Medtronic Grafton Crunch a) DBM S44105 GRAFTON 5CC CRUNCH, REF S44105; b) DBM S44115 GRAFTON 15CC CRUNCH, REF S44115; c) DBM T44105 5CC GRAFTON CRUNCH, REF T44105; d) DBM T44105INT 5CC GRAFTON CRUNCH, REF T44105INT; e) DBM T44115 15CC GRAFTON CRUNCH, REF T44115; f) DBM T44115INT 15CC GRAFTON CRUNCH, REF T44115INT; Demineralized Bone Matrix (DBM) product
Class I - DangerousWhat Should You Do?
- Check if you have this product: a) DBM S44105 GRAFTON 5CC CRUNCH, REF S44105, GTIN 00613994961099, Serial Numbers: A41055, A41991; b) DBM S44115 GRAFTON 15CC CRUNCH, REF S44115, GTIN 00613994961082, Serial Numbers: A46383, A43751, A43940, A44572; c) DBM T44105 5CC GRAFTON CRUNCH, REF T44105, GTIN 00643169123120, Serial Numbers: A38507-041, A38507-042, A38507-043, A38507-044, A38507-045, A38507-046, A38507-047, A38507-048, A38507-049, A38507-050, A40379-096, A40379-097, A40379-098, A40379-099, A40379-100, A40379-101, A40379-102, A40379-103, A40379-104, A40379-105, A40379-106, A40379-107, A40379-108, A40379-109, A40379-110, A40379-111, A40379-112, A40379-113, A40379-114, A40379-115, A40379-116, A41167-031, A41167-032, A41167-033, A41167-036, A41167-037, A41167-038, A41167-039, A41167-040, A41167-041, A41167-042, A41167-043, A41167-044, A41167-045, A41167-046, A41167-047, A41167-048, A41167-049, A41167-050, A41167-051, A41167-052, A41167-053, A41167-054, A41167-055, A41167-056, A41167-057, A41167-058, A41167-059, A41167-060, A41167-061, A41167-062, A41167-063, A41167-064, A41167-065, A41167-066, A41167-067, A41167-068, A41167-070, A41169-181, A41169-182, A41169-183, A41169-184, A41169-185, A41169-186, A41169-187, A41169-189, A41169-192, A41169-193, A41169-194, A41169-195, A41169-196, A41169-197, A41169-198, A41169-199, A41169-200, A41169-201, A41169-202, A41169-203, A41169-204, A41568-135, A41568-136, A41568-137, A41568-138, A41568-139, A41568-140, A41568-141, A41568-142, A41568-143, A41568-144, A41568-145, A41568-146, A41568-147, A41568-148, A41568-149, A41989-232, A41989-233, A41989-234, A41989-235, A41989-236, A41989-237, A41989-238, A41989-239, A41989-240, A41989-241, A41989-242, A41989-243, A41989-244, A41989-245, A41989-246, A41989-247, A41989-248, A41989-249, A41989-250, A41989-251, A41989-252, A41989-253, A41989-254, A41989-255, A42288-123, A42288-124, A42288-125, A42288-126, A42288-127, A42288-128, A42288-129, A42288-130, A42288-131, A42288-132, A42288-133, A42288-134, A42288-135, A42288-136, A42288-137, A42288-138, A42288-139, A42288-140, A42288-141, A42288-142, A42288-143, A42288-144, A42288-145, A42288-146, A42288-147, A42288-148, A42288-149, A42288-150, A42288-151, A42288-152, A42753-135, A42753-136, A42753-137, A42753-138, A42753-139, A42753-140, A42753-141, A42753-142, A42753-143, A42753-144, A42753-145, A42753-146, A42753-147, A42753-148, A42753-149, A42753-150, A42753-151, A42833-077, A42833-078, A42833-079, A42833-080, A42833-081, A42833-082, A42833-083, A42833-084, A42833-085, A42833-086, A42833-087, A42833-088, A42833-089, A42833-090, A42833-091, A42833-092, A42934-052, A42934-053, A42934-054, A42934-055, A42934-056, A42934-057, A42934-058, A42934-059, A42934-060, A42934-061, A42934-062, A42934-063, A42934-064, A42934-065, A42934-066, A42934-067, A42934-068, A42934-069, A42934-070, A42934-071, A42934-072, A42934-073, A42934-074, A42934-075, A42934-076, A42934-077, A42934-078, A42934-079, A42934-080, A42934-081, A42934-082, A42934-083, A42934-084, A42984-055, A42984-056, A42984-057, A42984-058, A42984-059, A42984-060, A42984-061, A43116-077, A43116-078, A43116-079, A43116-080, A43116-081, A43116-082, A43116-083, A43116-084, A43116-085, A43116-086, A43116-087, A43116-088, A43116-089, A43116-090, A43116-091, A43116-092, A43116-093, A43116-094, A43116-095, A43116-096, A43344-174, A43344-175, A43344-176, A43344-177, A43905-096, A43905-097, A43905-098, A43905-099, A43905-100, A43905-101, A43905-102, A43905-103, A43905-104, A43905-105, A43905-106, A43905-107, A43905-108, A43905-109, A43905-110, A43905-111, A43905-112, A43905-113, A43905-114, A43905-115, A43905-116, A43905-117, A44372-105, A44372-106, A44372-107, A44372-108, A44372-109, A44372-110, A44372-111, A44372-112, A44372-113, A44372-114, A44372-115, A44372-116, A44372-117, A44372-118, A44372-119, A44372-120, A44372-121, A44372-122, A44372-123, A44372-124, A44372-125, A44372-126, A44372-127, A44372-128, A44852-074, A44852-075, A44852-076, A44852-077, A44852-078, A44852-079, A44852-080, A44852-081, A44852-082, A44852-083, A44852-084, A44852-085, A44852-086, A44852-087, A44852-088, A44852-089, A44852-090, A45132-059, A45132-060, A45132-061, A45132-062, A45132-063, A45132-064, A45132-065, A45132-066, A45132-067, A45132-068, A45132-069, A45132-070, A45132-071, A45132-072, A45132-073, A45132-074, A45132-075, A45132-076, A45132-077, A45132-078, A45132-079, A45281-101, A45281-102, A45281-103, A45281-104, A45281-105, A45281-106, A45281-107, A45281-108, A45281-109, A45281-110, A45281-111, A45281-112, A45281-113, A45281-114, A45281-115, A45281-116, A45364-089, A45364-090, A45364-091, A45364-092, A45364-093, A45364-094, A45364-095, A45364-096, A45364-097, A45364-098, A45364-099, A45364-100, A45364-101, A45364-102, A45364-103, A45364-104, A45364-105, A45364-106, A45364-107, A45364-108, A45364-109, A45364-110, A45364-111, A45364-112, A45364-113, A45364-114, A45364-115, A45364-116, A45364-117, A45364-118, A45364-119, A45364-120, A45607-089, A45607-090, A45607-091, A45607-092, A45607-093, A45607-094, A45607-095, A45607-096, A45607-098, A45607-099, A45607-100, A45607-102, A45607-103, A45607-104, A45607-105, A45607-106, A45607-107, A45607-108, A45607-110, A45607-111, A45607-112, A45607-113, A45607-114, A45607-115, A45607-116, A45607-117, A45607-118, A45607-119, A46579-166, A46579-167, A46579-168, A46579-169, A46579-170, A46579-171, A46579-172, A46579-173, A46579-174, A46579-175, A46579-176, A46579-177, A46579-178, A46579-179, A46579-180, A46579-181, A46579-182, A46579-183, A46579-184, A46579-185, A46579-186, A46579-187, A46579-188, A46579-189, A47032-099, A47032-100, A47032-101, A47032-102, A47032-103, A47032-104, A47032-105, A47032-107, A47032-108, A47032-109, A47032-110, A47032-111, A47032-112, A47032-113, A47032-114, A47032-116, A47032-117, A47032-118, A47032-119, A47087-021, A47087-022, A47087-023, A47087-027, A47087-028, A47087-029, A47087-030, A47087-031, A47087-032, A47087-033, A47087-034, A47087-035, A47087-036, A47087-037, A47087-038, A47087-039, A47087-040, A47087-041, A47087-042, A47087-043, A47087-044, A47087-045, A47087-046, A47087-047, A47087-049, A47087-050, A47087-051, A47087-052, A47087-053, A47087-054, A47087-055, A47087-056, A47087-057, A47087-058, A47087-059, A47129-073, A47129-074, A47129-075, A47129-076, A47129-077, A47129-078, A47129-079, A47129-080, A47129-081, A47129-082, A47129-083, A47129-084, A47129-085, A47129-086, A47129-087, A47129-088, A47129-089, A47129-090, A47129-091, A47129-092, A47129-093, A47129-094, A47129-095, A47129-096, A47129-097, A47129-098, A47129-099, A47129-100, A47129-101, A47129-102, A47129-103, A47129-104, A48098-037, A48098-038, A48098-039, A48098-040, A48098-041, A48098-042, A48098-043, A48098-044, A48098-045, A48098-046, A48098-047, A48098-048, A48145-050, A48145-051, A48145-052, A48145-053, A48145-054, A48145-055, A48145-056, A48145-057, A48145-058, A48145-059, A48145-060, A48145-061, A48145-062, A48145-063, A48145-064, A48145-065, A48145-066, A48145-067, A48145-068, A48145-069, A48145-070, A48145-071, A48145-072, A48145-073, A48145-074, A48145-075, A48145-076, A48145-077, A48145-078, A48145-079, A48145-080, A48593-138, A48593-139, A48593-140, A48593-141, A48593-142, A48593-143, A48593-144, A48593-145, A48593-146, A48593-147, A48593-148, A48593-149, A48593-150, A48593-151, A48593-152, A48593-153, A48593-154, A48593-155, A48593-156, A48593-157, A48593-158, A48593-159, A48593-160, A48593-161, A48593-162, A48593-163, A48593-164, A48991-120, A48991-121, A48991-122, A48991-123, A48991-124, A48991-125, A48991-126, A48991-127, A48991-128, A48991-129, A48991-130, A48991-131, A48991-132, A48991-133, A48991-134, A48991-135, A48991-136, A48991-137, A48991-138, A48991-139, A48991-140, A48991-141, A48991-142, A48991-143, A48991-144, A48991-145, A48991-146, A48991-147, A48991-148, A48991-149, A48991-150, A48991-151, A48991-152, A48991-153, A48991-154, A48991-155, A48991-156, A48991-157, A