Medtronic Grafton Matrix Strips: a) DBM S42275 8MMX1CMX10CM 2EA MATRIX STRPS, REF S42275; b) DBM S42280 8MMX1CMX20CM 2EA MATRX STRPS, REF S42280; c) DBM T42275 8MMX1CMX10CM 2 EACH GRAFTON M, REF T42275; d) DBM T42275AUS 8MMX1CMX10CM 2 EA GRFTON S, REF T42275AUS; e) DBM T42275INT 8MMX1CMX10CM 2 EA GRFTON S, REF T42275INT; f) DBM T42275JPN GRAFTON MATRIX 1CM X 10CM, REF T42275JPN; g) DBM T42280 8MMX1CMX20CM 2 EACH GRAFTON M, REF T42280; h) DBM T42280AUS 8MMX1CMX20CM 2 EA GRFTON S, REF T42280AUS; i) DBM T42280INT 8MMX1CMX20CM 2 EA GRFTON S, REF T42280INT; j) DBM T42280JPN GRAFTON MATRIX 1CM X 20CM, REF T42280JPN; Demineralized Bone Matrix (DBM) product

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a) DBM S42275 8MMX1CMX10CM 2EA MATRIX STRPS, REF S42275, GTIN 00613994960931, Serial Numbers: A42914, A43126, A43182, A43185, A43460, A43613, A43715, A43896, A44262, A44388, A44561, A45997, A46401, A46690, A46699, A47001, A47719, A47942, A48136, A48371, A48792, A49045, A49089, A49108, A49790, A49952, A50039, A50528, A50688, A51827, A52539, A52601, A52788, A53401, A53671, A53788, A53881, A53904, A54204, A54988, A55032, A55132; b) DBM S42280 8MMX1CMX20CM 2EA MATRX STRPS, REF S42280, GTIN 00613994960924, Serial Numbers: A42850, A43731, A44343, A44529, A44821, A45022, A45035, A46734, A47604, A47621, A47701, A47971, A48051, A48578, A49616, A51011, A52580, A52809, A54208, A54369, A54440, A54597, A55584; c) DBM T42275 8MMX1CMX10CM 2 EACH GRAFTON M, REF T42275, GTIN 00643169123076, Serial Numbers: A39563-061, A39563-062, A39563-063, A39563-064, A39563-065, A39563-066, A39563-068, A39563-070, A39563-071, A39563-072, A39563-073, A39563-074, A39563-075, A39563-076, A39563-077, A40837-011, A41106-011, A41106-012, A41106-013, A41106-014, A41106-015, A41106-016, A41106-017, A41106-018, A41106-019, A41558-011, A41558-012, A41558-013, A41558-014, A41558-015, A41558-016, A41656-011, A41656-012, A41656-013, A41656-014, A41656-015, A41656-016, A41656-017, A41656-018, A41656-019, A41656-020, A41656-021, A41656-022, A41656-023, A41656-024, A41656-025, A41656-026, A41656-027, A41656-028, A41656-029, A41656-030, A41656-031, A41656-032, A41656-033, A41656-034, A41656-035, A41656-036, A41656-037, A41656-038, A41656-039, A41656-040, A41961-011, A42020-011, A42020-012, A42020-013, A42020-014, A42020-015, A42020-016, A42020-017, A42020-018, A42020-019, A42020-020, A42020-021, A42020-022, A42020-023, A42020-024, A42020-025, A42020-026, A42020-027, A42020-028, A42085-011, A42551-011, A42551-012, A42551-013, A42551-014, A42551-015, A42551-016, A42554-011, A42554-012, A42554-013, A42554-014, A42585-011, A42585-012, A42586-011, A42586-012, A42586-013, A42586-014, A42586-015, A42586-016, A42586-017, A42586-018, A42586-019, A42586-020, A42586-021, A42586-022, A42586-023, A42586-024, A42586-025, A42586-026, A42586-027, A42586-028, A42586-029, A42586-030, A42586-031, A42596-011, A42596-012, A42596-013, A42596-014, A42596-015, A42596-016, A42596-018, A42596-019, A42596-020, A42596-021, A42596-022, A42596-023, A42596-024, A42596-025, A42596-026, A42596-027, A42596-028, A42596-029, A42596-030, A42596-031, A42596-032, A42596-033, A42634-021, A42634-022, A42634-023, A42634-024, A42634-025, A42634-026, A42634-027, A42634-028, A42634-029, A42634-030, A42634-031, A42634-032, A42634-033, A42634-034, A42634-035, A42634-036, A42634-037, A42634-038, A42634-039, A42634-040, A42635-011, A42635-012, A42635-013, A42635-014, A42635-015, A42635-016, A42635-017, A42635-018, A42635-019, A42635-020, A42635-021, A42635-022, A42635-023, A42635-024, A42635-025, A42635-026, A42635-027, A42635-028, A42635-029, A42635-030, A42635-031, A42635-032, A42635-033, A42635-034, A42635-035, A42635-036, A42635-037, A42635-038, A42648-011, A42648-012, A42648-013, A42648-014, A42648-015, A42648-016, A42648-018, A42648-019, A42648-020, A42648-021, A42648-022, A42648-023, A42648-024, A42648-025, A42648-026, A42648-027, A42648-028, A42648-029, A42648-030, A42648-031, A42671-011, A42671-012, A42680-011, A42680-012, A42683-136, A42683-137, A42683-138, A42683-139, A42683-140, A42683-141, A42683-142, A42683-143, A42683-144, A42683-145, A42683-146, A42683-147, A42683-148, A42683-149, A42683-150, A42683-151, A42683-152, A42683-153, A42683-154, A42683-155, A42683-156, A42683-157, A42815-011, A42815-012, A42815-013, A42815-014, A42815-015, A42815-016, A42815-017, A42815-018, A42815-019, A42815-020, A42815-021, A42815-022, A42815-023, A42815-024, A42815-025, A42815-026, A42815-027, A42815-028, A42815-029, A42815-030, A42844-011, A42844-012, A42844-013, A42851-011, A42851-012, A42851-013, A42851-014, A42851-015, A42851-016, A42851-017, A42851-018, A42851-019, A42851-020, A42851-021, A42851-022, A42851-023, A42851-024, A42851-025, A42851-026, A42851-027, A42851-028, A42851-029, A42851-030, A42851-031, A42851-032, A42851-033, A42851-034, A42851-035, A42851-036, A42851-037, A42857-011, A42857-012, A42857-013, A42857-014, A42857-015, A42861-011, A42879-011, A42879-012, A42879-013, A42879-014, A42879-015, A42879-016, A42879-017, A42903-011, A42903-012, A42903-013, A42917-099, A42925-011, A42925-012, A42925-013, A42925-014, A42925-015, A42925-016, A42925-017, A42925-018, A42925-019, A42925-020, A42925-021, A42925-022, A42925-023, A42925-024, A42925-025, A42925-026, A42925-027, A42925-028, A42960-011, A43037-011, A43037-012, A43037-013, A43037-014, A43037-015, A43037-016, A43037-017, A43037-018, A43044-011, A43044-012, A43044-013, A43044-014, A43044-015, A43044-016, A43044-017, A43044-018, A43044-019, A43044-020, A43044-021, A43044-022, A43044-023, A43044-024, A43044-025, A43044-026, A43044-027, A43044-028, A43044-029, A43044-030, A43044-031, A43044-032, A43044-033, A43044-034, A43044-035, A43044-036, A43044-037, A43094-011, A43094-012, A43094-013, A43094-014, A43094-015, A43094-016, A43094-017, A43094-018, A43094-019, A43094-020, A43209-011, A43209-012, A43209-013, A43262-011, A43262-012, A43262-013, A43262-014, A43262-015, A43262-016, A43262-017, A43262-018, A43275-011, A43275-012, A43275-013, A43275-014, A43275-015, A43275-016, A43275-017, A43303-011, A43303-012, A43303-013, A43303-014, A43303-015, A43303-016, A43325-011, A43325-012, A43367-011, A43367-012, A43367-013, A43367-014, A43367-015, A43367-016, A43367-017, A43367-018, A43367-019, A43367-020, A43367-021, A43367-022, A43367-023, A43367-024, A43367-025, A43367-026, A43367-027, A43367-028, A43367-029, A43367-030, A43367-031, A43367-032, A43367-033, A43367-034, A43367-035, A43367-036, A43367-037, A43367-038, A43367-039, A43367-040, A43367-041, A43367-042, A43401-175, A43401-176, A43401-177, A43401-178, A43401-179, A43418-011, A43418-012, A43460-011, A43460-012, A43460-013, A43460-014, A43460-015, A43460-016, A43465-011, A43465-012, A43467-011, A43467-012, A43467-013, A43476-011, A43476-012, A43476-013, A43476-014, A43476-015, A43476-016, A43476-017, A43476-018, A43476-019, A43476-020, A43476-021, A43476-022, A43476-023, A43476-024, A43476-025, A43476-026, A43476-027, A43540-011, A43540-012, A43540-013, A43540-014, A43540-015, A43540-016, A43540-017, A43540-018, A43540-019, A43540-020, A43540-021, A43540-022, A43540-023, A43540-024, A43540-025, A43540-026, A43540-027, A43540-028, A43540-029, A43540-030, A43540-031, A43540-032, A43540-033, A43540-034, A43540-035, A43540-036, A43540-037, A43540-038, A43544-011, A43544-012, A43544-013, A43544-014, A43544-015, A43544-016, A43544-017, A43544-018, A43544-019, A43544-020, A43544-022, A43544-023, A43544-024, A43544-025, A43570-071, A43570-072, A43570-073, A43570-074, A43598-011, A43598-012, A43598-013, A43601-011, A43601-012, A43601-013, A43601-014, A43601-015, A43601-016, A43601-017, A43601-018, A43601-019, A43601-020, A43601-021, A43601-022, A43601-023, A43601-024, A43601-025, A43601-026, A43601-027, A43601-028, A43601-029, A43601-030, A43601-031, A43601-032, A43601-033, A43601-034, A43601-035, A43601-036, A43601-037, A43603-011, A43603-012, A43603-013, A43603-014, A43603-015, A43603-016, A43603-017, A43603-018, A43603-019, A43603-020, A43603-021, A43603-022, A43603-023, A43603-024, A43603-025, A43603-026, A43603-027, A43603-028, A43603-029, A43603-030, A43603-031, A43665-011, A43665-012, A43665-013, A43665-014, A43665-015, A43665-016, A43666-032, A43666-033, A43666-034, A43704-011, A43704-012, A43704-013, A43720-011, A43720-012, A43720-013, A43720-014, A43720-015, A43720-016, A43720-017, A43720-018, A43720-019, A43720-020, A43720-021, A43720-022, A43720-023, A43720-024, A43720-025, A43720-026, A43720-027, A43720-028, A43720-029, A43720-030, A43720-031, A43720-032, A43720-033, A43