Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    GTIN 07613327634914, Serial Number 2521008873
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Stryker Corporation
Reason for Recall:
It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System

Product Codes/Lot Numbers:

GTIN 07613327634914, Serial Number 2521008873

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2282-2026

Related Recalls

Due to nonconforming products being inadvertently distributed.

Apr 9, 2026 Implants & Prosthetics Nationwide View Details →

Due to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during target tissue removal procedures.

Jan 21, 2026 Surgical Instruments Nationwide View Details →