48991-158, A48991-159, A48991-160, A48991-161, A48991-162, A49650-046, A49650-047, A49650-048, A49650-049, A49650-050, A49650-051, A49650-052, A49650-053, A49650-054, A49650-055, A49650-056, A49650-057, A49650-058, A49650-059, A49650-060, A49650-061, A49650-062, A49650-063, A49650-064, A49650-065, A49650-066, A49650-067, A49650-068, A49650-069, A49650-070, A49650-071, A49710-104, A49710-105, A49710-106, A49710-107, A49710-108, A49710-109, A49710-110, A49710-111, A49747-113, A49747-114, A49747-115, A49747-116, A49747-117, A49747-118, A49747-119, A49747-120, A49747-121, A49747-122, A49747-123, A49747-124, A49747-125, A49747-126, A49747-127, A49747-128, A49747-129, A49747-130, A49747-131, A49747-132, A49747-133, A49747-134, A49747-135, A49747-136, A49864-050, A49864-051, A49864-052, A49864-053, A49864-054, A49864-055, A49864-056, A49864-057, A49864-058, A49864-059, A49864-060, A49864-061, A49864-062, A50081-080, A50081-081, A50081-082, A50081-083, A50081-084, A50081-085, A50081-086, A50081-087, A50081-088, A50081-089, A50081-090, A50081-091, A50081-092, A50081-093, A50081-094, A50081-095, A50081-096, A50081-097, A50081-098, A50081-099, A50081-100, A50081-101, A50580-026, A50580-027, A50580-028, A50580-029, A50580-030, A50580-031, A50580-032, A50580-033, A50580-034, A50580-035, A50580-036, A50580-037, A50580-038, A50580-039, A50580-040, A50580-041, A50580-042, A50580-043, A50580-044, A50580-045, A50580-046, A50580-047, A50844-064, A50844-065, A50844-066, A50844-067, A50844-068, A50844-069, A50844-070, A50844-071, A50844-072, A50844-073, A50844-074, A51585-064, A51585-065, A51585-066, A51585-067, A51585-068, A51585-069, A51585-070, A51585-071, A51585-072, A51585-073, A51585-074, A51585-075, A51585-076, A51585-077, A51585-078, A51585-079, A51585-080, A51585-081, A51585-082, A51585-083, A51585-084, A51585-085, A51585-086, A51585-087, A51585-088, A51585-089, A51585-090, A51585-091, A51585-092, A51585-093, A51585-094, A51585-095, A51585-096, A52421-050, A52421-051, A52421-052, A52421-053, A52421-054, A52421-055, A52421-056, A52421-057, A52421-058, A52421-059, A52421-060, A52421-061, A52421-062, A52421-063, A52421-064, A52421-065, A52421-066, A52421-067, A52421-068, A52421-069, A52421-070, A52421-071, A52421-072, A52421-073, A52421-074, A52421-075, A52421-076, A52421-077, A52421-078, A52421-079, A52421-080, A52421-081, A52421-082, A52421-083, A52421-084, A52421-085, A52421-086, A52421-087, A52421-088, A52421-089, A52421-090, A52460-181, A52460-182, A52460-183, A52460-184, A52460-185, A52460-186, A52460-187, A52460-188, A52460-189, A52460-190, A52711-031, A52711-032, A52711-033, A52711-034, A52711-035, A52711-036, A52711-037, A52711-038, A52711-039, A52711-040, A52711-041, A52711-042, A52711-043, A52711-044, A52711-045, A52711-046, A52711-047, A52711-048, A52711-049, A52711-050, A52711-051, A52711-052, A52711-053, A52711-054, A52711-055, A52711-056, A52711-057, A52711-058, A52711-059, A52711-060, A52711-061, A52711-062, A52711-063, A52711-064, A52711-065, A52711-066, A52711-067, A53585-026, A53585-027, A53585-028, A53585-029, A53585-030, A53585-031, A53585-032, A53585-033, A53585-034, A53585-035, A53585-036, A53585-037, A53585-038, A53585-039, A53585-040, A53585-041, A53585-042, A53585-043, A53585-044, A53585-045, A53585-046, A53585-047, A53585-048, A53585-049, A53585-050, A53585-051, A53585-053, A53751-053, A53751-054, A53751-055, A53751-056, A53751-057, A53751-058, A53751-059, A53751-060, A53751-061, A53751-062, A53751-063, A53751-064, A53751-065, A53751-066, A53751-067, A53751-068, A53751-069, A53751-070, A53751-071, A53751-072, A53751-073, A53751-074, A53751-075, A53751-076, A53751-077, A53751-078, A53751-079, A53751-080, A53751-081, A53756-127, A53756-128, A53756-129, A53756-130, A53756-131, A53756-132, A53756-133, A53756-134, A53756-135, A53756-136, A53756-137, A53756-138, A53756-139, A53756-140, A53756-141, A53756-142, A53756-143, A53756-144, A53756-145, A53756-146, A53756-147, A53756-148, A53756-149, A53756-150, A53756-151, A53756-152, A53756-153, A53756-154, A54021-026, A54021-027, A54021-028, A54021-029, A54021-030, A54021-031, A54021-032, A54021-033, A54021-034, A54021-035, A54021-036, A54021-037, A54021-038, A54021-039, A54021-040, A54021-041, A54021-042, A54021-043, A54021-044, A54119-097, A54119-098, A54119-099, A54119-100, A54119-101, A54119-102, A54119-103, A54119-104, A54119-105, A54119-106, A54119-107, A54119-108, A54119-109, A54464-113, A54464-114, A54464-115, A54675-026, A54675-027, A54675-028, A54675-029, A54675-030, A54675-031, A54675-032, A54675-033, A54675-034, A54675-035, A54675-036, A54675-037, A54675-038, A54675-039, A54675-040, A54675-041, A54675-042, A54675-043, A54675-044, A54675-045, A54675-046, A54675-047, A54675-048, A54675-049, A54675-050, A54675-051, A54675-052, A54675-053, A54675-054, A54675-055, A54675-056, A54675-057, A54675-058, A54675-059, A54675-060, A54675-061, A54675-062, A54675-063, A54675-064, A55091-031, A55091-032, A55091-033, A55091-034, A55091-036, A55091-037, A55091-038, A55091-039, A55091-040, A55091-041, A55091-042, A55091-043, A55091-044, A55091-045, A55091-046, A55091-047, A55091-048, A55091-049, A55091-050, A55091-051, A55091-052, A55091-053, A55091-054, A55091-055, A55091-056, A55091-058, A55091-059, A55091-060, A55091-061, A55091-062, A55091-063, A55091-064; d) DBM T44105INT 5CC GRAFTON CRUNCH, REF T44105INT, GTIN 00643169965034, Serial Numbers: A38507-051, A38507-052, A38507-053, A38507-054, A38507-055, A38507-056, A38507-057, A38507-058, A38507-059, A38507-060, A38507-061, A38507-062, A38507-063, A38507-064, A38507-065, A38507-066, A38507-067, A39972-046, A39972-047, A39972-048, A39972-049, A39972-050, A39972-051, A39972-052, A39972-053, A39972-054, A39972-055, A39972-056, A39972-057, A39972-058, A39972-059, A39972-060, A39972-061, A39972-062, A39972-063, A39972-064, A39972-065, A39972-066, A39972-067, A39972-068, A39972-069, A39972-070, A39972-071, A39972-072, A39972-073, A39972-074, A39972-075, A39972-076, A39972-077, A39972-078, A39972-079, A39972-080, A41213-031, A41213-032, A41213-033, A41213-034, A41213-035, A41213-036, A41213-037, A41213-038, A41213-039, A41213-040, A41213-041, A41213-042, A41213-043, A41213-044, A41213-045, A41213-046, A41213-047, A41213-048, A41213-049, A41213-050, A41213-051, A41213-052, A41213-053, A41213-054, A41213-055, A41516-026, A41516-027, A41516-028, A41516-029, A41516-030, A41516-031, A41516-032, A41516-033, A41516-034, A41516-035, A41516-036, A41516-037, A41516-038, A41516-039, A41516-040, A41516-041, A41516-042, A41516-043, A41516-045, A41516-047, A41516-048, A41516-049, A41516-050, A41516-052, A41516-053, A41516-056, A42221-281, A42221-282, A42221-283, A42323-034, A42323-035, A42323-036, A42323-037, A42323-038, A42323-039, A42323-040, A42323-041, A42323-042, A42323-043, A42323-044, A42323-045, A42323-046, A42323-047, A42323-048, A42323-049, A42323-050, A42323-051, A42323-052, A42323-053, A42323-054, A42323-055, A42323-056, A42323-057, A42323-058, A42323-059, A42753-152, A42753-153, A42753-154, A42753-155, A42753-156, A42753-157, A42760-066, A42760-067, A42760-068, A42760-069, A42760-071, A42760-073, A42760-075, A42760-076, A42760-077, A42760-078, A42760-081, A42760-082, A42760-084, A42760-085, A42760-086, A42760-087, A42760-088, A42760-089, A42760-090, A42760-091, A42760-092, A42760-093, A42760-094, A42934-087, A42934-088, A42934-089, A42934-090, A42984-033, A42984-034, A42984-035, A42984-036, A42984-037, A42984-038, A42984-039, A42984-040, A42984-041, A42984-042, A42984-043, A42984-044, A42984-045, A42984-046, A42984-047, A42984-048, A42984-049, A42984-050, A42984-051, A42984-052, A42984-053, A42984-054, A43103-183, A43103-184, A43103-185, A43103-186, A43103-187, A43103-188, A43103-189, A43103-190, A43103-191, A43103-192, A43103-193, A43103-194, A43103-195, A43103-196, A43103-197, A43103-198, A43103-199, A43103-200, A43103-201, A43103-202, A43103-203, A43103-204, A43103-205, A43103-206, A43103-207, A43103-208, A43103-209, A43103-210, A43344-155, A43344-156, A43344-157, A43344-158, A43344-159, A43344-160, A43344-161, A43344-163, A43344-164, A43344-165, A43344-168, A43344-169, A43344-170, A43344-171, A43344-172, A43344-173, A43546-101, A43546-102, A43546-103, A43546-104, A43546-105, A43546-106, A43546-107, A43546-108, A43546-109, A43546-110, A43546-111, A43546-112, A43546-113, A43