720-034, A43732-011, A43732-012, A43732-013, A43763-011, A43763-012, A43782-011, A43782-012, A43782-013, A43782-014, A43782-015, A43782-016, A43782-017, A43782-018, A43782-019, A43782-020, A43782-021, A43782-022, A43782-023, A43782-024, A43782-025, A43782-026, A43782-027, A43782-028, A43782-029, A43782-030, A43782-031, A43784-011, A43784-012, A43784-013, A43816-011, A43816-012, A43816-013, A43816-014, A43816-015, A43816-016, A43816-017, A43816-018, A43816-019, A43816-020, A43816-021, A43816-022, A43816-023, A43816-024, A43816-025, A43816-026, A43816-027, A43816-028, A43816-029, A43816-030, A43816-031, A43816-032, A43816-033, A43816-034, A43816-035, A43816-037, A43816-038, A43816-039, A43816-040, A43816-041, A43816-042, A43816-043, A43816-044, A43839-011, A43839-012, A43839-013, A43839-014, A43839-015, A43839-016, A43839-017, A43839-018, A43839-019, A43839-020, A43839-021, A43839-022, A43839-023, A43839-024, A43839-025, A43839-026, A43839-027, A43839-028, A43839-029, A43839-030, A43839-031, A43839-032, A43839-033, A43839-034, A43839-035, A43841-011, A43841-012, A43841-013, A43841-014, A43841-015, A43841-016, A43841-017, A43841-018, A43841-019, A43841-020, A43841-021, A43841-022, A43841-023, A43841-024, A43841-025, A43841-026, A43841-027, A43870-011, A43870-012, A43870-013, A43870-014, A43870-015, A43870-016, A43870-017, A43870-018, A43870-019, A43870-020, A43870-021, A43930-011, A43930-012, A43930-013, A43988-011, A43988-012, A43988-013, A43988-014, A43992-011, A43992-012, A43992-013, A43992-014, A43992-015, A43992-016, A43992-017, A43992-018, A43992-019, A43992-020, A43992-021, A43992-022, A43992-023, A43992-024, A43992-025, A43992-026, A43992-027, A43992-028, A44006-011, A44006-012, A44048-011, A44048-012, A44048-013, A44060-011, A44060-012, A44060-013, A44065-011, A44065-012, A44065-013, A44065-014, A44065-015, A44065-016, A44065-017, A44065-018, A44065-019, A44065-020, A44065-021, A44065-022, A44065-023, A44065-024, A44065-025, A44065-026, A44065-027, A44065-028, A44182-011, A44182-012, A44182-013, A44182-014, A44182-015, A44182-016, A44182-017, A44182-018, A44182-019, A44182-020, A44182-021, A44182-022, A44182-023, A44182-024, A44182-025, A44182-026, A44182-027, A44221-071, A44221-072, A44226-011, A44226-012, A44226-013, A44226-014, A44226-015, A44226-016, A44226-017, A44226-018, A44226-019, A44226-020, A44226-021, A44226-022, A44226-023, A44226-024, A44226-025, A44226-026, A44226-027, A44226-028, A44226-029, A44244-011, A44244-012, A44244-013, A44244-014, A44244-015, A44255-011, A44255-012, A44255-013, A44255-014, A44255-015, A44255-016, A44255-017, A44255-018, A44255-019, A44255-020, A44255-021, A44255-022, A44255-023, A44255-024, A44255-025, A44255-026, A44255-027, A44255-028, A44255-029, A44255-030, A44255-031, A44255-032, A44255-033, A44255-034, A44255-035, A44255-036, A44255-037, A44255-038, A44255-039, A44255-040, A44255-041, A44255-042, A44255-043, A44261-011, A44261-012, A44261-013, A44261-014, A44261-015, A44261-016, A44261-017, A44301-011, A44310-011, A44310-012, A44310-013, A44310-014, A44310-015, A44310-016, A44310-017, A44310-018, A44310-019, A44310-020, A44310-021, A44310-022, A44310-023, A44310-024, A44310-025, A44310-026, A44310-027, A44310-028, A44310-029, A44310-030, A44310-031, A44310-032, A44323-021, A44323-022, A44323-023, A44323-024, A44323-025, A44323-026, A44323-027, A44323-028, A44323-029, A44323-030, A44323-031, A44323-032, A44323-033, A44323-034, A44323-035, A44323-036, A44323-037, A44323-038, A44323-039, A44323-040, A44323-041, A44323-042, A44323-043, A44323-044, A44323-045, A44323-046, A44326-021, A44326-022, A44326-023, A44326-024, A44326-025, A44328-011, A44328-012, A44328-013, A44328-014, A44328-015, A44328-016, A44330-011, A44330-012, A44330-013, A44330-014, A44330-015, A44330-016, A44330-017, A44330-018, A44330-019, A44330-020, A44330-021, A44330-022, A44330-023, A44330-024, A44330-025, A44330-026, A44330-027, A44330-028, A44330-029, A44330-030, A44330-031, A44330-032, A44330-033, A44330-034, A44330-035, A44330-036, A44330-037, A44330-038, A44330-039, A44337-011, A44337-012, A44337-013, A44337-014, A44337-015, A44337-016, A44337-017, A44337-018, A44337-019, A44337-020, A44337-021, A44337-022, A44337-023, A44337-024, A44337-025, A44337-026, A44337-027, A44337-028, A44337-029, A44337-030, A44337-031, A44340-011, A44340-012, A44340-013, A44340-014, A44340-015, A44381-011, A44381-012, A44381-013, A44381-014, A44381-015, A44381-016, A44381-017, A44381-018, A44381-019, A44381-020, A44381-021, A44381-022, A44381-023, A44381-024, A44381-025, A44381-026, A44381-027, A44381-028, A44381-029, A44381-030, A44381-031, A44381-032, A44381-033, A44381-034, A44381-035, A44381-036, A44388-011, A44388-012, A44388-013, A44388-014, A44389-011, A44389-012, A44389-013, A44389-014, A44389-015, A44389-016, A44389-017, A44389-018, A44389-019, A44389-020, A44389-021, A44389-022, A44389-023, A44389-024, A44389-025, A44389-026, A44389-027, A44389-028, A44389-029, A44389-030, A44389-031, A44389-032, A44389-033, A44393-021, A44393-022, A44393-023, A44393-024, A44393-025, A44393-026, A44393-027, A44393-028, A44393-029, A44393-030, A44393-031, A44393-032, A44393-033, A44455-036, A44455-037, A44455-038, A44455-039, A44455-040, A44455-041, A44455-042, A44455-043, A44455-044, A44455-045, A44455-046, A44455-047, A44455-048, A44455-049, A44455-050, A44455-051, A44455-052, A44455-053, A44455-054, A44455-055, A44504-011, A44514-021, A44514-022, A44514-023, A44514-024, A44514-025, A44514-026, A44514-027, A44514-028, A44514-029, A44514-030, A44514-031, A44514-032, A44514-033, A44514-034, A44514-035, A44514-036, A44514-037, A44588-011, A44588-012, A44588-013, A44588-014, A44588-015, A44588-016, A44588-017, A44588-018, A44588-019, A44588-020, A44588-021, A44588-022, A44588-023, A44588-024, A44588-025, A44588-026, A44588-027, A44588-028, A44588-029, A44588-030, A44588-031, A44588-032, A44588-033, A44622-011, A44622-012, A44622-013, A44622-014, A44622-015, A44622-016, A44622-017, A44622-018, A44622-019, A44622-020, A44622-021, A44622-022, A44622-023, A44622-024, A44622-025, A44622-026, A44622-027, A44622-028, A44622-029, A44622-030, A44622-031, A44622-032, A44622-033, A44622-034, A44622-035, A44657-022, A44657-023, A44657-025, A44657-026, A44657-027, A44657-028, A44657-029, A44657-030, A44657-031, A44657-032, A44657-033, A44657-034, A44657-035, A44657-036, A44657-037, A44657-038, A44657-039, A44657-040, A44657-041, A44657-042, A44657-043, A44657-044, A44657-045, A44657-046, A44657-047, A44657-048, A44657-049, A44657-050, A44657-051, A44658-011, A44658-012, A44658-013, A44658-014, A44658-015, A44658-016, A44658-017, A44658-019, A44658-020, A44658-021, A44658-022, A44658-023, A44658-024, A44658-025, A44658-026, A44658-027, A44658-028, A44658-029, A44658-030, A44658-031, A44658-032, A44658-033, A44658-034, A44658-035, A44658-036, A44658-037, A44658-038, A44658-039, A44658-040, A44658-041, A44667-021, A44667-022, A44667-023, A44667-024, A44667-025, A44667-026, A44667-027, A44667-028, A44667-029, A44667-030, A44667-031, A44667-032, A44667-033, A44667-034, A44667-035, A44667-036, A44667-037, A44667-038, A44667-039, A44667-040, A44667-041, A44667-042, A44667-043, A44667-044, A44667-045, A44669-011, A44669-012, A44669-013, A44669-014, A44669-015, A44669-016, A44669-017, A44669-018, A44669-019, A44669-020, A44669-021, A44669-022, A44669-023, A44669-024, A44669-025, A44669-026, A44669-027, A44669-028, A44669-029, A44669-030, A44723-031, A44723-032, A44723-033, A44723-034, A44723-035, A44723-036, A44723-037, A44723-038, A44723-039, A44723-040, A44723-041, A44723-042, A44723-043, A44723-044, A44723-045, A44723-046, A44723-047, A44723-048, A44723-049, A44723-050, A44740-011, A44777-011, A44777-012, A44777-013, A44777-014, A44777-015, A44777-016, A44777-017, A44777-018, A44777-019, A44777-020, A44777-021, A44777-022, A44777-023, A44777-024, A44777-025, A44777-026, A44817-011, A44817-012, A44817-013, A44817-014, A44817-015, A44817-016, A44817-017, A44817-018, A44817-019, A44817-020, A44817-021, A44817-022, A44817-023, A44817-024, A44817-025, A44817-026, A44817-027, A44817-028, A44821-011, A44821-012, A44833-021, A44833-022, A44833-023, A44833-024, A44833-025, A44833-026, A44833-027, A44833-028, A44873-011, A44873-012, A44873-013, A44873-014, A44873-015, A44873-016, A44873-017, A44873-018, A44873-019, A44873-020, A44886-011, A44886-012, A44886-013, A44886-014, A44886-015, A44886-016, A44886-017, A44886-018, A44886-019, A44886-020, A44914-011, A44914-012, A44914-013, A44914-015, A44914-016, A44921-011, A44921