Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 AlluraXperFDlO C, 722002 Allura Xper FD1O F, 722003 Allura Xper FD1O, 722005 Allura Xper FD1O/10, 722006 Allura Xper FD2O, 722008 Allura Xper FD2O Biplane, 722010 Allura Xper FD1O, 722011 Allura Xper FD1O/10, 722012 Allura Xper FD2O, 722013 Allura XPER FD2O BIPLANE, 722014 Allura Xper FD1O OR Table, 722015 Allura Xper FD2O OR Table, 722019 Allura Xper FD1O/10 OR Table, 722020 Allura Xper FD2O Biplane OR Table, 722022 Allura Xper FD1O OR Table, 722023 Allura Xper FD2O OR Table, 722024 Allura Xper FD1O/10 OR Table, 722025 Allura Xper FD2O Biplane OR Table, 722026 Allura Xper FD1O, 722027 Allura Xper FD1O/10, 722028 Allura Xper FD2O, 722029 Allura Xper FD2O/10, 722031 Allura CV2O, 722033 Allura Xper FD1O OR Table, 722034 Allura Xper FD1O/10 OR Table, 722035 Allura Xper FD2O OR Table, 722036 Allura Xper FD2O/10 OR Table, 722038 Allura Xper FD2O/20, 722039 Allura Xper FD2O/20 OR Table, 722058 Allura Xper FD20115, 722059 Allura Xper FD2O/15 OR Table, 722123 Field extensions Xper cardio systems, 722124 Field extensions Xper vascular systems, 722126 SmartPath to AlluraClarity cardio, 722127 SmartPath to AlluraClarity vascular, 722133 Field ext. Xper cardio systems R7.6, 722134 Field ext. Xper vascular systems R7.6, 722400 Cardio Vascular-Allura Centron, 889006 Diamond Select Allura Xper FD1 0 The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Tis included, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper FD series is compatible with a hybrid Operating Room. FD10 is compatible with specific magnetic navigation systems

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Foot Switches: 452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m; 452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2; 459800415531 Wireless Footswitch 3P; 459800415561 Wireless footswitch set 3P.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Medical Systems, Inc.
Reason for Recall:
Loss of key image functionality due to a bent pedal of the Footswitch.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 AlluraXperFDlO C, 722002 Allura Xper FD1O F, 722003 Allura Xper FD1O, 722005 Allura Xper FD1O/10, 722006 Allura Xper FD2O, 722008 Allura Xper FD2O Biplane, 722010 Allura Xper FD1O, 722011 Allura Xper FD1O/10, 722012 Allura Xper FD2O, 722013 Allura XPER FD2O BIPLANE, 722014 Allura Xper FD1O OR Table, 722015 Allura Xper FD2O OR Table, 722019 Allura Xper FD1O/10 OR Table, 722020 Allura Xper FD2O Biplane OR Table, 722022 Allura Xper FD1O OR Table, 722023 Allura Xper FD2O OR Table, 722024 Allura Xper FD1O/10 OR Table, 722025 Allura Xper FD2O Biplane OR Table, 722026 Allura Xper FD1O, 722027 Allura Xper FD1O/10, 722028 Allura Xper FD2O, 722029 Allura Xper FD2O/10, 722031 Allura CV2O, 722033 Allura Xper FD1O OR Table, 722034 Allura Xper FD1O/10 OR Table, 722035 Allura Xper FD2O OR Table, 722036 Allura Xper FD2O/10 OR Table, 722038 Allura Xper FD2O/20, 722039 Allura Xper FD2O/20 OR Table, 722058 Allura Xper FD20115, 722059 Allura Xper FD2O/15 OR Table, 722123 Field extensions Xper cardio systems, 722124 Field extensions Xper vascular systems, 722126 SmartPath to AlluraClarity cardio, 722127 SmartPath to AlluraClarity vascular, 722133 Field ext. Xper cardio systems R7.6, 722134 Field ext. Xper vascular systems R7.6, 722400 Cardio Vascular-Allura Centron, 889006 Diamond Select Allura Xper FD1 0 The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Tis included, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper FD series is compatible with a hybrid Operating Room. FD10 is compatible with specific magnetic navigation systems

Product Codes/Lot Numbers:

Foot Switches: 452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m; 452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2; 459800415531 Wireless Footswitch 3P; 459800415561 Wireless footswitch set 3P.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1280-2016

Related Recalls

Class I - Dangerous

Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.

Mar 7, 2026 Other Medical Devices Nationwide View Details →

Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.

Sep 25, 2025 Infusion Pumps View Details →

Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.

Sep 25, 2025 Other Medical Devices View Details →