546-114, A43546-115, A43546-116, A43571-095, A43571-096, A43571-097, A43571-098, A43571-099, A43571-100, A43571-101, A43571-102, A43571-103, A43571-104, A43571-105, A43571-106, A43571-107, A43571-108, A43571-109, A43571-110, A43571-111, A43571-112, A43571-113, A43571-114, A44852-091, A44852-092, A44852-093, A44852-094, A44852-095, A44852-096, A45084-086, A45084-087, A45084-088, A45084-089, A45084-090, A45084-091, A45084-092, A45084-093, A45084-094, A45084-095, A45084-096, A45084-097, A45084-098, A45084-099, A45084-100, A45084-101, A45084-102, A45084-103, A45084-104, A45084-105, A45084-106, A45084-107, A45084-108, A45084-109, A45084-110, A45084-111, A45084-112, A45084-113, A45084-114, A45084-115, A46628-221, A46628-222, A46628-223, A46628-224, A46628-225, A46628-226, A46628-227, A46628-228, A46628-229, A46628-230, A46628-231, A46628-232, A46628-233, A46628-234, A46628-235, A46628-236, A46628-237, A46628-238, A46628-239, A46628-240, A46628-241, A46628-242, A46628-243, A46628-244, A46628-245, A46628-246, A46628-247, A46628-248, A46628-249, A46628-250, A46628-251, A47081-045, A47081-046, A47081-047, A47081-048, A47081-049, A47081-050, A47081-051, A47081-052, A47081-053, A47081-054, A47081-055, A47081-056, A47081-057, A47081-058, A47081-059, A47081-060, A47087-060, A47087-061, A47087-062, A47087-063, A47087-064, A47087-065, A47087-066, A47087-067, A47087-068, A47087-069, A47087-070, A47087-071, A47087-072, A47087-073, A47087-074, A48098-049, A48098-050, A48098-051, A48098-052, A48098-053, A48098-054, A48098-055, A48098-056, A48098-057, A48098-058, A48098-059, A48098-060, A48098-061, A48098-062, A49710-112, A49710-113, A49710-114, A49710-115, A49710-116, A49710-117, A49710-119, A49710-120, A49710-121, A49710-122, A49710-123, A49710-124, A49710-125, A49710-126, A49710-127, A49747-137, A49747-138, A49747-139, A49747-140, A49747-141, A49747-142, A49747-143, A49747-144, A49747-145, A50082-058, A50082-059, A50082-060, A50082-061, A50082-062, A50082-063, A50082-064, A50082-065, A50082-066, A50082-067, A50082-068, A50082-069, A50082-070, A50082-071, A50082-072, A50082-073, A50082-074, A50082-075, A50082-076, A50082-077, A50082-078, A50082-079, A50082-080, A50082-081, A50082-082, A50082-083, A50082-084, A50082-085, A50082-086, A50082-087, A50082-088, A50082-089, A50968-031, A50968-032, A50968-033, A50968-034, A50968-035, A50968-036, A50968-037, A50968-038, A50968-039, A50968-040, A50968-041, A50968-042, A50968-043, A50968-044, A50968-045, A50968-046, A50968-047, A50968-048, A50968-049, A50968-050, A50968-051, A50968-052, A50968-053, A50968-054, A50968-055, A51981-026, A51981-027, A51981-028, A51981-029, A51981-030, A51981-031, A51981-032, A51981-033, A51981-034, A51981-035, A51981-036, A51981-037, A51981-038, A51981-039, A51981-040, A51981-041, A51981-042, A51981-043, A51981-044, A51981-045, A51981-046, A51981-047, A51981-048, A52460-162, A52460-163, A52460-164, A52460-165, A52460-166, A52460-167, A52460-168, A52460-169, A52460-170, A52460-171, A52460-173, A52460-180, A54119-110, A54119-111, A54119-112, A54119-113, A54119-114, A54119-115, A54119-116, A54119-117, A54119-118, A54119-119, A54119-120, A54119-121, A54464-086, A54464-087, A54464-088, A54464-089, A54464-093, A54464-094, A54464-095, A54464-103, A54464-107; e) DBM T44115 15CC GRAFTON CRUNCH, REF T44115, GTIN 00643169123113, Serial Numbers: A26631-051, A26631-052, A26631-053, A26631-054, A26631-055, A26631-056, A26631-057, A38377-125, A38377-126, A38377-127, A38377-128, A38377-129, A38377-130, A38377-131, A38377-132, A38377-133, A38471-041, A38471-042, A38471-043, A38471-044, A38471-045, A38471-046, A38471-047, A38472-047, A38472-048, A38472-049, A38472-050, A38472-051, A38472-052, A38796-048, A38796-049, A38796-050, A38796-051, A38796-052, A38796-053, A38796-054, A38796-055, A38796-056, A38796-057, A38796-058, A38796-059, A38796-060, A38796-061, A38796-062, A38796-063, A38796-064, A38796-065, A38796-066, A38796-067, A39047-108, A39047-109, A39047-110, A39047-111, A39047-112, A39047-113, A39091-085, A39091-086, A39091-087, A39091-088, A39091-089, A39091-090, A39091-091, A39091-092, A39091-093, A39091-094, A39111-182, A39111-183, A39111-184, A39111-185, A39111-186, A39111-187, A39111-188, A39111-189, A39111-190, A39111-191, A39111-192, A39111-193, A39111-194, A39111-195, A39111-196, A39202-053, A39202-054, A39202-055, A39202-056, A39202-057, A39202-058, A39202-059, A39202-060, A39202-061, A39202-062, A39202-063, A39202-064, A39202-065, A39202-066, A39202-067, A39202-068, A39202-069, A39202-070, A39202-071, A39245-046, A39245-047, A39245-048, A39245-049, A39245-050, A39245-051, A39245-052, A39245-053, A39245-054, A39245-055, A39245-056, A39283-114, A39283-115, A39283-116, A39283-117, A39283-118, A39283-119, A39413-267, A39413-268, A39413-269, A39413-270, A39413-271, A39413-272, A39413-273, A39413-274, A39413-275, A39413-276, A39413-277, A39512-026, A39512-027, A39512-028, A39512-029, A39512-030, A39512-031, A39512-032, A39512-033, A39512-034, A39512-035, A39512-036, A39512-037, A39512-038, A39512-039, A39512-040, A39512-041, A39512-042, A39512-043, A39512-044, A39512-045, A39512-046, A39512-047, A39516-092, A39516-093, A39516-094, A39516-095, A39516-096, A39520-103, A39520-104, A39520-105, A39520-106, A39520-107, A39520-108, A39520-109, A39520-110, A39520-111, A39811-170, A39811-171, A39811-172, A39811-173, A39811-174, A39811-175, A39811-176, A39811-177, A39811-178, A39811-179, A39811-180, A39866-164, A39866-165, A39866-166, A39866-167, A39866-168, A39866-169, A39866-170, A39866-171, A39866-172, A39866-173, A39867-049, A39867-050, A39867-051, A39867-052, A39867-053, A39867-054, A39867-055, A39867-056, A39867-057, A39962-052, A39962-053, A39962-054, A39962-055, A39962-056, A39962-057, A39962-058, A40005-103, A40005-104, A40005-105, A40005-106, A40005-107, A40005-108, A40005-109, A40005-110, A40005-111, A40005-112, A40108-144, A40108-145, A40108-146, A40108-147, A40108-148, A40108-149, A40108-150, A40108-151, A40108-152, A40108-153, A40108-154, A40108-155, A40108-156, A40112-053, A40112-054, A40112-055, A40112-056, A40112-057, A40112-058, A40112-059, A40112-060, A40112-061, A40112-062, A40112-063, A40138-042, A40138-043, A40138-044, A40138-045, A40138-046, A40138-047, A40138-048, A40138-049, A40173-162, A40173-163, A40173-164, A40173-165, A40173-166, A40173-167, A40173-168, A40173-169, A40173-170, A40246-056, A40246-057, A40246-058, A40246-059, A40246-060, A40246-061, A40246-062, A40246-063, A40309-097, A40309-098, A40309-099, A40309-100, A40309-101, A40309-102, A40309-103, A40309-104, A40309-105, A40309-106, A40309-107, A40309-108, A40309-109, A40309-110, A40309-111, A40453-026, A40453-027, A40453-028, A40453-029, A40453-030, A40453-031, A40453-032, A40453-033, A40453-034, A40453-035, A40543-075, A40543-076, A40543-077, A40543-078, A40543-079, A40543-080, A40543-081, A40543-082, A40543-083, A40543-084, A40543-085, A40652-126, A40652-127, A40652-128, A40652-129, A40652-130, A40652-131, A40733-040, A40733-041, A40733-042, A40733-043, A40733-044, A40733-045, A40733-046, A40733-047, A40733-048, A40733-049, A40733-050, A40733-051, A40733-052, A40733-053, A40736-081, A40736-082, A40736-083, A40736-084, A40736-085, A40736-086, A40736-087, A40736-088, A40736-089, A40736-090, A40736-091, A40913-086, A40913-087, A40913-088, A40913-089, A40913-090, A40913-091, A40913-092, A40913-093, A40913-094, A40913-095, A40913-096, A40915-101, A40915-102, A40915-103, A40915-104, A40915-105, A40915-106, A40915-107, A40915-108, A40915-109, A40915-110, A40915-111, A40915-112, A40915-113, A40915-114, A40915-115, A40915-116, A40915-117, A40915-118, A40953-174, A40953-175, A40953-176, A40953-177, A40953-178, A40953-179, A40953-180, A40953-181, A40953-182, A40989-196, A40989-197, A40989-198, A40989-199, A40989-200, A40989-201, A40989-202, A40989-203, A40989-204, A40989-205, A40989-206, A40999-031, A40999-032, A40999-033, A40999-034, A40999-035, A40999-036, A40999-037, A40999-038, A41012-047, A41012-048, A41012-049, A41012-050, A41012-051, A41012-052, A41012-053, A41012-054, A41012-055, A41012-056, A41012-057, A41012-058, A41012-059, A41012-060, A41012-061, A41012-062, A41012-063, A41270-121, A41270-122, A41270-123, A41270-124, A41270-125, A41270-126, A41270-127, A41270-128, A41270-129, A41270-130, A41270-131, A41270-132, A41270-133, A41270-134, A41270-135