-012, A44921-013, A44921-014, A44921-015, A44921-016, A44921-017, A44921-018, A44921-019, A44921-020, A44921-021, A44921-022, A44921-023, A44923-011, A44923-012, A44923-013, A44923-014, A44923-015, A44923-016, A44923-017, A44923-018, A44923-019, A44923-020, A44923-021, A44923-022, A44923-023, A44923-024, A44923-025, A44923-026, A44923-027, A44923-028, A44923-029, A44923-030, A44923-031, A44923-032, A44923-033, A44923-034, A44923-035, A44923-036, A44923-037, A44923-038, A44923-039, A44931-011, A44931-012, A44931-013, A44945-011, A44945-012, A44945-013, A44945-014, A44945-015, A44945-016, A44945-017, A44945-018, A44950-011, A44950-012, A44950-013, A44950-014, A44950-015, A44950-016, A44950-017, A44950-018, A44950-019, A44950-020, A44950-021, A44950-022, A44950-023, A44950-024, A44950-025, A44950-026, A44950-027, A44950-028, A44950-029, A44950-030, A44950-031, A44950-032, A44950-033, A44952-011, A44952-012, A44988-011, A44988-012, A44988-013, A44988-014, A44988-015, A44988-016, A44988-017, A44988-018, A44988-019, A44988-020, A44988-021, A44990-011, A44990-012, A44990-013, A44990-014, A44990-015, A44990-016, A44990-017, A44990-018, A44990-019, A44990-020, A44990-021, A44990-022, A44990-023, A44990-024, A45036-011, A45036-012, A45036-013, A45061-011, A45061-012, A45061-013, A45078-011, A45078-012, A45078-013, A45143-011, A45143-012, A45143-013, A45143-014, A45143-015, A45143-016, A45143-017, A45143-018, A45143-019, A45143-020, A45143-021, A45143-022, A45143-023, A45143-024, A45143-025, A45143-026, A45143-027, A45143-028, A45143-029, A45143-030, A45143-031, A45143-032, A45143-033, A45145-011, A45145-012, A45145-013, A45145-014, A45145-015, A45145-016, A45145-017, A45145-018, A45145-019, A45145-020, A45145-021, A45145-022, A45145-023, A45145-024, A45145-025, A45145-026, A45145-027, A45145-028, A45145-029, A45145-030, A45145-031, A45145-032, A45145-033, A45145-034, A45145-035, A45145-036, A45191-011, A45191-012, A45191-013, A45191-014, A45191-015, A45191-016, A45191-017, A45191-018, A45191-019, A45191-020, A45191-021, A45191-022, A45191-023, A45191-024, A45191-025, A45191-026, A45191-027, A45191-028, A45191-029, A45191-030, A45191-031, A45191-032, A45191-033, A45228-011, A45228-012, A45228-013, A45228-014, A45228-015, A45228-016, A45228-017, A45228-018, A45228-019, A45228-020, A45228-021, A45228-022, A45228-023, A45228-024, A45228-025, A45228-026, A45228-027, A45228-028, A45232-011, A45232-012, A45232-013, A45232-014, A45232-015, A45232-016, A45232-017, A45232-018, A45232-019, A45232-020, A45232-021, A45232-022, A45232-023, A45232-024, A45232-025, A45232-026, A45232-027, A45232-028, A45232-029, A45232-030, A45232-031, A45232-032, A45232-033, A45232-034, A45232-035, A45232-036, A45246-011, A45246-012, A45246-013, A45246-014, A45246-015, A45246-016, A45246-017, A45246-018, A45246-019, A45246-020, A45246-021, A45246-022, A45246-023, A45246-024, A45246-025, A45246-026, A45246-027, A45246-028, A45246-029, A45246-030, A45246-031, A45246-032, A45246-033, A45246-034, A45246-035, A45246-036, A45246-037, A45246-038, A45246-039, A45246-040, A45246-041, A45246-042, A45252-011, A45252-012, A45252-013, A45252-014, A45252-015, A45252-016, A45252-017, A45252-018, A45252-019, A45252-020, A45252-021, A45252-022, A45252-023, A45252-024, A45252-025, A45252-026, A45252-027, A45252-028, A45252-029, A45252-030, A45252-031, A45252-032, A45252-033, A45252-034, A45252-035, A45252-036, A45252-037, A45252-038, A45252-039, A45255-011, A45255-012, A45255-013, A45255-014, A45255-015, A45255-016, A45255-017, A45255-018, A45255-019, A45255-020, A45255-021, A45255-022, A45255-023, A45255-024, A45255-025, A45255-026, A45255-027, A45255-028, A45255-029, A45255-030, A45255-031, A45255-032, A45255-033, A45255-034, A45255-035, A45263-011, A45263-012, A45263-013, A45263-014, A45263-015, A45263-016, A45263-017, A45263-018, A45263-019, A45263-020, A45263-021, A45263-022, A45263-023, A45263-024, A45263-025, A45263-026, A45263-027, A45263-028, A45263-029, A45263-030, A45263-031, A45263-032, A45284-011, A45284-012, A45284-013, A45284-014, A45284-015, A45284-016, A45284-017, A45284-018, A45284-019, A45284-020, A45284-021, A45284-022, A45284-023, A45284-024, A45284-025, A45284-026, A45284-027, A45284-028, A45284-029, A45284-030, A45284-031, A45284-032, A45284-033, A45284-034, A45284-035, A45284-036, A45284-037, A45284-038, A45290-011, A45290-012, A45290-013, A45290-014, A45290-015, A45290-016, A45290-017, A45290-018, A45290-019, A45290-020, A45290-021, A45290-022, A45290-023, A45290-024, A45290-025, A45290-026, A45290-027, A45290-028, A45290-029, A45290-030, A45290-031, A45290-032, A45290-033, A45290-034, A45290-035, A45290-036, A45290-037, A45290-038, A45290-039, A45290-040, A45290-041, A45290-042, A45290-043, A45303-011, A45303-012, A45303-013, A45303-014, A45303-015, A45303-016, A45303-017, A45303-018, A45303-019, A45303-020, A45303-021, A45303-022, A45303-023, A45303-024, A45303-025, A45303-026, A45303-027, A45303-028, A45303-029, A45309-011, A45309-012, A45309-013, A45309-014, A45309-015, A45309-016, A45309-017, A45309-018, A45309-019, A45309-020, A45309-021, A45309-022, A45345-011, A45345-012, A45345-013, A45345-014, A45345-015, A45345-016, A45345-017, A45345-018, A45345-019, A45345-020, A45345-021, A45345-022, A45345-023, A45345-024, A45345-025, A45345-026, A45345-027, A45345-028, A45345-029, A45345-030, A45345-031, A45345-032, A45345-033, A45345-034, A45380-011, A45380-012, A45380-013, A45380-014, A45380-015, A45380-016, A45380-017, A45380-018, A45380-019, A45380-020, A45380-021, A45380-022, A45380-023, A45380-024, A45380-025, A45380-026, A45380-027, A45380-028, A45380-029, A45380-030, A45380-031, A45380-032, A45380-033, A45380-034, A45380-035, A45380-036, A45380-037, A45380-038, A45380-039, A45380-040, A45385-011, A45385-012, A45385-013, A45385-014, A45385-015, A45385-016, A45385-017, A45385-018, A45385-019, A45385-020, A45385-021, A45385-022, A45385-023, A45385-024, A45385-025, A45385-026, A45385-027, A45385-028, A45385-029, A45385-030, A45385-031, A45385-032, A45385-033, A45385-034, A45385-035, A45385-036, A45396-011, A45396-012, A45396-013, A45396-014, A45396-015, A45396-016, A45396-017, A45396-018, A45409-011, A45409-012, A45409-013, A45409-014, A45409-015, A45409-016, A45409-017, A45409-018, A45409-019, A45409-020, A45409-021, A45409-022, A45409-023, A45409-024, A45409-025, A45409-026, A45409-027, A45409-028, A45409-029, A45409-030, A45409-031, A45409-032, A45409-033, A45409-034, A45435-011, A45435-012, A45435-013, A45435-014, A45435-015, A45435-016, A45435-017, A45435-018, A45435-019, A45435-020, A45435-021, A45435-022, A45435-023, A45435-024, A45435-025, A45435-026, A45435-027, A45435-028, A45435-029, A45435-030, A45435-031, A45435-032, A45435-033, A45435-034, A45435-035, A45435-036, A45435-037, A45435-038, A45435-039, A45435-040, A45435-041, A45435-042, A45461-011, A45461-012, A45461-013, A45461-014, A45461-015, A45461-016, A45461-017, A45461-018, A45461-019, A45461-020, A45461-021, A45461-022, A45461-023, A45461-024, A45461-025, A45461-026, A45461-027, A45461-028, A45461-029, A45461-030, A45463-011, A45463-012, A45463-013, A45463-014, A45463-015, A45463-016, A45463-017, A45463-018, A45463-019, A45463-020, A45463-021, A45463-022, A45463-023, A45463-024, A45463-025, A45463-026, A45463-027, A45463-028, A45463-029, A45463-030, A45463-031, A45463-032, A45463-033, A45490-011, A45498-011, A45498-012, A45498-013, A45498-014, A45498-015, A45498-016, A45498-017, A45498-018, A45498-019, A45498-020, A45498-021, A45498-022, A45498-023, A45498-024, A45498-025, A45566-011, A45567-011, A45567-012, A45567-013, A45567-014, A45567-015, A45567-016, A45567-017, A45567-018, A45567-019, A45567-020, A45567-021, A45567-022, A45567-023, A45567-024, A45567-025, A45567-026, A45567-027, A45567-028, A45567-029, A45567-030, A45567-031, A45569-011, A45569-012, A45569-013, A45569-014, A45569-015, A45569-016, A45569-017, A45569-018, A45569-019, A45569-020, A45569-021, A45569-022, A45569-023, A45569-024, A45569-025, A45701-011, A45701-012, A45701-013, A45701-014, A45701-015, A45701-016, A45701-017, A45701-018, A45701-019, A45701-020, A45701-021, A45701-022, A45701-023, A45701-024, A45701-025, A45701-026, A45701-027, A45701-028, A45701-029, A45701-030, A45701-031, A45701-032, A45701-033, A45701-034, A45701-035, A45701-036, A45703-027, A45703-028, A45703-029, A45703-030, A45703-031, A45703-032, A45703-033, A45703-034, A45703-035, A45733-036, A45733-037, A45733-038, A45733-039, A45733-040, A45733-041, A45733-042, A45733-043, A45733-044, A45733-045, A45733-