Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was distributed with the following sheath/probe: 0250070451-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Spatula-Tip, 0250070452-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, J-TIP, 0250070453-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, L-TIP, 0250070455-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Needle Tip
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers: 010818-02, 010818-05, 010818-06, 010818-07, 010818-08, 010818-09, 040816-01, 040816-02
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Stryker Corporation
- Reason for Recall:
- Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was distributed with the following sheath/probe: 0250070451-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Spatula-Tip, 0250070452-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, J-TIP, 0250070453-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, L-TIP, 0250070455-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Needle Tip
Product Codes/Lot Numbers:
Lot Numbers: 010818-02, 010818-05, 010818-06, 010818-07, 010818-08, 010818-09, 040816-01, 040816-02
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1509-2020
Related Recalls
It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.
Due to nonconforming products being inadvertently distributed.
Due to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during target tissue removal procedures.