, A41270-136, A41270-137, A41270-138, A41270-139, A41270-140, A41273-031, A41273-032, A41273-033, A41273-034, A41273-035, A41273-036, A41397-031, A41397-032, A41397-033, A41397-034, A41397-035, A41397-036, A41400-052, A41400-053, A41400-054, A41400-055, A41400-056, A41400-057, A41400-058, A41400-059, A41400-060, A41439-177, A41439-178, A41439-179, A41439-180, A41439-181, A41464-210, A41464-211, A41464-212, A41464-213, A41464-214, A41464-215, A41464-216, A41464-217, A41464-218, A41472-174, A41472-175, A41472-176, A41472-177, A41472-178, A41472-179, A41472-180, A41472-181, A41472-182, A41472-183, A41630-031, A41630-032, A41630-033, A41630-034, A41630-035, A41630-036, A41630-037, A41630-038, A41630-039, A41630-040, A41630-041, A41630-042, A41630-043, A41630-044, A41630-045, A41630-046, A41630-047, A41630-048, A41784-086, A41784-087, A41784-088, A41784-089, A41784-090, A41784-091, A41784-092, A41784-093, A41784-094, A41784-095, A41784-096, A41784-097, A41879-091, A41879-092, A41879-093, A41879-094, A41879-095, A41879-096, A41879-097, A41879-098, A41879-099, A41879-100, A41879-101, A41879-102, A41879-103, A41879-104, A41905-026, A41905-027, A41905-028, A41905-029, A41905-030, A41905-031, A41905-032, A41905-033, A41905-034, A41991-068, A41991-069, A41991-070, A41991-071, A41991-072, A41991-073, A41991-074, A41991-075, A41991-076, A41991-077, A41991-078, A41995-118, A41995-119, A41995-120, A41995-121, A41995-122, A41995-123, A41995-124, A41995-125, A41995-126, A41995-127, A41995-128, A41995-129, A41995-130, A41995-131, A41995-132, A42080-046, A42080-047, A42080-048, A42080-049, A42080-050, A42080-051, A42080-052, A42080-053, A42080-054, A42080-055, A42102-046, A42102-047, A42102-048, A42102-049, A42102-050, A42102-051, A42102-052, A42102-053, A42102-054, A42102-055, A42102-056, A42102-057, A42102-058, A42102-059, A42102-060, A42102-061, A42193-088, A42193-089, A42193-090, A42193-091, A42193-092, A42193-093, A42193-094, A42193-095, A42193-096, A42225-043, A42225-044, A42225-045, A42225-046, A42225-047, A42225-048, A42225-049, A42225-050, A42225-051, A42225-052, A42225-053, A42225-054, A42225-055, A42225-056, A42225-057, A42225-058, A42225-059, A42225-060, A42225-061, A42234-061, A42234-062, A42234-063, A42234-064, A42234-065, A42234-066, A42234-067, A42234-068, A42251-078, A42251-079, A42251-080, A42251-081, A42251-082, A42251-083, A42251-084, A42251-085, A42251-086, A42251-087, A42251-088, A42251-089, A42251-090, A42251-091, A42251-092, A42282-133, A42282-134, A42282-135, A42282-136, A42282-137, A42282-138, A42282-139, A42282-140, A42282-141, A42282-142, A42282-143, A42282-144, A42539-046, A42539-047, A42539-048, A42539-049, A42539-050, A42539-051, A42539-052, A42539-053, A42539-054, A42539-055, A42539-056, A42607-057, A42607-058, A42607-059, A42607-060, A42607-061, A42607-062, A42607-063, A42607-064, A42607-065, A42607-066, A42607-067, A42607-068, A42607-069, A42607-070, A42607-071, A42607-072, A42607-073, A42607-074, A42607-075, A42607-076, A42607-077, A42661-092, A42661-093, A42661-094, A42661-095, A42661-096, A42661-097, A42662-129, A42662-130, A42662-131, A42662-132, A42662-133, A42662-134, A42662-135, A42662-136, A42662-137, A42665-085, A42665-086, A42665-087, A42665-088, A42665-089, A42665-090, A42665-091, A42665-092, A42665-093, A42665-094, A42665-095, A42665-096, A42665-097, A42665-098, A42665-099, A42665-100, A42754-147, A42754-148, A42754-149, A42754-150, A42754-151, A42754-152, A42754-153, A42754-154, A42754-155, A42791-064, A42791-065, A42791-066, A42791-067, A42791-068, A42791-069, A42791-070, A42791-071, A42791-072, A42791-073, A42791-074, A42791-075, A42791-076, A42791-077, A42791-078, A42791-079, A42826-053, A42826-054, A42826-055, A42826-056, A42826-057, A42826-058, A42826-059, A42826-060, A42826-061, A42826-062, A42826-063, A42826-064, A42826-065, A42872-053, A42872-054, A42872-055, A42872-056, A42872-057, A42872-058, A42872-059, A42872-060, A42872-061, A42872-062, A42872-063, A42872-064, A42872-065, A42872-066, A42872-067, A42872-068, A42872-069, A42872-070, A42872-071, A42932-026, A42932-027, A42932-028, A42932-029, A42932-030, A42932-031, A42932-032, A42932-033, A42932-034, A42932-035, A42932-036, A42932-037, A42935-155, A42935-156, A42935-157, A42935-158, A42935-159, A42935-160, A42935-161, A42935-162, A42935-163, A42935-164, A42972-089, A42972-090, A42972-091, A42972-092, A42972-093, A42972-094, A42972-095, A42972-096, A42972-097, A42972-098, A42972-099, A42972-100, A42972-101, A42972-102, A42982-031, A42982-032, A42982-033, A42982-034, A42982-035, A42982-036, A42982-037, A42982-038, A43058-034, A43058-035, A43058-036, A43058-037, A43058-038, A43058-039, A43058-040, A43058-041, A43058-042, A43058-043, A43058-044, A43058-045, A43058-046, A43058-047, A43058-048, A43058-049, A43100-081, A43100-082, A43100-083, A43100-084, A43100-085, A43100-086, A43100-087, A43100-088, A43100-089, A43100-090, A43100-091, A43104-056, A43104-057, A43104-058, A43104-059, A43104-060, A43104-061, A43104-062, A43104-063, A43104-064, A43104-065, A43104-066, A43104-067, A43104-068, A43104-069, A43104-070, A43104-071, A43104-072, A43104-073, A43112-064, A43112-065, A43112-066, A43112-067, A43112-068, A43112-069, A43112-070, A43112-071, A43112-072, A43112-073, A43191-057, A43191-058, A43191-059, A43191-060, A43191-061, A43191-062, A43191-063, A43191-064, A43191-065, A43191-066, A43191-067, A43191-068, A43191-069, A43191-070, A43191-071, A43191-072, A43193-040, A43193-041, A43193-042, A43193-043, A43193-044, A43193-045, A43193-046, A43193-047, A43193-048, A43193-049, A43193-050, A43193-051, A43284-221, A43284-222, A43284-223, A43284-224, A43284-225, A43284-226, A43284-227, A43285-046, A43285-047, A43285-048, A43285-049, A43285-050, A43285-051, A43285-052, A43285-053, A43285-054, A43285-055, A43285-056, A43285-057, A43285-058, A43298-044, A43298-045, A43298-046, A43298-047, A43298-048, A43298-049, A43298-050, A43298-051, A43298-052, A43298-053, A43298-054, A43298-055, A43298-056, A43298-057, A43298-058, A43340-036, A43340-037, A43340-038, A43340-039, A43340-040, A43340-041, A43340-042, A43340-043, A43340-044, A43340-045, A43340-046, A43340-047, A43340-048, A43340-049, A43340-050, A43340-051, A43340-052, A43340-053, A43374-040, A43374-041, A43374-042, A43374-043, A43374-044, A43374-045, A43374-046, A43374-047, A43374-048, A43374-049, A43374-050, A43395-054, A43395-055, A43395-056, A43395-057, A43395-058, A43395-059, A43395-060, A43395-061, A43497-157, A43497-158, A43497-159, A43497-160, A43497-161, A43497-162, A43497-164, A43497-165, A43501-026, A43501-027, A43501-028, A43501-029, A43501-030, A43501-031, A43501-032, A43501-033, A43501-034, A43501-035, A43577-095, A43577-096, A43577-097, A43577-098, A43577-099, A43577-100, A43577-101, A43577-102, A43577-103, A43577-104, A43577-105, A43634-112, A43634-113, A43634-114, A43634-115, A43634-116, A43634-117, A43634-118, A43634-119, A43634-120, A43698-131, A43698-132, A43698-133, A43698-134, A43698-135, A43698-136, A43698-137, A43698-138, A43698-139, A43698-140, A43698-141, A43698-142, A43698-143, A43698-144, A43751-121, A43751-122, A43751-123, A43751-124, A43751-125, A43751-126, A43751-127, A43904-036, A43904-037, A43904-038, A43904-039, A43904-040, A43904-041, A43904-042, A43904-043, A43904-044, A43904-045, A43909-113, A43909-114, A43909-115, A43909-116, A43909-117, A43909-118, A43909-119, A43909-120, A43909-121, A43936-089, A43936-090, A43936-091, A43936-092, A43936-093, A43936-094, A43936-095, A43936-096, A43936-097, A43936-098, A43936-099, A43936-100, A43936-101, A43936-102, A43936-103, A43936-104, A43936-105, A43936-106, A43936-107, A43938-046, A43938-047, A43938-048, A43938-049, A43938-050, A43938-051, A43938-052, A43938-053, A43938-054, A43938-055, A43938-056, A43938-057, A43938-058, A43938-059, A43938-060, A43938-061, A43948-045, A43948-046, A43948-047, A43948-048, A43948-049, A43948-050, A43948-051, A43948-052, A43948-053, A43948-054, A43948-055, A43948-056, A43948-057, A43948-058, A43983-105, A43983-106, A43983-107, A43983-108, A43983-109, A43983-110, A439
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Sofamor Danek USA Inc
- Reason for Recall:
- Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Medtronic Grafton Crunch a) DBM S44105 GRAFTON 5CC CRUNCH, REF S44105; b) DBM S44115 GRAFTON 15CC CRUNCH, REF S44115; c) DBM T44105 5CC GRAFTON CRUNCH, REF T44105; d) DBM T44105INT 5CC GRAFTON CRUNCH, REF T44105INT; e) DBM T44115 15CC GRAFTON CRUNCH, REF T44115; f) DBM T44115INT 15CC GRAFTON CRUNCH, REF T44115INT; Demineralized Bone Matrix (DBM) product
Product Codes/Lot Numbers:
a) DBM S44105 GRAFTON 5CC CRUNCH, REF S44105, GTIN 00613994961099, Serial Numbers: A41055, A41991; b) DBM S44115 GRAFTON 15CC CRUNCH, REF S44115, GTIN 00613994961082, Serial Numbers: A46383, A43751, A43940, A44572; c) DBM T44105 5CC GRAFTON CRUNCH, REF T44105, GTIN 00643169123120, Serial Numbers: A38507-041, A38507-042, A38507-043, A38507-044, A38507-045, A38507-046, A38507-047, A38507-048, A38507-049, A38507-050, A40379-096, A40379-097, A40379-098, A40379-099, A40379-100, A40379-101, A40379-102, A40379-103, A40379-104, A40379-105, A40379-106, A40379-107, A40379-108, A40379-109, A40379-110, A40379-111, A40379-112, A40379-113, A40379-114, A40379-115, A40379-116, A41167-031, A41167-032, A41167-033, A41167-036, A41167-037, A41167-038, A41167-039, A41167-040, A41167-041, A41167-042, A41167-043, A41167-044, A41167-045, A41167-046, A41167-047, A41167-048, A41167-049, A41167-050, A41167-051, A41167-052, A41167-053, A41167-054, A41167-055, A41167-056, A41167-057, A41167-058, A41167-059, A41167-060, A41167-061, A41167-062, A41167-063, A41167-064, A41167-065, A41167-066, A41167-067, A41167-068, A41167-070, A41169-181, A41169-182, A41169-183, A41169-184, A41169-185, A41169-186, A41169-187, A41169-189, A41169-192, A41169-193, A41169-194, A41169-195, A41169-196, A41169-197, A41169-198, A41169-199, A41169-200, A41169-201, A41169-202, A41169-203, A41169-204, A41568-135, A41568-136, A41568-137, A41568-138, A41568-139, A41568-140, A41568-141, A41568-142, A41568-143, A41568-144, A41568-145, A41568-146, A41568-147, A41568-148, A41568-149, A41989-232, A41989-233, A41989-234, A41989-235, A41989-236, A41989-237, A41989-238, A41989-239, A41989-240, A41989-241, A41989-242, A41989-243, A41989-244, A41989-245, A41989-246, A41989-247, A41989-248, A41989-249, A41989-250, A41989-251, A41989-252, A41989-253, A41989-254, A41989-255, A42288-123, A42288-124, A42288-125, A42288-126, A42288-127, A42288-128, A42288-129, A42288-130, A42288-131, A42288-132, A42288-133, A42288-134, A42288-135, A42288-136, A42288-137, A42288-138, A42288-139, A42288-140, A42288-141, A42288-142, A42288-143, A42288-144, A42288-145, A42288-146, A42288-147, A42288-148, A42288-149, A42288-150, A42288-151, A42288-152, A42753-135, A42753-136, A42753-137, A42753-138, A42753-139, A42753-140, A42753-141, A42753-142, A42753-143, A42753-144, A42753-145, A42753-146, A42753-147, A42753-148, A42753-149, A42753-150, A42753-151, A42833-077, A42833-078, A42833-079, A42833-080, A42833-081, A42833-082, A42833-083, A42833-084, A42833-085, A42833-086, A42833-087, A42833-088, A42833-089, A42833-090, A42833-091, A42833-092, A42934-052, A42934-053, A42934-054, A42934-055, A42934-056, A42934-057, A42934-058, A42934-059, A42934-060, A42934-061, A42934-062, A42934-063, A42934-064, A42934-065, A42934-066, A42934-067, A42934-068, A42934-069, A42934-070, A42934-071, A42934-072, A42934-073, A42934-074, A42934-075, A42934-076, A42934-077, A42934-078, A42934-079, A42934-080, A42934-081, A42934-082, A42934-083, A42934-084, A42984-055, A42984-056, A42984-057, A42984-058, A42984-059, A42984-060, A42984-061, A43116-077, A43116-078, A43116-079, A43116-080, A43116-081, A43116-082, A43116-083, A43116-084, A43116-085, A43116-086, A43116-087, A43116-088, A43116-089, A43116-090, A43116-091, A43116-092, A43116-093, A43116-094, A43116-095, A43116-096, A43344-174, A43344-175, A43344-176, A43344-177, A43905-096, A43905-097, A43905-098, A43905-099, A43905-100, A43905-101, A43905-102, A43905-103, A43905-104, A43905-105, A43905-106, A43905-107, A43905-108, A43905-109, A43905-110, A43905-111, A43905-112, A43905-113, A43905-114, A43905-115, A43905-116, A43905-117, A44372-105, A44372-106, A44372-107, A44372-108, A44372-109, A44372-110, A44372-111, A44372-112, A44372-113, A44372-114, A44372-115, A44372-116, A44372-117, A44372-118, A44372-119, A44372-120, A44372-121, A44372-122, A44372-123, A44372-124, A44372-125, A44372-126, A44372-127, A44372-128, A44852-074, A44852-075, A44852-076, A44852-077, A44852-078, A44852-079, A44852-080, A44852-081, A44852-082, A44852-083, A44852-084, A44852-085, A44852-086, A44852-087, A44852-088, A44852-089, A44852-090, A45132-059, A45132-060, A45132-061, A45132-062, A45132-063, A45132-064, A45132-065, A45132-066, A45132-067, A45132-068, A45132-069, A45132-070, A45132-071, A45132-072, A45132-073, A45132-074, A45132-075, A45132-076, A45132-077, A45132-078, A45132-079, A45281-101, A45281-102, A45281-103, A45281-104, A45281-105, A45281-106, A45281-107, A45281-108, A45281-109, A45281-110, A45281-111, A45281-112, A45281-113, A45281-114, A45281-115, A45281-116, A45364-089, A45364-090, A45364-091, A45364-092, A45364-093, A45364-094, A45364-095, A45364-096, A45364-097, A45364-098, A45364-099, A45364-100, A45364-101, A45364-102, A45364-103, A45364-104, A45364-105, A45364-106, A45364-107, A45364-108, A45364-109, A45364-110, A45364-111, A45364-112, A45364-113, A45364-114, A45364-115, A45364-116, A45364-117, A45364-118, A45364-119, A45364-120, A45607-089, A45607-090, A45607-091, A45607-092, A45607-093, A45607-094, A45607-095, A45607-096, A45607-098, A45607-099, A45607-100, A45607-102, A45607-103, A45607-104, A45607-105, A45607-106, A45607-107, A45607-108, A45607-110, A45607-111, A45607-112, A45607-113, A45607-114, A45607-115, A45607-116, A45607-117, A45607-118, A45607-119, A46579-166, A46579-167, A46579-168, A46579-169, A46579-170, A46579-171, A46579-172, A46579-173, A46579-174, A46579-175, A46579-176, A46579-177, A46579-178, A46579-179, A46579-180, A46579-181, A46579-182, A46579-183, A46579-184, A46579-185, A46579-186, A46579-187, A46579-188, A46579-189, A47032-099, A47032-100, A47032-101, A47032-102, A47032-103, A47032-104, A47032-105, A47032-107, A47032-108, A47032-109, A47032-110, A47032-111, A47032-112, A47032-113, A47032-114, A47032-116, A47032-117, A47032-118, A47032-119, A47087-021, A47087-022, A47087-023, A47087-027, A47087-028, A47087-029, A47087-030, A47087-031, A47087-032, A47087-033, A47087-034, A47087-035, A47087-036, A47087-037, A47087-038, A47087-039, A47087-040, A47087-041, A47087-042, A47087-043, A47087-044, A47087-045, A47087-046, A47087-047, A47087-049, A47087-050, A47087-051, A47087-052, A47087-053, A47087-054, A47087-055, A47087-056, A47087-057, A47087-058, A47087-059, A47129-073, A47129-074, A47129-075, A47129-076, A47129-077, A47129-078, A47129-079, A47129-080, A47129-081, A47129-082, A47129-083, A47129-084, A47129-085, A47129-086, A47129-087, A47129-088, A47129-089, A47129-090, A47129-091, A47129-092, A47129-093, A47129-094, A47129-095, A47129-096, A47129-097, A47129-098, A47129-099, A47129-100, A47129-101, A47129-102, A47129-103, A47129-104, A48098-037, A48098-038, A48098-039, A48098-040, A48098-041, A48098-042, A48098-043, A48098-044, A48098-045, A48098-046, A48098-047, A48098-048, A48145-050, A48145-051, A48145-052, A48145-053, A48145-054, A48145-055, A48145-056, A48145-057, A48145-058, A48145-059, A48145-060, A48145-061, A48145-062, A48145-063, A48145-064, A48145-065, A48145-066, A48145-067, A48145-068, A48145-069, A48145-070, A48145-071, A48145-072, A48145-073, A48145-074, A48145-075, A48145-076, A48145-077, A48145-078, A48145-079, A48145-080, A48593-138, A48593-139, A48593-140, A48593-141, A48593-142, A48593-143, A48593-144, A48593-145, A48593-146, A48593-147, A48593-148, A48593-149, A48593-150, A48593-151, A48593-152, A48593-153, A48593-154, A48593-155, A48593-156, A48593-157, A48593-158, A48593-159, A48593-160, A48593-161, A48593-162, A48593-163, A48593-164, A48991-120, A48991-121, A48991-122, A48991-123, A48991-124, A48991-125, A48991-126, A48991-127, A48991-128, A48991-129, A48991-130, A48991-131, A48991-132, A48991-133, A48991-134, A48991-135, A48991-136, A48991-137, A48991-138, A48991-139, A48991-140, A48991-141, A48991-142, A48991-143, A48991-144, A48991-145, A48991-146, A48991-147, A48991-148, A48991-149, A48991-150, A48991-151, A48991-152, A48991-153, A48991-154, A48991-155, A48991-156, A48991-157, A48991-158, A48991-159, A48991-160, A48991-161, A48991-162, A49650-046, A49650-047, A49650-048, A49650-049, A49650-050, A49650-051, A49650-052, A49650-053, A49650-054, A49650