046, A45733-047, A45733-048, A45733-049, A45733-050, A45733-051, A45733-052, A45733-053, A45733-054, A45733-055, A45733-056, A45733-057, A45733-058, A45738-021, A45778-011, A45778-012, A45778-013, A45778-014, A45797-011, A45797-012, A45797-013, A45803-011, A45803-012, A45803-013, A45803-014, A45803-015, A45803-016, A45803-017, A45803-018, A45803-019, A45896-011, A45896-012, A45896-013, A45896-014, A45896-015, A45896-016, A45896-017, A45896-018, A45896-019, A45896-020, A45896-021, A45896-022, A45896-023, A45896-024, A45963-031, A45963-032, A45963-033, A45963-034, A45963-035, A45963-036, A45963-037, A45963-038, A45963-039, A45963-040, A45963-041, A45963-042, A45963-043, A45963-044, A45963-045, A45963-046, A45963-047, A45963-048, A45963-049, A45963-050, A45963-051, A45985-011, A45985-012, A45985-013, A45985-014, A45985-015, A45985-016, A45985-017, A45985-018, A45985-019, A45985-020, A45985-021, A45985-022, A45985-023, A45985-024, A45985-025, A45985-026, A45985-027, A45985-028, A45985-029, A45985-030, A45985-031, A45985-032, A45985-033, A45985-034, A45985-035, A45985-036, A45985-037, A45985-038, A45985-039, A45985-040, A45997-011, A45997-012, A45997-013, A45997-014, A45997-015, A45997-016, A45997-017, A45997-018, A45997-019, A45997-020, A45997-021, A45997-022, A45997-023, A45997-024, A45997-025, A45997-026, A45997-027, A45997-028, A45997-029, A46109-011, A46109-012, A46109-013, A46109-014, A46109-015, A46109-016, A46109-017, A46109-018, A46109-019, A46109-020, A46109-021, A46109-022, A46109-023, A46109-024, A46109-025, A46109-026, A46109-027, A46109-028, A46109-029, A46162-039, A46162-040, A46162-041, A46162-042, A46162-043, A46162-044, A46162-045, A46199-011, A46199-012, A46199-013, A46199-014, A46199-015, A46199-016, A46199-017, A46199-018, A46199-019, A46199-020, A46199-021, A46199-022, A46199-023, A46199-024, A46199-025, A46199-026, A46199-027, A46199-028, A46227-011, A46227-012, A46227-013, A46227-014, A46227-015, A46227-016, A46227-018, A46227-019, A46227-020, A46227-021, A46227-022, A46227-023, A46227-024, A46227-025, A46227-026, A46227-027, A46227-028, A46227-029, A46227-030, A46227-031, A46227-032, A46227-033, A46238-011, A46238-012, A46238-013, A46238-014, A46238-015, A46238-016, A46238-017, A46238-018, A46238-019, A46238-020, A46238-021, A46238-022, A46238-023, A46238-024, A46238-025, A46238-026, A46238-027, A46238-028, A46238-029, A46238-030, A46238-031, A46238-032, A46238-033, A46238-034, A46238-035, A46414-055, A46414-056, A46414-057, A46414-058, A46414-059, A46414-060, A46414-061, A46414-062, A46414-063, A46414-064, A46414-065, A46414-066, A46414-067, A46414-068, A46414-069, A46414-070, A46414-071, A46414-072, A46489-118, A46489-119, A46489-120, A46489-121, A46489-122, A46489-123, A46522-011, A46522-012, A46522-013, A46522-014, A46522-015, A46522-016, A46522-017, A46522-018, A46522-019, A46522-020, A46522-021, A46522-022, A46522-023, A46522-024, A46522-025, A46522-026, A46522-027, A46522-028, A46522-029, A46522-030, A46522-031, A46522-032, A46522-033, A46522-034, A46522-035, A46531-011, A46531-012, A46531-013, A46531-014, A46531-015, A46531-016, A46531-018, A46531-019, A46531-020, A46531-021, A46531-022, A46531-023, A46531-024, A46531-025, A46531-026, A46531-027, A46531-028, A46531-029, A46531-030, A46531-031, A46531-032, A46549-082, A46549-083, A46549-084, A46549-085, A46549-086, A46549-087, A46552-027, A46570-011, A46570-012, A46570-013, A46570-014, A46590-011, A46590-012, A46590-013, A46590-014, A46590-015, A46590-016, A46590-017, A46590-018, A46590-019, A46590-020, A46590-021, A46590-022, A46590-023, A46590-024, A46590-025, A46590-026, A46590-027, A46590-028, A46590-029, A46590-030, A46590-031, A46590-032, A46590-033, A46590-034, A46590-035, A46590-036, A46590-037, A46590-038, A46590-039, A46613-011, A46613-012, A46613-013, A46627-031, A46627-032, A46627-033, A46627-034, A46627-035, A46627-036, A46627-037, A46627-038, A46627-039, A46627-040, A46627-041, A46627-042, A46627-043, A46627-044, A46627-045, A46627-046, A46627-047, A46627-048, A46627-049, A46627-050, A46627-051, A46627-052, A46679-011, A46679-012, A46679-013, A46679-014, A46679-015, A46679-016, A46679-017, A46679-018, A46679-019, A46679-020, A46679-021, A46679-022, A46679-023, A46679-024, A46679-025, A46679-026, A46679-027, A46679-028, A46679-029, A46679-030, A46679-031, A46679-032, A46679-033, A46679-034, A46679-035, A46679-036, A46679-037, A46681-011, A46681-012, A46681-013, A46681-014, A46681-015, A46681-016, A46681-017, A46681-018, A46681-019, A46681-020, A46681-021, A46681-022, A46681-023, A46681-024, A46681-025, A46694-011, A46694-012, A46694-013, A46694-014, A46694-015, A46694-016, A46694-017, A46694-018, A46694-019, A46694-020, A46694-021, A46694-022, A46694-023, A46694-024, A46694-025, A46694-026, A46694-027, A46694-028, A46694-029, A46694-030, A46694-031, A46694-032, A46706-011, A46706-012, A46706-013, A46706-014, A46706-015, A46706-016, A46706-017, A46706-018, A46706-019, A46706-020, A46706-021, A46706-022, A46706-023, A46706-024, A46706-025, A46706-026, A46706-027, A46706-028, A46706-029, A46706-030, A46706-031, A46720-039, A46720-040, A46720-041, A46720-042, A46720-043, A46720-044, A46720-045, A46720-046, A46720-047, A46740-011, A46740-012, A46740-013, A46740-014, A46740-015, A46740-016, A46740-017, A46740-018, A46740-019, A46740-020, A46740-021, A46740-022, A46740-023, A46740-024, A46740-025, A46740-026, A46740-027, A46740-028, A46740-029, A46740-030, A46740-031, A46740-033, A46740-034, A46740-035, A46740-036, A46740-037, A46744-011, A46744-012, A46744-013, A46744-014, A46744-015, A46744-016, A46744-017, A46744-018, A46744-019, A46744-020, A46744-021, A46744-022, A46744-023, A46744-024, A46744-025, A46744-026, A46744-027, A46744-028, A46744-029, A46744-030, A46744-031, A46744-032, A46744-033, A46744-034, A46744-035, A46744-036, A46744-037, A46768-022, A46768-023, A46768-024, A46768-025, A46768-026, A46768-027, A46768-028, A46768-029, A46768-030, A46800-011, A46800-012, A46800-013, A46800-014, A46800-015, A46800-016, A46800-017, A46800-018, A46800-019, A46800-020, A46800-021, A46800-022, A46800-023, A46800-024, A46800-025, A46800-026, A46800-027, A46800-028, A46800-029, A46800-030, A46800-031, A46800-032, A46800-033, A46807-029, A46807-030, A46807-031, A46842-031, A46842-032, A46842-033, A46842-034, A46842-035, A46842-036, A46842-037, A46842-038, A46842-039, A46842-040, A46842-041, A46842-042, A46842-043, A46842-044, A46842-045, A46842-046, A46842-047, A46842-048, A46857-011, A46857-012, A46857-013, A46857-014, A46857-015, A46857-016, A46857-017, A46857-018, A46857-019, A46857-020, A46857-021, A46857-022, A46857-023, A46857-024, A46857-025, A46868-011, A46868-012, A46868-013, A46868-014, A46868-015, A46868-016, A46868-017, A46868-018, A46868-019, A46869-011, A46869-012, A46874-011, A46874-012, A46874-013, A46874-014, A46874-015, A46874-016, A46874-017, A46874-018, A46874-019, A46874-020, A46874-021, A46874-022, A46874-023, A46874-024, A46874-025, A46874-026, A46874-027, A46874-028, A46874-029, A46875-011, A46875-012, A46875-013, A46875-014, A46875-015, A46875-016, A46875-017, A46875-018, A46875-019, A46875-020, A46875-021, A46875-022, A46875-023, A46875-024, A46875-025, A46875-026, A46875-027, A46875-028, A46875-029, A46875-030, A46875-031, A46875-032, A46875-033, A46875-034, A46875-035, A46875-036, A46875-037, A46883-148, A46883-149, A46883-150, A46930-011, A46930-012, A46930-013, A46930-014, A46930-015, A46930-016, A46930-017, A46930-018, A46930-019, A46930-020, A46930-021, A46930-022, A46930-023, A46930-024, A46930-025, A46930-026, A46930-027, A46930-028, A46930-029, A46930-030, A46930-031, A46930-032, A46930-033, A46930-034, A469
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Sofamor Danek USA Inc
Reason for Recall:
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Medtronic Grafton Matrix Strips: a) DBM S42275 8MMX1CMX10CM 2EA MATRIX STRPS, REF S42275; b) DBM S42280 8MMX1CMX20CM 2EA MATRX STRPS, REF S42280; c) DBM T42275 8MMX1CMX10CM 2 EACH GRAFTON M, REF T42275; d) DBM T42275AUS 8MMX1CMX10CM 2 EA GRFTON S, REF T42275AUS; e) DBM T42275INT 8MMX1CMX10CM 2 EA GRFTON S, REF T42275INT; f) DBM T42275JPN GRAFTON MATRIX 1CM X 10CM, REF T42275JPN; g) DBM T42280 8MMX1CMX20CM 2 EACH GRAFTON M, REF T42280; h) DBM T42280AUS 8MMX1CMX20CM 2 EA GRFTON S, REF T42280AUS; i) DBM T42280INT 8MMX1CMX20CM 2 EA GRFTON S, REF T42280INT; j) DBM T42280JPN GRAFTON MATRIX 1CM X 20CM, REF T42280JPN; Demineralized Bone Matrix (DBM) product

Product Codes/Lot Numbers:

a) DBM S42275 8MMX1CMX10CM 2EA MATRIX STRPS, REF S42275, GTIN 00613994960931, Serial Numbers: A42914, A43126, A43182, A43185, A43460, A43613, A43715, A43896, A44262, A44388, A44561, A45997, A46401, A46690, A46699, A47001, A47719, A47942, A48136, A48371, A48792, A49045, A49089, A49108, A49790, A49952, A50039, A50528, A50688, A51827, A52539, A52601, A52788, A53401, A53671, A53788, A53881, A53904, A54204, A54988, A55032, A55132; b) DBM S42280 8MMX1CMX20CM 2EA MATRX STRPS, REF S42280, GTIN 00613994960924, Serial Numbers: A42850, A43731, A44343, A44529, A44821, A45022, A45035, A46734, A47604, A47621, A47701, A47971, A48051, A48578, A49616, A51011, A52580, A52809, A54208, A54369, A54440, A54597, A55584; c) DBM T42275 8MMX1CMX10CM 2 EACH GRAFTON M, REF T42275, GTIN 00643169123076, Serial Numbers: A39563-061, A39563-062, A39563-063, A39563-064, A39563-065, A39563-066, A39563-068, A39563-070, A39563-071, A39563-072, A39563-073, A39563-074, A39563-075, A39563-076, A39563-077, A40837-011, A41106-011, A41106-012, A41106-013, A41106-014, A41106-015, A41106-016, A41106-017, A41106-018, A41106-019, A41558-011, A41558-012, A41558-013, A41558-014, A41558-015, A41558-016, A41656-011, A41656-012, A41656-013, A41656-014, A41656-015, A41656-016, A41656-017, A41656-018, A41656-019, A41656-020, A41656-021, A41656-022, A41656-023, A41656-024, A41656-025, A41656-026, A41656-027, A41656-028, A41656-029, A41656-030, A41656-031, A41656-032, A41656-033, A41656-034, A41656-035, A41656-036, A41656-037, A41656-038, A41656-039, A41656-040, A41961-011, A42020-011, A42020-012, A42020-013, A42020-014, A42020-015, A42020-016, A42020-017, A42020-018, A42020-019, A42020-020, A42020-021, A42020-022, A42020-023, A42020-024, A42020-025, A42020-026, A42020-027, A42020-028, A42085-011, A42551-011, A42551-012, A42551-013, A42551-014, A42551-015, A42551-016, A42554-011, A42554-012, A42554-013, A42554-014, A42585-011, A42585-012, A42586-011, A42586-012, A42586-013, A42586-014, A42586-015, A42586-016, A42586-017, A42586-018, A42586-019, A42586-020, A42586-021, A42586-022, A42586-023, A42586-024, A42586-025, A42586-026, A42586-027, A42586-028, A42586-029, A42586-030, A42586-031, A42596-011, A42596-012, A42596-013, A42596-014, A42596-015, A42596-016, A42596-018, A42596-019, A42596-020, A42596-021, A42596-022, A42596-023, A42596-024, A42596-025, A42596-026, A42596-027, A42596-028, A42596-029, A42596-030, A42596-031, A42596-032, A42596-033, A42634-021, A42634-022, A42634-023, A42634-024, A42634-025, A42634-026, A42634-027, A42634-028, A42634-029, A42634-030, A42634-031, A42634-032, A42634-033, A42634-034, A42634-035, A42634-036, A42634-037, A42634-038, A42634-039, A42634-040, A42635-011, A42635-012, A42635-013, A42635-014, A42635-015, A42635-016, A42635-017, A42635-018, A42635-019, A42635-020, A42635-021, A42635-022, A42635-023, A42635-024, A42635-025, A42635-026, A42635-027, A42635-028, A42635-029, A42635-030, A42635-031, A42635-032, A42635-033, A42635-034, A42635-035, A42635-036, A42635-037, A42635-038, A42648-011, A42648-012, A42648-013, A42648-014, A42648-015, A42648-016, A42648-018, A42648-019, A42648-020, A42648-021, A42648-022, A42648-023, A42648-024, A42648-025, A42648-026, A42648-027, A42648-028, A42648-029, A42648-030, A42648-031, A42671-011, A42671-012, A42680-011, A42680-012, A42683-136, A42683-137, A42683-138, A42683-139, A42683-140, A42683-141, A42683-142, A42683-143, A42683-144, A42683-145, A42683-146, A42683-147, A42683-148, A42683-149, A42683-150, A42683-151, A42683-152, A42683-153, A42683-154, A42683-155, A42683-156, A42683-157, A42815-011, A42815-012, A42815-013, A42815-014, A42815-015, A42815-016, A42815-017, A42815-018, A42815-019, A42815-020, A42815-021, A42815-022, A42815-023, A42815-024, A42815-025, A42815-026, A42815-027, A42815-028, A42815-029, A42815-030, A42844-011, A42844-012, A42844-013, A42851-011, A42851-012, A42851-013, A42851-014, A42851-015, A42851-016, A42851-017, A42851-018, A42851-019, A42851-020, A42851-021, A42851-022, A42851-023, A42851-024, A42851-025, A42851-026, A42851-027, A42851-028, A42851-029, A42851-030, A42851-031, A42851-032, A42851-033, A42851-034, A42851-035, A42851-036, A42851-037, A42857-011, A42857-012, A42857-013, A42857-014, A42857-015, A42861-011, A42879-011, A42879-012, A42879-013, A42879-014, A42879-015, A42879-016, A42879-017, A42903-011, A42903-012, A42903-013, A42917-099, A42925-011, A42925-012, A42925-013, A42925-014, A42925-015, A42925-016, A42925-017, A42925-018, A42925-019, A42925-020, A42925-021, A42925-022, A42925-023, A42925-024, A42925-025, A42925-026, A42925-027, A42925-028, A42960-011, A43037-011, A43037-012, A43037-013, A43037-014, A43037-015, A43037-016, A43037-017, A43037-018, A43044-011, A43044-012, A43044-013, A43044-014, A43044-015, A43044-016, A43044-017, A43044-018, A43044-019, A43044-020, A43044-021, A43044-022, A43044-023, A43044-024, A43044-025, A43044-026, A43044-027, A43044-028, A43044-029, A43044-030, A43044-031, A43044-032, A43044-033, A43044-034, A43044-035, A43044-036, A43044-037, A43094-011, A43094-012, A43094-013, A43094-014, A43094-015, A43094-016, A43094-017, A43094-018, A43094-019, A43094-020, A43209-011, A43209-012, A43209-013, A43262-011, A43262-012, A43262-013, A43262-014, A43262-015, A43262-016, A43262-017, A43262-018, A43275-011, A43275-012, A43275-013, A43275-014, A43275-015, A43275-016, A43275-017, A43303-011, A43303-012, A43303-013, A43303-014, A43303-015, A43303-016, A43325-011, A43325-012, A43367-011, A43367-012, A43367-013, A43367-014, A43367-015, A43367-016, A43367-017, A43367-018, A43367-019, A43367-020, A43367-021, A43367-022, A43367-023, A43367-024, A43367-025, A43367-026, A43367-027, A43367-028, A43367-029, A43367-030, A43367-031, A43367-032, A43367-033, A43367-034, A43367-035, A43367-036, A43367-037, A43367-038, A43367-039, A43367-040, A43367-041, A43367-042, A43401-175, A43401-176, A43401-177, A43401-178, A43401-179, A43418-011, A43418-012, A43460-011, A43460-012, A43460-013, A43460-014, A43460-015, A43460-016, A43465-011, A43465-012, A43467-011, A43467-012, A43467-013, A43476-011, A43476-012, A43476-013, A43476-014, A43476-015, A43476-016, A43476-017, A43476-018, A43476-019, A43476-020, A43476-021, A43476-022, A43476-023, A43476-024, A43476-025, A43476-026, A43476-027, A43540-011, A43540-012, A43540-013, A43540-014, A43540-015, A43540-016, A43540-017, A43540-018, A43540-019, A43540-020, A43540-021, A43540-022, A43540-023, A43540-024, A43540-025, A43540-026, A43540-027, A43540-028, A43540-029, A43540-030, A43540-031, A43540-032, A43540-033, A43540-034, A43540-035, A43540-036, A43540-037, A43540-038, A43544-011, A43544-012, A43544-013, A43544-014, A43544-015, A43544-016, A43544-017, A43544-018, A43544-019, A43544-020, A43544-022, A43544-023, A43544-024, A43544-025, A43570-071, A43570-072, A43570-073, A43570-074, A43598-011, A43598-012, A43598-013, A43601-011, A43601-012, A43601-013, A43601-014, A43601-015, A43601-016, A43601-017, A43601-018, A43601-019, A43601-020, A43601-021, A43601-022, A43601-023, A43601-024, A43601-025, A43601-026, A43601-027, A43601-028, A43601-029, A43601-030, A43601-031, A43601-032, A43601-033, A43601-034, A43601-035, A43601-036, A43601-037, A43603-011, A43603-012, A43603-013, A43603-014, A43603-015, A43603-016, A43603-017, A43603-018, A43603-019, A43603-020, A43603-021, A43603-022, A43603-023, A43603-024, A43603-025, A43603-026, A43603-027, A43603-028, A43603-029, A43603-030, A43603-031, A43665-011, A43665-012, A43665-013, A43665-014, A43665-015, A43665-016, A43666-032, A43666-033, A43666-034, A43704-011, A43704-012, A43704-013, A43720-011, A43720-012, A43720-013, A43720-014, A43720-015, A43720-016, A43720-017, A43720-018, A43720-019, A43720-020, A43720-021, A43720-022, A43720-023, A43720-024, A43720-025, A43720-026, A43720-027, A43720-028, A43720-029, A43720-030, A43720-031, A43720-032, A43720-033, A43720-034, A43732-011, A43732-012, A43732-013, A43763-011, A43763-012, A43782-011, A43782-012, A43782-013, A43782-014, A43782-015, A43782-016, A43782-017, A43782-018, A43782-019, A43782-020, A43782-021, A43782-022, A43782-023, A43782-024, A43782-025, A43782-026, A43782-027, A43782-028, A43782-029, A43782-030, A43782-031, A43784-011, A43784-012, A43784-013, A43816-011, A43816-012, A43816-013, A43816-014, A43816-015, A43816-016, A43816-017, A43816-018, A43816-019, A43816-020, A43816-021, A43816-022, A43816-023, A43816-024, A43816-025, A43816-026, A43816-027, A43816-028, A43816-029, A43816-030, A43816-031, A43816-032, A43816-033, A43816-034, A43816-035, A43816-037, A43816-038, A43816-039, A43816-040, A43816-041, A43816-042, A43816-043, A43816-044, A43839