VGB002290 VGC000494 VENGO000705 VGD000318 VGD002401 VGB001149 VGB001153 VGB002056 VGB002057 VGB002058 VGB002059 VGD001561 VGD001567 VGD001583 VENGO000463 VENGO000464 VGC000512 VENGO000396 VGD000527 VGD001452 VGD002229 VGC007592 RVGC006685 VGB000698 VGD000556 VGD002356 VGD002207 VENGO000526 VENGO000565 VENGO000576 VGC004510 VGB002013 VGB002730 VGB002733 VGC004003 VGC004007 VGB000994 VGB002326 VGC004486 VGC004492 VGD000521 VGD001880 VGD002195 VGC000942 VGD001552 VGC004499 RVGB000829 RVGC000596 RVGB000197 VGC000743 VGC006327 RVGB000162 RVGB000176 RVGD000012 VGB000609 VGB002549 VGB002551 VGC006701 VGC007147 VGB000507 VGB000530 VGD000569 VGC000812 VGD000295 VGD001348 VGC004402 VGD001403 VGD001407 VGC000954 VGC007486 VGD000243 VGC000360 VGC004139 VGC004262 VGC004265 RVGC006836 VGC007057 VGC007058 RVGD000132 VGD001228 VGD001908 VGD001912 VGD001942 VGD001809 VGD001814 VGD001822 VGD001887 VGD001907 VGC000493 VGB001010 VGC007109 VGC006766 VGD000066 VGD002422 VGC006277 VGB000986 VGB002562 VGC000497 VGC000938 VGC000956 VGC007111 VGC007451 VGD000334 VGC000211 VGD000972 VGD002037 VGC000405 RVGB000471 VGD001618 VENGO000270 VGC000134 VGC000502 VGC000516 VGC004282 VGD000549 VGC006802 VGD001306 VGD001311 VGD001312 VGD002141 VGB002598 VGD001208 VGD001945 VGD002050 VGB000836 VGC004225 VGC004272 VGC007743 VGD002002 VGC006046 VGC006059 VGD002409 VGC000107 VGC000933 VGC007024 VGD000036 VGD000051 VGD001920 VGD001922 VGD002044 VGC007748 VENGO001008 VGB002951 RVGD000134 RVGB001722 VGC000215 VGC006066 VGD001546 VGD002246 VENGO000391 VGC000525 VGC004523 VGD002267 VGC004090 VGC007351 RVGD000139 VGC006049 VGD001642 VGC000434 VGD002274 VGC000508 VGC006219 VGD002248 VGD002078 VGD002385 VGC004304 VGC007440 VGD002296 VGC007144 VGD001229 VGC004243 VGC004259 VGC004415 VGC000079 VGC000126 VGC000214 VGC004137 VGC006670 VGC006823 VGC007103 VGC007106 VGC007108 VGD000918 VGD000991 VGD000548 VGC007436 VGC007494 VGC007496 VGC007507 VGC007507 VGC000330 VGC000474 VGC007534 VGC000330 VGC000474 VGC006087 VGD001386 VGC004482 VGC004484 VGC006102 VGD000547 VGD002043 VGD002066 VGC000276 VGC006107 VGB001931 VGC007125 VGD001877 RVGB000045 VGD000519 VGD000523 VGD000979 VGD000984 RVGC000065 RVGC000583 VGD001898 RVGC000042 RVGD000018 VGD001087 VGC006643 RVGB000478 RVGB002340 RVGD002199 RVGD002199 RVGC006108 VGD002060 VGD002071 VGD002060 VGD002071 VGB001164 VGB001165 VENGO000995 VGB002835 VGC000917 VGC000968 VGD000398 VGD002118 VGC000231 VGC000232 VGC000234 VGD000771 VGD000771 VGC004154 VGD002240 VGD002245 VGD002302 VGC000063 VGC000048 VGD002029 RVGB000042 RVGB000236 VGC006564 VGC007028 VGD002128 VGD001999 VGC007018 VGB000990 VGD002069 RVGB000134 VGC006090 VGD001063 VGD001555 VGD002418 VGC006500 VGD000059 RVGC006332 VGC000061 VGC006343 VGC007800 VGC006989 VGD001381 RVGC000579 VGC000676 VGD000913 VGB000562 VGC007128 VGC006051 VGC006053 VGC006054 VGC000182 VGB000650 VGB000681 VGB000682 VGC007065 VGD001620 VGD001668 VGB002545 VGD001927 VGC007428 VGC000153 VGC004145 VGD001634 VGD001813 VGC000645 VGC000428 VGC004242 VGC000757 RVGB000780 VGB002636 VGB002637 VGD002353 VGB002566 VGB002572 VGB002578 VGB002580 VGC007099 VGD000233 VGD000236 VGD000237 VGD000236 VGD001829 VGD002355 VGD002358 VGC000973 VGC007120 VGB002225 VGC000679 VGD001712 VGC007142 VGC000482 VGC007117 VGD000798 VGD002020 VGD000230 VGD001617 VGD001633 VGB000489 VGB002697 RVGC007202 VGD002284 VGC000402 VGC006040 VGC006042 VGC006104 VGD001113 RVGC000566 VGC000905 VGC000907 VGC000915 VGC000922 VGC004364 VGD001116 14616 VGB002115 VGD000864 VGD001411 VGB002522 VGC
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
GE Medical Systems, LLC
Reason for Recall:
The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Venue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (Model Numbers H45135VG, H45135VGW), R3 Surgery for China (Model Number H45113VGSU), R3 PeriOP for China (Model Number H45113VGPO), R3 Expert for China (Model Number H45113VGEX ), R3 Pro for China (Model Number H45113VGPR), R3 Performance for China, R3 Intervention for China, R3 Surgery for China (Model Number H45113VGSU) diagnostic ultrasound systems

Product Codes/Lot Numbers:

UDI-DI: Venue Go R2 system (GTIN 00840682138840); Venue Go R3 system (GTIN 00840682190503 / GTIN 00195278358011); Venue Go R4 system (GTIN 00195278516510 / GTIN 00195278516527); Venue Go R3 Surgery for China (GTIN 00195278547866) Venue Go R3 PeriOP for China (GTIN 00195278548344) Venue Go R3 Expert for China (GTIN 00195278547262) Venue Go R3 Pro for China (GTIN 00195278547828) Venue Go R3 Performance for China (GTIN 00195278547842) Venue Go R3 Intervention for China (GTIN 00195278548061) Venue Go R5 system (GTIN 00195278450319 / GTIN 00195278451774); System ID: VENGO000382 VGB002128 VGB003021 VGD000325 VGD002313 VGD000452 VGD000488 VGD002390 VGC006411 VGB002040 VGD000354 VGC000685 VGD001444 VGC006698 VGC007776 RVGC000034 RVGC000035 VGD002256 VGD002533 VGD002535 VGD002536 VGB001162 VGC007292 VGC007119 VGC006083 VGC006841 VGB001011 VGC000147 VGD001510 RVGC000559 RVGC000562 VGD000044 VGD001371 VGD001372 VGD001375 VGD001374 VGD001373 VGD001625 VGD001638 VGC000526 RVGD002184 VGD002513 VGB002239 VGD001705 VGD002307 VGC000746 VGC004373 VGC004378 VGC006707 RVGC007218 RVGC000591 VGD001899 RVGC004147 VGC004351 VGC007098 VGC007102 VGC007105 RVGC000024 RVGC004165 RVGD001551 RVGB000480 RVGB002675 VGC007113 VGD000091 RVGD001076 RVGB000467 VENGO000545 VGC007350 VGD001731 VGB002526 VGC000227 VGD001964 VGD002225 RVGC000303 RVGB000757 RVGB000797 RVGC000302 RVGC000554 VGC000799 RVGC006372 RVGC006771 VGC007879 VGC007880 VGD000524 RVGB000791 VGD001873 VGD002221 VGD002227 VGD002467 VGD000730 VGD001872 VGC007607 VGC007611 RVGC007204 VGC006763 VGD000731 VGD000843 RVGB000694 VGD000553 VGD000554 VGD000558 VGD001336 VGC000898 VGD000200 VGC006715 VGD002460 VGD002462 VGD002427 RVGB000032 RVGC000594 RVGC000617 VGD001738 VGD001747 RVGC006758 RVGD001055 VGB000796 VGB000901 RVGD001795 VGD001400 VGD001401 VGB002539 VGB002613 VGB001125 VGD001200 VGD001257 VGD001258 VGD001258 VGD001257 VGD001261 VGD001257 VGC000386 RVGD001071 VGD002275 VGD001701 VGD001704 VENGO000317 VGB002122 VGD001687 VGD001884 VGD001885 VGD001889 VGC006031 VGC006036 VGC004366 VGC004521 RVGD001797 VGB001150 VGB000766 VGC000947 VGC000948 VGC006790 VGD000142 VGD001602 VGD000814 VGC006847 VGD001463 VGD001475 VGD001584 VGD002463 VGB000701 VGC000109 VGD001673 VGB000737 VGC000379 VGC004001 VGD001058 VGD001923 VGD001982 VGD001984 VGC007456 VGC007643 VGC007644 VGD002283 VGD002285 VGB002527 VGB002531 VGD001569 VGD002397 VGD002506 VGB000720 VGB000722 VGB000723 VGD001090 VGB002102 VGC000435 VGC000951 VGC004160 VGC007136 VGB002574 VGC007639 VGD002403 VGD002407 VGD002411 RVGB000122 VGC000866 VGC006812 VGC006814 VGC000082 VGC000084 VGC000086 VGC000085 VGC006402 VGC006406 VGC006652 VGC007049 VGC007053 VGC006409 VGD000035 VGD002484 VGD001037 VGD001150 VGD001900 VGD002237 VGB000165 VGC000943 RVGB000184 VENGO000700 RVGC000185 RVGC007163 VGC000321 RVGB000265 VGB000674 RVGC006341 VGD000399 VGD002328 VGB001415 RVGC000488 14540 VGB001140 VGB001916 VGB002688 RVGB000226 RVGB000472 VGD000932 RVGD001065 RVGD001427 VGC000671 VGC006086 VGC000156 VGC006113 VGC006127 RVGB000425 RVGB000171 VGC006710 VGD000603 VGD002018 VGC000879 RVGB000073 VGD002487 VGD002375 VGD000969 VGD000970 VGC007104 VGB001132 14617 14280 VGD002377 VENGO000703 VGB000710 VGB000707 VGB002548 VGD001005 VGD002230 VGD001573 VGB001934 VGB002489 VGB002565 VGC006105 VGD001043 VGD001482 VGC000407 VGD002126 VGB001009 VGD002509 VGD000914 VGC006958 VGD000350 VGD000353 VGD002170 VGC000099 VGC000115 VGC000099 VGC000146 VGD001298 VGD001301 VGB002073 VGC000269 VGD001275 RVGB000420 VGD001933 VGB001168 VGD002234 VGC006278 VGC007636 VGD001719 VGC006661 VGB002734 VGC006294 VGC006296 VGB002469 VGD002189 VGC000718 VGD002362 VGC000521 VGB000466 VGC000324 VGC000319 VGC000322 VGC000322 VGC000324 VGC000327 RVGC000019 VGB000998 VGC000278 VGD000138 VGC006956 VGD001161 VGD001163 RVGB000437 VGB000911 VGB001876 VGD002280 RVGC000570 VGD000513 VGC006534 VGD000902 RVGC000041 VGC006995 VGD001390 RVGB002663 RVGB002674 VGC006562 VGC006563 VGC006567 VGC006569 VGC006570 VGC006574 VGD002329 RVGC000040 VGB002970 VGD002036 RVGB000116 RVGB000160 RVGB000832 VGD002466 RVGD000137 VGB003006 VGD000307 VGD002178 VGB004063 VGC004102 VGC004110 VGD001382 VGD001749 VGD001656 VGC006697 VGC007044 VGC006837 VGD001949 VGC000487 VGD002003 VGD002062 VGD000753 VGD000761 VGC006816 VGB002524 VGC000279 VGC006665 VGD002056 VGD001051 VGD001052 VGD001639 VGB000684 VENGO000686 VGB000688 VGC007135 VGB000712 VGD002516 VGB000865 VGC000130 VGC000480 VGC000703 VGC000704 VGC007107 VGD000536 VGB002772 VGC006340 VGC000346 VGD002370 VGC006702 VGD002017 RVGB000066 RVGB000793 RVGB002244 VGD001517 VGD001526 VGB001268 VGB002573 VGC000323 VGC000364 VGC000741 VGC000750 VGC000753 VGC006443 RVGC007168 VGC007421 VGC007422 VGC004000 VGC006634 VGC007326 VGD002350 VGC006672 VGD000306 VGC004205 VGC004206 VGD000176 VGD001514 VGB004017 VGB004019 VGC000081 VGC000083 VGC000090 VGC000089 VGC000092 VGC000091 VGC000087 VGC000093 VGC007032 VGC007037 VGC007032 VGC007053 VGC007037 VGC007032 VGC007048 VGC007051 VGC007052 VGD001713 VGD001718 VGC000086 VGD001897 VGC007741 VGD001194 VGD002172 VGC007455 VGD002346 VGC007023 RVGB000383 VGD000541 VGD000546XX VGD000955 VGD001377 VGD002233 VGC000094 