-055, A49650-056, A49650-057, A49650-058, A49650-059, A49650-060, A49650-061, A49650-062, A49650-063, A49650-064, A49650-065, A49650-066, A49650-067, A49650-068, A49650-069, A49650-070, A49650-071, A49710-104, A49710-105, A49710-106, A49710-107, A49710-108, A49710-109, A49710-110, A49710-111, A49747-113, A49747-114, A49747-115, A49747-116, A49747-117, A49747-118, A49747-119, A49747-120, A49747-121, A49747-122, A49747-123, A49747-124, A49747-125, A49747-126, A49747-127, A49747-128, A49747-129, A49747-130, A49747-131, A49747-132, A49747-133, A49747-134, A49747-135, A49747-136, A49864-050, A49864-051, A49864-052, A49864-053, A49864-054, A49864-055, A49864-056, A49864-057, A49864-058, A49864-059, A49864-060, A49864-061, A49864-062, A50081-080, A50081-081, A50081-082, A50081-083, A50081-084, A50081-085, A50081-086, A50081-087, A50081-088, A50081-089, A50081-090, A50081-091, A50081-092, A50081-093, A50081-094, A50081-095, A50081-096, A50081-097, A50081-098, A50081-099, A50081-100, A50081-101, A50580-026, A50580-027, A50580-028, A50580-029, A50580-030, A50580-031, A50580-032, A50580-033, A50580-034, A50580-035, A50580-036, A50580-037, A50580-038, A50580-039, A50580-040, A50580-041, A50580-042, A50580-043, A50580-044, A50580-045, A50580-046, A50580-047, A50844-064, A50844-065, A50844-066, A50844-067, A50844-068, A50844-069, A50844-070, A50844-071, A50844-072, A50844-073, A50844-074, A51585-064, A51585-065, A51585-066, A51585-067, A51585-068, A51585-069, A51585-070, A51585-071, A51585-072, A51585-073, A51585-074, A51585-075, A51585-076, A51585-077, A51585-078, A51585-079, A51585-080, A51585-081, A51585-082, A51585-083, A51585-084, A51585-085, A51585-086, A51585-087, A51585-088, A51585-089, A51585-090, A51585-091, A51585-092, A51585-093, A51585-094, A51585-095, A51585-096, A52421-050, A52421-051, A52421-052, A52421-053, A52421-054, A52421-055, A52421-056, A52421-057, A52421-058, A52421-059, A52421-060, A52421-061, A52421-062, A52421-063, A52421-064, A52421-065, A52421-066, A52421-067, A52421-068, A52421-069, A52421-070, A52421-071, A52421-072, A52421-073, A52421-074, A52421-075, A52421-076, A52421-077, A52421-078, A52421-079, A52421-080, A52421-081, A52421-082, A52421-083, A52421-084, A52421-085, A52421-086, A52421-087, A52421-088, A52421-089, A52421-090, A52460-181, A52460-182, A52460-183, A52460-184, A52460-185, A52460-186, A52460-187, A52460-188, A52460-189, A52460-190, A52711-031, A52711-032, A52711-033, A52711-034, A52711-035, A52711-036, A52711-037, A52711-038, A52711-039, A52711-040, A52711-041, A52711-042, A52711-043, A52711-044, A52711-045, A52711-046, A52711-047, A52711-048, A52711-049, A52711-050, A52711-051, A52711-052, A52711-053, A52711-054, A52711-055, A52711-056, A52711-057, A52711-058, A52711-059, A52711-060, A52711-061, A52711-062, A52711-063, A52711-064, A52711-065, A52711-066, A52711-067, A53585-026, A53585-027, A53585-028, A53585-029, A53585-030, A53585-031, A53585-032, A53585-033, A53585-034, A53585-035, A53585-036, A53585-037, A53585-038, A53585-039, A53585-040, A53585-041, A53585-042, A53585-043, A53585-044, A53585-045, A53585-046, A53585-047, A53585-048, A53585-049, A53585-050, A53585-051, A53585-053, A53751-053, A53751-054, A53751-055, A53751-056, A53751-057, A53751-058, A53751-059, A53751-060, A53751-061, A53751-062, A53751-063, A53751-064, A53751-065, A53751-066, A53751-067, A53751-068, A53751-069, A53751-070, A53751-071, A53751-072, A53751-073, A53751-074, A53751-075, A53751-076, A53751-077, A53751-078, A53751-079, A53751-080, A53751-081, A53756-127, A53756-128, A53756-129, A53756-130, A53756-131, A53756-132, A53756-133, A53756-134, A53756-135, A53756-136, A53756-137, A53756-138, A53756-139, A53756-140, A53756-141, A53756-142, A53756-143, A53756-144, A53756-145, A53756-146, A53756-147, A53756-148, A53756-149, A53756-150, A53756-151, A53756-152, A53756-153, A53756-154, A54021-026, A54021-027, A54021-028, A54021-029, A54021-030, A54021-031, A54021-032, A54021-033, A54021-034, A54021-035, A54021-036, A54021-037, A54021-038, A54021-039, A54021-040, A54021-041, A54021-042, A54021-043, A54021-044, A54119-097, A54119-098, A54119-099, A54119-100, A54119-101, A54119-102, A54119-103, A54119-104, A54119-105, A54119-106, A54119-107, A54119-108, A54119-109, A54464-113, A54464-114, A54464-115, A54675-026, A54675-027, A54675-028, A54675-029, A54675-030, A54675-031, A54675-032, A54675-033, A54675-034, A54675-035, A54675-036, A54675-037, A54675-038, A54675-039, A54675-040, A54675-041, A54675-042, A54675-043, A54675-044, A54675-045, A54675-046, A54675-047, A54675-048, A54675-049, A54675-050, A54675-051, A54675-052, A54675-053, A54675-054, A54675-055, A54675-056, A54675-057, A54675-058, A54675-059, A54675-060, A54675-061, A54675-062, A54675-063, A54675-064, A55091-031, A55091-032, A55091-033, A55091-034, A55091-036, A55091-037, A55091-038, A55091-039, A55091-040, A55091-041, A55091-042, A55091-043, A55091-044, A55091-045, A55091-046, A55091-047, A55091-048, A55091-049, A55091-050, A55091-051, A55091-052, A55091-053, A55091-054, A55091-055, A55091-056, A55091-058, A55091-059, A55091-060, A55091-061, A55091-062, A55091-063, A55091-064; d) DBM T44105INT 5CC GRAFTON CRUNCH, REF T44105INT, GTIN 00643169965034, Serial Numbers: A38507-051, A38507-052, A38507-053, A38507-054, A38507-055, A38507-056, A38507-057, A38507-058, A38507-059, A38507-060, A38507-061, A38507-062, A38507-063, A38507-064, A38507-065, A38507-066, A38507-067, A39972-046, A39972-047, A39972-048, A39972-049, A39972-050, A39972-051, A39972-052, A39972-053, A39972-054, A39972-055, A39972-056, A39972-057, A39972-058, A39972-059, A39972-060, A39972-061, A39972-062, A39972-063, A39972-064, A39972-065, A39972-066, A39972-067, A39972-068, A39972-069, A39972-070, A39972-071, A39972-072, A39972-073, A39972-074, A39972-075, A39972-076, A39972-077, A39972-078, A39972-079, A39972-080, A41213-031, A41213-032, A41213-033, A41213-034, A41213-035, A41213-036, A41213-037, A41213-038, A41213-039, A41213-040, A41213-041, A41213-042, A41213-043, A41213-044, A41213-045, A41213-046, A41213-047, A41213-048, A41213-049, A41213-050, A41213-051, A41213-052, A41213-053, A41213-054, A41213-055, A41516-026, A41516-027, A41516-028, A41516-029, A41516-030, A41516-031, A41516-032, A41516-033, A41516-034, A41516-035, A41516-036, A41516-037, A41516-038, A41516-039, A41516-040, A41516-041, A41516-042, A41516-043, A41516-045, A41516-047, A41516-048, A41516-049, A41516-050, A41516-052, A41516-053, A41516-056, A42221-281, A42221-282, A42221-283, A42323-034, A42323-035, A42323-036, A42323-037, A42323-038, A42323-039, A42323-040, A42323-041, A42323-042, A42323-043, A42323-044, A42323-045, A42323-046, A42323-047, A42323-048, A42323-049, A42323-050, A42323-051, A42323-052, A42323-053, A42323-054, A42323-055, A42323-056, A42323-057, A42323-058, A42323-059, A42753-152, A42753-153, A42753-154, A42753-155, A42753-156, A42753-157, A42760-066, A42760-067, A42760-068, A42760-069, A42760-071, A42760-073, A42760-075, A42760-076, A42760-077, A42760-078, A42760-081, A42760-082, A42760-084, A42760-085, A42760-086, A42760-087, A42760-088, A42760-089, A42760-090, A42760-091, A42760-092, A42760-093, A42760-094, A42934-087, A42934-088, A42934-089, A42934-090, A42984-033, A42984-034, A42984-035, A42984-036, A42984-037, A42984-038, A42984-039, A42984-040, A42984-041, A42984-042, A42984-043, A42984-044, A42984-045, A42984-046, A42984-047, A42984-048, A42984-049, A42984-050, A42984-051, A42984-052, A42984-053, A42984-054, A43103-183, A43103-184, A43103-185, A43103-186, A43103-187, A43103-188, A43103-189, A43103-190, A43103-191, A43103-192, A43103-193, A43103-194, A43103-195, A43103-196, A43103-197, A43103-198, A43103-199, A43103-200, A43103-201, A43103-202, A43103-203, A43103-204, A43103-205, A43103-206, A43103-207, A43103-208, A43103-209, A43103-210, A43344-155, A43344-156, A43344-157, A43344-158, A43344-159, A43344-160, A43344-161, A43344-163, A43344-164, A43344-165, A43344-168, A43344-169, A43344-170, A43344-171, A43344-172, A43344-173, A43546-101, A43546-102, A43546-103, A43546-104, A43546-105, A43546-106, A43546-107, A43546-108, A43546-109, A43546-110, A43546-111, A43546-112, A43546-113, A43546-114, A43546-115, A43546-116, A43571-095, A43571-096, A43571-097, A43571-098, A43571-099, A43571-100, A43571-101, A43571-102, A43571-103, A43571-104, A43571-105, A43571-106