-011, A43839-012, A43839-013, A43839-014, A43839-015, A43839-016, A43839-017, A43839-018, A43839-019, A43839-020, A43839-021, A43839-022, A43839-023, A43839-024, A43839-025, A43839-026, A43839-027, A43839-028, A43839-029, A43839-030, A43839-031, A43839-032, A43839-033, A43839-034, A43839-035, A43841-011, A43841-012, A43841-013, A43841-014, A43841-015, A43841-016, A43841-017, A43841-018, A43841-019, A43841-020, A43841-021, A43841-022, A43841-023, A43841-024, A43841-025, A43841-026, A43841-027, A43870-011, A43870-012, A43870-013, A43870-014, A43870-015, A43870-016, A43870-017, A43870-018, A43870-019, A43870-020, A43870-021, A43930-011, A43930-012, A43930-013, A43988-011, A43988-012, A43988-013, A43988-014, A43992-011, A43992-012, A43992-013, A43992-014, A43992-015, A43992-016, A43992-017, A43992-018, A43992-019, A43992-020, A43992-021, A43992-022, A43992-023, A43992-024, A43992-025, A43992-026, A43992-027, A43992-028, A44006-011, A44006-012, A44048-011, A44048-012, A44048-013, A44060-011, A44060-012, A44060-013, A44065-011, A44065-012, A44065-013, A44065-014, A44065-015, A44065-016, A44065-017, A44065-018, A44065-019, A44065-020, A44065-021, A44065-022, A44065-023, A44065-024, A44065-025, A44065-026, A44065-027, A44065-028, A44182-011, A44182-012, A44182-013, A44182-014, A44182-015, A44182-016, A44182-017, A44182-018, A44182-019, A44182-020, A44182-021, A44182-022, A44182-023, A44182-024, A44182-025, A44182-026, A44182-027, A44221-071, A44221-072, A44226-011, A44226-012, A44226-013, A44226-014, A44226-015, A44226-016, A44226-017, A44226-018, A44226-019, A44226-020, A44226-021, A44226-022, A44226-023, A44226-024, A44226-025, A44226-026, A44226-027, A44226-028, A44226-029, A44244-011, A44244-012, A44244-013, A44244-014, A44244-015, A44255-011, A44255-012, A44255-013, A44255-014, A44255-015, A44255-016, A44255-017, A44255-018, A44255-019, A44255-020, A44255-021, A44255-022, A44255-023, A44255-024, A44255-025, A44255-026, A44255-027, A44255-028, A44255-029, A44255-030, A44255-031, A44255-032, A44255-033, A44255-034, A44255-035, A44255-036, A44255-037, A44255-038, A44255-039, A44255-040, A44255-041, A44255-042, A44255-043, A44261-011, A44261-012, A44261-013, A44261-014, A44261-015, A44261-016, A44261-017, A44301-011, A44310-011, A44310-012, A44310-013, A44310-014, A44310-015, A44310-016, A44310-017, A44310-018, A44310-019, A44310-020, A44310-021, A44310-022, A44310-023, A44310-024, A44310-025, A44310-026, A44310-027, A44310-028, A44310-029, A44310-030, A44310-031, A44310-032, A44323-021, A44323-022, A44323-023, A44323-024, A44323-025, A44323-026, A44323-027, A44323-028, A44323-029, A44323-030, A44323-031, A44323-032, A44323-033, A44323-034, A44323-035, A44323-036, A44323-037, A44323-038, A44323-039, A44323-040, A44323-041, A44323-042, A44323-043, A44323-044, A44323-045, A44323-046, A44326-021, A44326-022, A44326-023, A44326-024, A44326-025, A44328-011, A44328-012, A44328-013, A44328-014, A44328-015, A44328-016, A44330-011, A44330-012, A44330-013, A44330-014, A44330-015, A44330-016, A44330-017, A44330-018, A44330-019, A44330-020, A44330-021, A44330-022, A44330-023, A44330-024, A44330-025, A44330-026, A44330-027, A44330-028, A44330-029, A44330-030, A44330-031, A44330-032, A44330-033, A44330-034, A44330-035, A44330-036, A44330-037, A44330-038, A44330-039, A44337-011, A44337-012, A44337-013, A44337-014, A44337-015, A44337-016, A44337-017, A44337-018, A44337-019, A44337-020, A44337-021, A44337-022, A44337-023, A44337-024, A44337-025, A44337-026, A44337-027, A44337-028, A44337-029, A44337-030, A44337-031, A44340-011, A44340-012, A44340-013, A44340-014, A44340-015, A44381-011, A44381-012, A44381-013, A44381-014, A44381-015, A44381-016, A44381-017, A44381-018, A44381-019, A44381-020, A44381-021, A44381-022, A44381-023, A44381-024, A44381-025, A44381-026, A44381-027, A44381-028, A44381-029, A44381-030, A44381-031, A44381-032, A44381-033, A44381-034, A44381-035, A44381-036, A44388-011, A44388-012, A44388-013, A44388-014, A44389-011, A44389-012, A44389-013, A44389-014, A44389-015, A44389-016, A44389-017, A44389-018, A44389-019, A44389-020, A44389-021, A44389-022, A44389-023, A44389-024, A44389-025, A44389-026, A44389-027, A44389-028, A44389-029, A44389-030, A44389-031, A44389-032, A44389-033, A44393-021, A44393-022, A44393-023, A44393-024, A44393-025, A44393-026, A44393-027, A44393-028, A44393-029, A44393-030, A44393-031, A44393-032, A44393-033, A44455-036, A44455-037, A44455-038, A44455-039, A44455-040, A44455-041, A44455-042, A44455-043, A44455-044, A44455-045, A44455-046, A44455-047, A44455-048, A44455-049, A44455-050, A44455-051, A44455-052, A44455-053, A44455-054, A44455-055, A44504-011, A44514-021, A44514-022, A44514-023, A44514-024, A44514-025, A44514-026, A44514-027, A44514-028, A44514-029, A44514-030, A44514-031, A44514-032, A44514-033, A44514-034, A44514-035, A44514-036, A44514-037, A44588-011, A44588-012, A44588-013, A44588-014, A44588-015, A44588-016, A44588-017, A44588-018, A44588-019, A44588-020, A44588-021, A44588-022, A44588-023, A44588-024, A44588-025, A44588-026, A44588-027, A44588-028, A44588-029, A44588-030, A44588-031, A44588-032, A44588-033, A44622-011, A44622-012, A44622-013, A44622-014, A44622-015, A44622-016, A44622-017, A44622-018, A44622-019, A44622-020, A44622-021, A44622-022, A44622-023, A44622-024, A44622-025, A44622-026, A44622-027, A44622-028, A44622-029, A44622-030, A44622-031, A44622-032, A44622-033, A44622-034, A44622-035, A44657-022, A44657-023, A44657-025, A44657-026, A44657-027, A44657-028, A44657-029, A44657-030, A44657-031, A44657-032, A44657-033, A44657-034, A44657-035, A44657-036, A44657-037, A44657-038, A44657-039, A44657-040, A44657-041, A44657-042, A44657-043, A44657-044, A44657-045, A44657-046, A44657-047, A44657-048, A44657-049, A44657-050, A44657-051, A44658-011, A44658-012, A44658-013, A44658-014, A44658-015, A44658-016, A44658-017, A44658-019, A44658-020, A44658-021, A44658-022, A44658-023, A44658-024, A44658-025, A44658-026, A44658-027, A44658-028, A44658-029, A44658-030, A44658-031, A44658-032, A44658-033, A44658-034, A44658-035, A44658-036, A44658-037, A44658-038, A44658-039, A44658-040, A44658-041, A44667-021, A44667-022, A44667-023, A44667-024, A44667-025, A44667-026, A44667-027, A44667-028, A44667-029, A44667-030, A44667-031, A44667-032, A44667-033, A44667-034, A44667-035, A44667-036, A44667-037, A44667-038, A44667-039, A44667-040, A44667-041, A44667-042, A44667-043, A44667-044, A44667-045, A44669-011, A44669-012, A44669-013, A44669-014, A44669-015, A44669-016, A44669-017, A44669-018, A44669-019, A44669-020, A44669-021, A44669-022, A44669-023, A44669-024, A44669-025, A44669-026, A44669-027, A44669-028, A44669-029, A44669-030, A44723-031, A44723-032, A44723-033, A44723-034, A44723-035, A44723-036, A44723-037, A44723-038, A44723-039, A44723-040, A44723-041, A44723-042, A44723-043, A44723-044, A44723-045, A44723-046, A44723-047, A44723-048, A44723-049, A44723-050, A44740-011, A44777-011, A44777-012, A44777-013, A44777-014, A44777-015, A44777-016, A44777-017, A44777-018, A44777-019, A44777-020, A44777-021, A44777-022, A44777-023, A44777-024, A44777-025, A44777-026, A44817-011, A44817-012, A44817-013, A44817-014, A44817-015, A44817-016, A44817-017, A44817-018, A44817-019, A44817-020, A44817-021, A44817-022, A44817-023, A44817-024, A44817-025, A44817-026, A44817-027, A44817-028, A44821-011, A44821-012, A44833-021, A44833-022, A44833-023, A44833-024, A44833-025, A44833-026, A44833-027, A44833-028, A44873-011, A44873-012, A44873-013, A44873-014, A44873-015, A44873-016, A44873-017, A44873-018, A44873-019, A44873-020, A44886-011, A44886-012, A44886-013, A44886-014, A44886-015, A44886-016, A44886-017, A44886-018, A44886-019, A44886-020, A44914-011, A44914-012, A44914-013, A44914-015, A44914-016, A44921-011, A44921-012, A44921-013, A44921-014, A44921-015, A44921-016, A44921-017, A44921-018, A44921-019, A44921-020, A44921-021, A44921-022, A44921-023, A44923-011, A44923-012, A44923-013, A44923-014, A44923-015, A44923-016, A44923-017, A44923-018, A44923-019, A44923-020, A44923-021, A44923-022, A44923-023, A44923-024, A44923-025, A44923-026, A44923-027, A44923-028, A44923-029, A44923-030, A44923-031, A44923-032, A44923-033, A44923-034, A44923-035, A44923-036, A44923-037, A44923-038, A44923-039, A44931-011, A44931-012, A44931-013, A44945-011, A44945-012, A44945-013, A44945-014, A44945-015, A44945-016, A44945-017, A44945-018, A44950-011, A44950-012, A44950-013, A44950-014, A44950-015, A44950-016, A44950-017, A44950-018, A44950-019, A44950-020, A44950-021, A44950-