VGD000297 RVGB000808 VGD000777 RVGD001755 VGB002547 VGD001023 VGC006506 VGC006508 VGD000938 RVGC006368 RVGC006370 VGD000901 VGB000992 VGB001020 VGB001246 VGB001999 VGB002696 VGB002281 VGB002554 VGB002555 VGB002556 VGB002557 VGB002559 VGB002561 VGC000911 VGC000970 VGD002132 VGD002134 VGD002273 RVGC006682 VGC000680 VGC006650 VGC006669 VGD000424 VGD000445 VGB000666 VGB002462 RVGB002692 VGB002781 VGC004144 VGD001694 VGD001695 VGB002098 RVGD001798 VGD002340 VGB000699 VGB002117 VGC004161 VGC004263 VGC004266 VGC007387 VGD002082 VGD002181 VGC006093 VGD000235 VGD001217 VGB000756 VGB002870 RVGC000045 VGC004458 VGD002041 VGD002051 VGD002041 VGD000247 RVGD001450 RVGB000470 VGD001645 VGD001847 VGC000760 VGC000763 VGC000785 VGC007041 VGD000391 VGC007400 VGC006723 VGD000040 RVGC000168 VGC004038 VGD000045 RVGB000427 RVGD001537 VGB002100 VGB002104 VGB002106 VGB002108 VGC006103 VGC006678 VGC007039 VGB001414 VGD002205 VGD002279 VGD002315 VGC007060 VGC004141 VGC000072 VENGO000293 VGD000310 VGD000317 VGD000310 VGC000461 VGC006070 VGC007596 VGD000985 VGC004078 RVGC000585 RVGC004182 VGD002473 RVGC000001 RVGC006764 VGB000655 VGD000343 VGB000649 VGD001830 VGC000936 VGC007431 VGC000004 VGC004463 VGC004512 VGC006575 VGD000538 VGD000555 VGC000004 VGC000098 VGC000097 VGC000106 VGC000105 VGC004513 VGC004514 VGC007064 VGC007069 VGC007073 VGD000234 VGB001004 VGB002039 VGC006675 VGD000342 VGB002170 VGD001710 VGC004497 VGC006062 VGD001857 VGD002033 VGC000742 VGC000756 VGC000903 VGC004116 VGC004118 VGC004506 VGC009011 VGD002337 VGD002338 VGC000143 RVGB001984 VGC006052 VGB002004 VGC000688 VGC000701 VGC000720 VGC007210 VGD002300 VGC000670 VGD000925 VGB001015 RVGB000252 VGC000417 VGB000692 VGB001480 VGB001491 VGB002016 VGB002052 VGB002123 VGC000533 VGC000921 VGD000715 VGD001031 VGD001337 VGD001337 VGD001337 VGD001337 VGD001611 VGD001953 VGD001955 RVGB000354 VGC000530 VGC000200 VGC000219 VGC000213 VGC000219 VGC000219 VGC000219 VGC000219 VGC004107 VGC004490 VGC006338 VGD000790 VGD001160 VGD001782 VGB000693 RVGC006684 VGC007562 VGC006759 VGD002209 VGD000884 VGD001481 VGC000602 VGD001457 VGD001468 VGB002311 VGC000481 VGC000489 VGC006041 VGC006043 VGC006094 VGC006124 VGC006126 VGC006913 VGC007631 VGC007635 VGD002402 VGD002405 VGD002408 VGD001118 VGD002376 VGC000645 VGC007671 VGC006794 VGC006169 VGC004086 RVGC004150 VGC004403 VGD000155 VGD001737 VGD001986 VGD002005 VGD002016 VGD002025 VGD002030 VGD002034 VENGO000403 VGB000503 VGB000504 VGB000510 VGC004516 VGC006378 VGC007027 VGD002507 VENGO000339 VGB002157 VGB000805 VGC007707 VGD000917 VGB000805 VGB000671 VGB000680 VGB000685 VGC000138 VGC000140 VGC000142 VGC000151 VGC000152 VGC004518 VGC004518 VGC000103 VGD002323 VGD002326 VGD002327 VGD002392 VGB001982 VGC006720 VGD001486 VGB001917 VGC007733 VGC007233 VGD000057 RVGB000835 VGD000302 RVGB000199 VENGO000385 VGC000077 VGC000759 VGC000762 VGC006030 VGC006035 VGD000926 VGB002643 VGD002086 VGD002086 VGC004479 VGC004461 VGC004027 VENGO000459 VGB002383 VGD002013 VGD001388 VGD001918 VGC004030 RVGC007198 VGD001085 VGD001266 VGD001278 VGB001269 VGC000403 VGC004488 VGC004493 VGB002029 VGC000385 VGC006029 VGC006034 VGC006444 VGC006450 VGC006967 VGC006988 VGC000485 VGC007781 VGB002129 VGB002130 VGC000144 VGC000934 VGC000935 VGC004026 VGC004508 VGC004526 VGC007063 VGC000953 VGC004515 VGC006071 VGB000667 VGC004509 VGC000950 VGB002101 VGC000912 VGC000932 VGC000965 VGC007450 VGD002414 VGD000935 VGD002434 VGD002450 VGB002431 VGC004197 VGC004199 VGC004201 VGC004203 VGB000873 VGB001002 VGB001994 VGD001004 VGB002520 VGC000421 VGC000413 VGD002530 VGD002538 VGB002030 VGD000832 VGD001197 VGD001378 VGB000673 VGD002265 VGC000522 VGC006243 VGC006252 VGC000897 VGB002140 VGB002141 VGB002492 VGB002512 VGB002626 VGB002641 VGC000073 VGC000074 VGC000074 VGC000478 VGC000479 VGC000539 VGC000895 VGC004143 VGC006084 VGC006097 VGC006139 VGC007075 VGC007594 VGC007601 VGC007637 VGD000797 VGD001459 VGD001460 VGD001540 VGD001823 VGD002341 VGD002511 VGC004034 VGC006140 VGC007002 VGB002532 VGC004023 VGC004517 RVGC007169 VGC000463 RVGB000402 VGC000924 RVGC006848 RVGC007194 VGD000590 VGC004088 VGC007618 RVGC000066 RVGD000338 RVGD001066 RVGD001067 RVGD001068 RVGD001070 RVGD001466 RVGD001506 RVGD001550 RVGD001622 RVGD001624 RVGD001789 RVGD001801 VGD000341 VGD000362 RVGC000012 VGD000311 VGD001032 VGD001034 VGD001958 VGD001483 VGB002103 VGC004559 VGC007835 VGD002011 VGD002021 VGB001104 VGC000929 VGD002413 VGC007630 VGC000392 VGD001493 VGB001174 VGC000532 VGB002125 VGB002533 VGC006419 VGC007395 VGD001120 VGD001121 VGD001151 VGD001152 VGB002267 VENGO000493 RVGC006687 VGC004066 RVGC007160 VGC006593 RVGB002253 RVGC004124 3061702 VENGO000444 RVGC006608 RVGB000130 