, A43571-107, A43571-108, A43571-109, A43571-110, A43571-111, A43571-112, A43571-113, A43571-114, A44852-091, A44852-092, A44852-093, A44852-094, A44852-095, A44852-096, A45084-086, A45084-087, A45084-088, A45084-089, A45084-090, A45084-091, A45084-092, A45084-093, A45084-094, A45084-095, A45084-096, A45084-097, A45084-098, A45084-099, A45084-100, A45084-101, A45084-102, A45084-103, A45084-104, A45084-105, A45084-106, A45084-107, A45084-108, A45084-109, A45084-110, A45084-111, A45084-112, A45084-113, A45084-114, A45084-115, A46628-221, A46628-222, A46628-223, A46628-224, A46628-225, A46628-226, A46628-227, A46628-228, A46628-229, A46628-230, A46628-231, A46628-232, A46628-233, A46628-234, A46628-235, A46628-236, A46628-237, A46628-238, A46628-239, A46628-240, A46628-241, A46628-242, A46628-243, A46628-244, A46628-245, A46628-246, A46628-247, A46628-248, A46628-249, A46628-250, A46628-251, A47081-045, A47081-046, A47081-047, A47081-048, A47081-049, A47081-050, A47081-051, A47081-052, A47081-053, A47081-054, A47081-055, A47081-056, A47081-057, A47081-058, A47081-059, A47081-060, A47087-060, A47087-061, A47087-062, A47087-063, A47087-064, A47087-065, A47087-066, A47087-067, A47087-068, A47087-069, A47087-070, A47087-071, A47087-072, A47087-073, A47087-074, A48098-049, A48098-050, A48098-051, A48098-052, A48098-053, A48098-054, A48098-055, A48098-056, A48098-057, A48098-058, A48098-059, A48098-060, A48098-061, A48098-062, A49710-112, A49710-113, A49710-114, A49710-115, A49710-116, A49710-117, A49710-119, A49710-120, A49710-121, A49710-122, A49710-123, A49710-124, A49710-125, A49710-126, A49710-127, A49747-137, A49747-138, A49747-139, A49747-140, A49747-141, A49747-142, A49747-143, A49747-144, A49747-145, A50082-058, A50082-059, A50082-060, A50082-061, A50082-062, A50082-063, A50082-064, A50082-065, A50082-066, A50082-067, A50082-068, A50082-069, A50082-070, A50082-071, A50082-072, A50082-073, A50082-074, A50082-075, A50082-076, A50082-077, A50082-078, A50082-079, A50082-080, A50082-081, A50082-082, A50082-083, A50082-084, A50082-085, A50082-086, A50082-087, A50082-088, A50082-089, A50968-031, A50968-032, A50968-033, A50968-034, A50968-035, A50968-036, A50968-037, A50968-038, A50968-039, A50968-040, A50968-041, A50968-042, A50968-043, A50968-044, A50968-045, A50968-046, A50968-047, A50968-048, A50968-049, A50968-050, A50968-051, A50968-052, A50968-053, A50968-054, A50968-055, A51981-026, A51981-027, A51981-028, A51981-029, A51981-030, A51981-031, A51981-032, A51981-033, A51981-034, A51981-035, A51981-036, A51981-037, A51981-038, A51981-039, A51981-040, A51981-041, A51981-042, A51981-043, A51981-044, A51981-045, A51981-046, A51981-047, A51981-048, A52460-162, A52460-163, A52460-164, A52460-165, A52460-166, A52460-167, A52460-168, A52460-169, A52460-170, A52460-171, A52460-173, A52460-180, A54119-110, A54119-111, A54119-112, A54119-113, A54119-114, A54119-115, A54119-116, A54119-117, A54119-118, A54119-119, A54119-120, A54119-121, A54464-086, A54464-087, A54464-088, A54464-089, A54464-093, A54464-094, A54464-095, A54464-103, A54464-107; e) DBM T44115 15CC GRAFTON CRUNCH, REF T44115, GTIN 00643169123113, Serial Numbers: A26631-051, A26631-052, A26631-053, A26631-054, A26631-055, A26631-056, A26631-057, A38377-125, A38377-126, A38377-127, A38377-128, A38377-129, A38377-130, A38377-131, A38377-132, A38377-133, A38471-041, A38471-042, A38471-043, A38471-044, A38471-045, A38471-046, A38471-047, A38472-047, A38472-048, A38472-049, A38472-050, A38472-051, A38472-052, A38796-048, A38796-049, A38796-050, A38796-051, A38796-052, A38796-053, A38796-054, A38796-055, A38796-056, A38796-057, A38796-058, A38796-059, A38796-060, A38796-061, A38796-062, A38796-063, A38796-064, A38796-065, A38796-066, A38796-067, A39047-108, A39047-109, A39047-110, A39047-111, A39047-112, A39047-113, A39091-085, A39091-086, A39091-087, A39091-088, A39091-089, A39091-090, A39091-091, A39091-092, A39091-093, A39091-094, A39111-182, A39111-183, A39111-184, A39111-185, A39111-186, A39111-187, A39111-188, A39111-189, A39111-190, A39111-191, A39111-192, A39111-193, A39111-194, A39111-195, A39111-196, A39202-053, A39202-054, A39202-055, A39202-056, A39202-057, A39202-058, A39202-059, A39202-060, A39202-061, A39202-062, A39202-063, A39202-064, A39202-065, A39202-066, A39202-067, A39202-068, A39202-069, A39202-070, A39202-071, A39245-046, A39245-047, A39245-048, A39245-049, A39245-050, A39245-051, A39245-052, A39245-053, A39245-054, A39245-055, A39245-056, A39283-114, A39283-115, A39283-116, A39283-117, A39283-118, A39283-119, A39413-267, A39413-268, A39413-269, A39413-270, A39413-271, A39413-272, A39413-273, A39413-274, A39413-275, A39413-276, A39413-277, A39512-026, A39512-027, A39512-028, A39512-029, A39512-030, A39512-031, A39512-032, A39512-033, A39512-034, A39512-035, A39512-036, A39512-037, A39512-038, A39512-039, A39512-040, A39512-041, A39512-042, A39512-043, A39512-044, A39512-045, A39512-046, A39512-047, A39516-092, A39516-093, A39516-094, A39516-095, A39516-096, A39520-103, A39520-104, A39520-105, A39520-106, A39520-107, A39520-108, A39520-109, A39520-110, A39520-111, A39811-170, A39811-171, A39811-172, A39811-173, A39811-174, A39811-175, A39811-176, A39811-177, A39811-178, A39811-179, A39811-180, A39866-164, A39866-165, A39866-166, A39866-167, A39866-168, A39866-169, A39866-170, A39866-171, A39866-172, A39866-173, A39867-049, A39867-050, A39867-051, A39867-052, A39867-053, A39867-054, A39867-055, A39867-056, A39867-057, A39962-052, A39962-053, A39962-054, A39962-055, A39962-056, A39962-057, A39962-058, A40005-103, A40005-104, A40005-105, A40005-106, A40005-107, A40005-108, A40005-109, A40005-110, A40005-111, A40005-112, A40108-144, A40108-145, A40108-146, A40108-147, A40108-148, A40108-149, A40108-150, A40108-151, A40108-152, A40108-153, A40108-154, A40108-155, A40108-156, A40112-053, A40112-054, A40112-055, A40112-056, A40112-057, A40112-058, A40112-059, A40112-060, A40112-061, A40112-062, A40112-063, A40138-042, A40138-043, A40138-044, A40138-045, A40138-046, A40138-047, A40138-048, A40138-049, A40173-162, A40173-163, A40173-164, A40173-165, A40173-166, A40173-167, A40173-168, A40173-169, A40173-170, A40246-056, A40246-057, A40246-058, A40246-059, A40246-060, A40246-061, A40246-062, A40246-063, A40309-097, A40309-098, A40309-099, A40309-100, A40309-101, A40309-102, A40309-103, A40309-104, A40309-105, A40309-106, A40309-107, A40309-108, A40309-109, A40309-110, A40309-111, A40453-026, A40453-027, A40453-028, A40453-029, A40453-030, A40453-031, A40453-032, A40453-033, A40453-034, A40453-035, A40543-075, A40543-076, A40543-077, A40543-078, A40543-079, A40543-080, A40543-081, A40543-082, A40543-083, A40543-084, A40543-085, A40652-126, A40652-127, A40652-128, A40652-129, A40652-130, A40652-131, A40733-040, A40733-041, A40733-042, A40733-043, A40733-044, A40733-045, A40733-046, A40733-047, A40733-048, A40733-049, A40733-050, A40733-051, A40733-052, A40733-053, A40736-081, A40736-082, A40736-083, A40736-084, A40736-085, A40736-086, A40736-087, A40736-088, A40736-089, A40736-090, A40736-091, A40913-086, A40913-087, A40913-088, A40913-089, A40913-090, A40913-091, A40913-092, A40913-093, A40913-094, A40913-095, A40913-096, A40915-101, A40915-102, A40915-103, A40915-104, A40915-105, A40915-106, A40915-107, A40915-108, A40915-109, A40915-110, A40915-111, A40915-112, A40915-113, A40915-114, A40915-115, A40915-116, A40915-117, A40915-118, A40953-174, A40953-175, A40953-176, A40953-177, A40953-178, A40953-179, A40953-180, A40953-181, A40953-182, A40989-196, A40989-197, A40989-198, A40989-199, A40989-200, A40989-201, A40989-202, A40989-203, A40989-204, A40989-205, A40989-206, A40999-031, A40999-032, A40999-033, A40999-034, A40999-035, A40999-036, A40999-037, A40999-038, A41012-047, A41012-048, A41012-049, A41012-050, A41012-051, A41012-052, A41012-053, A41012-054, A41012-055, A41012-056, A41012-057, A41012-058, A41012-059, A41012-060, A41012-061, A41012-062, A41012-063, A41270-121, A41270-122, A41270-123, A41270-124, A41270-125, A41270-126, A41270-127, A41270-128, A41270-129, A41270-130, A41270-131, A41270-132, A41270-133, A41270-134, A41270-135, A41270-136, A41270-137, A41270-138, A41270-139, A41270-140, A41273-031, A41273-032, A41273-033, A41273-034, A41273-035, A41273-036, A41397-031, A41397-032, A41397-033, A41