022, A44950-023, A44950-024, A44950-025, A44950-026, A44950-027, A44950-028, A44950-029, A44950-030, A44950-031, A44950-032, A44950-033, A44952-011, A44952-012, A44988-011, A44988-012, A44988-013, A44988-014, A44988-015, A44988-016, A44988-017, A44988-018, A44988-019, A44988-020, A44988-021, A44990-011, A44990-012, A44990-013, A44990-014, A44990-015, A44990-016, A44990-017, A44990-018, A44990-019, A44990-020, A44990-021, A44990-022, A44990-023, A44990-024, A45036-011, A45036-012, A45036-013, A45061-011, A45061-012, A45061-013, A45078-011, A45078-012, A45078-013, A45143-011, A45143-012, A45143-013, A45143-014, A45143-015, A45143-016, A45143-017, A45143-018, A45143-019, A45143-020, A45143-021, A45143-022, A45143-023, A45143-024, A45143-025, A45143-026, A45143-027, A45143-028, A45143-029, A45143-030, A45143-031, A45143-032, A45143-033, A45145-011, A45145-012, A45145-013, A45145-014, A45145-015, A45145-016, A45145-017, A45145-018, A45145-019, A45145-020, A45145-021, A45145-022, A45145-023, A45145-024, A45145-025, A45145-026, A45145-027, A45145-028, A45145-029, A45145-030, A45145-031, A45145-032, A45145-033, A45145-034, A45145-035, A45145-036, A45191-011, A45191-012, A45191-013, A45191-014, A45191-015, A45191-016, A45191-017, A45191-018, A45191-019, A45191-020, A45191-021, A45191-022, A45191-023, A45191-024, A45191-025, A45191-026, A45191-027, A45191-028, A45191-029, A45191-030, A45191-031, A45191-032, A45191-033, A45228-011, A45228-012, A45228-013, A45228-014, A45228-015, A45228-016, A45228-017, A45228-018, A45228-019, A45228-020, A45228-021, A45228-022, A45228-023, A45228-024, A45228-025, A45228-026, A45228-027, A45228-028, A45232-011, A45232-012, A45232-013, A45232-014, A45232-015, A45232-016, A45232-017, A45232-018, A45232-019, A45232-020, A45232-021, A45232-022, A45232-023, A45232-024, A45232-025, A45232-026, A45232-027, A45232-028, A45232-029, A45232-030, A45232-031, A45232-032, A45232-033, A45232-034, A45232-035, A45232-036, A45246-011, A45246-012, A45246-013, A45246-014, A45246-015, A45246-016, A45246-017, A45246-018, A45246-019, A45246-020, A45246-021, A45246-022, A45246-023, A45246-024, A45246-025, A45246-026, A45246-027, A45246-028, A45246-029, A45246-030, A45246-031, A45246-032, A45246-033, A45246-034, A45246-035, A45246-036, A45246-037, A45246-038, A45246-039, A45246-040, A45246-041, A45246-042, A45252-011, A45252-012, A45252-013, A45252-014, A45252-015, A45252-016, A45252-017, A45252-018, A45252-019, A45252-020, A45252-021, A45252-022, A45252-023, A45252-024, A45252-025, A45252-026, A45252-027, A45252-028, A45252-029, A45252-030, A45252-031, A45252-032, A45252-033, A45252-034, A45252-035, A45252-036, A45252-037, A45252-038, A45252-039, A45255-011, A45255-012, A45255-013, A45255-014, A45255-015, A45255-016, A45255-017, A45255-018, A45255-019, A45255-020, A45255-021, A45255-022, A45255-023, A45255-024, A45255-025, A45255-026, A45255-027, A45255-028, A45255-029, A45255-030, A45255-031, A45255-032, A45255-033, A45255-034, A45255-035, A45263-011, A45263-012, A45263-013, A45263-014, A45263-015, A45263-016, A45263-017, A45263-018, A45263-019, A45263-020, A45263-021, A45263-022, A45263-023, A45263-024, A45263-025, A45263-026, A45263-027, A45263-028, A45263-029, A45263-030, A45263-031, A45263-032, A45284-011, A45284-012, A45284-013, A45284-014, A45284-015, A45284-016, A45284-017, A45284-018, A45284-019, A45284-020, A45284-021, A45284-022, A45284-023, A45284-024, A45284-025, A45284-026, A45284-027, A45284-028, A45284-029, A45284-030, A45284-031, A45284-032, A45284-033, A45284-034, A45284-035, A45284-036, A45284-037, A45284-038, A45290-011, A45290-012, A45290-013, A45290-014, A45290-015, A45290-016, A45290-017, A45290-018, A45290-019, A45290-020, A45290-021, A45290-022, A45290-023, A45290-024, A45290-025, A45290-026, A45290-027, A45290-028, A45290-029, A45290-030, A45290-031, A45290-032, A45290-033, A45290-034, A45290-035, A45290-036, A45290-037, A45290-038, A45290-039, A45290-040, A45290-041, A45290-042, A45290-043, A45303-011, A45303-012, A45303-013, A45303-014, A45303-015, A45303-016, A45303-017, A45303-018, A45303-019, A45303-020, A45303-021, A45303-022, A45303-023, A45303-024, A45303-025, A45303-026, A45303-027, A45303-028, A45303-029, A45309-011, A45309-012, A45309-013, A45309-014, A45309-015, A45309-016, A45309-017, A45309-018, A45309-019, A45309-020, A45309-021, A45309-022, A45345-011, A45345-012, A45345-013, A45345-014, A45345-015, A45345-016, A45345-017, A45345-018, A45345-019, A45345-020, A45345-021, A45345-022, A45345-023, A45345-024, A45345-025, A45345-026, A45345-027, A45345-028, A45345-029, A45345-030, A45345-031, A45345-032, A45345-033, A45345-034, A45380-011, A45380-012, A45380-013, A45380-014, A45380-015, A45380-016, A45380-017, A45380-018, A45380-019, A45380-020, A45380-021, A45380-022, A45380-023, A45380-024, A45380-025, A45380-026, A45380-027, A45380-028, A45380-029, A45380-030, A45380-031, A45380-032, A45380-033, A45380-034, A45380-035, A45380-036, A45380-037, A45380-038, A45380-039, A45380-040, A45385-011, A45385-012, A45385-013, A45385-014, A45385-015, A45385-016, A45385-017, A45385-018, A45385-019, A45385-020, A45385-021, A45385-022, A45385-023, A45385-024, A45385-025, A45385-026, A45385-027, A45385-028, A45385-029, A45385-030, A45385-031, A45385-032, A45385-033, A45385-034, A45385-035, A45385-036, A45396-011, A45396-012, A45396-013, A45396-014, A45396-015, A45396-016, A45396-017, A45396-018, A45409-011, A45409-012, A45409-013, A45409-014, A45409-015, A45409-016, A45409-017, A45409-018, A45409-019, A45409-020, A45409-021, A45409-022, A45409-023, A45409-024, A45409-025, A45409-026, A45409-027, A45409-028, A45409-029, A45409-030, A45409-031, A45409-032, A45409-033, A45409-034, A45435-011, A45435-012, A45435-013, A45435-014, A45435-015, A45435-016, A45435-017, A45435-018, A45435-019, A45435-020, A45435-021, A45435-022, A45435-023, A45435-024, A45435-025, A45435-026, A45435-027, A45435-028, A45435-029, A45435-030, A45435-031, A45435-032, A45435-033, A45435-034, A45435-035, A45435-036, A45435-037, A45435-038, A45435-039, A45435-040, A45435-041, A45435-042, A45461-011, A45461-012, A45461-013, A45461-014, A45461-015, A45461-016, A45461-017, A45461-018, A45461-019, A45461-020, A45461-021, A45461-022, A45461-023, A45461-024, A45461-025, A45461-026, A45461-027, A45461-028, A45461-029, A45461-030, A45463-011, A45463-012, A45463-013, A45463-014, A45463-015, A45463-016, A45463-017, A45463-018, A45463-019, A45463-020, A45463-021, A45463-022, A45463-023, A45463-024, A45463-025, A45463-026, A45463-027, A45463-028, A45463-029, A45463-030, A45463-031, A45463-032, A45463-033, A45490-011, A45498-011, A45498-012, A45498-013, A45498-014, A45498-015, A45498-016, A45498-017, A45498-018, A45498-019, A45498-020, A45498-021, A45498-022, A45498-023, A45498-024, A45498-025, A45566-011, A45567-011, A45567-012, A45567-013, A45567-014, A45567-015, A45567-016, A45567-017, A45567-018, A45567-019, A45567-020, A45567-021, A45567-022, A45567-023, A45567-024, A45567-025, A45567-026, A45567-027, A45567-028, A45567-029, A45567-030, A45567-031, A45569-011, A45569-012, A45569-013, A45569-014, A45569-015, A45569-016, A45569-017, A45569-018, A45569-019, A45569-020, A45569-021, A45569-022, A45569-023, A45569-024, A45569-025, A45701-011, A45701-012, A45701-013, A45701-014, A45701-015, A45701-016, A45701-017, A45701-018, A45701-019, A45701-020, A45701-021, A45701-022, A45701-023, A45701-024, A45701-025, A45701-026, A45701-027, A45701-028, A45701-029, A45701-030, A45701-031, A45701-032, A45701-033, A45701-034, A45701-035, A45701-036, A45703-027, A45703-028, A45703-029, A45703-030, A45703-031, A45703-032, A45703-033, A45703-034, A45703-035, A45733-036, A45733-037, A45733-038, A45733-039, A45733-040, A45733-041, A45733-042, A45733-043, A45733-044, A45733-045, A45733-046, A45733-047, A45733-048, A45733-049, A45733-050, A45733-051, A45733-052, A45733-053, A45733-054, A45733-055, A45733-056, A45733-057, A45733-058, A45738-021, A45778-011, A45778-012, A45778-013, A45778-014, A45797-011, A45797-012, A45797-013, A45803-011, A45803-012, A45803-013, A45803-014, A45803-015, A45803-016, A45803-017, A45803-018, A45803-019, A45896-011, A45896-012, A45896-013, A45896-014, A45896-015, A45896-016, A45896-017, A45896-018, A45896-019, A45896-020, A45896-021, A45896-022, A45896-023, A45896-024, A45963-031, A45963-032, A45963-033, A45963-034, A45963-035, A45963-036, A45963-037, A45963-038, A45963-039, A45963-040, A45963-041, A45963-042, A45963-043, A45963-044, A45963-045, A45963-046, A45963-047, A45963-048, A45963-049, A45963-050