VGC007121 VGB002025 RVGC000592 VGD002260 VGB001183 VGB001187 VGB001189 VGB002706 VGC000916 VGC006653 VGC006664 VGD001667 VGC000515 VGD000345 VGC006818 VGB000813 VENGO000528 VGB000842 VGC000484 VGC000890 VGC004481 VGD000330 VGD002004 VGD002503 VGD002508 VGD000336 VGC000958 VGD001836 VGC004248 VGC004255 RVGC006352 RVGC006774 RVGC006775 RVGC007154 RVGC007155 RVGC007166 RVGC007192 VGD000111 VGD000113 VGD000116 VGD001429 VGD001518 VGD002136 VGD002254 VENGO000687 VENGO000695 VGD001740 VGD001741 VGD002075 VGC004467 VGC004487 VGC007725 VGD001089 VGC000308 VGC000312 VGC000314 VGD000229 RVGD000081 VGD001515 3137004 VGD001686 VGC007599 VGB000390 VGB001021 RVGC000044 RVGC000046 RVGC000047 RVGC000062 RVGC000064 RVGC000248 RVGC000043 VGD000765 VGC007006 VGD001564 VENGO000422 VENGOB000327 VGB002082 VGB002378 VGB002391 VGB002681 VGB002715 VGC004472 VGC004473 VGD000998 VGD001008 VGD001008 VGD001396 VGD001541 VGD002147 VGD002168 VGD001226 VGD000899 RVGB000047 VENGO000439 VGD000117 VGD001729 VENGO000245 RVGC000015 VGC007031 VGC007035 VGC007036 RVGC006769 VGC000289 VGC000382 VGC004480 VGC004485 VGD001812 VGC007641 VGD001643 VGC004094 VGC004560 VGD000054 VGD000055 VGD000049 VGB002673 VGD002053 VGB000060 VGB000658 RVGC004135 RVGC006765 VENGO000394 VGD001972 VGD001478 VGD001479 VGD000272 VGB002174 RVGB000369 VGB001184 VGB001191 VGB001901 RVGB002261 RVGB002679 VGC000110 RVGC000616 RVGC006545 RVGD001387 VGD001806 VGD002336 VGD001967 RVGB002788 VGD002475 RVGC007157 VGD002173 VGD002175 VENGO000389 VGB001899 VGC000436 VGC000080 RVGB000765 VGC007056 VGD001576 VGC006996 VGC006999 VGB001903 VGC006358 VGC006830 VGD000038 VGD000115 VGD000149 VGD001489 VGD001501 VGD001502 VGD001505 VGD001508 VGB000817 VGB001568 VGD002259 VGD000344 VGB002568 VGC004507 VGC004520 VGD001030 VGB001897 RVGB000768 RVGC000576 VGD000561 VGB002523 VGC007443 RVGB002097 VGC000408 VGC000416 VGC006044 VGD000539 VGD000545 VGD002187 VGD000324 VGC000154 VGC000636 RVGB000743 VGD001756 VGD002135 VGC004064 VGC006641 VGD000332 VGD000333 VGD000421 VGD000436 VGD000443 VGD001934 VGD002295 VGD002330 VGD002332 VGD002334 VGC007271 VGC007275 VGC007277 VGC007271 VGC006683 VGC006727 VGD000359 VGB002525 VGB002731 VGC007663 VGD002373 VGD000741 VGC006783 VGD001536 VGC004238 VGD001745 VGB000406 VGC000277 RVGC007164 VGC007389 VGB002120 VGB002133 VGB002138 RVGB000173 VGD001808 VGD001948 VGD001608 VGD001610 VGD001613 VGB002486 VGC000376 VGD001079 VGD000752 VGD000754 VGC006668 VGC000078 RVGB002271 1000000926 VGD000997 VGD002235 VGC006696 VGC006700 VGC006704 VGC006957 1000003449 VGC006122 VGB002543 VGC006074 VGD001978 VGC000171 VGC000171 VGB003051 VGC007348 VGC007363 VGB002766 VGD000245 VGD001050 VGC000055 VGB004004 VGB004005 VGB001870 RVGB000809 VGC006114 VGD000232 RVGC000030 RVGD001666 VGC006260 VGC007132 RVGB000150 RVGC000587 VENGO000076LO VENGO0002407LO VENGO000241LO VENGO001981LO VENGO001992LO VENGO000233LO VENGO000720LO VENGO004236LO VENGO006099LO VENGO007263LO VENGO000559LO VENGO000907LO VENGO001750LO VENGO002398LO VENGO000981LO VENGO002054LO VENGO000085LO VENGO002100LO VENGO004196LO VENGO000084LO VGD001708 VGD002252 VGB000807 VGD001981 VGC000519 VGD000892 VGD002065 RVGB000818 RVGD000017 VGC006525 VGC006528 VGD000608 VGD000721 VGD000722 VGD001080 VGD001084 VGD002400 VGC007597 VGD001653 VGD002144 VGC000906 VGC004494 VGC004496 VGC006321 VGD002268 VGC004264 VGC004268 VGC007775 VGD000610 VGB002066 VGC004350 VGC000558 VGC006231 VGD001285 RVGB002260 VGC006908 VGC007114 VGD001419 RVGB000170 VGB000855 VGB002331 VGB002332 VGB002335 VGC000909 VGC000967 VGC000969 VGD001240 VGB002365 VGC004037 RVGC0000521 RVGC007213 VGD000058 VENGO000290 RVGB000177 VGB002208 VGD002048 VGC007533 RVGB002668 VGD002133 VGC007001 RVGC000586 RVGC004114 RVGC004136 RVGC004146 VENGO000381 VGD001547 VGC000491 VGC007110 VGB002118 VGB002121 VGB001885 RVGC000228 VGD001316 RVGB002273 VGC000520 RVGB002249 RVGC000028 VGC000184 VGC000187 VGD000355 VGD002420 VGD000606 VGD000607 VGD002179 VGC000358 VGD001714 RVGB002270 RVGC000226 RVGC000238 VGC006785 RVGB000178 RVGB000186 VGC000766 VGC006531 VGC006537 RVGB000108 RVGB000167 RVGB000029 RVGB000125 RVGB000137 RVGB000426 RVGB000144 VGB001288 RVGC004134 VGD002076 VGD002130 VGC004005 RVGC000249 VGD001669 VGD002289 VGD002299 VENGO000642 VGC007632 RVGB000198 RVGB000828 RVGB002258 VGD001635 VGB000871 VGB001926 VENGO000278 VENGO000281 VENGO000286 VENGO000289 VENGO000296 RVGB000117 VGD000923 VGD001392 RVGB000138 RVGB000142 RVGC006777 VGC000167 RVGB000143 VGD001439 VGD001535 VGC006147 VGC006148 VGC006149 VGC006150 VGD000540 VGD001413 VGD002476 VGD001243 VGB000732 VGB000754 RVGC000013 VGC007138 RVGC007179 RVGB002256 RVGB002275 RVGB002661 RVGB002662 