397-034, A41397-035, A41397-036, A41400-052, A41400-053, A41400-054, A41400-055, A41400-056, A41400-057, A41400-058, A41400-059, A41400-060, A41439-177, A41439-178, A41439-179, A41439-180, A41439-181, A41464-210, A41464-211, A41464-212, A41464-213, A41464-214, A41464-215, A41464-216, A41464-217, A41464-218, A41472-174, A41472-175, A41472-176, A41472-177, A41472-178, A41472-179, A41472-180, A41472-181, A41472-182, A41472-183, A41630-031, A41630-032, A41630-033, A41630-034, A41630-035, A41630-036, A41630-037, A41630-038, A41630-039, A41630-040, A41630-041, A41630-042, A41630-043, A41630-044, A41630-045, A41630-046, A41630-047, A41630-048, A41784-086, A41784-087, A41784-088, A41784-089, A41784-090, A41784-091, A41784-092, A41784-093, A41784-094, A41784-095, A41784-096, A41784-097, A41879-091, A41879-092, A41879-093, A41879-094, A41879-095, A41879-096, A41879-097, A41879-098, A41879-099, A41879-100, A41879-101, A41879-102, A41879-103, A41879-104, A41905-026, A41905-027, A41905-028, A41905-029, A41905-030, A41905-031, A41905-032, A41905-033, A41905-034, A41991-068, A41991-069, A41991-070, A41991-071, A41991-072, A41991-073, A41991-074, A41991-075, A41991-076, A41991-077, A41991-078, A41995-118, A41995-119, A41995-120, A41995-121, A41995-122, A41995-123, A41995-124, A41995-125, A41995-126, A41995-127, A41995-128, A41995-129, A41995-130, A41995-131, A41995-132, A42080-046, A42080-047, A42080-048, A42080-049, A42080-050, A42080-051, A42080-052, A42080-053, A42080-054, A42080-055, A42102-046, A42102-047, A42102-048, A42102-049, A42102-050, A42102-051, A42102-052, A42102-053, A42102-054, A42102-055, A42102-056, A42102-057, A42102-058, A42102-059, A42102-060, A42102-061, A42193-088, A42193-089, A42193-090, A42193-091, A42193-092, A42193-093, A42193-094, A42193-095, A42193-096, A42225-043, A42225-044, A42225-045, A42225-046, A42225-047, A42225-048, A42225-049, A42225-050, A42225-051, A42225-052, A42225-053, A42225-054, A42225-055, A42225-056, A42225-057, A42225-058, A42225-059, A42225-060, A42225-061, A42234-061, A42234-062, A42234-063, A42234-064, A42234-065, A42234-066, A42234-067, A42234-068, A42251-078, A42251-079, A42251-080, A42251-081, A42251-082, A42251-083, A42251-084, A42251-085, A42251-086, A42251-087, A42251-088, A42251-089, A42251-090, A42251-091, A42251-092, A42282-133, A42282-134, A42282-135, A42282-136, A42282-137, A42282-138, A42282-139, A42282-140, A42282-141, A42282-142, A42282-143, A42282-144, A42539-046, A42539-047, A42539-048, A42539-049, A42539-050, A42539-051, A42539-052, A42539-053, A42539-054, A42539-055, A42539-056, A42607-057, A42607-058, A42607-059, A42607-060, A42607-061, A42607-062, A42607-063, A42607-064, A42607-065, A42607-066, A42607-067, A42607-068, A42607-069, A42607-070, A42607-071, A42607-072, A42607-073, A42607-074, A42607-075, A42607-076, A42607-077, A42661-092, A42661-093, A42661-094, A42661-095, A42661-096, A42661-097, A42662-129, A42662-130, A42662-131, A42662-132, A42662-133, A42662-134, A42662-135, A42662-136, A42662-137, A42665-085, A42665-086, A42665-087, A42665-088, A42665-089, A42665-090, A42665-091, A42665-092, A42665-093, A42665-094, A42665-095, A42665-096, A42665-097, A42665-098, A42665-099, A42665-100, A42754-147, A42754-148, A42754-149, A42754-150, A42754-151, A42754-152, A42754-153, A42754-154, A42754-155, A42791-064, A42791-065, A42791-066, A42791-067, A42791-068, A42791-069, A42791-070, A42791-071, A42791-072, A42791-073, A42791-074, A42791-075, A42791-076, A42791-077, A42791-078, A42791-079, A42826-053, A42826-054, A42826-055, A42826-056, A42826-057, A42826-058, A42826-059, A42826-060, A42826-061, A42826-062, A42826-063, A42826-064, A42826-065, A42872-053, A42872-054, A42872-055, A42872-056, A42872-057, A42872-058, A42872-059, A42872-060, A42872-061, A42872-062, A42872-063, A42872-064, A42872-065, A42872-066, A42872-067, A42872-068, A42872-069, A42872-070, A42872-071, A42932-026, A42932-027, A42932-028, A42932-029, A42932-030, A42932-031, A42932-032, A42932-033, A42932-034, A42932-035, A42932-036, A42932-037, A42935-155, A42935-156, A42935-157, A42935-158, A42935-159, A42935-160, A42935-161, A42935-162, A42935-163, A42935-164, A42972-089, A42972-090, A42972-091, A42972-092, A42972-093, A42972-094, A42972-095, A42972-096, A42972-097, A42972-098, A42972-099, A42972-100, A42972-101, A42972-102, A42982-031, A42982-032, A42982-033, A42982-034, A42982-035, A42982-036, A42982-037, A42982-038, A43058-034, A43058-035, A43058-036, A43058-037, A43058-038, A43058-039, A43058-040, A43058-041, A43058-042, A43058-043, A43058-044, A43058-045, A43058-046, A43058-047, A43058-048, A43058-049, A43100-081, A43100-082, A43100-083, A43100-084, A43100-085, A43100-086, A43100-087, A43100-088, A43100-089, A43100-090, A43100-091, A43104-056, A43104-057, A43104-058, A43104-059, A43104-060, A43104-061, A43104-062, A43104-063, A43104-064, A43104-065, A43104-066, A43104-067, A43104-068, A43104-069, A43104-070, A43104-071, A43104-072, A43104-073, A43112-064, A43112-065, A43112-066, A43112-067, A43112-068, A43112-069, A43112-070, A43112-071, A43112-072, A43112-073, A43191-057, A43191-058, A43191-059, A43191-060, A43191-061, A43191-062, A43191-063, A43191-064, A43191-065, A43191-066, A43191-067, A43191-068, A43191-069, A43191-070, A43191-071, A43191-072, A43193-040, A43193-041, A43193-042, A43193-043, A43193-044, A43193-045, A43193-046, A43193-047, A43193-048, A43193-049, A43193-050, A43193-051, A43284-221, A43284-222, A43284-223, A43284-224, A43284-225, A43284-226, A43284-227, A43285-046, A43285-047, A43285-048, A43285-049, A43285-050, A43285-051, A43285-052, A43285-053, A43285-054, A43285-055, A43285-056, A43285-057, A43285-058, A43298-044, A43298-045, A43298-046, A43298-047, A43298-048, A43298-049, A43298-050, A43298-051, A43298-052, A43298-053, A43298-054, A43298-055, A43298-056, A43298-057, A43298-058, A43340-036, A43340-037, A43340-038, A43340-039, A43340-040, A43340-041, A43340-042, A43340-043, A43340-044, A43340-045, A43340-046, A43340-047, A43340-048, A43340-049, A43340-050, A43340-051, A43340-052, A43340-053, A43374-040, A43374-041, A43374-042, A43374-043, A43374-044, A43374-045, A43374-046, A43374-047, A43374-048, A43374-049, A43374-050, A43395-054, A43395-055, A43395-056, A43395-057, A43395-058, A43395-059, A43395-060, A43395-061, A43497-157, A43497-158, A43497-159, A43497-160, A43497-161, A43497-162, A43497-164, A43497-165, A43501-026, A43501-027, A43501-028, A43501-029, A43501-030, A43501-031, A43501-032, A43501-033, A43501-034, A43501-035, A43577-095, A43577-096, A43577-097, A43577-098, A43577-099, A43577-100, A43577-101, A43577-102, A43577-103, A43577-104, A43577-105, A43634-112, A43634-113, A43634-114, A43634-115, A43634-116, A43634-117, A43634-118, A43634-119, A43634-120, A43698-131, A43698-132, A43698-133, A43698-134, A43698-135, A43698-136, A43698-137, A43698-138, A43698-139, A43698-140, A43698-141, A43698-142, A43698-143, A43698-144, A43751-121, A43751-122, A43751-123, A43751-124, A43751-125, A43751-126, A43751-127, A43904-036, A43904-037, A43904-038, A43904-039, A43904-040, A43904-041, A43904-042, A43904-043, A43904-044, A43904-045, A43909-113, A43909-114, A43909-115, A43909-116, A43909-117, A43909-118, A43909-119, A43909-120, A43909-121, A43936-089, A43936-090, A43936-091, A43936-092, A43936-093, A43936-094, A43936-095, A43936-096, A43936-097, A43936-098, A43936-099, A43936-100, A43936-101, A43936-102, A43936-103, A43936-104, A43936-105, A43936-106, A43936-107, A43938-046, A43938-047, A43938-048, A43938-049, A43938-050, A43938-051, A43938-052, A43938-053, A43938-054, A43938-055, A43938-056, A43938-057, A43938-058, A43938-059, A43938-060, A43938-061, A43948-045, A43948-046, A43948-047, A43948-048, A43948-049, A43948-050, A43948-051, A43948-052, A43948-053, A43948-054, A43948-055, A43948-056, A43948-057, A43948-058, A43983-105, A43983-106, A43983-107, A43983-108, A43983-109, A43983-110, A439
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1324-2023
Related Recalls
Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
Grafton DBM (Flex), Model Numbers: T42100, Demineralized Bone Matrix
Medtronic Sofamor Danek USA
Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.
Grafton DMB Matrix, Model Numbers: a) T42200, b) T42210; Demineralized Bone Matrix
Medtronic Sofamor Danek USA
Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.