, A45963-051, A45985-011, A45985-012, A45985-013, A45985-014, A45985-015, A45985-016, A45985-017, A45985-018, A45985-019, A45985-020, A45985-021, A45985-022, A45985-023, A45985-024, A45985-025, A45985-026, A45985-027, A45985-028, A45985-029, A45985-030, A45985-031, A45985-032, A45985-033, A45985-034, A45985-035, A45985-036, A45985-037, A45985-038, A45985-039, A45985-040, A45997-011, A45997-012, A45997-013, A45997-014, A45997-015, A45997-016, A45997-017, A45997-018, A45997-019, A45997-020, A45997-021, A45997-022, A45997-023, A45997-024, A45997-025, A45997-026, A45997-027, A45997-028, A45997-029, A46109-011, A46109-012, A46109-013, A46109-014, A46109-015, A46109-016, A46109-017, A46109-018, A46109-019, A46109-020, A46109-021, A46109-022, A46109-023, A46109-024, A46109-025, A46109-026, A46109-027, A46109-028, A46109-029, A46162-039, A46162-040, A46162-041, A46162-042, A46162-043, A46162-044, A46162-045, A46199-011, A46199-012, A46199-013, A46199-014, A46199-015, A46199-016, A46199-017, A46199-018, A46199-019, A46199-020, A46199-021, A46199-022, A46199-023, A46199-024, A46199-025, A46199-026, A46199-027, A46199-028, A46227-011, A46227-012, A46227-013, A46227-014, A46227-015, A46227-016, A46227-018, A46227-019, A46227-020, A46227-021, A46227-022, A46227-023, A46227-024, A46227-025, A46227-026, A46227-027, A46227-028, A46227-029, A46227-030, A46227-031, A46227-032, A46227-033, A46238-011, A46238-012, A46238-013, A46238-014, A46238-015, A46238-016, A46238-017, A46238-018, A46238-019, A46238-020, A46238-021, A46238-022, A46238-023, A46238-024, A46238-025, A46238-026, A46238-027, A46238-028, A46238-029, A46238-030, A46238-031, A46238-032, A46238-033, A46238-034, A46238-035, A46414-055, A46414-056, A46414-057, A46414-058, A46414-059, A46414-060, A46414-061, A46414-062, A46414-063, A46414-064, A46414-065, A46414-066, A46414-067, A46414-068, A46414-069, A46414-070, A46414-071, A46414-072, A46489-118, A46489-119, A46489-120, A46489-121, A46489-122, A46489-123, A46522-011, A46522-012, A46522-013, A46522-014, A46522-015, A46522-016, A46522-017, A46522-018, A46522-019, A46522-020, A46522-021, A46522-022, A46522-023, A46522-024, A46522-025, A46522-026, A46522-027, A46522-028, A46522-029, A46522-030, A46522-031, A46522-032, A46522-033, A46522-034, A46522-035, A46531-011, A46531-012, A46531-013, A46531-014, A46531-015, A46531-016, A46531-018, A46531-019, A46531-020, A46531-021, A46531-022, A46531-023, A46531-024, A46531-025, A46531-026, A46531-027, A46531-028, A46531-029, A46531-030, A46531-031, A46531-032, A46549-082, A46549-083, A46549-084, A46549-085, A46549-086, A46549-087, A46552-027, A46570-011, A46570-012, A46570-013, A46570-014, A46590-011, A46590-012, A46590-013, A46590-014, A46590-015, A46590-016, A46590-017, A46590-018, A46590-019, A46590-020, A46590-021, A46590-022, A46590-023, A46590-024, A46590-025, A46590-026, A46590-027, A46590-028, A46590-029, A46590-030, A46590-031, A46590-032, A46590-033, A46590-034, A46590-035, A46590-036, A46590-037, A46590-038, A46590-039, A46613-011, A46613-012, A46613-013, A46627-031, A46627-032, A46627-033, A46627-034, A46627-035, A46627-036, A46627-037, A46627-038, A46627-039, A46627-040, A46627-041, A46627-042, A46627-043, A46627-044, A46627-045, A46627-046, A46627-047, A46627-048, A46627-049, A46627-050, A46627-051, A46627-052, A46679-011, A46679-012, A46679-013, A46679-014, A46679-015, A46679-016, A46679-017, A46679-018, A46679-019, A46679-020, A46679-021, A46679-022, A46679-023, A46679-024, A46679-025, A46679-026, A46679-027, A46679-028, A46679-029, A46679-030, A46679-031, A46679-032, A46679-033, A46679-034, A46679-035, A46679-036, A46679-037, A46681-011, A46681-012, A46681-013, A46681-014, A46681-015, A46681-016, A46681-017, A46681-018, A46681-019, A46681-020, A46681-021, A46681-022, A46681-023, A46681-024, A46681-025, A46694-011, A46694-012, A46694-013, A46694-014, A46694-015, A46694-016, A46694-017, A46694-018, A46694-019, A46694-020, A46694-021, A46694-022, A46694-023, A46694-024, A46694-025, A46694-026, A46694-027, A46694-028, A46694-029, A46694-030, A46694-031, A46694-032, A46706-011, A46706-012, A46706-013, A46706-014, A46706-015, A46706-016, A46706-017, A46706-018, A46706-019, A46706-020, A46706-021, A46706-022, A46706-023, A46706-024, A46706-025, A46706-026, A46706-027, A46706-028, A46706-029, A46706-030, A46706-031, A46720-039, A46720-040, A46720-041, A46720-042, A46720-043, A46720-044, A46720-045, A46720-046, A46720-047, A46740-011, A46740-012, A46740-013, A46740-014, A46740-015, A46740-016, A46740-017, A46740-018, A46740-019, A46740-020, A46740-021, A46740-022, A46740-023, A46740-024, A46740-025, A46740-026, A46740-027, A46740-028, A46740-029, A46740-030, A46740-031, A46740-033, A46740-034, A46740-035, A46740-036, A46740-037, A46744-011, A46744-012, A46744-013, A46744-014, A46744-015, A46744-016, A46744-017, A46744-018, A46744-019, A46744-020, A46744-021, A46744-022, A46744-023, A46744-024, A46744-025, A46744-026, A46744-027, A46744-028, A46744-029, A46744-030, A46744-031, A46744-032, A46744-033, A46744-034, A46744-035, A46744-036, A46744-037, A46768-022, A46768-023, A46768-024, A46768-025, A46768-026, A46768-027, A46768-028, A46768-029, A46768-030, A46800-011, A46800-012, A46800-013, A46800-014, A46800-015, A46800-016, A46800-017, A46800-018, A46800-019, A46800-020, A46800-021, A46800-022, A46800-023, A46800-024, A46800-025, A46800-026, A46800-027, A46800-028, A46800-029, A46800-030, A46800-031, A46800-032, A46800-033, A46807-029, A46807-030, A46807-031, A46842-031, A46842-032, A46842-033, A46842-034, A46842-035, A46842-036, A46842-037, A46842-038, A46842-039, A46842-040, A46842-041, A46842-042, A46842-043, A46842-044, A46842-045, A46842-046, A46842-047, A46842-048, A46857-011, A46857-012, A46857-013, A46857-014, A46857-015, A46857-016, A46857-017, A46857-018, A46857-019, A46857-020, A46857-021, A46857-022, A46857-023, A46857-024, A46857-025, A46868-011, A46868-012, A46868-013, A46868-014, A46868-015, A46868-016, A46868-017, A46868-018, A46868-019, A46869-011, A46869-012, A46874-011, A46874-012, A46874-013, A46874-014, A46874-015, A46874-016, A46874-017, A46874-018, A46874-019, A46874-020, A46874-021, A46874-022, A46874-023, A46874-024, A46874-025, A46874-026, A46874-027, A46874-028, A46874-029, A46875-011, A46875-012, A46875-013, A46875-014, A46875-015, A46875-016, A46875-017, A46875-018, A46875-019, A46875-020, A46875-021, A46875-022, A46875-023, A46875-024, A46875-025, A46875-026, A46875-027, A46875-028, A46875-029, A46875-030, A46875-031, A46875-032, A46875-033, A46875-034, A46875-035, A46875-036, A46875-037, A46883-148, A46883-149, A46883-150, A46930-011, A46930-012, A46930-013, A46930-014, A46930-015, A46930-016, A46930-017, A46930-018, A46930-019, A46930-020, A46930-021, A46930-022, A46930-023, A46930-024, A46930-025, A46930-026, A46930-027, A46930-028, A46930-029, A46930-030, A46930-031, A46930-032, A46930-033, A46930-034, A469

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1328-2023

Related Recalls

Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11MM, Model Number: 6068113. 6. SPACER 6068116 CATALYFT PL LONG 11MM, Model Number: 6068116. 7. SPACER 6069073 CATALYFT PL40 SHORT 7MM, Model Number: 6069073. 8. SPACER 6069076 CATALYFT PL40 LONG 7MM, Model Number: 6069076. 9. SPACER 6069093 CATALYFT PL40 SHORT 9MM, Model Number: 6069093. 10. SPACER 6069096 CATALYFT PL40 LONG 9MM, Model Number: 6069096. 11. SPACER 6069113 CATALYFT PL40 SHORT 11MM, Model Number: 6069113. 12. SPACER 6069116 CATALYFT PL40 LONG 11MM, Model Number: 6069116. 13. CATALYFT PL IMPLANT SET, Model Number: SPS03143. 14. CATALYFT PL 40 IMPLANT SET, Model Number: SPS03144. 15. KIT CATALYFT PL IMPLANTS, Model Number: SPS04012. 16. KIT CATALYFT PL 40 IMPLANTS, Model Number: SPS04013. 17. CATALYFT PL 40 IMP KIT, Model Number: SPSJEP40. 18. CATALYFT PL 40 HIGH IMP KIT, Model Number: SPSJEP4H. 19. CATALYFT PL 40 SHORT IMP KIT, Model Number: SPSJEP4S. 20. CATALYFT PL 40 SHORT HIGH IMP KIT, Model Number: SPSJEP4T. 21. CATALYFT PL STRAIGHT HIGH IMP KIT, Model Number: SPSJEPSH. 22. CATALYFT PL STRAIGHT IMP KIT, Model Number: SPSJEPSI. These systems utilize the following affected Instructions for Use (IFU), Document M333023W048E, and Surgical Technique Guide, Document M333023W217.

Medtronic Sofamor Danek USA

Class I - Dangerous

Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.

Sep 16, 2025 Implants & Prosthetics Nationwide View Details →
Class I - Dangerous

Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.

Sep 26, 2023 Other Medical Devices View Details →