VGC000281 VGC000281 VGC000422 VGC000500 VGC000499 VGC006911 VGD000473 VGD002412 VGD002421 VGC000499 VGC006337 VGB001262 RVGD000013 RVGD000016 RVGD000019 RVGC006826 VGC000764 VGC000771 VGC000774 VGC006631 VGD002384 RVGB002664 RVGC007180 VGD002180 VGC006067 VGB000647 VGC000161 VGC000161 VGD002514 VGD000072 VGB000856 VGB000868 VGB001001 RVGD001410 VGD001961 VGD001965 RVGC004130 VGD001854 VGD001041 VGD001255 VGD002393 VGC000772 VGC006660 VGC006660 VGD001297 VGD002023 VGD000885 VGC000937 VGC000393 VGC000523 RVGC006840 VGD000724 VGD000043 VGD000061 VGC000190 RVGD000130 VGD000557 VGC006538 VGD000289 VGD000394 RVGB000462 RVGB000468 VGD002430 RVGB000128 RVGB000133 RVGB000145 RVGB000152 RVGB000203 RVGB000231 RVGC006828 VGD002203 RVGB000159 VGD000766 VGC000957 RVGB000069 VGB000772 VGB000774 VGD000592 VGB001006 VGB001006 VGB002784 VGB002288 VGB001176 VGC007076 VGC007082 VGC007080 VGC007079 VGC000893 VGC000930 VGD001664 RVGB000039 VGD000047 RVGD001072 VGD001319 VGD001318 VGD002290 RVGD002537 VGD002198 VGD001696 RVGB000833 VGC007071 VGB002671 VGC006192 VGC006202 VGD000075 VGD000078 VGD000079 VGD000082 VGD000886 RVGC007186 VGC006123 VGB000755 VGD001578 RVGD000444 VGC000601 VGC007736 VGC006050 VGC006064 VGC006080 VGD000060 VGC006265 VGD001868 RVGB000482 RVGB000488 VGD000908 VGD000915 VGD000922 VGD000928 VGD000929 RVGD001527 VGD001630 VGB004012 VGC007480 VGD000062 VGC004379 VGC007068 VGD002363 RVGC000571 VGD002019 VGC000163 VGC000896 VGC007541 VGC006850 VGD002143 VGD002312 VGC000390 VGB002425 VGC000258 VGC007115 VGC000765 VGD000357 RVGC004140 VGC006861 VGC006863 VGD001790 VGB002310 VGC004455 VGC007025 VGC000945 VGC006673 VGB001197 VGB001951 VGB002139 VGB002538 VGD000986 VENGO000646 RVGB002124 VGD001671 RVGB002657 VGD002009 VGB002546 VGB000999 RVGD000241 VGB000690 RVGC006548 VGD000204 VGC007434 VGB002224 VGB002230 VGD000323 VGC000889 VGC000891 VGC006676 VGC006679 VGC007600 VGC000514 VGC006693 RVGD001406 RVGB000428 VGD002357 VGB001998 VGB002041 VGC000401 VGD002495 VGD002182 RVGB000080 RVGB000110 VGB002597 VGC006566 VGD001511 RVGD000153 VGC000116 VGC000117 VGC000119 VGD001484 RVGB000798 RVGB002220 VGD001628 RVGC000020 VGD001647 VGC000702 VGD000221 VENGO000239 VGD000150 VGD002154 RVGB000135 RVGB000277 RVGB000831 VGB002250 VGC000952 VGC004055 RVGB000230 VGC006383 RVGC000573 RVGB002450 VGC000976 VGC004010 VGC006796 VGC006799 VGC006912 VGC006621 VGC006699 VGD002417 VGC000053 VGC000057 VGC000441 VGC007500 VGC007502 VGC007503 VGC007591 VGD000820 VGD001585 VGD002468 VGD002470 VGD002472 VENGO000536 VGB000672 RVGB000799 VGC000173 VGC000352 VGD001748 VGD001954 VGC006807 RVGC000501 VGD001286 VGD001519 VGB002698 VGC000527 VGC006335 VGC007546 VGD001881 VGD000143 RVGC007167 VGD001938 VGD002345 RVGD001440 VENGO000434 RVGC000229 RVGC000307 RVGD001391 VGD001492 VGC004142 VGC007623 VGD002429 RVGD000015 RVGD000135 RVGD001432 RVGB000773 VGC006659 VGD002152 VGC007409 VGC000207 VGC007837 VGD001902 VGD002064 VGB003022 VGD002499 VGD002497 VGD002498 VGD002490 VGC007805 RVGB000263 RVGC000584 VGD000304 VGD000304 VGD000320 VGC007061 VGB000664 VGB000981 VGB000982 VGB001017 VGB001019 VGC004498 VGC006032 RVGC000425 VGC004008 RVGC000578 VGD001204 RVGC000310 RVGB000164 VGD002293 VGD002415 VGB001564 VGC007806 VGC006401 RVGC000006 RVGC000016 RVGC000026 RVGC000021 RVGC000025 RVGC000027 RVGC000029 RVGB000779 VGC007134 RVGB000096 RVGB000124 RVGB000126 RVGB000136 VGB001932 VGB000691 VGD000063 RVGC004133 VGC009042 VGC009044 VGC009045 VGB000702 VGC006501 VGC007783 VGD001256 VGD001264 VGC000885 VENGO000811 VGD001919 VGD001921 VGD001925 VGD001929 VGD001930 VGD001951 VGC006038 VGD001996 VGD000789 VGC006624 VGD001343 VGD001344 VGD001345 VGD001346 VGD001347 VGD001349 VGD001352 VENGO000324 VGB002294 VGB000834 VGB000866 VGD001179 VGD001187 VGD001190 VGD001191 VGD002081 VGD002091 VGD001189 RVGC000070 VGC007687 VGC007687 VENGO000717 VENGO000718 VENGO000719 VENGO000721 VGC000137 VENGO000282 VENGO000283 VENGO000287 VENGO000288 VGB002356 VGD002459 VGD002433 VGC000177 VGC000178 VGC000177 VGC004002 VGC004002 VGD000817 VGD000819 VGC007126 VGD000738 VGD000727 VGB002725 VGC007638 RVGC006344 VGD002320 VGD002322 VGD002494 VGC004019 VGD001730 VGC006987 VGC006805 VGD001366 VGD002319 RVGC000398 VGB002362 VGB002519 VGC004348 VGC006620 VGC006671 VGC006817 VGD000896 VGD001716 VGD001891 VGB002038 VGB002042 VGD002032 VGB002303 VGB002303 VGB000713 VGB000753 RVGC000577 VGD002387 VGD002389 VGD001828 VGD002202 VGC009058 RVGC000023 VGD002007 VGC007429 VGD001332 VGD001783 VGC006039 VGD001294 VGC007211 VENGO000393 VENGO000404 RVGB000075 VENGO000386 VGC000483 VGC006511 VGC006757 VGD001454 VGD002059 VGC006435 RVGB000049 RVGB001900 RVGD001528 RVGB000031 VGC006512 VGB000689 VGB000714 VGC000715 VGC007660 VGC007670 VGD001906 VGD001909 VGD001910 VGD001914 VGC007669 VGB001441 VGB001570 RVGB000424 VGC009030 RVGC000049 RVGD000014 RVGD000124 VGB000854 VGC000315 VGC000925 VENGO000285 VGB002773 VGC007000 VGC007059 VGC000974 VGD002453 VGB002172 VGC000412 VGC000513 VGC000513 VGC000528 VGC000902 VGD001487 VGC000949 RVGB002272 RVGB002262 RVGC000556 VGC007122 RVGB000824 VGC007481 VGB002001 RVGB000284 RVGC006680 VGC006142 VGD001011 VGD000052 VGD002523 VGD000596 VGB000677 VGB000678 VGB002521 VGC000767 VGC004183 VGC004344 VGC000108 VGC007721 VGB002158 VGC004251 VGC004257 VGD002291 VGD002304 VGD002305 RVGB000786 RVGB000821 VGB001986 VGC009013 VGC004438 VGD002368 VGD001467 VGB001962 VGB000683 VGD001183 VGD002047 RVGC000575 VGC000325 VGC000326 VGD002258 VGC006129 VGC007452 VGC006712 VGC006138 VGC004080 VGC006106 VGC006136 VGC006146 VGB000665 VGB000670 VGD000046 VGC000265 RVGD001420 VGB000668 VENGO000733 VGB002359 VGB002361 VGD001621 RVGD001600 RVGB000041 VGC006786 VGD002491 RVGB000792 RVGB002269 VGC000972 VGD000199 VGD000393 VGC000420 VGC006918 VGD001874 VGD001875 VGC000170 VGC006824 VGC006825 VGD001959 VGC007442 VGC007810 VGB001991 VGB001997 VGB002490 VGC000113 VGC000357 VGC000357 VGC007123 VGD001047 VGB001192 VGD001640 VGC000328 VGC007077 VGC007077 VGC007100 VGC007100 VGD001650 VENGO000295 RVGB002264 VGC004245 VGC004249 VGC004525 VGC004527 VGD000909 VGD000910 VENGOB000337 VGB001166 RVGC007158 VGD001042 RVGB000156 RVGB000421 VGB000452 VGB000456 VGB000474 VGC000359 VGC004016 VGC007070 VGC007140 VGC007330 VGC007338 VGC007339 VGC007340 VGC007745 VGC007764 VGC007769 VGD000944 VGD001678 VGD001980 VGC007074 VGC007081 VGC007336 VGD001173 VGD001198 RVGB000123 RVGB000193 VGC000939 RVGD001056 RVGD001525 RVGD001543 VGD000801 VGD001581 VGD001768 VGD001353 VGD001016 RVGC000036 VGC006045 VGC006058 VGC006057 VGC006058 VGD001123 VGD001534 VGD002114 VGD001751 VGD002115 VGD002108 VGD002109 VGD002115 VGD002109 VGD002106 VGD002114 VGD002115 VGD000846 VGD001832 VGC000975 VGC004180 VGC006834 VGD000583 VGD000584 VGD000888 VGD000903 VGD000988 VGD001835 VGD001837 VGB001118 VGD000795 VGD001572 VGC006061 VGD001876 RVGC000071 VGC006557 VGC006565 VGC006568 VGC006584 VGC009006 VGC009005 VGC009008 VGC009004 VGD002084 VGC000928 VGC000923 VGC000910 VGC000918 VGC007150 VGC007152 VGD000159 VGD000882 VGD000883 VGD000937 VGD001102 VGD001108 VGD001815 VGD001817 VGD001818 VGD001819 VGD001826 VGD001826 VGD001819 VGD001831 VGC007072 VGD002372 VGD002378 VGC004542 VGC006477 VGC006477 VGC006481 VGC006498 VGD000879 VGD000880 VGD000881 VGD000912 VGC004541 VGD000351 VGC004408 VGD000591 VGC006130 VGC006141 VGC006448 VGC006452 VGC006452 VGC006791 VGC006795 VGD000126 VGD000127 VGD000126 VGD000818 VGD000821 VGD000859 VGD000861 VGD000862 VGD000865 VGD000990 VGD001081 VGD001101 VGD001125 VGD001339 VGD001341 VGD001341 VGD002212 VGD002218 VGD002219 VGD002212 VGD002232 VGD002218 VGD002232 VGD002218 VGB001902 RVGC000581 VGC006654 RVGC000305 VGB001005 VGB002142 VGB002147 VGC004454 VGC004511 VGC006717 VGD001878 RVGC006835 RVGC006369 RVGC006677 VGC006651 VGD000887 VGD000889 VGD000889 VGD001003 VGD001279 VGD001283 VGD002211 VGD002216 VGD002211 VGD002211 VGC007765 VGD000594 VGC006657 VGD001376 VGD001469 VGD001470 VGD002303 VGD002324 VGD001851 VGC007272 VGC007814 VGD001252 VGC006947 VGD002122 VGD002123 RVGB000036 RVGB000046 VGB000654 VGC007127 VGC007231 VGD000601 VGD000609 VGD001549 VGD001632 VGD001800 VGB002530 VGD000890 VGD001024 VGD001317 VGD001335 VGD001869 RVGB000129 VGC007603 VGD000595 VGD000597 VGD000598 VGD000599 VGD000602 VGD001883 VGD000597 VGD002171 RVGB000155 VGD000600 VGC006120 VGC006125 VGC006133 VGC006135 VGD002145 VGD000957 VGD000960 VGD000961 VGD000964 VGB001022 VGC006096 VGC006109 VGC006112 VGC006131 VGC006360 VGC006361 VGB002534 VGC006527 VGD001082 VGD000292 VGD000314 VGD001009 VGD001015 VGD001098 VGD001099 VGD001103 VGD001106 VGD001110 VGD001111 VGD001119 VGD001122 VGD001271 VGD001288 VGD001289 VGD001293 VGD001342 VGD002022 VGD002057 VGD002093 VGD002094 VGD002096 VGD002097 VGD002098 VGD002101 VGD002102 VGD002103 VGD002104 VGD001879 VGD001882 VGD001888 VGD001890 VGD001894 RVGB000146 RVGB000147 RVGB000148 RVGB000166 RVGB000172 RVGB000151 RVGC006345 VGD001155 VGD001615 RVGB000481 RVGC000022 VGD001464 VGC000744 VGC000745 VGC000748 VGC000749 VGC000752 VGC000754 VGC007449 VGD002185 VGD001557 VGC007619 VGC007629 VENGO000297 RVGC006773 RVGC007212 VENGO000266 RVGC000418 RVGC006689 RVGD001195 RVGC000039 VGD000329 RVGB000140 RVGC007224 VGD002015 RVGB002665 RVGB002669 VGC000913 VGC006709 VGC007029 VGC007030 VGC007033 VGC007034 VGC007038 VGC007040 VGC007045 VGD000924 VGD001046 VGD001054 VGD001046 VGD001231 VGD001296 VGD001303 VGD001310 VGD001307 VGD001304 VGD001313 RVGB002268 VGC004462 VGD002479 VGD000231 VGD000231 VGD001892 VGD002186 VGD002190 VGD000965 RVGB000027 RVGB000054 RVGB000056 RVGB000059 RVGB000062 VGB000067 RVGB000154 RVGB000183 RVGB000229 VGB000514 VGB000661 RVGB000750 RVGC000051 RVGC000056 RVGC000059 RVGD000348 RVGD001078 RVGD001178 RVGD001399 RVGD001415 RVGD001425 RVGD001430 RVGD001431 RVGD001438 RVGD001441 RVGD001442 RVGD001521 RVGD001522 RVGD001532 RVGD001788 RVGD001792 RVGD001794 RVGD001799 RVGD001803 RVGD001810 RVGD002192 RVGD002194 RVGB000157 VGB000715 VENGO000770 RVGB000051 VGD000782 VGB000679 RVGB000068 VGC007768 RVGD001520 VGD001636 VGC000509 VGC000770 RVGC006829 VGD002382 VGD002383 VGD002382 RVGB000781 RVGB000781 RVGC000242 VGC004386 VGC006326 VGB000566 VGB000675 VGB001950 RVGB000052 RVGB000153 VGB002131 VGB002137 RVGC000593 RVGC000598 RVGC000611 VGC000946 VGC006662 RVGC006770 VGC004369 VGD000833 VGD000528 RVGC000239 VGB000711 RVGC000607 VGD002269 VGD001253 VGD002239 RVGC004121 VGC007062 VGD002469 VENGO000345 RVGC006968 VGC006299 RVGC0000541 VGC006069 VGC006072 VGC006073 VGC006085 VGC006076 VGC006078 VGC006079 VGC006076 VGC006075 VGC006098 VGC006100 VGC006101 VGC006110 VGC006639 VGC007129 VGC007130 VGC007521 VGC007538 VGC007539 VGC007540 VGD000313 VGD000315 VGD000319 VGD000905 VGD000418 RVGD001074 RVGC000017 VGB000104 VGC007197 VGC006821 RVGB000175 RVGB000763 RVGC000003 VGD000904 VGD002210 VENGO000508 RVGB000376 VGC000220 RVGB000127 VGB001185 VGB002360 VGC007228 VGC007388 RVGD001732 RVGC000608 RVGC004115 VGD002505 VGC004185 RVGB000174 RVGB000761 RVGB000761 VENGO000262LO VENGO000446 RVGC007172 VGD002471 VGC006572 VGC006718 VGC007278 VGC007808 VGC007809 VGB002062 VGB002064 RVGD001416 VGC006648 VGC004553 VGD000321 VGB002544 VGB002583 VGB002672 VGB002960 VGC004014 VGD001320 VGD002451 VGB002126 VGB002854 VGC006077 VGD001735 VGD001833 VENGO000228 VGD000823 VGD001654 VGD001674 VGD001676 VGD001677 VGD001781 VGC006820 RVGD001421 RVGB000457 VGC000400 RVGB000822 VGD002072 VGB001877 VGC007374 VGD001616 VGC000507 VGC006729 VGD000831 VGD001471 RVGD001796 VGC000920 VGB002542 VGB001113 VGC000268 RVGB000120 VGD001370 RVGB000826 VGB001178 VGD001991 VENGO000380 RVGC000564 VGD001988 RVGC006365 RVGB000196 VGD001273 VGD001603 VGB002055 VGD000331 VGD000337 VGD001985 VGD001993 RVGC006367 RVGC007165 RVGC007171 VGD001230 VGD001232 VGD001233 VGD001235 VGD001239 VGD001233 VGD001244 VGD002519 RVGC007184 VGD002464 VGD002517 VGC006509 VGD000449 VGD000587 VGD000588 VGD000605 RVGD001791 VGD001895 VGC006091 VGC004240 VGC004241 VGC004156 RVGB000070 RVGB000072 RVGB002670 VGB001270 VGB000657 VGD001631 VGD000023 VGD002481 VGD002481 VENGO000461 RVGC006741 VGD002014 VGC004009 VGB001890 VGC000329 VGD002354 VGB002145 VGB002146 VGD002454 VGD001963 RVGC009064 VGB002511 VGD000546 VGC004452 VGB002536 VGB002541 VGB004102 RVGB000819 VGC006355 VGB000708 VGC004331 VGC006688 VGD001029 VENGO000407 VGC007141 VGC007139 VGC007131 VGC007143 VGC007145 VGC007146 VGC007151 VGC007779 VGD000074 VGD000076 VGD000919 VGD000920 VGD000927 VGD000943 VGD001292 VGD001689 VGD001689 VGD001707 VGD001709 VGD000039 VGD000911 VGD001451 VGC000397 VGD000982 VGC000919 VGD000589 VGD001952 VGD002174 VGD002406 RVGC007195 VGC007288 VGC006815 VGC006980 VGB001919 VGB002299 VGB002367 VGB002370 VGB002371 VGC006629 VGD000571 VGB002373 RVGC000008 RVGC000009 VGC000658 VGC000658 VGC000661 VGC000662 VGC006111 VGC000506 VGD001461 VGB002560 VGB002563 VGC000747 VGC000955 VGC006092 VGD001966 VGD001968 VGD001969 VGD001968 VGD001990 VGD001992 VGD001994 VGD001995 VGD001966 VGD001969 VGC000188 VGC004370 VGC004375 VGC004435 VGD002366 VGD002416 VGB002591 RVGC004152 VENGO000541 VGC007042 VGD000048 VGD001839 VGD001840 RVGB000827 VGD001793 RVGD001646 VGC007133 VGC000561 RVGC004173 VGD001935 VGC007007 VGC007009 VGD000895 VGD000897 VGD000895 VGD001094 VGD001104 VGOVGC007882 VGOVGB002883 VGB000077 RVGB000025 VGC000095 VGC006618 VGB000506 RVGC007156 RVGC007161 VGD001725 VGB000659 RVGB000131 VGB001154 VGD001380 VGC006089 VENGO000233 VGC004464 VGC004495 VGD001485 VGD001488 RVGC006633 VGD001774 VGC006792 RVGB000163 RVGB000168 VGC006623 VGC006922 VGD000298 VGD002226 VGC004020 VGC007423 VGC006716 VENGO000048 VGC004105 VGC000668 VGD002243 VGB000975 VGC000595 VGC007393 VGB002372 VGC004261 VGC004466 VGC004470 VGC006137 VGC006946 VGD000404 VGC000894 VGD002396 VGD002214 VGC000695 VGD001597 VGD000767 VGB001018 VGC000971 VGC006417 VGC006674 VGC006761 VGC007604 VGD000776 VGD000780 VGD000784 VGD000785 VGD002127 VGD002125 VGD002127 VGB002399 VGD000284 VENG0000311 VENGO000325 VGC000768 VGC006063 VGC006068 VGD002215 VGD002253 VGD001225 VGD000269 VGD002482 VGC006691 RVGC007149 RVGC007182 VGD001049 VGD000303 VGD000305 VGD002318 VGB001995 VGD002465 VGD000768 VGD000770 VGC006760 VGD000544 VGD002297 VGC004006 VGD001105 VGC000334 VGC000331 RVGC006772 RVGC007181 VGC006128 VGC004387 VGB000663 VENGO000476 VGC000531 VGC000536 VGC000537 VGC000538 VGC000541 VGC000529 VGB000812 RVGB000200 RVGB002263 VENGO000492 VGD001648 VENGO000400 VENGO000405 VGC000096 VGC007112 RVGC000580 VGC006961 VGC000381 VGD002213 VGD001064 VGD001649 VGD002301 VGC006810 VGD000403 VGD001559 VGB001271 VGB001012 VGB001560 VGB002112 VGB002252 VGB002279 VGB002283 VGB002285 VGB002291 VGC000155 VGC000157 VGC000158 VGC000160 VGC000162 VGC004082 VGC000159 VGB002290 VGC000494 VENGO000705 VGD000318 VGD002401 VGB001149 VGB001153 VGB002056 VGB002057 VGB002058 VGB002059 VGD001561 VGD001567 VGD001583 VENGO000463 VENGO000464 VGC000512 VENGO000396 VGD000527 VGD001452 VGD002229 VGC007592 RVGC006685 VGB000698 VGD000556 VGD002356 VGD002207 VENGO000526 VENGO000565 VENGO000576 VGC004510 VGB002013 VGB002730 VGB002733 VGC004003 VGC004007 VGB000994 VGB002326 VGC004486 VGC004492 VGD000521 VGD001880 VGD002195 VGC000942 VGD001552 VGC004499 RVGB000829 RVGC000596 RVGB000197 VGC000743 VGC006327 RVGB000162 RVGB000176 RVGD000012 VGB000609 VGB002549 VGB002551 VGC006701 VGC007147 VGB000507 VGB000530 VGD000569 VGC000812 VGD000295 VGD001348 VGC004402 VGD001403 VGD001407 VGC000954 VGC007486 VGD000243 VGC000360 VGC004139 VGC004262 VGC004265 RVGC006836 VGC007057 VGC007058 RVGD000132 VGD001228 VGD001908 VGD001912 VGD001942 VGD001809 VGD001814 VGD001822 VGD001887 VGD001907 VGC000493 VGB001010 VGC007109 VGC006766 VGD000066 VGD002422 VGC006277 VGB000986 VGB002562 VGC000497 VGC000938 VGC000956 VGC007111 VGC007451 VGD000334 VGC000211 VGD000972 VGD002037 VGC000405 RVGB000471 VGD001618 VENGO000270 VGC000134 VGC000502 VGC000516 VGC004282 VGD000549 VGC006802 VGD001306 VGD001311 VGD001312 VGD002141 VGB002598 VGD001208 VGD001945 VGD002050 VGB000836 VGC004225 VGC004272 VGC007743 VGD002002 VGC006046 VGC006059 VGD002409 VGC000107 VGC000933 VGC007024 VGD000036 VGD000051 VGD001920 VGD001922 VGD002044 VGC007748 VENGO001008 VGB002951 RVGD000134 RVGB001722 VGC000215 VGC006066 VGD001546 VGD002246 VENGO000391 VGC000525 VGC004523 VGD002267 VGC004090 VGC007351 RVGD000139 VGC006049 VGD001642 VGC000434 VGD002274 VGC000508 VGC006219 VGD002248 VGD002078 VGD002385 VGC004304 VGC007440 VGD002296 VGC007144 VGD001229 VGC004243 VGC004259 VGC004415 VGC000079 VGC000126 VGC000214 VGC004137 VGC006670 VGC006823 VGC007103 VGC007106 VGC007108 VGD000918 VGD000991 VGD000548 VGC007436 VGC007494 VGC007496 VGC007507 VGC007507 VGC000330 VGC000474 VGC007534 VGC000330 VGC000474 VGC006087 VGD001386 VGC004482 VGC004484 VGC006102 VGD000547 VGD002043 VGD002066 VGC000276 VGC006107 VGB001931 VGC007125 VGD001877 RVGB000045 VGD000519 VGD000523 VGD000979 VGD000984 RVGC000065 RVGC000583 VGD001898 RVGC000042 RVGD000018 VGD001087 VGC006643 RVGB000478 RVGB002340 RVGD002199 RVGD002199 RVGC006108 VGD002060 VGD002071 VGD002060 VGD002071 VGB001164 VGB001165 VENGO000995 VGB002835 VGC000917 VGC000968 VGD000398 VGD002118 VGC000231 VGC000232 VGC000234 VGD000771 VGD000771 VGC004154 VGD002240 VGD002245 VGD002302 VGC000063 VGC000048 VGD002029 RVGB000042 RVGB000236 VGC006564 VGC007028 VGD002128 VGD001999 VGC007018 VGB000990 VGD002069 RVGB000134 VGC006090 VGD001063 VGD001555 VGD002418 VGC006500 VGD000059 RVGC006332 VGC000061 VGC006343 VGC007800 VGC006989 VGD001381 RVGC000579 VGC000676 VGD000913 VGB000562 VGC007128 VGC006051 VGC006053 VGC006054 VGC000182 VGB000650 VGB000681 VGB000682 VGC007065 VGD001620 VGD001668 VGB002545 VGD001927 VGC007428 VGC000153 VGC004145 VGD001634 VGD001813 VGC000645 VGC000428 VGC004242 VGC000757 RVGB000780 VGB002636 VGB002637 VGD002353 VGB002566 VGB002572 VGB002578 VGB002580 VGC007099 VGD000233 VGD000236 VGD000237 VGD000236 VGD001829 VGD002355 VGD002358 VGC000973 VGC007120 VGB002225 VGC000679 VGD001712 VGC007142 VGC000482 VGC007117 VGD000798 VGD002020 VGD000230 VGD001617 VGD001633 VGB000489 VGB002697 RVGC007202 VGD002284 VGC000402 VGC006040 VGC006042 VGC006104 VGD001113 RVGC000566 VGC000905 VGC000907 VGC000915 VGC000922 VGC004364 VGD001116 14616 VGB002115 VGD000864 VGD001411 VGB002522 VGC

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1555-2025

Related Recalls

Class I - Dangerous

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Apr 20, 2026 Other Medical Devices Nationwide View Details →
Class I - Dangerous

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Apr 20, 2026 Other Medical Devices Nationwide View Details →
Class I - Dangerous

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.